(28 days)
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring Catheter is indicated for delivery of contrast medium in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressure measurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring catheter is intended for use as a pressure measurement catheter and for delivery of contrast media during angiographic studies. The DLP catheter consists of the Merit Medical Softouch Diagnostic Intravascular Catheter (K943739) as the inner lumen and an extruded outer lumen designed by Vascular Solutions. The inner lumen and hub assembly are purchased from Merit Medical in a non-sterile state. The Merit Softouch catheter and outer lumen are joined together using an adapter junction placed near the proximal end of the Merit Softouch catheter. The adapter junction incorporates a side port fitted with a stopcock/tube assembly for fluid flow and pressure measurement within the outer tube. The distal end of the outer tube is perforated with side holes to allow pressure measurement simultaneously with the sideholes at the tip of the catheter. The dual lumen catheter is deployed through standard guide catheters and will accommodate standard 0.038'' diameter guidewires.
The provided text describes a 510(k) submission for the Langston™ Dual Lumen Pressure Monitoring Catheter. This device is an Intravascular Diagnostic Catheter and is an unclassified device. The submission focuses on demonstrating substantial equivalence to predicate devices rather than providing a study that assesses its performance against specific acceptance criteria in a clinical setting.
Therefore, much of the requested information cannot be extracted directly from this document because it outlines a regulatory submission process for a medical device rather than for an AI/machine learning device that would undergo rigorous performance evaluation against established reference standards.
Here's a breakdown of what can and cannot be answered based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
No specific "acceptance criteria" for device performance (e.g., accuracy, sensitivity, specificity) and corresponding reported performance are provided in the document. The submission focuses on demonstrating substantial equivalence to predicate devices based on design verification and biocompatibility.
2. Sample Size Used for the Test Set and Data Provenance:
No test set for evaluating the device's performance (in the context of AI/ML) is mentioned. The document refers to "design verification" non-clinical testing.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
Not applicable. The document does not describe a study involving expert-established ground truth for device performance evaluation.
4. Adjudication Method:
Not applicable.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No MRMC study was conducted or reported.
6. Standalone Performance:
No standalone (algorithm-only) performance is reported, as this is a physical medical device, not an AI algorithm.
7. Type of Ground Truth Used:
Not applicable in the context of an AI/ML device. For a physical medical device, "ground truth" would refer to its physical and functional characteristics verified through engineering and biocompatibility testing. The text states: "Testing conducted included assessments of the design verification of the Dual Lumen Pressure Monitoring Catheter along with biocompatibility assessments."
8. Sample Size for the Training Set:
Not applicable; no AI/ML model was trained.
9. How the Ground Truth for the Training Set Was Established:
Not applicable; no AI/ML model was trained.
Summary of Information from the Provided Text:
The document describes a 510(k) premarket notification for a medical device, not an AI/ML product. The regulatory pathway for this device relies on demonstrating substantial equivalence to existing predicate devices, primarily through non-clinical testing (design verification and biocompatibility) rather than clinical performance studies against specific acceptance criteria.
Key points from the document regarding the device's qualification:
- Non-Clinical Testing: "Testing conducted included assessments of the design verification of the Dual Lumen Pressure Monitoring Catheter along with biocompatibility assessments. The results of this battery of tests confirmed the suitability of the Dual Lumen Pressure Monitoring Catheter for its intended use."
- Clinical Testing: "No clinical evaluations of this product have been conducted." (This explicitly states the absence of human clinical trials).
- Predicate Devices: The device is deemed substantially equivalent to:
- Langston Dual Lumen Pigtail Catheters
- Cordis Dual Lumen Pressure Monitoring Catheter
- Merit Medical Softouch Diagnostic Intravascular Catheter
- Conclusion: "The testing performed confirms that the Langston Dual Lumen Pressure Monitoring Catheter will perform as intended." This statement refers to the non-clinical design verification and biocompatibility testing.
{0}------------------------------------------------
JUN 2 4 2005
510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
| Common/Usual Name: | Intravascular Diagnostic Catheter |
|---|---|
| Product Trade Name: | LangstonTM Dual Lumen Pressure Monitoring Catheter |
| Classification Name: | UnclassifiedProduct Code: DQO |
| Manufacturer: | Vascular Solutions, Inc.6464 Sycamore CourtMinneapolis, Minnesota 55369USA |
| Establishment Registration: | 2134812 |
| Contact: | Sara L. CoonSenior Regulatory Affairs Associate(763) 656-4300 phone(763) 656-4200 fax |
| Performance Standards: | No performance standards have been developed undersection 514 for this device. |
Device Description:
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring catheter is intended for use as a pressure measurement catheter and for delivery of contrast media during angiographic studies. The DLP catheter consists of the Merit Medical Softouch Diagnostic Intravascular Catheter (K943739) as the inner lumen and an extruded outer lumen designed by Vascular Solutions. The inner lumen and hub assembly are purchased from Merit Medical in a non-sterile state. The Merit Softouch catheter and outer lumen are joined together using an adapter junction placed near the proximal end of the Merit Softouch catheter. The adapter junction incorporates a side port fitted with a stopcock/tube assembly for fluid flow and pressure measurement within the outer tube. The distal end of the outer tube is perforated with side holes to allow pressure measurement simultaneously with the sideholes at the tip of the catheter. The dual lumen catheter is deployed through standard guide catheters and will accommodate standard 0.038'' diameter guidewires.
Intended Use:
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring Catheter is indicated for delivery of contrast medium in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.
{1}------------------------------------------------
Summary of Non-Clinical Testing:
Testing conducted included assessments of the design verification of the Dual Lumen Pressure Monitoring Catheter along with biocompatibility assessments. The results of this battery of tests confirmed the suitability of the Dual Lumen Pressure Monitoring Catheter for its intended use.
Summary of Clinical Testing:
No clinical evaluations of this product have been conducted.
Predicate Device
The Dual Lumen Pressure Monitoring catheter is similar in intended use and function to the Langston Dual Lumen Pigtail Catheters, the Cordis Dual Lumen Pressure Monitoring Catheter, and the Merit Medical Softouch Diagnostic Intravascular Catheter.
Conclusions:
The Langston Dual Lumen Pressure Monitoring Catheter is substantially equivalent to the Langston Dual Lumen Pigtail Catheters, the Cordis Dual Lumen Pressure Monitoring Catheter and the Merit Medical Softouch Diagnostic Intravascular Catheter. The testing performed confirms that the Langston Dual Lumen Pressure Monitoring Catheter will perform as intended.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem consisting of three curved shapes that resemble a bird in flight or flowing ribbons. The emblem is positioned in the center of the seal.
JUN 2 4 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Vascular Solutions, Inc. c/o Ms. Sara L. Coon Sr. Regulatory Affairs Associate 6464 Sycamore Court Minneapolis, MN 55369
Re: K051395
Langston™ Dual Lumen Pressure Monitoring Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II (two) Product Code: DQO Dated: May 26, 2005 Received: May 27, 2005
Dear Ms. Coon:
We have reviewed your Section 510(k) premarket notification of intent to market the indicati We have reviewed your Section - 10(tr) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosure) to regarly manated producal Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food Track commerce prior to May 28, 1970, the chacinent with the provisions of the Federal Food. Drug, devices that have been reclassined in accerdance was a premarket approval application (PMA).).
and Cosmetic Act (Act) that do not require approval of a presidions of the Act and Cosment Act (Act) that do not require subject to the general controls provisions of the Act. The You may, therefore, market the do rest, soloject to the month registration, listing of
general controls provisions of the Act include requirements for annual registration, l general controls provisions of the free labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is Classified (secure) into existing major regulations affecting your device can
may be subject to such additional controls. Existing major may be subject to such additional controller Entralians, Title 21, Parts 800 to 898. In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to be found in the Code of Poderal Registerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Sara L. Coon
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please of advised that IDA s issualled of a succession with other requirements of the Act that IDA has made a dolorimisations administered by other Federal agencies. You must or any Federal statutes and regulations adminities out not limited to: registration and listing (21 comply with an the Act 31equirements, newslager, which and the requirements as set CFR Part 807); labeling (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Declions of a described in your Section 510(k)
This letter will allow you to begin marketing your device as described in your section 51 I his letter will anow you to begin maneting your and equivalence of your device to a legally premarket notification. The PDA midning of backers and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac not the case note the regulation entitled, and Contact the Office of Comphalled at (217) = if & = = = = 807.97). You may obtain "Misbranding by reference to premation inculties. Included the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Consumer Provinces of contraction of the many of the status of the status
Sincerely yours,
Bhimima foi
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
510(k) Number:
Device Name:
Langston™ Dual Lumen Pressure Monitoring Catheter
Indications for Use:
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring Catheter is indicated for delivery of contrast medium in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressure measurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blimmerman
(Division Sign-Off)
n of Cardlovascular Devices
Page 1 of 1
§ 870.1200 Diagnostic intravascular catheter.
(a)
Identification. An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.(b)
Classification. Class II (performance standards).