K Number
K053308
Device Name
ABX MICROS CRP 200, OPTION IM
Manufacturer
Date Cleared
2005-12-29

(31 days)

Product Code
Regulation Number
864.5220
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The ABX MICROS CRP 200 is an open tube, automated (microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood & plasma specimens. The device operates in complete blood count (CBC) mode or or in CBC & Creactive protein (CRP) mode. Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders. For the CRP mode, the MICROS CRP 200 uses dedicated HoribaABX reagents (ABX CRP REA), controls (ABX CRP TROL I & III) and calibrator (ABX CRP STD). The MICROS CRP 200 may be coupled, on option, with an information management (iM) system.
Device Description
ABX MICROS CRP 200 is an automated hematology analyzer that was developed by Horiba ABX (Montpellier, France). It utilizes cytochemistry, focused flow imredance, light scattering, and turbidity to provide quantitative information on complete blood cell and differential counts, in addition to other hematology parameters and conventional CRP. CRP levels are measured in patients by reacting anti-CRP antibody coated latex particles with lysed blood, and determining the rate of the turbidimetric reaction in the near infrared spectrum.
More Information

Not Found

No
The summary describes a standard automated hematology analyzer using established technologies like impedance, light scattering, and turbidimetry. There is no mention of AI, ML, or related concepts in the device description, intended use, or performance studies.

No
This device is an in vitro diagnostic (IVD) device used for testing whole blood and plasma specimens to provide quantitative information on complete blood cell and differential counts, and conventional CRP levels. It is used for diagnosis and monitoring, not for therapy.

Yes

The "Intended Use / Indications for Use" section explicitly states that the device is "used for the in vitro diagnostic testing of whole blood & plasma specimens."

No

The device description clearly states it is an automated hematology analyzer that utilizes physical methods like cytochemistry, impedance, light scattering, and turbidity, indicating it is a hardware device with integrated software.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use/Indications for Use: The very first sentence explicitly states it is "used for the in vitro diagnostic testing of whole blood & plasma specimens."
  • Device Description: It describes how the device analyzes blood and plasma samples to provide quantitative information on various parameters, including CRP levels, which are used for diagnostic purposes.
  • Reagents, Controls, and Calibrator: The description mentions the use of dedicated reagents, controls, and a calibrator, which are typical components of an IVD system used for testing biological samples.
  • Measurement of Conventional CRP: The intended use states that the measurement of conventional CRP "aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders," which are all diagnostic applications.

N/A

Intended Use / Indications for Use

The ABX MICROS CRP 200 is an open tube, automated (microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood & plasma specimens. The device operates in complete blood count (CBC) mode or or in CBC & Creactive protein (CRP) mode.

Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.

For the CRP mode, the MICROS CRP 200 uses dedicated HoribaABX reagents (ABX CRP REA), controls (ABX CRP TROL I & III) and calibrator (ABX CRP STD).

The MICROS CRP 200 may be coupled, on option, with an information management (iM) system.

Product codes

GKZ, DCK, JJX, JIS, JQP

Device Description

ABX MICROS CRP 200 is an automated hematology analyzer that was developed by Horiba ABX (Montpellier, France). It utilizes cytochemistry, focused flow imredance, light scattering, and turbidity to provide quantitative information on complete blood cell and differential counts, in addition to other hematology parameters and conventional CRP.

CRP levels are measured in patients by reacting anti-CRP antibody coated latex particles with lysed blood, and determining the rate of the turbidimetric reaction in the near infrared spectrum.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

The data presented in this 510K Pre-market Notification demonstrate good precision in accordance with EP5-A (NCCLS guidelines) and is entirely acceptable for all evailable parameters.

The linearity claim for the parameters WBC (0 - 80 x 103/μL), RBC (0 - 7.5 x 109/μL), HGB (0 - 23g/dl), HCT (0 - 62.4%), PLT (0 - 900 x 103/pL), conventional CRP for whole blood (0 - 200 mg/L), and plasma (0 -- 150 mg/L) are entirely supported by the clinical data provided in this submission.

Accuracy (Inter-procedural Correlation) showed no evidence of significant bias between the ABX MICROS CRP 200 and the predicate devices.

The sample stability for leukocyte counts and the three part leukocyte differential is acceptable for over a 48 hour period from the time of blood collection at both room temperature and 4°C. For C-Reactive Protein determinations data assures there is sample stability over a 72 hour period at both room temperature and 4°C.

No effect of contamination of the instrument was dissimulated by the clinical data of this study, supporting a carryover claim of

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

0

DEC 2 9 2005

Premarket Notification [510(k)] Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is : 10533308

  • Company: Horiba ABX Parc Euromédecine Rue du Caducée - BP 7290 34184 Montpellier cedex 4 FRANCE Telephone: + (33) 4 67 14 73 20 Fax: + (33) 4 67 14 15 17
    Contact Person: Tim Lawton (tlawton@fr.abx.fr)

Date Prepared: 22nd November 2005

Instrument :

CRP

Trade/Proprietary Name:ABX MICROS CRP 200
Common or Usual Name:Automated cell counter and
Automated differential cell counter
Device ClassClass II : Special Controls Guidance Document
Classification Name:Automated cell counter (§864.5200) and
Automated differential cell counter (§864.5220)
Product Code:GKZ : for the hematology analyzer
CRP Reagent
Trade/Proprietary Name:ABX CRP REA
Common or Usual Name:Conventional CRP Reagent
Device ClassClass II : Special Controls Guidance Document
Classification Name:C-reactive protein immunological test system
(§866.5270)
Product Code:DCK :
C-reactive protein, antigen, antiserum, and control

1

CRP Control

Trade/Proprietary Name:ABX CRP TROL I-III
Common or Usual Name:CRP Control
Device ClassClass I : exempt general conditions
Classification Name:Quality Control Material (assayed & unassayed)
(§862.1660)
Product Code:JJX
Single (specified) analyte controls (assayed and
unassayed)
CRP Calibrator
Trade/Proprietary Name:ABX CRP STD
Common or Usual Name:CRP Calibrator
Device ClassClass II

ice Class Class II Classification Name: : Quality Control Material (assayed & unassayed) (§862.1150) Product Code: JIS : Calibrator, primary

Optional device name :

Trade/Proprietary Name:iM (Information Management)
Common or Usual Name:Laboratory Information System
Device ClassClass I : exempt
Classification Name:Calculator/data processing module for clinical use
($862.2100)
Product Code:JQP

Substantial Equivalence:

The ABX MICROS CRP 200 can be considered substantially equivalent to the predicate device ABX MICROS CRP cleared to market under K002646.

The fundamental scientific technology for the analyzer itself has not changed. All parameters for complete blood count, differential leucocyte count, the reagents and controls, measuring principles, and the principles of operation are the same as previously cleared by the FDA.

2

The dedicated CRP reagents used on the ABX MICROS CRP 200 can be considered substantially equivalent to those previously used on the ABX MICROS CRP.

In addition, an optional add-on, to provide an Information Management System (iM) through which the ABX Micros CRP 200 ensures file management.

The overall effects on the Safety and Effectiveness of the Micros CRP 200 and the associated CRP reagents, controls and calibrators have been evaluated in internal & external clinical studies and internal validation procedures.

There is no known risk to the device Safety and Effectiveness.

Description:

ABX MICROS CRP 200 is an automated hematology analyzer that was developed by Horiba ABX (Montpellier, France). It utilizes cytochemistry, focused flow imredance, light scattering, and turbidity to provide quantitative information on complete blood cell and differential counts, in addition to other hematology parameters and conventional CRP.

CRP levels are measured in patients by reacting anti-CRP antibody coated latex particles with lysed blood, and determining the rate of the turbidimetric reaction in the near infrared spectrum.

Intended Use:

The ABX MICROS CRP 200 is an open tube, automated (microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood & plasma specimens. The device operates in complete blood count (CBC) mode or or in CBC & Creactive protein (CRP) mode.

Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.

For the CRP mode, the MICROS CRP 200 uses dedicated HoribaABX reagents (ABX CRP REA), controls (ABX CRP TROL I & III) and calibrator (ABX CRP STD).

The MICROS CRP 200 may be coupled, on option, with an information management (iM) system.

Discussion of Performance Data:

The studies and data analysis were carried out in accordance with appropriate indications given by the FDA guidelines.

3

The data presented in this 510K Pre-market Notification demonstrate good precision in accordance with EP5-A (NCCLS guidelines) and is entirely acceptable for all evailable parameters.

The linearity claim for the parameters WBC (0 - 80 x 103/μL), RBC (0 - 7.5 x 109/μL), HGB (0 - 23g/dl), HCT (0 - 62.4%), PLT (0 - 900 x 103/pL), conventional CRP for whole blood (0 - 200 mg/L), and plasma (0 -- 150 mg/L) are entirely supported by the clinical data provided in this submission.

Accuracy (Inter-procedural Correlation) showed no evidence of significant bias between the ABX MICROS CRP 200 and the predicate devices.

The sample stability for leukocyte counts and the three part leukocyte differential is acceptable for over a 48 hour period from the time of blood collection at both room temperature and 4°C. For C-Reactive Protein determinations data assures there is sample stability over a 72 hour period at both room temperature and 4°C.

No effect of contamination of the instrument was dissimulated by the clinical data of this study, supporting a carryover claim of Trade/Device Name: ABX MICROS CRP 200 (option: iM) ABX CRP REA ABX CRP TOL I & III ABX CRP STD

Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ, DCK, JJX, JIS, JQP Dated: November 22, 2005 Received: November 29, 2005

Dear Mr. Lawton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

Image /page/5/Picture/12 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the wings of a bird. The caduceus is surrounded by a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

6

Page 2 - Mr. Tim Lawton

notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours.

Robert Becker

Robert L. Becker, Jr., MD, Ph.D Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

7

510(k) Number (if known): K053308

Device Name: ABX MICROS CRP 200 (option : iM)

Indications For Use:

The ABX MICROS CRP 200 is an open tube, automated (microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood & plasma specimens. The device operates in complete blood count (CBC) mode or in CBC & Creactive protein (CRP) mode.

Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.

The MICROS CRP 200 may be coupled, on option, with a information management (iM) system.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Robert Becker

Division Sign-Off

Page 1 of 1____1_

Office of In Vitro Diagnostic Device Evaluation and Safe 510(k)

8

510(k) Number (if known): K053308

Device Name: ABX CRP STD

Indications For Use:

The ABX CRP STD is intended to be used for the calibration of the ABX CRP REA reagent on the ABX MICROS CRP 200 hematology analyzer for quantitative determination of C-reactive protein in human whole blood or plasma serum.

Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Robert R. Beckerleaf

Division Sign-Off

Page 1 of 1_____________________________________________________________________________________________________________________________________________________________

Office of In Vitro Diagnostic Device Evaluation and Safety

K053308

9

510(k) Number (if known): K053308

Device Name: ABX CRP TROL I & III

Indications For Use:

The ABX CRP TROL I & III are intended to be used for the control of the ABX CRP REA reagent on the ABX MICROS CRP 200 hematology analyzer for quantitative determination of C-reactive protein in human whole blood or plasma serum.

Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Robert L. Reed

Division Sign-Off

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

Office of In Vitro Diagnostig Devico Evaluation and Safer

510(k) KC53308

10

510(k) Number (if known): K053308

Device Name: ABX CRP REA

Indications For Use:

The ABX CRP REA is a quantitative assay for use with the ABX MICROS CRP 200 hematology analyzer for the determination of C-reactive protein in human whole blood or plasma serum.

Measurement of conventional CRP aids in the evaluation of infection, tissue injury and therapy & monitoring of inflammatory disorders.

Prescription Use X (Part 21 CFR 801 Subpart D)


AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Robert H. Beckef

Page 1 of _1

Office of In Vitro Dia Evaluation and Sa

510(k) K053308