K Number
K050168
Device Name
LANGSTON DUAL LUMEN PRESSURE MONITORING CATHETER
Date Cleared
2005-05-18

(112 days)

Product Code
Regulation Number
870.1200
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring Catheter is indicated for delivery of contrast media in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressure measurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.
Device Description
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring catheter is intended for use as a pressure measurement catheter and for delivery of contrast media during angiographic studies. The DLP catheter consists of the Merit Medical Softouch Diagnostic Intravascular Catherer (K943739) as the inner lumen and an extruded outer lumen designed by Vascular Solutions. The inner lumen and hub assembly are purchased from Merit Medical in a non-sterile state. The Merit Softouch catheter and outer lumen are joined together using an adapter junction placed near the proximal end of the Merit Softouch catherer. The adapter junction incorporates a side port fitted with a stopcock/tube assembly for fluid flow and pressure measurement within the outer tube. The distal end of the outer tube is perforated with multiple holes to allow pressure measurement simultaneously with the inner catheter. The dual lumen catheter is deployed through standard introducers and will accommodate standard 0.038'' diameter guidewires.
More Information

No
The description focuses on the mechanical design and function of a catheter for pressure measurement and contrast delivery. There is no mention of AI or ML in the intended use, device description, or performance studies.

No
The device is indicated for diagnostic purposes (delivery of contrast for angiographic studies and simultaneous pressure measurement) rather than for treating a disease or condition.

Yes
The device is indicated for "simultaneous pressure measurement from two sites," which is useful in determining "transvalvular, intravascular and intraventricular pressure gradients." This measurement capability is for diagnostic purposes.

No

The device description clearly describes a physical catheter with lumens, an adapter junction, a side port, and perforations, indicating it is a hardware device.

Based on the provided information, the Vascular Solutions Langston™ Dual Lumen Pressure Monitoring Catheter is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is for "delivery of contrast media in angiographic studies and for simultaneous pressure measurement from two sites." This involves direct interaction with the patient's body (intravascular, intraventricular) for diagnostic purposes in vivo.
  • Device Description: The description details a catheter designed to be inserted into the body for fluid delivery and pressure measurement.
  • Lack of IVD Characteristics: IVD devices are designed to examine specimens (like blood, urine, tissue) outside the body (in vitro) to provide information about a patient's health. This device does not perform any analysis on specimens outside the body.

Therefore, this device falls under the category of an invasive diagnostic device used within the patient's body, not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

The Vascular Solutions Langston™ Dual Lumen Pigtail Catheter is indicated for delivery of The vacouxi oor in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressure measurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.

Product codes

DQO

Device Description

The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring catheter is intended for use as a pressure measurement catheter and for delivery of contrast media during angiographic studies. The DLP catheter consists of the Merit Medical Softouch Diagnostic Intravascular Catherer (K943739) as the inner lumen and an extruded outer lumen designed by Vascular Solutions. The inner lumen and hub assembly are purchased from Merit Medical in a non-sterile state. The Merit Softouch catheter and outer lumen are joined together using an adapter junction placed near the proximal end of the Merit Softouch catherer. The adapter junction incorporates a side port fitted with a stopcock/tube assembly for fluid flow and pressure measurement within the outer tube. The distal end of the outer for trail 110 ware possible holes to allow pressure measurement simultaneously with the tube is personed was as a catheter. The dual lumen catheter is deployed through standard sidences at are and will accommodate standard 0.038'' diameter guidewires.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Summary of Non-Clinical Testing:
Testing conducted included assessments of the design verification of the Dual Lumen I esuits conducted mended assessments biocompatibility assessments. The results of this Fressure Monitoning Catherer assist "" battery of tests confirmed the suitability of the Dual Lumen Pressure Monitoring Catheter for its intended use.

Summary of Clinical Testing:
No clinical evaluations of this product have been conducted.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K943739

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.1200 Diagnostic intravascular catheter.

(a)
Identification. An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.(b)
Classification. Class II (performance standards).

0

K050168

MAY 1 8 2005

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

Common/Usual Name:Intravascular Diagnostic Catheter
Product Trade Name:Langston™ Dual Lumen Pressure Monitoring Catheter
Classification Name:Class II
21 CFR 870.1200
Product Code: DQO
Manufacturer:Vascular Solutions, Inc.
6464 Sycamore Court
Minneapolis, Minnesota 55369
USA
Establishment Registration:2134812
Contact:Sara L. Coon
Senior Regulatory Affairs Associate
(763) 656-4300 phone
(763) 656-4200 fax
Performance Standards:No performance standards have been developed under
section 514 for this device.

Device Description:

Device Desemplosi.
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring catheter is intended for use as a pressure measurement catheter and for delivery of contrast media during angiographic studies. The DLP catheter consists of the Merit Medical Softouch Diagnostic Intravascular Catherer (K943739) as the inner lumen and an extruded outer lumen designed by Vascular Solutions. The inner lumen and hub assembly are purchased from Merit Medical in a non-sterile state. The Merit Softouch catheter and outer lumen are joined together using an adapter junction placed near the proximal end of the Merit Softouch catherer. The adapter junction incorporates a side port fitted with a stopcock/tube assembly for fluid flow and pressure measurement within the outer tube. The distal end of the outer for trail 110 ware possible holes to allow pressure measurement simultaneously with the tube is personed was as a catheter. The dual lumen catheter is deployed through standard sidences at are and will accommodate standard 0.038'' diameter guidewires.

Intended Use:

The Vascular Solutions Langston™ Dual Lumen Pigtail Catheter is indicated for delivery of The vacouxi oor in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressure measurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.

1

Summary of Non-Clinical Testing:

Sunnillary of Non-Gillicar 2 council.
Testing conducted included assessments of the design verification of the Dual Lumen I esuits conducted mended assessments biocompatibility assessments. The results of this Fressure Monitoning Catherer assist ""
battery of tests confirmed the suitability of the Dual Lumen Pressure Monitoring Catheter for its intended use.

Summary of Clinical Testing:

No clinical evaluations of this product have been conducted.

Predicate Device

I Iculcate Device I The Dual Lumen Pigtail Catheter, the Cordis Dual Lumen Pressure Monitoring Catheter, and the Merit Medical Softouch Diagnostic Intravascular Catheter.

Conclusions:

The Langston Dual Lumen Pressure Monitoring Catheter is substantially equivalent to the I The Langston Dual Lumen Pigtail Catheter , the Cordis Dual Lumen Pressure Monitoring Latheter and the Merit Medical Softouch Diagnostic Intravascular Catheter. The testing Catheter and the Heather the Langston Dual Lumen Pressure Monitoring Catheter will perform as intended.

2

Image /page/2/Picture/1 description: The image is a black and white emblem for the U.S. Department of Health & Human Services. The emblem features the department's logo, which consists of three stylized human figures connected by flowing lines, representing the department's mission to protect the health of all Americans. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 8 2005

Ms. Sara L. Coon Senior Regulatory Affairs Associate Vascular Solutions, Inc. 6464 Sycamore Court Minepolis, MN 55369

Re: K050168

Trade/Device Name: Langston™ Dual Lumen Pressure Monitoring Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: II Product Code: DQO Dated: May 3, 2005 Received: May 4, 2005

Dear Ms. Coon:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nevers for was your we determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass surfed to May 28, 1976, the enactment date of the Medical Device Amendments, or to conninered prox conner 2012/ 2018 11:12 provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, therefered, issions of the Act include requirements for annual registration, listing of general concess profile to proctice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

3

Page 2 -- Ms. Sara L. Coon

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 100-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050 product radiation control provisions (Sections 991 ceries in your Section 510(k)
This letter will allow you to begin marketing your device of your daying to Jeggl This letter will allow you to begin makemig your avvalence of your device to a legally premarket notification. The PDA intallight substanced by and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please e If you desire specific advice for your aconce to and one one the regulation entitled, the regulation entitled, comact the Office of Comphalloo at (210) 276 off (21CFR Part 807.97). You may obtain "Misbranding by reference to premarket to premailities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or 1 Manufacturers, International and Consalher physical on the many of the new learners himl.

Sincerely yours,

Bhimman for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

Indications for Use Statement

510(k) Number:

K050168

Device Name:

Vascular Solutions Langston™ Dual Lumen Pressure Monitoring Catheter

Indications for Use:

The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring Catheter is indicated for The Vascular Solutions Earingershic studies and for simultaneous pressure measurement ticavely of containt modeline in angest of the good intravascular and intraventricular pressure gradients.

Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bhimman

(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K050168

Page 1 of