K Number
K040259
Manufacturer
Date Cleared
2004-04-22

(78 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Masimo SET® Intellivue Pulse Oximeter Module is intended for the continuous noninvasive monitoring of functional oxygen saturerial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals and hospital-type facilities.

The IntelliVue Pulse Oximeter Module is indicated for the continuous nonitoring of functional organization of financed for I ne ment vice culse oximicle moduced by an Spo-sensor). The Intellivie Pulse Oximeter Module is indicated for are with adult, pediatric and neonatal patients during both motion conditions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities.

Device Description

The Masimo SET® IntelliVue Pulse Oximeter Module is a continuous noninvasive, arterial oxygen saturation and pulse rate montor. The Masimo SET Intellivue Module features Masimo SET algorithms in a Philips single-width Intellivae compatible module.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (from "Specifications and Operating Ranges")Reported Device Performance (from "Clinical Tests" and "Nonclinical Tests")
Saturation (% SpO2) - No Motion
Adults, Pediatrics70% - 100% ± 2 digits± 2 digits (when compared to CO-Oximeter)
Neonates70% - 100% ± 3 digits± 3 digits (when compared to CO-Oximeter)
Saturation (% SpO2) - Motion
Adults, Pediatrics70% - 100% ± 3 digits± 3 digits (when compared to CO-Oximeter)
Neonates70% - 100% ± 3 digits± 3 digits (when compared to CO-Oximeter)
Pulse Rate (bpm) - No Motion
All Patients25 to 240 ± 3 digits± 3 digits (when compared to ECG)
Pulse Rate (bpm) - Motion
All Patients25 to 240 ± 5 digits± 5 digits (when compared to ECG)
Low Perfusion Performance
Saturation (% SpO2)± 2 digits (> 0.02% Pulse Amplitude and % Transmission > 5%)± 2 digits (when compared to simulators for bench testing)
Pulse Rate (bpm)± 3 digits (> 0.02% Pulse Amplitude and % Transmission > 5%)± 3 digits (when compared to simulators for bench testing)

Note on "0% - 69% unspecified": For saturation below 70%, the device's accuracy is not specified in the "Specifications and Operating Ranges" table for any patient group or condition. The text later reiterates this as "0% - 69% unspecified".

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Description:
    • Motion and No Motion Studies: "healthy adult volunteer subjects during no motion and motion conditions who were subjected to a progressive induced hypoxia"
    • Low Perfusion Studies: "healthy adult volunteer subjects to low perfusion conditions and to a progressive induced hypoxia"
    • Neonates (Accuracy Basis): "clinical studies on adult volunteers and 1% was added to account for the properties of fetal hemoglobin"
  • Sample Size: The exact number of subjects for the clinical studies (test set) is not specified in the provided text.
  • Data Provenance: The studies were conducted on "healthy adult volunteer subjects" and involved "human blood studies." The country of origin is not explicitly stated, but the context of an FDA submission (K040259, U.S. Department of Health & Human Services) strongly suggests the studies were conducted in the United States or overseen by US regulations. The studies were prospective, involving induced hypoxia and real-time monitoring.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not applicable in the traditional sense for this type of medical device study.
  • Qualifications of Experts: The ground truth was established by objective medical instruments, not by expert interpretation.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. The ground truth was established by direct measurement with reference devices.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is more common for diagnostic imaging devices where human interpretation is a primary component of the diagnostic process.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, a standalone performance assessment was done. The studies explicitly compare the device's readings (algorithm only output) to established reference measurements (CO-Oximeter and ECG) without human intervention in interpreting the device's output. The device itself is an automated measurement tool.

7. The Type of Ground Truth Used

  • Ground Truth Type:
    • For Saturation (% SpO2): Arterial blood samples measured by a CO-Oximeter. This is an objective laboratory measurement.
    • For Pulse Rate (bpm): Measurements from an ECG monitor. This is an objective physiological measurement.
    • For Low Perfusion (Bench Testing): The "Biotek Index 2 simulator and Masimo's simulator" were used, which are controlled bench test environments designed to simulate physiological conditions with known outputs.

8. The Sample Size for the Training Set

  • Training Set for Empirical Data: The text states, "The values in the look-up table are based upon human blood studies against a mu the Masmo 021 - miedir as betwarers in induced hypoxia states during motion and non-motion conditions."
  • Sample Size: The exact sample size for the training set (the "human blood studies" used to derive the look-up table) is not specified in the provided text.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: The ground truth for the empirical look-up table was established through "human blood studies against a CO-Oximeter" in "induced hypoxia states during motion and non-motion conditions." This indicates that, similar to the test set, the reference measurements for oxygen saturation were obtained from a CO-Oximeter. The mechanism for grounding the pulse rate data in the training set is not explicitly detailed but would logically follow similar objective physiological measurements.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text 'K040259' in a handwritten style. The text is rotated diagonally, with the 'K' at the bottom left and the '9' at the top right. The numbers and letters are written in a dark ink, contrasting with the white background.

APR 2 2 2004

Submitted by:Masimo Corporation2852 Kelvin AveIrvine, CA 92614-5826(714) 250-9688FAX (714) 250-9686
Company Contact:James J. Cronin, Vice President, Regulatory Affairs/Quality Assurance
Date Summary Prepared:February 2, 2004
Trade NameMasimo SET® Intellivue Pulse Oximeter Module
Common NamePulse Oximeter
Classification NameOximeter (74DQA) (870.2700)
Substantially Equivalent DevicesMasimo SET Radical Pulse Oximeter with SatShare™ and LNOP series of Sensors andCables 510(k) Number - K031330Philips Medizin Systeme MP40, MP50, MP60, MP70 and MP90 Intellivue PatieMonitor - K032858

The Masimo SET® IntelliVue Pulse Oximeter Module is a continuous noninvasive, arterial oxygen saturation and pulse rate montor. The Masimo SET Intellivue Module features Masimo SET algorithms in a Philips single-width Intellivae compatible module.

The following list outlines the key features and benefits of the Masimo SET IntelliVue Module

Features and Benefits

Clinically proven Masimo SET'1M technology performance Applicable for use on neonate, pediatric and adult patients Proven for accurate monitoring in motion and low perfusion environments SpO2, pulse rate, alarm, and signal quality displays BP27 police with Philips IntelliVue models MP 40/50, MP 60/70 and MP90 must be equipped with Philips Intellivue Flexible Module Server (FMS) with at least one available module slot. Compatible with all Masimo LNOP SpO2 sensors (except LNOP DCSC)

Intended use

The Masimo SET® Intelli Vue Pulse Oximeter Module is intended for the continuous noninvasive monitoring of functional oxygen saturerial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals and hospital-type facilities.

{1}------------------------------------------------

Indications for use

The IntelliVue Pulse Oximeter Module is indicated for the continuous nonitoring of functional organization of financed for I ne ment vice culse oximicle moduced by an Spo-sensor). The Intellivie Pulse Oximeter Module is indicated for are with adult, pediatric and neonatal patients during both motion conditions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities.

Principles of Operation

The principles of operation of the Masimo SET® Intellivue pulse oximeter Module are that oxyhemoglobin and deoxyhemoglobin I he principes of operation of the Masmo BDF - intentrade ers), the volume of arterial blood in tissue (and hence, light ative in then absorphon of real and minuted in (specifysmocraphy), and that arterio-venous shunting is highly variable and that also putin by that 0100) changed anning in paise during the pulse. Because oxyhemoglobin and deoxyhemoglobin differ in light absorption, the amount of red and infrared light absorbed by the blood is related to hemoglobin oxygen saturation. The Masimo SET® Intellivue pulse oximeter module decomposes the red and infraced pulsatile absorbance signal into an arterial signal plus a noise component and calculates the ratio of the arterial signals without noise. The ratio of the two arterial pulse-added absorbance signals and is used to find the SpO2 saturation in an empirically derived equation into the Masimo SET® Intellivue software. The values in the look-up table are based upon human blood studies against a mu the Masmo 021 - miedir as betwarers in induced hypoxia states during motion and non-motion conditions.

Method of Operation

The Masimo SET® Intellivue pulse oximeter Module is installed in an empty slot in a compatible Philips Intelli Vue Patient I homoring System. An oximetery sensor is attached to a patient's finger and one end of a patient cable is connected to the sensor and the other end connected to the Masimo SET® Intellivue pulse oximeter Module.

The monitor will begin continuously displaying the patient's pulse rate, and SpO2 value. The practitioner can then use the I no motion wat is continuously displayed on the monitor to help assess the condition of the patient and as an aide in determining if any intervention is required by the practitioner.

Once the practitioner determines the patient no longer requires monitoring, the cable is disconnected from the oximetry sensor is removed (and disposed of if it is a single use device), and the power to the monitor is turned off.

Power Source

The Masimo SET® Intellivue pulse oximeter module is powered by the Philips IntelliVue Patient Monitoring System.

Specifications and Operating Ranges

Range
Saturation (% SpO2)1% - 100%
Pulse Rate (bpm)25 - 240
Perfusion0.02% - 20%
Accuracy
Saturation (% SpO2) - During No Motion Conditions1
Adults, Pediatrics70% - 100% ± 2 digits
0% - 69% unspecified
Neonates70% - 100% ± 3 digits
0% - 69% unspecified
Saturation (% SpO2) - During Motion Conditions2,3
Adults, Pediatrics270% - 100% ± 3 digits
0% - 69% unspecified
Neonates370% - 100% ± 3 digits
Pulse Rate (bpm) - During No Motion Conditions1Adults, Pediatric, Neonates25 to 240 ± 3 digits
Pulse Rate (bpm) - During Motion Conditions2,3Adults, Pediatric, Neonates25 to 240 ± 5 digits
Resolution
Saturation (% SpO2)Pulse Rate (bpm)1%1
Low Perfusion Performance4
> 0.02% Pulse Amplitudeand % Transmission > 5%Saturation (% SpO2) ± 2 digitsPulse Rate ± 3 digits
Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to theamount of carboxyhemoglobin present. Dyes, or any substance containing dyes, that change usual arterialpigmentation may cause erroneous readings.
Isolation
Patient Leakage currentDielectric Withstand (mains to patient)Less than 100 μAmp>4000 VAC
Environmental
Operating TemperatureStorage Temperature32°F to + 131°F (0°C to +55°C)-40°F to + 158°F (-40°C to +70°C)
Relative Humidity5% to 95% noncondensing
Operating Altitudeup to 15,000 ft
Circuitry
Microprocessor controlledAutomatic self-test of oximeter when powered onAutomatic setting of parametersAutomatic alarm messages
Audio indicators
Controlled by Philips IntelliVue Patient Monitoring System
Physical characteristics
Dimensions:Weight:1.4" x 3.9" x 3.8" (36 mm x 99.6 mm x 97.5 mm)7.4oz. (209 g)
Modes
Averaging mode:Sensitivity5, 10, 20 secondsNormal

{2}------------------------------------------------

0% - 69% unspecified

  • The Masimo SET Technology with LNOP Adt sensors has been validated for no motion accuracy in human blood 1 studies on healthy adult volunteers in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory cooximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.

{3}------------------------------------------------

  • 2 The Masimo SET Technology with LNOP Adt sensors has been validated for motion accuracy in human blood The Husening OD - Feet.net By - I counters while performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
  • The specified saturation accuracy from 70% 100% for neonates is based on the results from clinical studies on adult I its specifics saturation account for the properties of fetal hemoglobin to be within ± 3 digits during both motion and no motion conditions when compared to the CO-Oximeter, and the pulse rate accuracy values for neonates to be within ± 3 digits during no motion and ± 5 digits during motion conditions when compared to the ECG.. This variation equals plus or minus one standard deviation which encompasses 68% of the population
  • The Masimo SET Technology has been validated for low perfusion accuracy in bench top testing against a Biotek র্ণ Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation which encompasses 68% of the population.

Environmental Testing

Applicable environmetal testing per the Reviewers Guidance for Premarket Submissions - November 1993, i.e. electrical, mechanical and environmental were performed and all tests passed.

Nonclinical tests performed that support a determination of substantial equivalence.

The Masimo SET® Intellivue Pulse Oximeters Module was subjected to bench testing using a simulator that determined the performance accuracy of the instruments against the range of saturation and pulse rates that both devices specify.

The results of the bench testing showed that the Masimo SET® Intellivue Pulse Oximeter Module returned the same saturation accuracy values within ± 2 digits and pulse rate values within ± 3 digits when compared to the simulators used.

Clinical tests performed that support a determination of substantial equivalence.

Clinical studies were performed using the Masimo SET® technology on healthy adult volunteer subjects during no motion and motion conditions who were subjected to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.

Clinical studies were performed using the Masimo SET® technology on healthy adult volunteer subjected to low perfusion conditions and to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.

Clinical studies were conducted following regulations under Title 21 of the Code of Federal Regulations (21CFR), Part 812 -Investigational Device Exemptions, Part 50 - Protection of Human Subjects and Part 56 - Institutional Review Boards.

The results from the clinical studies show that the Masimo SET® technology saturation accuracy values and pediatrics within ± 2 digits during no motion conditions and ± 3 digits during motion conditions when compared to the CO-Oximeter and the pulse rate accuracy values within ± 3 digits during no motion conditions and ± 5 digits during motion conditions when compared to the ECG.

The specified saturation accuracy from 70% - 100% for neonates is based on the results from clinical studies on adult volunteers and 1% was added to account for the properties of fetal hemoglobin to be within ± 3 digits during both motion and no motion conditions when compared to the CO-Oximeter, and the pulse rate accuracy values for neonates to be within ± 3 digits during no motion and ± 5 digits during motion conditions when compared to the ECG.

{4}------------------------------------------------

Conclusions

The results of the environmetal testing demonstrated that the Masimo SET® Intellivue Pulse Oximeter Module met the requirements of Reviewers Guidance for Premarket Submissions - November 1993.

The results of the bench testing demonstrates that the Masimo SET® Intellivue Pulse Oximeters Module met its performance requirements.

The results of the clinical testing demonstrates that the Masimo SET® Intellivue Pulse Oximeters its performance requirements during no motion and motion conditions and low perfusion conditions.

The non-clinical and clinical testing performed demonstrates that the Masimo SET® Intellivue Pulse Oximeters Module is safe, effective.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 2 2 2004

Mr. James Cronin Vice President, Regulatory Affairs/Quality Assurance Masimo Corporation 2852 Kelvin Ave. Irvine, CA 92614-5826

Re: K040259

Trade/Device Name: Masimo SET Intellivue Pulsc Oximeter Module Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA, DPZ Dated: February 2, 2004 Reccived: February 4, 2004

Dear Mr. Cronin:

We have reviewed your Section 510(k) premarket notification of intent to market the device wf nave roviewed your becamined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use autod in the encreations of the enactment date of the Medical Device Amendments, or to conniored prior to rial 2011a) 2011 devices that have been roomstined in assess approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The rod may, mercrove, include requirements for annual registration, listing of general controls provisions of ractice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your de rice is exassino (sional controls. Existing major regulations affecting your device can may be subject to back adden addess, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act that I Dri has Intaliations and regulations administered by other Federal agencies. You must or any I edolia surated and sequirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{6}------------------------------------------------

Page 2 - Mr. James Cronin

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Actily of in veur Section S product radiation control provisions (occiolis 501 Clescribed in your Section 510(k)
This letter will allow you to begin marketing your device of your device to legal This letter will anow you to begin manxoning your antial equivalence of your device to a legally premarket notification. The PDA micing of basification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac rise of the more the regulation entitled, Contiact the Office of Compullios at (001) of Station" (21CFR Part 807.97). You may obtain Misbrailding by terefected to premarket nother the Act from the Division of Small other geleral information on your responsible at its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at the local development treal Manufacturers, International and Consembers: http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Chris Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Section 3 - Indications for Use

510(k) Number (if known):

Masimo SET Intellivue Pulse Oximeter Module Device Name:

Indications For Use:

The Masino SET® Intellivue Pulse Oximeter Module is indicated for the continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The Masimo SET® Intellivue Pulse Oximeter is indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities.

Prescription Use _ X (Per 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use J (Per 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NE

Concurrence of CDRH, Office of Device Evaluation (ODE)

seth

(Owision Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices

Int., Dental Devices
510(k) Number: K040756

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).