K Number
K043080
Manufacturer
Date Cleared
2004-12-03

(25 days)

Product Code
Regulation Number
864.5680
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Heparin Dose Response (HDR) controls are used to verify the performance of the HDR cartridges and the HMS Plus instrument.

Device Description

The HDR Control set is an in vitro diagnostic device. The primary function is to identify if each pair of heparinized channels of the HDR cartridge has normal or abnormal function by measuring if the clotting time ratios between heparinized and unheparinized channels are within or outside a specified range. HDR control 1 is used to verify that channels 1,2, 5 and 6 of the HDR cartridge are functioning correctly. HDR control 2 is used to verify that channels 3, 4, 5 and 6 of the HDR cartridge are functioning correctly. The HDR Control set is similar in design and use (vial size and reconstitution method) to the existing liquid controls being used in the HMS platform. The HDR controls are single use, non-sterile, point of care, in-vitro diagnostic plasma controls for use with the HMS Plus instrument.

AI/ML Overview

This 510(k) summary does not contain detailed information about acceptance criteria and specific study results in the format requested. It is a high-level summary confirming substantial equivalence to a predicate device.

However, based on the provided text, here's what can be extracted and inferred:

1. A table of acceptance criteria and the reported device performance:

The document states, "The primary function is to identify if each pair of heparinized channels of the HDR cartridge has normal or abnormal function by measuring if the clotting time ratios between heparinized and unheparinized channels are within or outside a specified range." This implies that the acceptance criteria are related to these clotting time ratios falling within or outside a predefined "normal" or "abnormal" range.

The phrase "Validation testing was used to establish the performance characteristic of the modifications of this device from the previously marketed device" suggests that the device did meet these established criteria. However, the specific numerical acceptance criteria and the actual reported performance values are not provided in this summary.

Inferred Table:

Acceptance Criterion (Inferred)Reported Device Performance (Specificity and Sensitivity Inferred)
Clotting time ratio between heparinized and unheparinized channels falls within a specified range (Normal function)Device accurately identifies normal function of HDR cartridge channels 1, 2, 5, 6 (using HDR Control 1) and 3, 4, 5, 6 (using HDR Control 2). Specific metrics (e.g., accuracy, sensitivity, specificity) are not provided.
Clotting time ratio between heparinized and unheparinized channels falls outside a specified range (Abnormal function)Device accurately identifies abnormal function of HDR cartridge channels 1, 2, 5, 6 (using HDR Control 1) and 3, 4, 5, 6 (using HDR Control 2). Specific metrics are not provided.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

This information is not provided in the summary. The summary refers to "Validation testing" but does not detail the size or nature of the test set, nor the data provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

This information is not provided in the summary. The HDR Controls are an in vitro diagnostic device for verifying instrument performance, so "expert ground truth" in the clinical imaging sense is not directly applicable. The "ground truth" would likely be established by the expected chemical/biochemical reactions within the controls themselves, verified through analytical methods.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not provided in the summary.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device is an in vitro diagnostic control for instrument performance, not an AI-assisted diagnostic tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The device described is an "in vitro diagnostic device" (HDR Control set) designed to verify the performance of the "HMS Plus instrument" and "HDR cartridges." This implies a standalone performance evaluation where the control's ability to trigger expected responses from the instrument/cartridge is assessed. The "algorithm" here would be the expected chemical reaction and the instrument's detection capabilities. The summary states, "The primary function is to identify if each pair of heparinized channels of the HDR cartridge has normal or abnormal function by measuring if the clotting time ratios between heparinized and unheparinized channels are within or outside a specified range." This heavily suggests a standalone performance evaluation of the control and the instrument's interaction with it.

However, the specific details of this "standalone" study (e.g., how many controls were tested, how many times, under what conditions) are not provided in this summary.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For an in vitro diagnostic control device, the ground truth would be established by the inherent, expected chemical and physical properties of the control material itself, which are designed to produce a specific, known response when tested by the HMS Plus instrument and HDR cartridges. This "ground truth" is typically ensured by rigorous manufacturing, analytical testing of the control material, and established scientific principles of coagulation. It's not "expert consensus" or "pathology" in the typical clinical sense, but rather a known, engineered outcome.

8. The sample size for the training set:

This is not applicable in the context of an in vitro diagnostic control device for instrument verification. There is no AI model being "trained" in the conventional sense. The "training" would be the initial development and optimization of the control formulation to consistently produce the desired clotting time ratios.

9. How the ground truth for the training set was established:

As mentioned above, there isn't a "training set" in the AI sense. The "ground truth" for the control materials would be established during their design, formulation, and manufacturing processes, where the concentrations of active ingredients are precisely controlled to ensure they yield specific, predictable results (i.e., normal or abnormal clotting time ratios) when interacting with the HMS Plus instrument and HDR cartridges according to established biochemical principles and validated analytical methods.

{0}------------------------------------------------

K043080

DEC - 3 2004

(Optional Format 3-10-98)

:

510(k) Summary

Date Prepared:November 04, 2004
Submitter:Medtronic Perfusion Systems7611 Northland BoulevardBrooklyn Park, MN 55428
Contact Person:Bruce BacklundSenior Regulatory Affairs Specialist
Phone: (763)-391-9183Fax: (763) 391-9603

Device Name and Classification:

Trade Name:Heparin Dose Response Control
Common Name:Analyzer, Heparin, Automated
Classification:Class II
Predicate Devices:HMS Controls K894317

{1}------------------------------------------------

Device Description

The HDR Control set is an in vitro diagnostic device. The primary function is to identify if each pair of heparinized channels of the HDR cartridge has normal or abnormal function by measuring if the clotting time ratios between heparinized and unheparinized channels are within or outside a specified range. HDR control 1 is used to verify that channels 1,2, 5 and 6 of the HDR cartridge are functioning correctly. HDR control 2 is used to verify that channels 3, 4, 5 and 6 of the HDR cartridge are functioning correctly.

The HDR Control set is similar in design and use (vial size and reconstitution method) to the existing liquid controls being used in the HMS platform. The HDR controls are single use, non-sterile, point of care, in-vitro diagnostic plasma controls for use with the HMS Plus instrument.

Indication for Use

The Heparin Dose Response (HDR) controls are used to verify the performance of the HDR cartridges and the HMS Plus instrument.

Comparison to Predicate Device

The predicate device 510(k) K894317 was cleared on September 5, 1989. This 510(k) introduced the entire Hepcon® HMS product family, including cartridges and controls.

The HDR cartridge marketed under K042070 (previously K894317, see table for chronology), Instructions for Use is included in Appendix A.

Summary of Performance Data

Validation testing was used to establish the performance characteristic of the modifications of this device from the previously marketed device.

Conclusion

Medtronic Perfusion Systems has demonstrated that the HDR Controls are substantially equivalent to the predicate devices based upon design, test results, and indications for use.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle's head and three horizontal lines above it, representing the department's mission to promote health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Bruce Backlund Senior Regulatory Affairs Specialist Medtronic Cardiac Surgery 7611 Northland Drive N Minneapolis, MN 55428-1088

Re: K043080 Trade/Device Name: Heparin Dose Response Control Regulation Number: 21 CFR § 864.5680 Regulation Name: Automated Heparin Analyzer Regulatory Class: II Product Code: JOX, GGC Dated: November 4, 2004 Received: November 8, 2004

Dear Mr. Backlund:

We have reviewed your Section 510(k) premarket notification of intent to market the device w o nove and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

DFC - 3 2004

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r four ove actived a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny I edetal statuated and states and limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice OFF Part 887), as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirements as set form in keting your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In of questions on all proxie Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I ou may obtain other general messantional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Robert L. Becker Jr.

Robert L. Becker, Jr., M.D., Ph. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K04308080 Device Name: Heparin Dose Response Controls

Indications For Use:

The Heparin Dose Response (HDR) controls are used to verify the performance of HDR cartridges and the HMS Plus instrument.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR Over-The-Counter Use __________________________________________________________________________________________________________________________________________________

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Josephine Bautista

Page 1 of 1_

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safe

510(k) K043080

§ 864.5680 Automated heparin analyzer.

(a)
Identification. An automated heparin analyzer is a device used to determine the heparin level in a blood sample by mixing the sample with protamine (a heparin-neutralizing substance) and determining photometrically the onset of air-activated clotting. The analyzer also determines the amount of protamine necessary to neutralize the heparin in the patient's circulation.(b)
Classification. Class II (special controls).