(31 days)
Not Found
No
The summary describes a standard automated hematology analyzer using established technologies like impedance, light scattering, and turbidimetry. There is no mention of AI, ML, or related concepts in the device description, intended use, or performance studies.
No
This device is an in vitro diagnostic (IVD) device used for testing whole blood and plasma specimens to provide quantitative information on complete blood cell and differential counts, and conventional CRP levels. It is used for diagnosis and monitoring, not for therapy.
Yes
The "Intended Use / Indications for Use" section explicitly states that the device is "used for the in vitro diagnostic testing of whole blood & plasma specimens."
No
The device description clearly states it is an automated hematology analyzer that utilizes physical methods like cytochemistry, impedance, light scattering, and turbidity, indicating it is a hardware device with integrated software.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use/Indications for Use: The very first sentence explicitly states it is "used for the in vitro diagnostic testing of whole blood & plasma specimens."
- Device Description: It describes how the device analyzes blood and plasma samples to provide quantitative information on various parameters, including CRP levels, which are used for diagnostic purposes.
- Reagents, Controls, and Calibrator: The description mentions the use of dedicated reagents, controls, and a calibrator, which are typical components of an IVD system used for testing biological samples.
- Measurement of Conventional CRP: The intended use states that the measurement of conventional CRP "aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders," which are all diagnostic applications.
N/A
Intended Use / Indications for Use
The ABX MICROS CRP 200 is an open tube, automated (microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood & plasma specimens. The device operates in complete blood count (CBC) mode or or in CBC & Creactive protein (CRP) mode.
Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.
For the CRP mode, the MICROS CRP 200 uses dedicated HoribaABX reagents (ABX CRP REA), controls (ABX CRP TROL I & III) and calibrator (ABX CRP STD).
The MICROS CRP 200 may be coupled, on option, with an information management (iM) system.
Product codes
GKZ, DCK, JJX, JIS, JQP
Device Description
ABX MICROS CRP 200 is an automated hematology analyzer that was developed by Horiba ABX (Montpellier, France). It utilizes cytochemistry, focused flow imredance, light scattering, and turbidity to provide quantitative information on complete blood cell and differential counts, in addition to other hematology parameters and conventional CRP.
CRP levels are measured in patients by reacting anti-CRP antibody coated latex particles with lysed blood, and determining the rate of the turbidimetric reaction in the near infrared spectrum.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
The data presented in this 510K Pre-market Notification demonstrate good precision in accordance with EP5-A (NCCLS guidelines) and is entirely acceptable for all evailable parameters.
The linearity claim for the parameters WBC (0 - 80 x 103/μL), RBC (0 - 7.5 x 109/μL), HGB (0 - 23g/dl), HCT (0 - 62.4%), PLT (0 - 900 x 103/pL), conventional CRP for whole blood (0 - 200 mg/L), and plasma (0 -- 150 mg/L) are entirely supported by the clinical data provided in this submission.
Accuracy (Inter-procedural Correlation) showed no evidence of significant bias between the ABX MICROS CRP 200 and the predicate devices.
The sample stability for leukocyte counts and the three part leukocyte differential is acceptable for over a 48 hour period from the time of blood collection at both room temperature and 4°C. For C-Reactive Protein determinations data assures there is sample stability over a 72 hour period at both room temperature and 4°C.
No effect of contamination of the instrument was dissimulated by the clinical data of this study, supporting a carryover claim of
§ 864.5220 Automated differential cell counter.
(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
0
DEC 2 9 2005
Premarket Notification [510(k)] Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
The assigned 510(k) number is : 10533308
- Company: Horiba ABX Parc Euromédecine Rue du Caducée - BP 7290 34184 Montpellier cedex 4 FRANCE Telephone: + (33) 4 67 14 73 20 Fax: + (33) 4 67 14 15 17
Contact Person: Tim Lawton (tlawton@fr.abx.fr)
Date Prepared: 22nd November 2005
Instrument :
CRP
Trade/Proprietary Name: | ABX MICROS CRP 200 |
---|---|
Common or Usual Name: | Automated cell counter and |
Automated differential cell counter | |
Device Class | Class II : Special Controls Guidance Document |
Classification Name: | Automated cell counter (§864.5200) and |
Automated differential cell counter (§864.5220) | |
Product Code: | GKZ : for the hematology analyzer |
CRP Reagent | |
Trade/Proprietary Name: | ABX CRP REA |
Common or Usual Name: | Conventional CRP Reagent |
Device Class | Class II : Special Controls Guidance Document |
Classification Name: | C-reactive protein immunological test system |
(§866.5270) | |
Product Code: | DCK : |
C-reactive protein, antigen, antiserum, and control |
1
CRP Control
Trade/Proprietary Name: | ABX CRP TROL I-III |
---|---|
Common or Usual Name: | CRP Control |
Device Class | Class I : exempt general conditions |
Classification Name: | Quality Control Material (assayed & unassayed) |
(§862.1660) | |
Product Code: | JJX |
Single (specified) analyte controls (assayed and | |
unassayed) | |
CRP Calibrator | |
Trade/Proprietary Name: | ABX CRP STD |
Common or Usual Name: | CRP Calibrator |
Device Class | Class II |
ice Class Class II Classification Name: : Quality Control Material (assayed & unassayed) (§862.1150) Product Code: JIS : Calibrator, primary
Optional device name :
Trade/Proprietary Name: | iM (Information Management) |
---|---|
Common or Usual Name: | Laboratory Information System |
Device Class | Class I : exempt |
Classification Name: | Calculator/data processing module for clinical use |
($862.2100) | |
Product Code: | JQP |
Substantial Equivalence:
The ABX MICROS CRP 200 can be considered substantially equivalent to the predicate device ABX MICROS CRP cleared to market under K002646.
The fundamental scientific technology for the analyzer itself has not changed. All parameters for complete blood count, differential leucocyte count, the reagents and controls, measuring principles, and the principles of operation are the same as previously cleared by the FDA.
2
The dedicated CRP reagents used on the ABX MICROS CRP 200 can be considered substantially equivalent to those previously used on the ABX MICROS CRP.
In addition, an optional add-on, to provide an Information Management System (iM) through which the ABX Micros CRP 200 ensures file management.
The overall effects on the Safety and Effectiveness of the Micros CRP 200 and the associated CRP reagents, controls and calibrators have been evaluated in internal & external clinical studies and internal validation procedures.
There is no known risk to the device Safety and Effectiveness.
Description:
ABX MICROS CRP 200 is an automated hematology analyzer that was developed by Horiba ABX (Montpellier, France). It utilizes cytochemistry, focused flow imredance, light scattering, and turbidity to provide quantitative information on complete blood cell and differential counts, in addition to other hematology parameters and conventional CRP.
CRP levels are measured in patients by reacting anti-CRP antibody coated latex particles with lysed blood, and determining the rate of the turbidimetric reaction in the near infrared spectrum.
Intended Use:
The ABX MICROS CRP 200 is an open tube, automated (microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood & plasma specimens. The device operates in complete blood count (CBC) mode or or in CBC & Creactive protein (CRP) mode.
Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.
For the CRP mode, the MICROS CRP 200 uses dedicated HoribaABX reagents (ABX CRP REA), controls (ABX CRP TROL I & III) and calibrator (ABX CRP STD).
The MICROS CRP 200 may be coupled, on option, with an information management (iM) system.
Discussion of Performance Data:
The studies and data analysis were carried out in accordance with appropriate indications given by the FDA guidelines.
3
The data presented in this 510K Pre-market Notification demonstrate good precision in accordance with EP5-A (NCCLS guidelines) and is entirely acceptable for all evailable parameters.
The linearity claim for the parameters WBC (0 - 80 x 103/μL), RBC (0 - 7.5 x 109/μL), HGB (0 - 23g/dl), HCT (0 - 62.4%), PLT (0 - 900 x 103/pL), conventional CRP for whole blood (0 - 200 mg/L), and plasma (0 -- 150 mg/L) are entirely supported by the clinical data provided in this submission.
Accuracy (Inter-procedural Correlation) showed no evidence of significant bias between the ABX MICROS CRP 200 and the predicate devices.
The sample stability for leukocyte counts and the three part leukocyte differential is acceptable for over a 48 hour period from the time of blood collection at both room temperature and 4°C. For C-Reactive Protein determinations data assures there is sample stability over a 72 hour period at both room temperature and 4°C.
No effect of contamination of the instrument was dissimulated by the clinical data of this study, supporting a carryover claim of Trade/Device Name: ABX MICROS CRP 200 (option: iM) ABX CRP REA ABX CRP TOL I & III ABX CRP STD
Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ, DCK, JJX, JIS, JQP Dated: November 22, 2005 Received: November 29, 2005
Dear Mr. Lawton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket
Image /page/5/Picture/12 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the wings of a bird. The caduceus is surrounded by a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
6
Page 2 - Mr. Tim Lawton
notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours.
Robert Becker
Robert L. Becker, Jr., MD, Ph.D Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
7
510(k) Number (if known): K053308
Device Name: ABX MICROS CRP 200 (option : iM)
Indications For Use:
The ABX MICROS CRP 200 is an open tube, automated (microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood & plasma specimens. The device operates in complete blood count (CBC) mode or in CBC & Creactive protein (CRP) mode.
Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.
The MICROS CRP 200 may be coupled, on option, with a information management (iM) system.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Robert Becker
Division Sign-Off
Page 1 of 1____1_
Office of In Vitro Diagnostic Device Evaluation and Safe 510(k)
8
510(k) Number (if known): K053308
Device Name: ABX CRP STD
Indications For Use:
The ABX CRP STD is intended to be used for the calibration of the ABX CRP REA reagent on the ABX MICROS CRP 200 hematology analyzer for quantitative determination of C-reactive protein in human whole blood or plasma serum.
Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Robert R. Beckerleaf
Division Sign-Off
Page 1 of 1_____________________________________________________________________________________________________________________________________________________________
Office of In Vitro Diagnostic Device Evaluation and Safety
9
510(k) Number (if known): K053308
Device Name: ABX CRP TROL I & III
Indications For Use:
The ABX CRP TROL I & III are intended to be used for the control of the ABX CRP REA reagent on the ABX MICROS CRP 200 hematology analyzer for quantitative determination of C-reactive protein in human whole blood or plasma serum.
Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Robert L. Reed
Division Sign-Off
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Office of In Vitro Diagnostig Devico Evaluation and Safer
510(k) KC53308
10
510(k) Number (if known): K053308
Device Name: ABX CRP REA
Indications For Use:
The ABX CRP REA is a quantitative assay for use with the ABX MICROS CRP 200 hematology analyzer for the determination of C-reactive protein in human whole blood or plasma serum.
Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Robert H. Beckef
Page 1 of _1
Office of In Vitro Dia Evaluation and Sa
510(k) K053308