Search Results
Found 2 results
510(k) Data Aggregation
(76 days)
D130 Ph.I.S.I.O. Dideco Kids with 40 micron screen phosphoryIcholine coated is recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardidpulmony in the altenal life of the extraction of our dailing any processions particulate debrish that maybe introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.
The D130 Ph.I.S.I.O. is a sterile, non-pyrogenic disposable filter for use in arterial line of the cardiopulmonary bypass circuit with the flow rate not exceeding 0.7 liters/minute. The D130 Ph.I.S.I.O. is a Neonatal Arterial Filters with 40 micron filter screen designed to remove potentially harmful gaseus embeli, aggregated blood constituents, and particulate debris greater than 40 microns from the arterial line perfusate. The overall dimensions have been reduced in the modified version of the D736 Ph.I.S.I.O. unmodified device resulting in decreased priming volume and maximum blood flow to 0.7 liters/minute. The internal modifications which mainly involve the filter screen geometry, make the new filter esently a downscaled and simplified version of the D736 Ph.I.S.I.O.. The modifications to some of the external features result in enhanced ergonomics.
D130 Ph.I.S.I.O. Dideco Kids Neonatal Arterial Filter Study Summary
This document describes the acceptance criteria and the study performed to demonstrate the substantial equivalence of the D130 Ph.I.S.I.O. Dideco Kids Neonatal Arterial Filter to its predicate device, the D736 Ph.I.S.I.O. The study focused on in vitro evaluations and biocompatibility testing.
1. Table of Acceptance Criteria and Reported Device Performance
The device performance was evaluated against established specifications, primarily demonstrating equivalence or improvement compared to the predicate device, the D736 Ph.I.S.I.O.
Acceptance Criteria Category | Specific Test/Characteristic | Acceptance Criteria (implied, relative to predicate) | Reported Device Performance (D130 Ph.I.S.I.O.) |
---|---|---|---|
Biocompatibility | Hemolysis | Meet ISO 10993-1:2002 and FDA 1995 Memo requirements for raw materials and aged device. | Met established specifications. Performance was equivalent to the D736 Ph.I.S.I.O. unmodified device. |
Hemocompatibility | Meet ISO 10993-1:2002 and FDA 1995 Memo requirements for raw materials and aged device. | Met established specifications. Performance was equivalent to the D736 Ph.I.S.I.O. unmodified device. | |
Cytotoxicity | Meet ISO 10993-1:2002 and FDA 1995 Memo requirements for raw materials and aged device. | Met established specifications. Performance was equivalent to the D736 Ph.I.S.I.O. unmodified device. | |
Irritation | Meet ISO 10993-1:2002 and FDA 1995 Memo requirements for raw materials and aged device. | Met established specifications. Performance was equivalent to the D736 Ph.I.S.I.O. unmodified device. | |
Acute Systemic Toxicity | Meet ISO 10993-1:2002 and FDA 1995 Memo requirements for raw materials and aged device. | Met established specifications. Performance was equivalent to the D736 Ph.I.S.I.O. unmodified device. | |
Mutagenicity | Meet ISO 10993-1:2002 and FDA 1995 Memo requirements for raw materials and aged device. | Met established specifications. Performance was equivalent to the D736 Ph.I.S.I.O. unmodified device. | |
Sterility | Sterile, non-pyrogenic (demonstrated via ETO sterilization techniques). | Demonstrated effectiveness of sterilization techniques, assuring the device is sterile and non-pyrogenic. | |
Pyrogenicity | Non-pyrogenic. | Demonstrated effectiveness of sterilization techniques, assuring the device is sterile and non-pyrogenic. | |
ETO residuals | Acceptable levels of Ethylene Oxide residuals. | Met established specifications. | |
Package integrity | Maintain sterility and device integrity. | Met established specifications. | |
In Vitro Performance | Operating Blood Volume | Reduced compared to predicate D736 Ph.I.S.I.O. | Reduced priming volume compared to the D736 Ph.I.S.I.O. unmodified device. |
Structural Integrity | Comparable to predicate D736 Ph.I.S.I.O., integrity when pressurized over maximum recommended pressure. | Comparable structural integrity to the D736 Ph.I.S.I.O. unmodified device. Demonstrated integrity when pressurized over the maximum recommended pressure. | |
Pressure Integrity | Comparable to predicate D736 Ph.I.S.I.O. | Met established specifications. | |
Pressure Drop | Reduced compared to predicate D736 Ph.I.S.I.O. | Reduced pressure drop compared to the D736 Ph.I.S.I.O. unmodified device. | |
Filter Flow Rate Capacity | Effective at all blood flow rates (up to 0.7 L/min). | Effective filter flow rate capacity at all blood flow rates (up to 0.7 L/min). | |
In Vitro Hemolysis/Cell Depletion | Comparable to predicate D736 Ph.I.S.I.O. | Comparable hemolysis/cell depletion to the D736 Ph.I.S.I.O. unmodified device. | |
Filtration Efficiency | Comparable to predicate D736 Ph.I.S.I.O., overall mean percent removal greater than 80% for 40µ particles. | Comparable filtration efficiency to the D736 Ph.I.S.I.O. unmodified device, with an overall mean percent removal greater than 80% for particles equal to the nominal pore size of the filter (40µ). Removal capability of microbubbles as well as air bolus was comparable. | |
Leaching of Coating | No analytical evidence of leaching of phosphorylcholine coating. | No analytical evidence of the possibility of leaching of phosphorylcholine coating from the D130 Ph.I.S.I.O. was evidenced in the leaching test. | |
Air Handling Characteristics | Comparable to predicate D736 Ph.I.S.I.O. (specifically, removal capability of microbubbles and air bolus). | Removal capability of microbubbles as well as air bolus introduced in the circuit was comparable to the D736 Ph.I.S.I.O. unmodified device. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the numerical sample size for the test set. However, it indicates that "A complete battery of tests were carried out" and "Testing was performed on the D130 Ph.I.S.I.O." and "all tests, when applicable, were performed on sterilized aged devices comparing the D130 Ph.I.S.I.O. vs. the D736 Ph.I.S.I.O." This implies a sufficiently representative number of devices were tested to draw conclusions on performance, aging, and comparison.
The data provenance is from in vitro testing and laboratory experiments conducted by Sorin Group Italia S.r.l. in Italy. The results likely represent prospective data generation specifically for this 510(k) submission, as these tests were carried out to meet current regulatory guidance.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of medical device (arterial filter) relies on objective, measurable physical and chemical properties rather than subjective interpretation of diagnostic images or clinical assessments by human experts. Therefore, the concept of "experts establishing ground truth" in the traditional sense (e.g., radiologists interpreting images) is not directly applicable here. The "ground truth" or "established specifications" were likely based on:
- Engineering and Biomedical Scientists: Professionals with expertise in fluid dynamics, materials science, biocompatibility, and medical device design.
- Regulatory Standards: ISO 10993-1:2002 and FDA 1995 Memorandum, and "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submission" Final Guidance for Industry, dated November 29, 2000. These standards themselves are developed through expert consensus and scientific validation.
- Predicate Device Performance: The established performance of the D736 Ph.I.S.I.O. served as a benchmark for substantial equivalence.
No specific number or qualifications of individual human experts for ground truth establishment are provided because the evaluation method is primarily objective testing against predefined technical standards.
4. Adjudication Method for the Test Set
Not applicable. The tests are objective measurements (e.g., flow rate, pressure drop, particle count, chemical analysis) that do not require human adjudication in the typical sense of a clinical or image-based study. The results are compared directly to pre-established specifications or the predicate device's performance.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical medical device (an arterial filter) and not an AI or software-as-a-medical-device (SaMD) that involves human readers interpreting data.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is an arterial filter and does not involve an algorithm.
7. The Type of Ground Truth Used
The ground truth used was a combination of:
- Quantitative measurements and physical properties: Measured values for operating blood volume, pressure drop, flow rate, filtration efficiency, and integrity tests.
- Biological and chemical assays: Results from biocompatibility tests (hemolysis, cytotoxicity, etc.), sterility tests, pyrogenicity, and leaching tests.
- Established Regulatory Standards: Compliance with ISO 10993-1:2002, FDA Guidance documents, and the D736 Ph.I.S.I.O. 510(K) (K002493) data.
This represents objective scientific and technical data, often rooted in established scientific principles and regulatory benchmarks.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device that requires a training set. The device is a physical product designed through engineering principles.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as no training set was used.
Ask a specific question about this device
(16 days)
The Dideco D735 MICRO 20 with 20 micron screen and the Dideco D736 MICRO 40 with 40 micron screen are recommended for use in the arterial line of the extracorporeal circuit during any procedure that requires cardiopulmonary bypass. The filters are used to trap and remove gaseous emboli as well as particulate debris that maybe introduced through the arterial line. The device should not be used longer than 6 hours. Contact with blood for longer periods is not advised.
The D735 MICRO 20 and D736 MICRO 40 are sterile, non-pyrogenic disposable filters for use in arterial line of the cardiopulmonary bypass circuit with the flow rate not exceeding 2.5 liters/minute. The D735 MICRO 20 and D736 MICRO 40 are Newborn-Infant Arterial Filters with 20 and 40 micron filter screens designed to remove potentially harmful gaseous emboli, aggregated blood constituents, and particulate debris greater than 20 and 40 microns respectively from the arterial line perfusate. The bypass connector has been eliminated in the modified versions of the D736 predicate devices resulting in enhanced ergonomics. In addition, the maximum blood flow rate has been increased to 2.5 liters/minute, in order to be consistent with the maximum flow rate of the cleared D736 MICRO Ph.I.S.I.O. modified device (K002493).
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Hemolysis | Met established specifications | Met established specifications |
Cytotoxicity | Met established specifications | Met established specifications |
Irritation | Met established specifications | Met established specifications |
Acute Systemic Toxicity | Met established specifications | Met established specifications |
Mutagenicity | Met established specifications | Met established specifications |
Sterility | Met established specifications | Met established specifications |
Pyrogenicity | Met established specifications | Met established specifications |
ETO Residuals | Met established specifications | Met established specifications |
Package Integrity | Met established specifications | Met established specifications |
Priming Volume | Met established specifications | Met established specifications |
Filter Integrity | Met established specifications | Met established specifications |
Pull Strength | Met established specifications | Met established specifications |
Blood Side Pressure Drop | Met established specifications | Met established specifications |
In vitro Hemolysis/Cell Depletion | Met established specifications | Met established specifications |
Filtration Efficiency | Met established specifications | Met established specifications |
Air Handling Characteristics | Met established specifications | Met established specifications |
2. Sample Size Used for the Test Set and Data Provenance
The provided text does not specify exact sample sizes for each in-vitro test. However, it states that "all tests were performed on sterilized aged devices" comparing the D735 MICRO 20 (operated at 2.5 LPM) against its D735 predicate device (operated at 2.0 LPM), and the D736 MICRO 40 (operated at 2.5 LPM) against its D736 predicate device (operated at 2.0 LPM).
The data provenance is retrospective, as the testing was performed on aged devices, and the entire document is a 510(k) submission, which is a premarket notification process based on demonstrating substantial equivalence to a legally marketed predicate device. The country of origin for the submitter is Italy (Dideco S.p.A., Mirandola, MO, Italy).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This device is a medical filter, and its performance is evaluated through objective physical and biological in-vitro tests, not through expert-based assessment of diagnostic images or clinical outcomes. Therefore, there is no "ground truth" established by experts in the typical sense of a clinical study.
4. Adjudication Method for the Test Set
Not applicable. As described above, the evaluation relies on objective in-vitro test results against established specifications, not human adjudication of subjective data.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not a diagnostic device involving human readers or AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithmic device. The device itself is a physical filter.
7. The Type of Ground Truth Used
The "ground truth" for the performance evaluation of this medical device is based on pre-defined, objective experimental specifications for physical and biocompatibility characteristics, aligned with relevant guidance documents (e.g., "Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submission" and ISO 10993-1:1995). The comparison is against the performance of legally marketed predicate devices.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device, so there is no concept of a "training set."
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set.
Ask a specific question about this device
Page 1 of 1