(7 days)
Not Found
No
The summary describes a physical filter device for removing emboli and debris from blood during cardiopulmonary bypass. There is no mention of software, algorithms, or any computational processing that would suggest the use of AI or ML. The modifications are related to material coating for improved biocompatibility.
Yes
The device is used to remove harmful substances (gaseous emboli, blood constituents, and particulate debris) from the blood during cardiopulmonary bypass to prevent complications, which is a therapeutic function.
No
The device is a filter used to trap and remove emboli and particulate debris from the arterial line during cardiopulmonary bypass. It does not perform any diagnostic function.
No
The device description clearly indicates it is a physical filter with a screen and coating, designed to be used in an extracorporeal circuit. This is a hardware device, not software.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the device is used in the arterial line of an extracorporeal circuit during cardiopulmonary bypass to trap and remove gaseous emboli and particulate debris. This is a direct intervention on the blood flowing through the circuit, not a test performed in vitro (outside the body) on a sample of blood or other biological material to diagnose a condition.
- Device Description: The description reinforces its function as a filter within the extracorporeal circuit.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples, detecting analytes, or providing diagnostic information. The device's purpose is mechanical filtration.
Therefore, the Dideco D731 MICRO 20 Ph.I.S.I.O. and Dideco D733 MICRO 40 Ph.I.S.I.O. are medical devices used in a therapeutic/supportive capacity during surgery, not IVDs.
N/A
Intended Use / Indications for Use
The Dideco D731 MICRO 20 Ph.I.S.I.O. with 20 micron screen with phosphoryIcholine coating and the Dideco D733 MICRO 40 Ph.I.S.I.O. with 40 micron screen with phosphoryIcholine coating are recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filters are used to trap and remove gaseous emboli as well as particulate debris that maybe introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.
Product codes
DTM
Device Description
The D731 MICRO 20 Ph.I.S.I.O. and D733 MICRO 40 Ph.I.S.I.O. are sterile, non-pyrogenic disposable filters for use in the arterial line of the cardiopulmonary bypass circuit with the flow rate not exceeding 5.0 liters/minute. The D731 MICRO 20 Ph.I.S.I.O. and D733 MICRO 40 Ph.I.S.I.O. are Pediatric Arterial Filters with 20 and 40 micron filter screens designed to remove potentially harmful gaseous emboli, aggregated blood constituents, and particulate debris greater than 20 and 40 microns respectively from the arterial line perfusate. The D731 MICRO 20 Ph.J.S.I.O. and D733 MICRO 40 Ph.I.S.I.O. are a modified version of the currently marketed D731/D733 MICRO. The modification consists of coating all blood contact surfaces with phosphoryIcholine additive that improves the blood compatibility of the substrate materials. Other than this change the D 731/D733 MICRÓ Ph.I.S.I.O. and the D 731/D733 MICRO are identical in design, materials, and manufacturing processes.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Pediatric
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
NON CLINICAL TEST RESULTS:
A complete battery of tests were carried out in accordance with the requirements of ISO 10993-1:1995 and the FDA May 1, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing on the raw materials. Testing was performed on the D 733 MICRO Ph.I.S.I.O. (accelerated aging). The devices were aged up to three years and tested for Hemolysis, Cytotoxicity, Irritation, Acre Systemic Toxicity and Mutagenicity, Sterility, Pyrogenicity and ETO residuals. Package integrity testing was also conducted. The results of the testing met established specifications. As no new materials are used in the D731 MICRO 20 Ph.I.S.I.O. pediatric arterial filter with respect to the D733 MICRO 40 Ph.I.S.I.O data collected are considered applicable to both MICRO Ph.I.S.I.O filters.
IN VITRO TEST RESULTS:
In vitro testing was carried out in accordance with the relevant requirements of in viro couling was carterial Line Blood Filter Blood Filter 510(K) Submission" Final Guidance for Industry, dated November 29, 2000. These data demonstrate substantial equivalence with the unmodified devices and show that the devices are compliant with safety and effectiveness requirements. The device was aged up to 3 years and tested for structural integrity, mechanical integrity, was aged ap te syre drop, filter flow rate capacity, in vitro hemolysis/cell depletion. For comparative purposes all tests were performed on sterilized aged devices comparing the D731 MICRO 20 Ph.I.S.I.O. vs. the D733 non aged unmodified device and D731 MICRO 40 Ph.I.S.I.O vs. D733 non aged unmodified device operated at 5.0 LPM when applicable. The results of these tests met established specifications. The modifications being made to the MICRO Ph.I.O. arterial filter do not affect the performance of the device; therefore the filtration efficiency and air handling characteristics exhibited by the D731/D733 MICRO unmodified device apply also to the MICRO Ph.I.S.I.O. arterial filters.
The results of the study showed that the device characteristics of the D731 MICRO 20 Ph.I.S.I.O. vs. D731 unmodified device and D733 MICRO 40 Ph.I.S.I.O. vs. D733 unmodified device were comparable.
CONCLUSIONS:
The results of in vitro studies demonstrate that the D731 MICRO 20 Ph.I.S.I.O. and D733 MICRO Ph.I.S.I.O. devices perform in a manner substantially equivalent to the unmodified devices. Biocompatibility and functional tests demonstrate that their performance is equivalent to the D731 and D733 unmodified devices, according to their intended use. Additional testing has demonstrated the effectiveness of production techniques assuring that the newborn-infant arterial filters are sterile and non-pyrogenic.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.4260 Cardiopulmonary bypass arterial line blood filter.
(a)
Identification. A cardiopulmonary bypass arterial line blood filter is a device used as part of a gas exchange (oxygenator) system to filter nonbiologic particles and emboli (blood clots or pieces of foreign material flowing in the bloodstream which will obstruct circulation by blocking a vessel) out of the blood. It is used in the arterial return line.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions.”
0
.
Special 510(k): Device Modification May 12, 2005
APPENDIX I
MAY 2 0 2005
510(K) SUMMARY
1
510(k) SUMMARY Dideco S.r.l. SUBMITTER: 86, Via Statale 12 Nord 41037 Mirandola (MO) Italy Luigi Vecchi CONTACT PERSON: Phone: 011 39 0535 29811 Fax: 011 39 0535 25229 May 12, 2005 DATE PREPARED: D731 MICRO 20 Ph.I.S.I.O.: Dideco DEVICE TRADE NAME: D731 Micro 20 Pediatric Arterial Filter screen with micron with 20 phosphoryIcholine coating (hereafter referred to as D731 MICRO 20 Ph.I.S.I.O.) and D733 MICRO 40 Ph.I.S.I.O.: Dideco D733 MICRO 40 Pediatric Arterial Filter micron screen with 40 with coating coating coating of (hereafter phopshorilcoline D733 MICRO 40 referred to as Ph.I.S.I.O) Arterial Filter COMMON NAME: Cardiopulmonary Bypass Arterial Line CLASSIFICATION NAME: Blood Filter D731 MICRO 20: Dideco D731 Micro 20 UNMODIFIED DEVICES Pediatric Arterial Filter with 20 micron screen and D733 MICRO 40: Dideco D733 MICRO 40 Pediatric Arterial Filter with 40 micron screen (hereafter referred to as D731 and D733 respectively) (K945198 and K041061). D736 MICRO 40 Ph.I.S.I.O.: Dideco PREDICATE DEVICES: Ph.I.S.I.O. D736 Micro 40 -Newborn/Infant Arterial Filter with 40 biocompatible screen with micron treatment surface (K002493) (hereafter referred to as D736 MICRO Ph.I.S.I.O.
2
DEVICE DESCRIPTION:
The D731 MICRO 20 Ph.I.S.I.O. and D733 MICRO 40 Ph.I.S.I.O. are sterile, non-pyrogenic disposable filters for use in the arterial line of the cardiopulmonary bypass circuit with the flow rate not exceeding 5.0 liters/minute. The D731 MICRO 20 Ph.I.S.I.O. and D733 MICRO 40 Ph.I.S.I.O. are Pediatric Arterial Filters with 20 and 40 micron filter screens designed to remove potentially harmful gaseous emboli, aggregated blood constituents, and particulate debris greater than 20 and 40 microns respectively from the arterial line perfusate. The D731 MICRO 20 Ph.J.S.I.O. and D733 MICRO 40 Ph.I.S.I.O. are a modified version of the currently marketed D731/D733 MICRO. The modification consists of coating all blood contact surfaces with phosphoryIcholine additive that improves the blood compatibility of the substrate materials. Other than this change the D 731/D733 MICRÓ Ph.I.S.I.O. and the D 731/D733 MICRO are identical in design, materials, and manufacturing processes.
INDICATION FOR USE:
The Dideco D731 MICRO 20 Ph.I.S.I.O. with 20 micron screen with phosphorylcholine coating and the Dideco D733 MICRO 40 Ph.I.S.I.O. with 40 micron screen with phosphoryIcholine coating are recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filters are used to trap and remove gaseous emboli as well as particulate debris that maybe introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.
TECHNOLOGICAL CHARACTERISTICS:
The D731 Micro 20 Ph.I.S.I.O. and the D733 MICRO 40 Ph.I.S.I.O. have the same design features, operating principles and control mechanisms when compared to the D731/D733 predicate devices. The D731 Micro 20 Ph.I.S.I.O. and the D733 MICRO 40 Ph.I.S.I.O. utilize the same materials (with the exception of addition of the phosphorylcholine coating), filtering media with the same filter pore size (20 and 40 micron respectively) and the same main blood flow path as the unmodified devices.
The D731 MICRO 20 Ph.I.S.I.O. and of the D733 MICRO 40 Ph.I.S.I.O. are identical to the current MICRO Pediatric series unmodified devices in design, operating principles, control mechanisms and fundamental scientific technology. No change to the intended use has been made as result of the addition of the
3
phosphoryIcholine coating to all blood contact surfaces. Both devices share the identical manufacturing process. The Ph.I.S.I.O. coating solutions used for the D731 MICRO 20 Ph.I.S.I.O. and the D733 MICRO 40 Ph.I.S.I.O. as well as manufacturing process for applying the coating, is identical to that used for the manalableming 40 Ph.I.S.I.O. (K002493) predicate device. There are no differences in packaging type and material between the D731/D733 MICRO Ph.I.S.I.O. and the MICRO unmodified devices. The arterial filters are ethylene oxide sterilized and have a nonpyrogenic fluid path. They are for single use only.
NON CLINICAL TEST RESULTS:
A complete battery of tests were carried out in accordance with the requirements of ISO 10993-1:1995 and the FDA May 1, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing on the raw materials. Testing was performed on the D 733 MICRO Ph.I.S.I.O. (accelerated aging). The devices were aged up to three years and tested for Hemolysis, Cytotoxicity, Irritation, Acre Systemic Toxicity and Mutagenicity, Sterility, Pyrogenicity and ETO residuals. Package integrity testing was also conducted. The results of the testing met established specifications. As no new materials are used in the D731 MICRO 20 Ph.I.S.I.O. pediatric arterial filter with respect to the D733 MICRO 40 Ph.I.S.I.O data collected are considered applicable to both MICRO Ph.I.S.I.O filters.
IN VITRO TEST RESULTS:
In vitro testing was carried out in accordance with the relevant requirements of in viro couling was carterial Line Blood Filter Blood Filter 510(K) Submission" Final Guidance for Industry, dated November 29, 2000. These data demonstrate substantial equivalence with the unmodified devices and show that the devices are compliant with safety and effectiveness requirements. The device was aged up to 3 years and tested for structural integrity, mechanical integrity, was aged ap te syre drop, filter flow rate capacity, in vitro hemolysis/cell depletion. For comparative purposes all tests were performed on sterilized aged devices comparing the D731 MICRO 20 Ph.I.S.I.O. vs. the D733 non aged unmodified device and D731 MICRO 40 Ph.I.S.I.O vs. D733 non aged unmodified device operated at 5.0 LPM when applicable. The results of these tests met established specifications. The modifications being made to the MICRO Ph.I.O. arterial filter do not affect the performance of the device; therefore the filtration efficiency and air handling characteristics exhibited by the D731/D733 MICRO unmodified device apply also to the MICRO Ph.I.S.I.O. arterial filters.
The results of the study showed that the device characteristics of the D731 MICRO 20 Ph.I.S.I.O. vs. D731 unmodified device and D733 MICRO 40 Ph.I.S.I.O. vs. D733 unmodified device were comparable.
4
CONCLUSIONS:
The results of in vitro studies demonstrate that the D731 MICRO 20 Ph.I.S.I.O. and D733 MICRO Ph.I.S.I.O. devices perform in a manner substantially equivalent to the unmodified devices. Biocompatibility and functional tests demonstrate that their performance is equivalent to the D731 and D733 unmodified devices, according to their intended use. Additional testing has demonstrated the effectiveness of production techniques assuring that the newborn-infant arterial filters are sterile and non-pyrogenic.
5
Image /page/5/Picture/11 description: The image shows a logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract symbol that resembles a person embracing another person.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 0 2005
Dideco S.R.L. c/o Mr. Barry Sall Senior Regulatory Consultant Parexel International 195 West Street Waltham, MA 02451-1163
Re: K051232
R051232
D731 Micro 20 Ph.I.S.I.O. Pediatric Arterial Filter, D733M Regulation Number: 21 CFR 870.4260 Regulation Name: Cardiopulmonary Bypass Arterial Blood Filter Regulatory Class: Class II (Two) Product Code: DTM Dated: May 12, 2005 Received: May 13, 2005
Dear Mr. Sall:
く
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Securit >10(x) promation is substantially equivalent (for the indications
referenced above and have determined the devices marketed in interstate referenced above and nave determined the article predicate devices marketed in interstate for use stated in the enclosure) to legally marketed produced Device Amendments. Or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior to May 28, 1976, the enactinent and of the Federal Food. Drug.
devices that have been reclassified in accordance with the provisions of the PMA devices that have been reclassified in accordance while approval application (PMA).
and Cosmetic Act (Act) that do not require approval as traisions of the Act . The and Cosmetic Act (Act) that do not require approvate or a provisions of the Act. The Act. The
You may, therefore, market the device, subject to the general consussion, lis You may, therefore, market the device, subject to the gentralion, listing of
general controls provisions of the Act include requirements for annual registration, and general controls provisions of the Act mendo roquilibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). in If your device is classified (see above) mo chiler cass if expering your device can
may be subject to such additional controls. Existing major regulation FDA may may be subject to such additional controlis. Exibling migres of to 898. In addition. FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 Register be found in the Code of Federal Regarities in the Federal Register.
6
Page 2 - Mr. Barry Sall
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA 3 issuance of a successions in ther requirements of the Act that FDA has made a determination that your access of other Federal agencies. You must or any Federal statutes and regulanents and united to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as school and as a control leadly also alsot CPK Fatt 607), labeling (21 CFR Part 820); and if applicable, the clectronic Torth in the quality systems (QS) regulations (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Doctors vour device as described in your Section 510(k) I mis letter witi anow you to ocgin manening your and equivalence of your device to a legally premarket notification. The PDA mining of basification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please If you desire specific acvice for Jour do Novel 2011-01-20120. Also, please note the regulation entitled. Contact the Office of Comphalled at (210) 21 CFR Part 807.97). You may obtain Misoralluming by reference to premaritoritibilities under the Act from the Division of Small other general Information on your responsive Assistance at its toll-free number (800) 638-2041 or by J Manufacturers, International and Octisal http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blommarfor
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
7
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: D731 MICRO 20 Ph.I.S.I.O., Dideco D731 Micro 20 Pediatrio Device Name. Droom screen with phosphorylcholine coating and for the D733 MICRO 40 Ph.I.S.I.O., Dideco D733 Micro 40 Pediatric Arterial Filter with 40 micron screen with phosphorylcholine coating
Indications for Use:
The Dideco D731 MICRO 20 Ph.I.S.I.O., with 20 micron screen with The Dideco Drow and the Dideco D733 MICRO 40 Ph.I.S.I.O., with proophorylondine obating and with the coating are recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires the arenanials of the oxtraosiported. Inc. used to trap and remove gaseous emboli as well as particulate debris that maybe introduced through the arterial emooil as well as well as well longer than 6 hours. Contact with blood for longer periods is not advised.
Prescription Use X (Part 21 CFR 801 Subpart D) Over-the-Counter Use_ (21 CFR 807 Subpart C) AND/OR
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
05 1232 Bloomington
Division of Cardiovascular Devices 510(k) Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .