(167 days)
The PM-9000 Express Patient Monitor is a vital signs monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PM-9000 Express Patient Monitor has many features and functions, yet is easy to use through an integrated keypad, knob and an intuitive menu system.
The patient parameters that can be monitored by PM-9000 Express Patient Monitor are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Arterial Hemoglobin Oxygen Saturation(SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP), Carbon Dioxide (CO2), and Anaesthetic Gases(AG). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The PM-9000 Express Patient Monitor is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.
The PM-9000 Express Patient Monitor is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.
The PM-9000 Express Patient Monitor is a battery or line-powered patient monitor. The Patient Monitor acquires the physiological signals such as ECG respiration (RESP), non-invasive blood pressure (NIBP), saturation of pulse oxygen (SpO2), temperature (TEMP), invasive pressure (IBP), carbon dioxide(CO2) and anaesthetic gases(AG). These physiological signals are converted into digital data and processed. The PM-9000 Express Patient Monitor examines the data for alarm conditions and presents them on the color TFT display. The Patient Monitor also provides advantageous operating control for the user.
The optional built-in recorder, the optional CF memory card provides hard copies of all digital data and waveforms as well as Tabular and Graphic Trend Information, and storage the previous monitoring data information when power off accidentally.
The provided text describes a 510(k) premarket notification for the PM-9000 Express Patient Monitor, comparing it to predicate devices. However, it does not contain a table of acceptance criteria with reported device performance or information about a specific study designed to prove the device meets said acceptance criteria.
Instead, the document focuses on demonstrating substantial equivalence to predicate devices through compliance with various industry standards and general testing.
Here's an attempt to extract relevant information based on your request, noting where information is explicitly missing:
Acceptance Criteria and Study Information for PM-9000 Express Patient Monitor
The provided 510(k) summary does not include a direct table of acceptance criteria with corresponding device performance metrics. Instead, it asserts compliance with various national and international standards as evidence of safety and effectiveness, and substantial equivalence to predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance
No specific table of acceptance criteria with corresponding performance metrics (e.g., sensitivity, specificity, accuracy) is presented in the provided document. The device broadly claims compliance with established standards for various physiological measurements.
However, the document states:
- "all these parameters of the pending device and the predicate device comply with ANSI/AAMI standards, IEC standards, EN standards and ISO standards."
- "The PM-9000 Express Patient Monitor provides two kinds of Temperature accessories applicable. YSI series and MR series. Both of them have same measurement range and accuracy, and both comply with EN12470-4."
- "For CO2, both the pending device and the predicate device with EN864, ISO9918; for AQ, both the pending device and the predicate device comply with EN864, ISO11196, EN12598, EN7767."
- "Testing of the non-invasive blood pressure portion of the system was conducted according to the requirements outlined in the ANSI/AAMI Standards SP10 'Electronic automated sphygmomanometers.'"
Based on these statements, the "acceptance criteria" are implied to be the performance and safety requirements specified by these standards. The "reported device performance" is implicitly that the device meets these standards.
2. Sample Size Used for the Test Set and Data Provenance
This information is not provided in the document. The testing described is "Laboratory testing" and does not specify a "test set" in terms of patient data samples, nor does it mention data provenance (country, retrospective/prospective).
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not provided in the document. The summary focuses on engineering and regulatory compliance testing rather than clinical validation involving expert adjudicated ground truth.
4. Adjudication Method
This information is not provided in the document.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This information is not provided in the document. The document describes laboratory testing and compliance with standards, not a comparative effectiveness study involving human readers with and without AI assistance. The device itself is a patient monitor, not explicitly an AI-driven diagnostic tool in the context of this summary.
6. Standalone (Algorithm Only) Performance
The device is a "Patient Monitor" and its performance is described in the context of its overall function of acquiring, processing, and displaying physiological signals. The testing described focuses on the device as a whole and its compliance with relevant standards for patient monitoring equipment. It does not isolate an "algorithm only" performance separate from the integrated device.
7. Type of Ground Truth Used
The "ground truth" implicitly used for the testing described are the specifications and requirements defined by the various industry standards (e.g., ANSI/AAMI, IEC, EN, ISO). For instance, NIBP testing was against the requirements of ANSI/AAMI SP10. This would typically involve reference measurement techniques defined within those standards. There is no mention of pathology, outcomes data, or expert consensus in a clinical validation sense.
8. Sample Size for the Training Set
This information is not provided in the document. The document does not describe the use of machine learning algorithms requiring a distinct "training set" in the context of AI.
9. How Ground Truth for the Training Set Was Established
This information is not provided as there is no mention of a "training set" or AI-specific development.
Summary of Testing:
The PM-9000 Express Patient Monitor underwent:
- Laboratory testing to validate and verify design specifications.
- Environmental testing as identified in the FDA's DCRND November 1993 "Reviewer Guidance Document for Premarket notification Submissions" Draft Guidance Document.
- Compliance testing against:
- ANSI/AAMI standards (EC13-2002 for Cardiac monitors; SP10 for NIBP).
- IEC standards (IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-1-4, IEC60601-2-27, IEC60601-2-30).
- EN standards (EN864, EN865, EN12470-4 for Temperature, EN11196, EN12598, EN7767 for AG).
- ISO standards (ISO14971 for risk management, ISO9918 for CO2, ISO11196 for AG).
- Hazard analysis of the system and its software, with subsequent testing to validate overall operation.
- Safety testing performed by third-party agencies.
The conclusion drawn from this testing was that the device is "as safe, as effective, and performs as well as or better than the legally marketed predicate device."
{0}------------------------------------------------
MAY - 4 2006
Exhibit #1 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
The assigned 510(k) number is: K053234
Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
Contact Person:
- Li Dongling
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Date Prepared: .
Oct. 25, 2005
Name of the device:
Trade/Proprietary Name: PM-9000 Express Patient Monitor �
- Common Name: Patient Monitor ●
● Classification
| 21 CFR 870.2300 | Cardiac monitor (including cardiotachometer and rate alarm) | Class II |
|---|---|---|
| 21 CFR 870.1130 | Non-Invasive blood pressure measurement System | Class II |
| 21 CFR 870.1110 | Blood pressure computer | Class II |
| 21 CFR 880.2910 | Clinical Electronic Thermometers –Temperature Monitor with Probe | Class II |
| 21 CFR 870.2700 | Oximeter, Pulse | Class II |
| 21 CFR 870.2710 | Ear Oximeter, Pulse | Class II |
| 21 CFR 868.1400 | Carbon Dioxide Gas Analyzer | Class II |
PM-9000 Express Patient Monitor '
0051
{1}------------------------------------------------
510 (k) Summary
| 21 CFR 868.1500 Enflurane gas analyzer | Class II |
|---|---|
| 21 CFR 868.1620 Halothane gas analyzer | Class II |
| 21 CFR 868.1700 Nitrous Oxide gas analyzer | Class II |
| 21 CFR 868.1720 Oxygen gas analyzer | Class II |
Legally Marketed Predicate Device:
| K043348 | PM-8000 Patient Monitor (by Mindray Co., Ltd.) |
|---|---|
| K030038 | IntelliVue Models MP60, MP70 and MP90 Patient Monitors with portableTechnology and M3185 opt. C41 Wireless Bedside(by Philips Co., Ltd.) |
Description:
The PM-9000 Express Patient Monitor is a battery or line-powered patient monitor. The Patient Monitor acquires the physiological signals such as ECG respiration (RESP), non-invasive blood pressure (NIBP), saturation of pulse oxygen (SpO2), temperature (TEMP), invasive pressure (IBP), carbon dioxide(CO2) and anaesthetic gases(AG). These physiological signals are converted into digital data and processed. The PM-9000 Express Patient Monitor examines the data for alarm conditions and presents them on the color TFT display. The Patient Monitor also provides advantageous operating control for the user.
The optional built-in recorder, the optional CF memory card provides hard copies of all digital data and waveforms as well as Tabular and Graphic Trend Information, and storage the previous monitoring data information when power off accidentally.
Statement of intended Use:
The PM-9000 Express Patient Monitor is a vital signs monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PM-9000 Express Patient Monitor has many features and functions, yet is easy to use through an integrated keypad, knob and an intuitive menu system.
The patient parameters that can be monitored by PM-9000 Express Patient Monitor are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Saturation of Pulse Oxygen(SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP), Carbon Dioxide (CO2), and Anaesthetic Gases(AG). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The PM-9000 Express Patient Monitor is intended for use in health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.
005
{2}------------------------------------------------
The PM-9000 Express Patient Monitor is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.
Comparison of Technological Characteristics;
The PM-9000 Express Patient Monitor is substantially equivalent to systems currently marketed predicate devices. The design, components, storage technology and energy source of the PM-9000 Express Patient Monitor are similar to the predicate device named PM-8000 Patient Monitor. Both PM-9000 Express and PM-8000 provide a means for interfacing with a patient, collecting parameter specific physiological data, and processing the data for alarm generation and display of numeric values and waveforms on a bedside or central monitoring system. The parameters' specification of the PM-9000 Express Patient Monitor, including ECG, RESP, HR, PR, RESP, NIBP, SpO2, TEMP, IBP, is similar to the predicate PM-8000 Patient Monitor. The PM-9000 Express's specification of CO2 and AG is similar to IntelliVue MP70.
The notable difference between the technical specifications of the PM-9000 Express and PM-8000 is shown as following:
The PM-9000 Express Patient Monitor provides optional CF memory card, which can be used as a mass storage for monitoring patient information.
For the monitoring parameters, while the specification of some parameters is different, but all these parameters of the pending device and the predicate device comply with ANSI/AAMI standards, IEC standards, EN standards and ISO standards.
The PM-9000 Express Patient Monitor provides two kinds of Temperature accessories applicable. YSI series and MR series. Both of them have same measurement range and accuracy, and both comply with EN12470-4.
For CO2, both the pending device and the predicate device with EN864, ISO9918; for AQ, both the pending device and the predicate device comply with EN864, ISO11196, EN12598, EN7767.
These technological differences do not affect the safety or efficacy of the device. Any safety issues that may be raised by a software controlled medical device are the same issues already addressed by the predicate device and are addressed in the systems hazard analysis and in the system validation.
Testing:
Laboratory testing was conducted to validate and verify that the PM-9000 Express
MINDRAY®PROPRIETARY
{3}------------------------------------------------
Patient Monitor met all design specifications and was substantially equivalent to the predicate devices. This testing consisted of all environmental testing identified in the FDA's DCRND November 1993 "Reviewer Guidance Document for Premarket notification Submissions" Draft Guidance Document. Additional testing was performed to demonstrate compliance with the ANSUAAMI standards EC13-2002, "Cardiac monitors, heart rate meters, and alarms". Finally, a hazard analysis of the system and its software was performed and testing was conducted to validate the systems overall operation. Some safety testing has been performed by third party agencies to ensure the device complies with applicable industry and safety standards. The PM-9000 Express Patient Monitor has also been tested to assure compliance with the requirements of various published standards. including EN865, IEC60601-1, IEC60601-1-1. IEC60601-1-2, IEC60601-1-4, IEC60601-2-27, IEC60601-2-30, and ISO14971.
Testing of the non-invasive blood pressure portion of the system was conducted according to the requirements outlined in the ANSI/AAMI Standards SP10 "Electronic automated sphygmomanometers."
Although the device is neither life supporting nor life sustaining, diagnostic information derived from the use of the device and alarms generated by the device may be critical to the proper management of the patient. So, the areas of risk for this device are the same as other devices in this class, and the following:
- . Electrical shock
- Excessive electrical chassis leakage current can disturb the normal electrophysiology of the heart.
- Misdiagnosis
- Inadequate design of the signal processing and measurement circuitry or ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: program can lead generation of inaccurate diagnostic data. If inaccurate diagnostic data are used in managing the patient, the physician may prescribe a course of treatment that places the patient at risk unnecessarily.
- Inadequate design of the device's software, used to make various -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------measurements, can lead to generation of inaccurate diagnostic data. If inaccurate diagnostic data are used in managing the patient, the physician may prescribe a course of treatment that places the patient at risk unnecessarily.
- Inadequate design of the systems ability to alert the users through audible and visual indicators, can lead to user mistrust and/or inadequate response to the patient's condition. If an inadequate response to the patient's condition should occur the patient may unnecessarily be placed at risk.
Conclusion:
The conclusions drawn from clinical and laboratory testing of the PM-9000 Express Patient Monitor demonstrates that the device is as safe, as effective, and
MINDRAY*PROPRIETARY
A can a sua come a come of
0053
{4}------------------------------------------------
K053234 Page 5
510 (k) Summary
performs as well as or better than the legally marketed predicate device, the PM-8000 Patient Monitor numbered K#043348(by Mindray Co., Ltd) and the Intelli Vue model MP70 numbered K# 030038(by Philips Co., Ltd.).
.
005/
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 4 0000
Shenzhen Mindray Bio-medical Electronics Co., Ltd. c/o Ms. Susan Goldstein Falk 16 Electronics Avenue Danvers, MA 01923
Re: K053234
Trade Name: PM-9000 Express Patient Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: II (two) Product Code: MWI Dated: April 26, 2006 Received: April 27, 2006
Dear Ms. Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Susan Goldstein Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Bestimmt for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Exhibit #F
510(k) Number (if known): _ |< 05 32 3
Device Name: PM-9000 Express Patient Monitor
Indications For Use:
The PM-9000 Express Patient Monitor is a vital signs monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PM-9000 Express Patient Monitor has many features and functions, yet is easy to use through an integrated keypad, knob and an intuitive menu system.
The patient parameters that can be monitored by PM-9000 Express Patient Monitor are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Arterial Hemoglobin Oxygen Saturation(SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP), Carbon Dioxide (CO2), and Anaesthetic Gases(AG). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The PM-9000 Express Patient Monitor is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.
The PM-9000 Express Patient Monitor is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.
Prescription Use X (Per 21 CFR 801 Subpart D)
Over-The Counter Use OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
slinia
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K053234
concurrence of CDRH, Office of Device Evaluation (ODE)
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).