(141 days)
Not Found
No
The device description and performance studies focus on traditional immunoassay methods and analytical performance, with no mention of AI or ML.
No
The device is an in vitro diagnostic (IVD) intended for the quantitative measurement of C-Reactive Protein (CRP) to aid in the evaluation of infection, tissue injury, and inflammatory disorders. It does not directly treat or prevent a disease, but rather provides information for diagnosis and monitoring.
Yes
The device is intended for the quantitative measurement of C-Reactive Protein (CRP) concentration in serum or plasma, which is useful for determining the existence of inflammatory lesions and evaluating infection, tissue injury, and inflammatory disorders. This function aligns with the definition of a diagnostic device as it provides information for the diagnosis or evaluation of medical conditions.
No
The device is a reagent kit for in vitro diagnostic use, consisting of liquid reagents and intended for use on clinical chemistry analyzers. It is not a software-only device.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use/Indications for Use: The document explicitly states the device is "intended for the quantitative measurement of C-Reactive Protein (CRP) concentration in serum or plasma." It also mentions its use in "evaluation of infection, tissue injury, and inflammatory disorders in combination with a complete clinical evaluation." These are all typical uses for in vitro diagnostic devices, which are used to examine specimens from the human body to provide information for the diagnosis, treatment, or prevention of disease.
- Intended User / Care Setting: The document states "For in vitro diagnostic use." This is a clear indication that the device is intended for use in a laboratory or clinical setting for diagnostic purposes.
- Device Description: The description details the reagents used to perform the assay on "general clinical chemistry analyzers," which are common laboratory instruments used for in vitro diagnostic testing.
- Performance Studies: The document includes details about performance studies such as precision, linearity, detection limit, analytical specificity, and method comparison. These types of studies are required for IVD devices to demonstrate their analytical performance.
The information provided strongly aligns with the definition and characteristics of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The Nanopia Wide Range CRP Reagent is intended for the quantitative measurement of C-Reactive Protein (CRP) concentration in serum or plasma.
Measurement of CRP is useful for determining the existence of inflammatory lesions and to monitor treatment.
The Nanopia Wide Range CRP Calibrator is intended for the calibration of the Nanopia Wide Range CRP assay.
The assay is intended for use in the evaluation of infection, tissue injury, and inflammatory disorders in combination with a complete clinical evaluation.
Product codes (comma separated list FDA assigned to the subject device)
DCK, JIS
Device Description
The Nanopia Wide Range CRP assay consists of two liquid reagents. Reagent 1 is a buffering solution and Reagent 2 contains latex beads coated with mouse monoclonal anti-human CRP antibodies. The assay is for use on general clinical chemistry analyzers.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Precision/Reproducibility: Device imprecision was evaluated according to NCCLS EP5-A. Spiked serum samples were run in triplicate daily for 20 days.
Control 1: Mean 0.886 mg/L, Within Run SD 0.16, %CV 1.83, Total SD 0.022, %CV 2.46
Control 2: Mean 6.71 mg/L, Within Run SD 0.05, %CV 0.76, Total SD 0.089, %CV 1.31
Control 3: Mean 38.88 mg/L, Within Run SD 0.24, %CV 0.61, Total SD 0.430, %CV 1.11
Linearity/assay reportable range: The useable range of this device is 0.10 - 400 mg/L. Linearity was assessed using a sample with CRP values of 400 mg/L. Dilutions were prepared to create samples with theoretical concentrations from 40 to 400 mg/L in increments of 40 mg/L, plus a zero concentration sample. % recovery from 0 to 400 mg/L ranged from 97.5 to 102.7%. The acceptable criteria was 100% ± 5%. The results indicate that the assay is linear across the measuring range.
Detection limit: Analytical sensitivity is 0.10 mg/L, derived by extrapolation of plot of theoretical concentration vs. recovered concentration, assessed by diluting a serum sample with a concentration of 0.5 mg/L to 0.0 mg/L and measuring the mAbs in replicates of 10.
Analytical specificity: Interference of hemoglobin (up to 500 mg/dL), ascorbic acid (up to 100 mg/dL), free bilirubin (up to 50 mg/dL), conjugated bilirubin (up to 50 mg/dl), lipid emulsion (up to 5000 turbidity), Rheumatoid factor (up to 500 IU/mL), and ascorbic acid (up to 100 mg/dL) was evaluated on serum samples containing a nominal concentration of 3.7 mg/L. No interference was noted. Interference was defined as a result +/- 5 % of the control.
Method comparison with predicate device: Serum samples (n=98) ranging from 0.0 to 293 mg/L were measured using the Nanopia method and compared to the predicate method. Results: Nanopia = 1.015(Predicate) - 0.0349; R2 = 0.9992.
Matrix comparison: Paired samples of serum and EDTA plasma were collected concurrently from 50 individuals and measured using the Nanopia method. Results: Plasma = 0.994(Serum) + 0.03; r = 0.999. Additional serum pools were spiked with 110 mM of sodium citrate, 100 mM sodium oxalate, 25 mM EDTA, and 100 U/mL of sodium heparin and assayed. No significant differences observed between sample types.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 866.5270 C-reactive protein immunological test system.
(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).
0
Clinical Data, Inc.
USA: One Gateway Center, Suite 415 Newton, MA 02158 1-617-527-9933, Ext. 22 Fax: 1-617-527-8230
510(k) Summary
Nanopia Wide Range CRP
Koszsql The assigned 510(k) Number is:________________________________________________________________________________________________________________________________________________
- A. Analyte: C-Reactive Protein (CRP)
- B. Type of Test: Turbidimetric immunoassay
- C. Applicant:
One Gateway Center, Suite 415 Newton, MA 02158 Phone: 617-527-9933, Ext. 41 Fax: 617-527-8230
D. Proprietary and Established Names: Nanopia Wide Range CRP Reagent Nanopia Wide Range CRP Calibrator
E. Regulatory Information:
-
- Regulation section: 21 CFR § 866.5270, C-reactive protein immunological test system 21 CFR § 862.1150, Calibrator
-
- Classification: Class II
-
- Product Code: DCK, C-reactive protein, antigen, antiserum, and control JIS, Calibrator, primary
-
- Panel: Immunology (82)
1
F. Intended Use:
-
- Intended use(s):
The Nanopia Wide Range CRP Reagent is intended for the quantitative measurement of C-Reactive Protein (CRP) concentration in serum or plasma.
- Intended use(s):
Measurement of CRP is useful for determining the existence of inflammatory lesions and to monitor treatment.
The Nanopia Wide Range CRP Calibrator is intended for the calibration of the Nanopia Wide Range CRP assay.
-
- Indication(s) for use: See Intended Use section.
-
- Special condition for use statement(s): For in vitro diagnostic use.
Increases in CRP values are non-specific and should not be interpreted without a complete clinical history.
-
- Special instrument Reguirements: Clinical chemistry analyzers (testing performed on Roche Hitachi 917 analyzer)
G. Device Description:
The Nanopia Wide Range CRP assay consists of two liquid reagents. Reagent 1 is a buffering solution and Reagent 2 contains latex beads coated with mouse monoclonal anti-human CRP antibodies. The assay is for use on general clinical chemistry analyzers.
H. Substantial Equivalence Information:
-
- Predicate device name(s): N-Geneous Wide Range CRP Reagent.
-
- Predicate K number(s): K040241
2
3. Comparison with predicate:
Item | Device | Predicate |
---|---|---|
Intended Use | Quantitative measurement | |
of CRP in serum or plasma | Quantitative measurement | |
of CRP in serum or plasma | ||
Sample Matrix | Serum or plasma | Serum or plasma |
Antibody | Mouse monoclonal anti- | |
human CRP | Mouse monoclonal anti- | |
human CRP | ||
Assay Range | 0.1 to 400 mg/L | 0.04 to 320 mg/L |
Antibody substrate | Latex | Latex |
Number of | ||
calibrators | 5 | 5 |
The N-Geneous Wide Range CRP Reagent and the Nanopia Wide Range CRP are essentially the same product both produced by Daiichi Pure Chemicals Co., Ltd. for distribution by the respective companies_
I. Standard/Guidance Document Referenced (if applicable):
| NCCLS Guideline EP9-A - Method Comparison and Bias Estimation Using Patient
Samples |
---|
NCCLS Guideline EP5-A - Evaluation of Precision Performance of Clinical Chemistry |
Devices |
NCCLS Guideline EP6-A - Evaluation of the Linearity of Quantitative Analytical |
Methods |
NCCLS Guideline EP7-A - Interference Testing in Clinical Chemistry |
J. Test Principle:
Sample is mixed with the buffer solution and the anti-CRP antibody-coated beads. CRP in the sample binds the antibody-coated beads and agglutinates. The light scattering caused by an increase in particle size is measured. The amount of light scattering is proportional to the concentration of CRP in the sample.
3
K. Performance Characteristics (if/when applicable):
-
- Analytical performance:
- a. Precision/Reproducibility:
Device imprecision was evaluated according to NCCLS EP5-A. Spiked serum samples were run in triplicate daily for 20 days (units = mg/L).
Control | Mean | Within Run | Total | ||
---|---|---|---|---|---|
SD | % CV | SD | % CV | ||
1 | 0.886 | 0.16 | 1.83 | 0.022 | 2.46 |
2 | 6.71 | 0.05 | 0.76 | 0.089 | 1.31 |
3 | 38.88 | 0.24 | 0.61 | 0.430 | 1.11 |
b. Linearity/assay reportable range:
The useable range of this device is 0.10 - 400 mg/L (the Limit of Quantification to the upper end of the linear range).
Linearity was assessed using a sample with CRP values of 400 mg/L. Using a serum pool free of CRP dilutions were prepared to create samples with theoretical concentrations from 40 to 400 mg/L in increments of 40 mg/L. A sample from a serum pool free from CRP was included for a zero concentration. These samples were run in duplicate and the actual mean values were compared to the theoretical values. The % recovery from 0 to 400 mg/L ranged from 97.5 to 102.7%. Results are summarized below (units = mg/L).
(mg/L) | ||||
---|---|---|---|---|
Theoretical | ||||
CRP value | measurement value | % Recovery | ||
1 | 2 | mean | ||
0 | 0.00 | 0.00 | 0.00 | --- |
40 | 39.33 | 40.20 | 39.77 | 99.4 |
80 | 79.01 | 78.42 | 78.72 | 98.4 |
120 | 119.26 | 116.14 | 117.70 | 98.1 |
160 | 155.44 | 156.48 | 155.96 | 97.5 |
200 | 196.46 | 194.97 | 195.72 | 97.9 |
240 | 236.89 | 236.10 | 236.50 | 98.5 |
280 | 280.76 | 278.09 | 279.43 | 99.8 |
320 | 322.15 | 323.21 | 322.68 | 100.8 |
360 | 374.30 | 365.09 | 369.70 | 102.7 |
400 | 401.12 | 406.17 | 403.65 | 100.9 |
The acceptable criteria was 100% ± 5%
The results above indicate that the assay is linear across the measuring range of the assay
4
c. Traceability (controls, calibrators, or method):
The device is calibrated by a set of 5 calibrators (sold separately) with targeted CRP concentrations of 3, 6, 30, 180 and 360 mg/L. The calibrator is traceable to CRM470. A master calibrator is prepared and value assigned by multiple measurements of multiple lots using the device. The calibrators are traceable to this master calibrator and are value assigned using the device. Stability testing is described.
d. Detection limit:
Analytical sensitivity, defined as the lowest concentration at which the assay performs with + 2 SD as assessed by diluting a serum sample with a concentration of 0.5 mg/L to 0.0 mg/L and measuring the mAbs in replicates of 10. By this method, the functional sensitivity of this assay is said to be 0.10 mg/L (derived by extrapolation of plot of theoretical concentration vs. recovered concentration.
-
e. Analytical specificity:
Interference of endogenous substances was evaluated by testing for the effect of hemoglobin (up to 500 mg/dL), ascorbic acid (up to 100 mg/dL), free bilirubin (up to 50 mg/dL), conjugated bilirubin (up to 50 mg/dl, lipid emulsion (up to 5000 turbidity), Rheumatoid factor (up to 500 IU/mL), and ascorbic acid (up to 100 mg/dL) on serum samples containing a nominal concentration of 3.7 mg/L. No interference was noted on any of the above. Interference was defined as a result +/- 5 % of the control. This information is in the package insert. -
f. Assay cut-off:
Not applicable -
- Comparison studies:
- a. Method comparison with predicate device:
Serum samples (n=98) ranging from 0.0 to 293 mg/L were measured using the Nanopia method and comparing those measurement to those made with the predicate method. Results are summarized below.
Serum Samples: Nanopia = 1.015(Predicate) - 0.0349; R2 = 0.9992
- b. Matrix comparison:
5
Paired samples of serum and EDTA plasma were collected concurrently from 50 individuals and measured using the Nanopia method. Results are summarized below:
Plasma = 0.994(Serum) + 0.03; r = 0.999
Additional serum pools were spiked with 110 mM of sodium citrate, 100 mM sodium oxalate, 25 mM EDTA, and 100 U/mL of sodium heparin and assayed. There were no significant differences observed between sample types.
-
- Clinical studies:
- a. Clinical sensitivity: Not applicable
- b. Clinical specificity: Not applicable
- c. Other clinical supportive data (when a and b are not applicable): Not applicable
-
- Clinical cut-off: Not applicable
-
- Expected values/Reference range: Less than 3.0 mg/L.
L. Conclusion:
Clinical Data concludes that the Nanopia Wide Range CRP assay has a similar intended use, a similar technological principle, and clinically acceptable performance comparable to similar devices currently in commercial distribution and is substantially equivalent to the predicate device.
6
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its wings or feathers. The bird is positioned above a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
2006 FEB 9
2006
Israel M. Stein, MD President Clinical Data, Inc. One Gateway Center Suite 411 Newton, MA 02146
K052591 Re:
Trade/Device Name: Nanopia Wide Range CRP Reagent Regulation Number: 21 CFR 866.5270 Regulation Name: C-reactive protein immunological test system Regulatory Class: Class II Product Code: DCK, JIS Dated: January 12, 2006 Received: January 13, 2006
Dear Dr. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
7
Page 2 --
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally president redicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I ou may of annovaturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Alberto Gutierrez, Ph.D.
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
8
Indications for Use
510(k) Number (if known): K052591
Nanopia Wide Range CRP Reagent Device Name:
Indications For Use: The Nanopia Wide Range CRP Reagent is intended for the quantitative measurement of C-Reactive Protein (CRP) in serum or plasma. The assay is intended for use in the evaluation of infection, tissue injury, and inflammatory disorders in combination with a complete clinical evaluation.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Cinn Chapper
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of 1
FOSZSYI .