K Number
K052308
Manufacturer
Date Cleared
2006-05-10

(259 days)

Product Code
Regulation Number
866.5870
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ARCHITECT® Anti-Tg is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of the IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma on the ARCHITECT i System. The ARCHITECT Anti-Tg assay is intended for use as an aid in the diagnosis of thyroid disease.

The ARCHITECT® Anti-Tg Calibration of the calibration of the ARCHITECT® i System when used for the quantitative determination of IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma.

The ARCHITECT® Anti-Tg Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT® i System (reagents, calibrators and instrument) when used for the quantitative determination of the IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma.

Device Description

The ARCHITECT Anti-Tg assay is a two-step immunoassay for the quantitative determination of the IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex®.

In the first step, sample, assay diluent and Tg coated paramagnetic microparticles are combined and incubated. Anti-Tg present in the sample binds to the Tg coated microparticles. After washing, anti-human IgG acridinium labeled conjugate is added in the second step. Following another incubation and wash, pre-trigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLU). A direct relationship exists between the amount of anti-Tg in the sample and the RLUs detected by the ARCHITECT i system optics.

AI/ML Overview

The ARCHITECT® Anti-Tg assay is a device for the quantitative determination of IgG class thyroglobulin autoantibodies (anti-Tg) in human serum and plasma, intended as an aid in the diagnosis of thyroid disease.

1. Table of Acceptance Criteria and Reported Device Performance:

The provided document does not explicitly state numerical acceptance criteria in a table format. However, it indicates that the ARCHITECT® Anti-Tg assay demonstrated substantial equivalence to the predicate device, the Nichols Advantage Thyroglobulin Autoantibodies Assay, based on various performance characteristics.

Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
PrecisionSubstantially equivalent to predicateSubstantially equivalent to predicate
LinearitySubstantially equivalent to predicateSubstantially equivalent to predicate
InterferencesSubstantially equivalent to predicateSubstantially equivalent to predicate
StabilitySubstantially equivalent to predicateSubstantially equivalent to predicate
Method Concordance (NCCLS Standard EP-12A)Substantially equivalent to predicateSubstantially equivalent to predicate
Sample Stability (Lithium Heparin & Serum Separator tubes)No systematic gain or loss of detectabilityNo systematic gain or loss of detectability

2. Sample size used for the test set and the data provenance:

The document mentions a "summary clinical performance" study and a "method concordance using the NCCLS Standard (EP-12A)". However, the sample size for the test set is not explicitly stated. The data provenance (e.g., country of origin of the data, retrospective or prospective) is also not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. The study focuses on comparing the new device against a predicate device, rather than establishing a de novo ground truth with expert consensus.

4. Adjudication method for the test set:

This information is not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This device is an immunoassay system for laboratory analysis, not an AI-assisted diagnostic imaging or interpretation device that would involve human readers. Therefore, an MRMC comparative effectiveness study with human readers and AI assistance is not applicable and was not done.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The performance described is inherently a "standalone" or "algorithm-only" performance in the context of an automated immunoassay system. The document reports non-clinical performance (precision, linearity, interferences, stability) and clinical performance (method concordance, sample stability) where the device's output is compared to the predicate device. There is no human-in-the-loop interaction for interpretation in the same way as, for example, an AI-powered image analysis tool.

7. The type of ground truth used:

The "ground truth" for the comparison studies was implicitly established by the results of the legally marketed predicate device, the Nichols Advantage Thyroglobulin Autoantibodies Assay. The goal was to demonstrate substantial equivalence to this existing device.

8. The sample size for the training set:

This information is not applicable as this is an immunoassay system, not a machine learning or AI-based device that typically requires a separate training set. The device uses established biochemical principles for measurement.

9. How the ground truth for the training set was established:

This question is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Fisher Diagnostics. The logo consists of a circle with the letter 'F' inside it, followed by the words 'Fisher Diagnostics' in bold, sans-serif font. The logo is simple and professional, and it is likely used to identify the company's products and services.

8.3 510(k) Summary

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

052308

Applicant Name:

M. Heather Cameron Regulatory Compliance Specialist Fisher Diagnostics 8365 Valley Pike P.O. Box 307 Middletown, VA 22645 Phone: 540-869-8163 Fax: 540-869-8129

Establishment Registration Number: 1181121

ldentification of Device:

Device Name: ARCHITECT® Anti-Tg Proprietary/Trade Name: ARCHITECT® Anti-Tg Immunoassay, ARCHITECT® Anti-Tg Calibrators & Controls Common Name: Anti-Tg test system Device Classification: Class II Governing Regulation: 21 CFR 866.5870, 862.1660, 862.1150 FDA Panel: Immunology, Clinical Chemistry Product Code: JZO, JJX, JIT

Identification of Predicate Device:

Nichols Advantage Thyroglobulin Autoantibodies Assay (K983992) Nichols Advantage Chemiluminescence Tri-Level Controls (K972070)

Intended Use of the Device:

ARCHITECT® Anti-Tg is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of the IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma on the ARCHITECT i System. The ARCHITECT Anti-Tg assay is intended for use as an aid in the diagnosis of thyroid disease.

The ARCHITECT® Anti-Tg Calibration of the calibration of the ARCHITECT® i System when used for the quantitative determination of IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma.

The ARCHITECT® Anti-Tg Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT® i System (reagents, calibrators and instrument) when used for the quantitative determination of the IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma.

Description of the Device:

The ARCHITECT Anti-Tg assay is a two-step immunoassay for the quantitative determination of the IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex®

In the first step, sample, assay diluent and Tg coated paramagnetic microparticles are combined and incubated. Anti-Tg present in the sample binds to the Tg coated microparticles. After washing, anti-

Confidential

Page 68 of 95

MAY 10 1006

{1}------------------------------------------------

ARCHITECT® Anti-Tq Assay Auqust 23, 2005

Image /page/1/Picture/1 description: The image shows the logo for Fisher Diagnostics. The logo consists of a circle with the letter F inside, followed by the words "Fisher Diagnostics" in bold, sans-serif font. The logo is simple and professional, and it is likely used to identify the company's products and services.

human IqG acridinium labeled conjugate is added in the second step. Following another incubation and wash, pre-trigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLU). A direct relationship exists between the amount of anti-Tg in the sample and the RLUs detected by the ARCHITECT i system optics.

8.4 Comparison of Technological Characteristics:

The ARCHITECT® Anti-Tg and the Nichols Advantage Thyroglobulin Autoantibodies assays use a microparticle immunoassay method for the quantitative determination of Anti-Tg in human serum or plasma. Anti-microbial agent is used as a preservative for all reagent components (microparticles and conjugate) of the Nichols Advantage Thyroglobulin Autoantibodies assay as well as the ARCHITECT Anti-Tg. Both assays have magnetic microparticles as the solid support. The ARCHITECT® Anti-Tg and the Nichols Advantage Thyroglobulin Autoantibodies assays both use acridinium labeled conjugates to generate the chemiluminescent in the assay.

8.5 Summary of Non-Clinical Performance:

The ARCHITECT® Anti-Tg assay is substantially equivalent to the Nichols Advantage Thyroglobulin Autoantibodies assay in terms of precision, linearity, interferences, and stability as demonstrated in nonclinical performance data in this 510(k) submission.

8.6 Summary Clinical Performance

The ARCHITECT Anti-Tg assay demonstrated substantially equivalent to the Nichols Advantage Thyroglobulin Autoantibodies Assay. The sample stability study evaluated ARCHITECT Anti-Ta assay using Lithium Heparin and Serum Separator collection tubes. There was no systematic gain or loss of the delectability of Anti-Tg in serum or plasma samples under any of the storage conditions evaluated in this study. A method concordance using the NCCLS Standard (EP-12A) was also conducted with the ARCHITECT Anti-Tg and the Nichols Advantage Thyroglobulin Autoantibodies assays, and as a result. the two systems demonstrated substantial equivalence as indicated by clinical data in this 510(k) submission.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAY 10 2006

Fisher Diagnostics c/o Ms. M. Heather Cameron Regulatory Compliance Specialist 8365 Valley Pike P.O. Box 307 Middletown, VA 22645

Re: K052308

Trade/Device Name: ARCHITECT® Anti-Tg, ARCHITECT® Anti-Tg Calibrators and ARCHITECT® Anti-Tg Controls Regulation Number: 21 CFR 866.5870 Regulation Name: Thyroid autoantibody immunological test system Regulatory Class: Class II Product Code: JZO, JIT, JJX Dated: August 23, 2005 Received: August 24, 2005

Dear Ms. Cameron:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 --

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Robert A. Beckerf

Robert L. Becker, Jr., MD, PhD - Director

Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Fisher Diagnostics. The logo consists of a circle with the letter "F" inside it, followed by the words "Fisher Diagnostics" in bold, sans-serif font. The logo is simple and professional, and it is likely used to identify the company's products and services.

APPENDICES

Indications for Use 8.1

Reagent Intended Use 8.1.1

ARCHITECT® Anti-Tg is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of the IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma on the ARCHITECT i System. The ARCHITECT Anti-Tg assay is intended for use as an aid in the diagnosis of thyroid disease.

8.1.2 Calibrators Intended Use

The ARCHITECT® Anti-Tq Calibrators are for the calibration of the ARCHITECT® i System when used for the quantitative determination of IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma.

8.1.3 Controls Intended Use

The ARCHITECT® Anti-Tg Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT® i System (reagents, calibrators and instrument) when used for the quantitative determination of the IgG class of thyroglobulin autoantibodies (anti-Tg) in human serum and plasma.

510 (k) Number (if known):

Device Name: ARCHITECT® Anti-Tg Immunoassay, ARCHITECT® Anti-Tg Calibrators & Controls

052308

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mari Cham


Division Sign-Off

Office of In Vitro Diac

510(k) ka523 of

Confidential

Fisher Diagnostics

§ 866.5870 Thyroid autoantibody immunological test system.

(a)
Identification. A thyroid autoantibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the thyroid autoantibodies (antibodies produced against the body's own tissues). Measurement of thyroid autoantibodies may aid in the diagnosis of certain thyroid disorders, such as Hashimoto's disease (chronic lymphocytic thyroiditis), nontoxic goiter (enlargement of thyroid gland), Grave's disease (enlargement of the thyroid gland with protrusion of the eyeballs), and cancer of the thyroid.(b)
Classification. Class II (performance standards).