K Number
K051452
Manufacturer
Date Cleared
2005-10-25

(145 days)

Product Code
Regulation Number
862.1150
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Clinical Chemistry – The Sentinel Clin Chem Cal is a device intended for medical purposes for use in pancreatic amylase and cholinesterase assays to establish points of reference that are used in the determination of values in the measurement of pancreatic amylase and cholinesterase in human serum and plasma.

Device Description

The Sentinel Clin Chem Cal is a device intended for medical purposes for use in pancreatic amylase and cholinesterase assays to establish points of reference that are used in the determination of values in the measurement of pancreatic amylase and cholinesterase in human serum and plasma. Sentinel Clin Chem Cal contains the analytes in human serum matrix. The analytes consist of pancreatic amylase and cholinesterase.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the device (Sentinel Clin Chem Cal) meets the acceptance criteria for Calibrator Shelf-life Stability and Calibrator After Reconstitution Stability.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and the Reported Device Performance

ParameterAcceptance CriteriaReported Device Performance
Calibrator Shelf-life Stability95% to 105% recovery compared to Time zeroCalibrators showed stability for up to 33 months (See Table 1). For Pancreatic Amylase, % Recovery ranged from 97.2% to 103.6%. For Cholinesterase, % Recovery ranged from 97.9% to 101.8%. The claimed shelf-life is 24 months from the date of manufacture.
Calibrator After Reconstitution Stability (2-8°C)100% ± 7% recovery (i.e., 93-107%) between daysPancreatic Amylase (Table 2): After 2 days at 2-8°C, %R (2d vs To) for the three lots tested were 103%, 102%, and 101%, respectively. All within the 93-107% range. Cholinesterase (Table 3): After 2 days at 2-8°C, %R (2d vs To) for the three lots tested were 100%, 100%, and 100%, respectively. All within the 93-107% range.
Calibrator After Reconstitution Stability (-20°C)100% ± 7% recovery (i.e., 93-107%) between daysPancreatic Amylase (Table 2): After 14 days at -20°C, %R (30d vs To) for the three lots tested were 104%, 103%, and 102%, respectively. All within the 93-107% range. Cholinesterase (Table 3): After 14 days at -20°C, %R (30d vs To) for the three lots tested were 100%, 100%, and 100%, respectively. All within the 93-107% range.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Calibrator Shelf-life Stability: One lot of Sentinel Clin Chem Cal was used for the test. The stability was monitored prospectively by testing at multiple time points through a minimum of 30 months (and reported up to 33 months). The origin of the data is not explicitly stated but is implicitly from the manufacturer, Sentinel CH. Srl in Milan, Italy.
  • Calibrator After Reconstitution Stability: Three lots of calibrator were used for the test (one expired from 13 months, one expired from 8 months, and one still in validity). The study was prospective in nature, assessing stability at specific time points after reconstitution. The origin of the data is implicitly from the manufacturer, Sentinel CH. Srl in Milan, Italy.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This device is a calibrator for in-vitro diagnostic assays, not an AI or imaging device requiring expert ground truth for interpretation. The ground truth for this device's performance is based on analytical measurements against defined reference standards and the inherent property of a calibrator to maintain its assigned value over time and under specified conditions.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not an imaging or diagnostic device requiring adjudication of human-interpreted results.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is an IVD calibrator, not an algorithm. The "standalone" performance is equivalent to the reported analytical performance of the calibrator itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the calibrator's assigned values and stability assessments is based on:

  • Traceability to Reference Standards: The Pancreatic Amylase is traceable to IFCC EPS / 37 °C using $\varepsilon$ p-Nitrophenol. Cholinesterase is traceable to DGKC Butyrylthiocholine 37 °C using $\varepsilon$ Hexacyano-ferrate (III).
  • Methodology of Value Assignment: Assigned values are determined by taking the mean of measurements from multiple replicates across different instruments (Hitachi 717 and Abbott ARCHITECT® c8000), calibrated against a previous calibrator lot. This establishes the initial "true" or assigned value.
  • Recovery Method: For stability studies, the performance is measured as a percentage recovery compared to the "time zero" measurement, which serves as the reference or ground truth for that specific lot.

8. The sample size for the training set

Not applicable. This is not a machine learning or AI device that typically requires a 'training set'.

9. How the ground truth for the training set was established

Not applicable. There is no training set for this type of device.

{0}------------------------------------------------

001 25 2005

510(k) Summary

Submitter's Name/Address Sentinel CH. Srl Via Principe Eugenio 5 20155 Milan - ITALY

Contact Person

Mr. Davide Spada Application Specialist Telephone: +39 02 345514.1 Facsimile: +39 02 345514.64 E-Mail: spada@sentinel.it

Date of Preparation of this Summary:May, 28 2005
Device Trade or Proprietary Name:Sentinel Clin Chem Cal
Device Common/Usual Name or Classification Name:Calibrator
Classification Number/Class:JIX/Class II

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: _KOS 1452

Test Description:

The Sentinel Sentinel Clin Chem Cal is a device intended for medical purposes for use in pancreatic amylase and cholinesterase assays to establish points of reference that are used in the determination of values in the measurement of pancreatic amylase and cholinesterase in human serum and plasma.

Intended Use:

Sentinel Clin Chem Cal must be used only for the calibration of clinical chemistry tests.

{1}------------------------------------------------

Description of the Calibrator Material:

Sentinel Clin Chem Cal contains the analytes in human serum matrix. The analytes consist of pancreatic amylase and cholinesterase.

Assigned Values and Value Assignment Process:

Four vials of calibrator are reconstituted by weight (acceptability 2.970-3.030 g) following the Instruction for Use.

To evaluate vial variability, Pancreatic Amylase is assayed in three replicates on each vial and %CV is calculated for each vial. If %CV is equal to or greater than 4.5%, the vial is discarded from the analysis. At least 2 vials of calibrator can be used. If more than 3 vials are discarded, another set of four calibrator must be used.

The pool of all accepted vials is aliquoted in small volumes and stored at -20 ℃.

The Pancreatic Amylase and Cholinesterase assay are calibrated against the previous calibrator lot, stored at 2-8 ℃ and freshly reconstituted. During each testing run, two levels (normal and abnormal) of control materials are assayed to ensure the effectiveness of the measurements.

For three consecutive days (three runs), three replicates of Pancreatic Amylase and Cholinesterase are assayed on Hitachi 717. The average of all replicates is calculated (X1).

Pancreatic Amylase and Cholinesterase are also assayed on Abbott ARCHITECT® c8000 . Five vials of freshly reconstituted by weight vials are assayed in five replicates on c8000. The average of all replicates is calculated (X2).

The assigned value (Xm) is the mean of the two average: Xm=(X1+X2)/2

Directions for Use:

Refer to Draft Calibrator Labeling on page 17

{2}------------------------------------------------

Performance Characteristics:

    1. Precision/Reproducibility
      N/A

2. Linearity/assay reportable range

N/A

3. Traceability (controls, calibrators, or method)

The Sentinel Clin Chem Cal is traceable to the following referenced standards:

AnalyteMethodStandardization
AmyPPancreaticAmylaseIFCC EPS / 37 °C$\varepsilon$ p-Nitrophenol
ChECholinesteraseDGKC Butyrylthiocholine37 °C$\varepsilon$ Hexacyano-ferrate (III)

4. Detection limit (functional sensitivity)

N/A

5. Analytical specificity

N/A

6. Assay cut-off

N/A

7. Calibrator Shelf-life Stability

The calibrator shelf-life stability was determined by the recovery method on one lots of Sentinel Clin Chem Cal. One lot of test calibrator were stored at 4℃. At the time of analysis was reconstituted with distilled water. Percent recovery was calculated for each calibrator level by dividing the result of the test calibrator at

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows two handwritten phrases. The phrase on the left appears to say "Hand made". The phrase on the right is more difficult to read, but it seems to say "with love".

time zero by the result of control calibrators at every testing point. At each testing point, fresh reagent, control calibrators, and test calibrators were analyzed. Calibrators were tested at multiple test points through a minimum of 30 months. Target recovery was 95% to 105% for the test calibrators as compared to the Time zero. Results are found in Table 1.

The results show the calibrators are stable for up to 33 months. The resulting claim is 24 months from date of manufacture.

Since the calibrator is purchased from Roche which is the Predicate Device for ビ the Sentinel Clin Chem Cal, Shelf life of the calibrator is the same as the Predicate Device.

8. Calibrator After Reconstitution Stability

After Reconstitution stability was assessed on three lots of calibrator: one expired from 13 months (expiry date + 13 months), one expired from 8 months (expiry date + 8 months) and one still in validity.

The calibrator after reconstitution stability was determined by the recovery method. Percent recovery was calculated for each calibrator analyte by dividing the results of the reconstituted vial stored at 2 -- 8 ℃ after 2 days, and at -20 ℃ after 14 days by the results of a freshly reconstituted vial of the same lot. At each test point fresh reagents were used.

Acceptance criteria is 100 ± 7%.

Results are found in Table 2 and Table 3.

Stability of reconstituted material: 2 days at 2 to 8 °C or 14 days at -20 °C if aliquoted in small volumes.

{4}------------------------------------------------

Table 1 Sentinel Sentinel Clin Chem Cal Real Time Stability - Pancreatic Amylase and Cholinesterase

TimeCholinesterase(U/L)P - Amylase(U/L)
0 Monhts4581148.8
1 Mo4618150.6
% Rec100.8101.2
4 Mo4627151.3
% Rec101.0101.7
8 Mo4590152.4
% Rec100.2102.4
11 Mo4636150.6
% Rec101.2101.2
14 Mo4503144.6
% Rec98.397.2
17 Mo4595151.0
% Rec100.3101.5
20 Mo4485151.5
% Rec97.9101.8
23 Mo4508147.2
% Rec98.498.9
26 Mo4576150.0
% Rec99.9100.8
29 Mo4663154.2
% Rec101.8103.6
33 Mo4636148.7
% Rec101.299.9

{5}------------------------------------------------

Table 2

Sentinel Sentinel Clin Chem Cal After Reconstitution Stability - Pancreatic Amylase

product:Clin chem Calref 16550at 2-8°C and - 20°CSENTINEL
Used Material
Instrument.HITACHI 71717631lot31147exp : mar/2005
reagent lotPancreatic amylaseexp : jun/2006
Clin Chem CalREF 16550lotP0690
Clin Chem CalREF 16550lot3 differents lots
Calibration/Check system:Acceptability criteria1) The D% (factor and controls) between days must be less 5%2) all replicates of controls must be in the target range
Stability after recostitution Calibrator:1) The % Recovery (%R) between days must range 93-107%
Calibration /Check system (reagents+ instruments)time 02 days(2-8°C)14 days(-20°C)
samplesupplierlotExpired datetime ofvalutationexpectedvaluereplicatesvaluevalue%R(2d vs To)valueD%(30d vs To)
blankDabs0.00050.00020.0000
Dabs0.04730.04490.0455
calibratorSentinelP0690jun / 20062 months aftermanufacturing151factor322633784.7%33192.9%
Precinom URocheapr/2004still valid12 monthsaftermanufacturing39 ±71424043
2393937
3374138
Precipath Urochefeb/2004still valid14 monthsaftermanufacturing102 ± 181108103107
2106107102
3105105109
meancontrolPrecinom UPrecipath URocherocheapr/2004feb/200439.340.01.7%39.30.0%
106.3105.0-1.3%106.0-0.3%
The factor and controls meet the criteria of acceptabilityP YES* NO
3 lots calibrator stability after recostitutiontime 02 days(2-8°C)14 days(-20°C)
samplesupplierlotExpired datetime ofvalutationexpectedvaluereplicatesvaluevalue%R(2d vs To)value%R(30d vs To)
M0750sep/200213 monthsexpired183target NA1191197204
2184201198
3199194194
Clin Chem CalSentinelM0796jan/20038 monthsexpired172target NA1180185183
2173187188
3186178185
P0690jun/2006still valid151target NA1153155154
2150155157
3153150153
MeancalibratorClin Chem CalSentinelM0750M0796P0690191.3197.3103%198.7104%
179.7183.3102%185.3103%
152.0153.3101%154.7102%
ConclusionThe calibrator meet the acceptability criteria and the stability after recostitution is :2 days at 2-8°C or 14 days at -20°C
QC (date Signature)02/10/2003
Ezio MarelliTechnical ManagerSENTINEL CH.
SENTINELSENTINEL For - Via Pens no Cugemo 620135 Milan - 400V
1 . A14. 4.89 0.2 5455 141 Frex 139 0.2 (455 1464 - veww.tem*relle sentinellif

Sentinel Clin Chem Cal 510(k)
May 28, 2005

.

{6}------------------------------------------------

Table 3 Sentinel Sentinel Clin Chem Cal After Reconstitution Stability – Cholinesterase

STABILITY AFTER RECOSTITUTIONSENTINEL
product:Clin chem Calref 16550at 2-8°C and - 20°C
Instrument.HITACHI 717
reagent lotCholinesterase liquidREF 17019Alot30863exp : Oct/2004
Clin Chem CalREF 16550lotP0690exp : jul/2006
Clin Chem CalREF 16550lot3 differents lots
Calibration/Check system:Acceptability criteria1) The D% (factor and controls) between days must be less 5%2) all replicates of controls must be in the target range
Stability after recostitution Calibrator:1) The % Recovery (%R) between days must range 93-107%
Calibration /Check system (reagents+ instruments)time 02 days(2-8°C)14 days(-20°C)
samplesupplierlotExpired datetime ofvalutationexpectedvaluereplicatesvaluevalueD%(2d vs To)valueD%(30d vs To)
blankDabs-0.0036-0.0029-0.0031
calibratorSentinelP0690jun / 20062 months aftermanufacturing5920Dabsfactor0.0399-87774-0.0396-86817-1.1%-0.0407-84724-3.5%
Precinorm URocheapr/2004still valid12 monthsaftermanufactoring6700 ± 1100123672166416734664167016851663867036890
Precipath Urochefeb/2004still valid14 monthsaftermanufacturing6300 ± 1000123627862736185628663406204624562696265
meancontrolPrecinorm URocheapr/20046698.76731.00.5%6743.70.7%
Precipath Urochefeb/20046245.36276.70.5%6259.70.2%
The factor and controls meet the criteria of acceptabilityP YES* NO
3 lots calibrator stability after recostitutiontime 02 days(2-8°C)14 days(-20°C)
samplesupplierlotExpired datetime ofvalutationexpectedvaluereplicatevaluevalueD%(2d vs To)valueD%(30d vs To)
M0750sep/200213 monthsexpired4850target NA123467747854891467747854891486847544661
Clin Chem CalSentinelM0796jan/20038 monthsexpired5010target NA123500150145104500150145104510150074980
P0690jun/2006still valid5920target NA123582158115945582158115945590458015901
MeancalibratorClin Chem CalSentinelM0750M0796P06904784.34784.3100%4761.0100%
5039.75039.7100%5029.3100%
5859.05859.0100%5868.7100%
ConclusionThe calibrator meet the acceptability criteria and the stability after recostitution is :2 days at 2-8°C or 14 days at -20°C
QC (date Signature)02/10/2003Ezio MarelliTechnical ManagerSENTINEL CH.
SENTINELDACNUSENTINEL CHVia Principe Eugenio, 5 20155 Milano, ItalyPh. +39(02) 4655141 Fax +39(02) 46551454 www.sentinel.it sentinel@sentinel.it

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.

OCT 2 5 2005

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Davide Spada Application Specialist Sentinel CH S.r.l. Via Principe Eugenio, 5 20155 Milan-Italy

Re: K051452

Trade/Device Name: Sentinel Clinical Chemistry Calibrator Regulation Number: 21 CFR 862.1150 Regulation Name: Multi-Analyte Calibrator Regulatory Class: Class II Product Codc: JIX Dated: August 5, 2005 Received: August 19, 2005

Dear Mr. Spada:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.

If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{8}------------------------------------------------

Page 2 --

This letter will allow you to begin marketing your device as described in your Section 510(k) I ins letter whilanon you to ough indias of substantial equivalence of your device to a legally premation notification: "The a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or guestions on the promotion and advertising of your device, please contact the Office of In of quostions on the promotion and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I ou may oount other generers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincercly yours,

Alberto Gutierrez Ph.D.

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

510(k) Number (if known): ي 05 1452

Sentinel Clin Chem Cal Device Name:

Indications For Use:

Clinical Chemistry – The Sentinel Clin Chem Cal is a device intended for medical purposes for use in pancreatic amylase and cholinesterase assays to establish points of reference that are used in the determination of values in the measurement of pancreatic amylase and cholinesterase in human serum and plasma.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Lucy Phillips
Division Sign-Off

Page 1 of 1

Office of In Vitre Diagnostic Device Evaluation and Sulety

051450

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.