K Number
K050921
Device Name
EASYSTIM NMS-28
Date Cleared
2005-12-12

(243 days)

Product Code
Regulation Number
890.5850
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TENS stands for Transcutaneous Electrical Nerve Stimulation. This TENS system is used to provide symptomatic pain relief for chronic, acute or post operative pain.

PMS/EMS stands for Powered Muscles Stimulator/Electrical Muscles Stimulator. This PMS/EMS system is indicated for:

  • Relaxation of muscle spasm
  • Increasing local blood circulation
  • Muscle re-education
  • Prevention or retardation of disuse atrophy
  • Prevention of venous thrombosis of the calf muscles immediately after surgery
  • Maintaining or increasing range of motion
Device Description

For the TENS function A portable TENS device for pain relief.
For the EMS function A portable EMS device for the re-education of muscles.
Features:
· Innovative design
· Large LCD display
• Dual output channels
• 2 AA Alkaline Batteries
· Adjustable frequency, pulse width, and timing parameters
• 9 different modes
• Timer option
• Doctor lock/unlock facility
· Open circuit detectors
• Non-volatile

AI/ML Overview

The EasyMed EMS/TENS Model NMS-28 is a medical device that combines the functions of a Transcutaneous Electrical Nerve Stimulator (TENS) for pain relief and a Powered Muscle Stimulator (PMS) or Electrical Muscle Stimulator (EMS) for muscle re-education.

Here's an analysis based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

The document is a 510(k) summary for the EasyMed EMS/TENS Model NMS-28, which means it aims to demonstrate substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria in the same way a clinical trial would for a novel device. Therefore, a direct table of acceptance criteria and device performance based on quantitative metrics is not explicitly provided in the document.

Instead, the document asserts substantial equivalence based on:

  • Identical intended use and indications for use as predicate devices.
  • Technological comparison:
    • For TENS function: The software for the TENS function is a "subset" of the predicate EasyMed TENS Unit TN-28C (K040253). The new device shares the same enclosure, circuit diagram, PCB layout, components, and working principle.
    • For EMS function: The software for the EMS function is "very similar" to the predicate TensCare Ultima NMS/EMS Model XL-A3 (K023997), applying "similar control methods for muscle contraction-relaxation."
    • Safety and reliability considerations from the TN-28C are incorporated into the EMS function.
  • Safety information:
    • Design complies with relevant safety and recognized consensus standards.
    • Output energy is "well controlled" within safety and effectiveness ranges specified by FDA guidance.
    • "Detailed and strict" testing has been carried out.
    • Risk Analysis and FMEA (Failure Modes and Effects Analysis) show the unit is "safe with no any hazard."
    • Mechanical integrity ensures only one function (TENS or EMS) can be selected at a time.
  • Post-market experience (Europe): The device has been on the European market for two years, and a review of customer complaints, returned products, and post-market feedback demonstrated it "performed as Intended, to its Specified Requirements."

2. Sample Size Used for the Test Set and Data Provenance:

The document does not describe a specific clinical test set in terms of sample size or data provenance (e.g., country of origin, retrospective/prospective). The evidence supporting safety and effectiveness appears to be primarily based on:

  • Bench testing and engineering analysis to verify compliance with standards and equivalence to predicate devices.
  • Existing clinical evidence/regulatory filings for the predicate devices.
  • Post-market surveillance data from the European market.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

This information is not provided as the submission relies on demonstrating substantial equivalence to pre-existing, legally marketed devices, rather than establishing a new ground truth through an independent expert panel for a novel clinical claim.

4. Adjudication Method for the Test Set:

This information is not applicable/provided, as no clinical test set requiring expert adjudication for ground truth is described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

An MRMC study is not mentioned in the document. This type of study (evaluating human reader performance with and without AI assistance) is typically performed for AI-driven diagnostic or screening tools. The EasyMed NMS-28 is a physical electrical stimulation device, not an image-based diagnostic AI, so an MRMC study would generally not be relevant to its regulatory submission.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop):

A standalone performance study focused on an algorithm's performance is not described. As mentioned, this is an electrical stimulation device, not an AI algorithm in the typical sense. The "software" mentioned as part of the technological comparison refers to the embedded firmware controlling the device's electrical output, not a diagnostic or decision-making AI. The "performance" assessment focuses on the device's electrical output parameters and safety through engineering tests and comparison to predicate devices.

7. Type of Ground Truth Used:

The "ground truth" for the submission is based on the established safety and effectiveness of the predicate devices (EasyMed TENS Unit TN-28C and TensCare Ultima NMS/EMS Model XL-A3) and the assertion that the new device is technologically similar and meets relevant safety standards.

8. Sample Size for the Training Set:

This information is not applicable/provided. The device does not employ a machine learning model that requires a "training set" in the conventional sense. The "software" referred to is programmed control logic, not a learned algorithm.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable/provided, as there is no training set for a machine learning model described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "EasyMed" in bold, black font. The word is underlined with a thin, black line. The background is white.

依士文电子仪器有限公司

vour suffering .

EasyMed Instrument Co.,Ltd

6
7
8
9

K050921

DEC I 2 2005

510(k) Summary

Date of Summary prepared:January 21, 2005
Submitter:EasyMed Instrument Co., Ltd
Address :No.2 Bei Hai Da Rd Lunjiao, ShunDe, Foshan, Guangdong, China
Tel :+86-757-27727282 27727283
Fax :+86-757-27727868
E-Mail :info@EasyMed.com.cn
Contact person:Frances Lon (Long Shuming)

Jeffery Wu (Wu Tingjie)

Address of the manufacturing facility: The same as above

Address of American Representative:

Norman F. Estrin, PhD, RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac MD 20854 U.S.A Tel: 001-301-279-2899 Fax: 001-301-294-0126

Submitted Device:
Generic name:Powered Muscle Stimulator (PMS) (or Electrical Muscle Stimulator (EMS)) and Transcutaneous Electrical Nerve Stimulator (TENS)
Trade name:EasySTIM NMS-28
Common name:For the TENS function—TENS deviceFor the EMS function—Powered Muscle Stimulators
Classification name:For TENS functionsStimulator, Nerve, Transcutaneous, for Pain Relief - GZJ;21 CFR 882.5890.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "EasyMed" in a bold, sans-serif font. The word is black and appears to be underlined. The letters are closely spaced, giving the word a compact appearance.

Easy for the relief of your suffering

EasyMed Instrument Co.,Ltd

For EMS functions Powered Muscle Stimulator for re-education of muscles - IPF;

21 CFR 890.5850

Device Classification: For both TENS & EMS functions - Class II

Predicate Device: For the TENS function EasyMed TENS Unit TN-28C (K040253) For the EMS function TensCare Ultima NMS/EMS Model XL-A3 - (K023997)

The class of the predicate Device:

Class II

Device Description:

For the TENS function A portable TENS device for pain relief.

For the EMS function

A portable EMS device for the re-education of muscles.

Features:

· Innovative design

· Large LCD display

• Dual output channels

• 2 AA Alkaline Batteries

· Adjustable frequency, pulse width, and timing parameters

• 9 different modes

• Timer option

• Doctor lock/unlock facility

· Open circuit detectors

• Non-volatile

The intended use of the device:

  1. TENS stands for Transcutaneous Electrical Nerve Stimulation. This TENS system is used to provide symptomatic pain relief for chronic, acute or post-operative pain.

  2. For the EMS function, the NMS-28 is used for:

· Relaxation of muscle spasm

•Increasing local blood circulation

· Muscle re-education

· Prevention or retardation of disuse atrophy

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "EasyMed" in bold, black font. The word is split into two parts: "Easy" and "Med". A horizontal line underlines the entire word, emphasizing the brand name. The font is sans-serif and appears to be a logo or brand identifier.

依士文电子仪器有限公司

EasyMed Instrument Co.,Ltd suffering

•Prevention of venous thrombosis of the calf muscles immediately after surgery

· Maintaining or increasing range of motion

The intended use and indications for use of the new device are the same with that of the predicate devices.

Technological Comparison:

    1. The marketed device EasyMed TN-28C (K040253) is a digital TENS unit which has been well exploited the digital technology, supplying the user with full ranges of parameters, traditional TENS functions, and non-traditional new functions to be selected.
    1. In assembly, except software, the new device NMS-28 is exactly the same with the marketed device TN-28C, they have the same enclosure, the same circuit diagram, the same PCB layout, the same components and the same working principle;
    1. The marketed device TensCare Ultima NMS/EMS Model XL-A3 -(K023997) combines the functions of a TENS device and an EMS (Electrical Muscle Stimulator) or Powered Muscle Stimulator, into one package. It is also a digital unit with similar electronic working principle with TN-28C.
  • The new device NMS-28 also combines the functions of a TENS ধ device and an EMS into one package. It is also substantial equivalent to the marketed device TensCare Ultima NMS/EMS Model XL-A3 - (K023997), their features and performances in muscle controlling as well as pain relief therapy are very similar substantially.
    1. The software of the new device NMS-28 can be divided into two parts: one for TENS function, the another for EMS function. The part of software for TENS function of the new device NMS-28 is only a subset of that of the marketed device TN-28C, what is more,
  • the considerations on safety and reliability in TN-28C are also introduced into the EMS function of the new device; While the part of software for EMS function of the new device NMS-28 is very similar to that of the marketed device TensCare Ultima NMS/EMS Model XL-A3, by applying the similar control methods for muscle contraction-relaxation.
  • The accessories of the new device NMS-28 are almost the same ్.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "EasyMed" in a bold, sans-serif font. The letters are black, and there is a black line underneath the entire word. The word is likely a logo or brand name.

Easy for the relief of your suffering

EasyMed Instrument Co.,Ltd

with those of the marketed device TN-28C, and substantially equivalent to marketed device TensCare Ultima NMS/EMS Model XL-A3.

The Labeling is substantially equivalent to that of the predicate Labeling Comparison: devices.

Safety information:

Design to comply with relevant safety applicable recognized consensus standards; the output energy is well controlled in the safety and effectiveness ranges specified by relevant FDA guidance's. Testing has been carried out in very detailed and strict. · Test results, Risk Analysis, and FMEA analysis show that the new unit NMS-28 is safe with no any hazard.

NMS-28 combines the functions of TENS and EMS into one package, it is not possible to use the TENS and the EMS functions simultaneously. Mechanical integrity ensures that only one function can be selected at any one time.

NMS-28 has been on the European Market for the past two years. During this time a review of Customer Complaints, Returned Product and the results of Post Market Feedback, has demonstrated that the product has performed as Intended, to it's Specified Requirements.

Submitted times: It is the first submission to FDA for this new device

Conclusions:

The NMS-28 EMS/TENS unit is substantially equivalent to the predicate devices and any differences between the devices do not pose any new questions of safety and effectiveness.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure embracing a bird, which is meant to convey the department's mission of protecting and promoting the health and well-being of Americans.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 2 2005

Jeffrey Wu EasyMed Instrument Co., Ltd No. 2 Bei Hai Da Road LunJiao, ShunDe Guangdong, China

Re: K050921

K050921
Trade/Device Name: EasyMed EMS/TENS Model NMS-28 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: II Product Codes: IPF, GZJ Dated: June 17, 2005 Received: November 23, 2005

Dear Mr. Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 3 D(t), plenial is substant for the indications for the indications for referenced above and have delemined the devices marketed in interstate commerce
use stated in the enclosure) to legally marketed predicate or to devices that use stated in the enclosure) to tegally hiar Recol prece Amendments, or to devices that prior to May 28, 1976, the enactment date of the Federal Food, Drug, and Cosmetic
have been reclassified in accordance with the provisions of the Federal PMA) You may. have been reclassified in accordance with the provisions of the Act. The Act. The peneral Act (Act) that do not require approval or a prematic controls provisions of the Act. The general
therefore, market the device, subject to the general controls provisions of therefore, market the device, subject to the general on annual registration, listing of devices, good
controls provisions of the Act include requirements for anning mind and controls provisions of the Act mende requirements for and adulteration.
manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) into cliffer the stime aftecting your device can be
may be subject to such additional controls. Existing major regulations FDA may may be subject to such additions, Title 21, Parts 800 to 898 . In addition, FDA may publish
found in the Code of Federal Regulations, Title 21, Parts 800 to 898 . found in the Code of Federal resganaties in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that Please be advised that FDA's issualice of a substance requirements of the Act or any
FDA has made a determination that your device complex . You must comply with all FDA has made a determination inal your tee lice concess. You must comply with all
Federal statutes and regulations administerial and listing (21 CFR Part 807); Federal statutes and regulations administered or our registration and listing (21 CFR Part 807);
the Act's requirements, including, but not limited to: registration as set the Act s requirements, melading, but not maring practice requirements as set
labeling (21 CFR Part 801); good manufactures (200); and if annlice labeling (2) CFR Part 801); good inaliation (2) CFR Part 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR 1000-1050. forth in the quality systems (QS) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-

This letter will allow you to begin marketing your device as described in your Section 510(k)
. It is a legin and the first of a local started assissalence of your device to This letter will allow you to begin marketing your device as assisted by and one of your device to a legally

{5}------------------------------------------------

Page 2 -- Mr. Jeffrey Wu

marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs of the many of the many of 10, 2019 - Alancy lass mote the regulation e If you desire specific advice for your de vice on on our nooning note the regulation entitled,
contact the Office of Compliance at (240) 276-0120. Also, please note the regul contact the Office of Compilance at (240) 210 of 2019 Part 807.97). You may obtain other "Misbranding by reference to premaired nother the Act from the Division of Small Manufactures,
general information on your responsibilities under the Act 1 - 1000 - 3301 May general information on your responsionities under (800) 638-2041 or (301) 443-6597 or
International and Consumer Assistance at is this and works of the l international and Consumer Treaver Tra.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

2

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

Indications for Use

·· K050921 510(k) Number

EasyMed EMS/TENS Model NMS-28 Device Name :

Indications for Use :

TENS stands for Transcutaneous Electrical Nerve Stimulation. This TENS system is used to provide symptomatic pain relief for chronic, acute or post operative pain.

PMS/EMS stands for Powered Muscles Stimulator/Electrical Muscles Stimulator. This PMS/EMS system is indicated for:

  • · Relaxation of muscle spasm
  • · Increasing local blood circulation
  • Muscle re-education
  • · Prevention or retardation of disuse atrophy
  • · Prevention of venous thrombosis of the calf muscles immediately after surgery
  • Maintaining or increasing range of motion

Prescription Use_ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

page 1/1
Page 9 of 205

510(k) Number K050421

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).