(262 days)
Systems are intended for use with a hemodialysis system to remove organic and inorganic substances resulting in contaminants from water used to dilute dialysis concentrate to form dialysate, reprocessing of hemodialyzers, equipment rinse and disinfection.
The Acute PEDI System consists of a sediment filter, carbon filtration, mixed bed deionizers, and system monitoring with remote alarm functions. It is intended for use in a location with a portable dialysis machine for the dialysis of a single patient.
The Central PEDI System consists of a booster pump / pressure tank assembly, sediment filter, carbon filtration, mixed bed deionizers, resin trap filter, endotoxin filter and system monitoring with flow isolation/divert and remote alarm functions. It is intended for use in a central location for the dialysis of multiple patients.
The Back-up PEDI System consists of mixed bed deionizers, intermediate water quality (QC) indicator (light) and connection hoses. It is intended for use to provide a "back-up" for reverse osmosis filtration and system monitoring with flow isolation / divert and remote alarm functions are provided when required to meet ANSI/AAMI RD62:2001 requirements.
The PEDI systems are primary or temporary devices to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001.
The systems consist of a booster pump / pressure tank assembly, sediment filter, carbon filtration, mixed bed deionizers, resin trap filter, endotoxin filter and system monitoring with flow isolation/divert and remote alarm functions. Refer to Operating Manuals for flow diagrams.
The booster pump / pressure tank assembly (central system only) consists of a multistage pump, and pressure tank. The multistage pump construction has minimum 304 SST wetted parts with no copper or iron wetted parts. The pressure (bladder) tank construction is an airtight shell inside of coil rolled steel with a 1 27 approved butyl rubber diaphragm to separate water from tank inside. The assembly requires 115V/1P/60Hz power to operate a multistage pump to raise pressure up to 80 psig.
The sediment filter consists of a polypropylene housing with pressure relief valve and a replaceable filter cartridge. The filter cartridge consists of spun polypropylene fiber wound on a polypropylene core and with a filtration rating of 10 microns. (Back-up system uses existing sediment filtration).
The carbon filtration consists of ABS lined fiberglass pressure tank containing granular activated carbon and PVC distributors. The size and number of tanks is dependent on flowrate required with carbon and PVC distributors listed in Appendix C. Back-up system uses facilities existing carbon filters if available, Acute system single tank.
The mixed bed deionizers consist of ABS lined fiberglass pressure tank with PVC distributors containing mixed bed ion exchange resin. The size and number of tanks is dependent on flowrate required and inlet water conductivity maximum values listed in Operating Manual.
The resin trap filter consists of a polypropylene housing with pressure relief valve and a replaceable filter cartridge. The filter cartridge consists of spun polypropylene fiber wound on a polypropylene core and has a filtration rating of 5 microns. (Not used on acute systems).
The endotoxin filter consists of a polypropylene housing with a replaceable filter cartridge. The filter cartridge consists of an element with 222 o-ring/closed type seals and has a reduction factor of 10' for bacteria and 10 for endotoxins.
System monitoring consists of pressure gauges, quality control indicator and resistivity monitor. Pressure gauges prior to deionization minimum brass body with phosphorous bellows, gauge after deionization 304 SST stem with SST bellows. The resistivity monitor has a digital readout, indicators for above and below water quality, push buttons for alarm setpoint indication and alarm function test. Connections to the monitor provide under low water quality: remote alarm (horn), isolation of flow to use (valve) and diversion of flow to drain (valve). Inline sensor construction stainless element with polypropylene body. Divert to drain not supplied for acute systems located in patient room. Monitor and divert to drain not supplied for back-up systems with existing monitor/alarm/divert. Refer to Operating Manuals for Control Diagram.
This 510(k) summary (K042018) does not describe the acceptance criteria and the study that proves the device meets the acceptance criteria. It focuses on establishing substantial equivalence with a predicate device and describing the device's components and indications for use.
Therefore, I cannot provide the requested information from the provided text. The document does not contain:
- A table of acceptance criteria and reported device performance.
- Sample sizes used for test sets or their data provenance.
- Number of experts used to establish ground truth or their qualifications.
- Adjudication method for test sets.
- Information on a multi-reader multi-case (MRMC) comparative effectiveness study or effect sizes.
- Details on a standalone algorithm performance study.
- The type of ground truth used.
- Sample size for the training set.
- How ground truth for the training set was established.
The document is a regulatory submission for a water purification system for hemodialysis, which typically involves performance testing against industry standards (e.g., ANSI/AAMI RD62:2001 for water quality parameters) rather than clinical studies with expert readers or ground truth establishment in the way AI/imaging devices might. The "performance" in this context would likely refer to water quality parameters (e.g., conductivity, removal of specific contaminants) which are not detailed as specific acceptance criteria or study results here beyond general compliance with RD62:2001.
{0}------------------------------------------------
APR 1 5 2005
KO42018 1 :F 2 PAGE 000009
510(k) Summary
This 510(k) Summary is provided as part of this Premarket Notification to comply with the I mis J10(k) Summary is provided in part of this of 1990 requiring that either a summary be included in a provision or a statement that a summary is available upon request.
Submitter Raymond Kowalec Simply Clean Air & Water FDA Establishment No. 3004114902 28 Shepard Drive P.O. Box 310962 Newington, CT 06131-0962 Tel: (860) 231-0687 Fax:(860) 231-1425 March 4, 2005
Device Names Acute Portable Exchange Deionization (PEDI) System Central PEDI System Back-up PEDI System
Common or usual Name Connization system with pre & post treatment and water distribution components.
Classification Name Water purification systems for hemodialysis (21CFR 876.5665)
Predicate Device
The Simply Clean Air & Water PEDI Systems are substantially equivalent to US Filter Corporation's (K980182) predicate marketed water treatment systems for dialysis which use carbon and deionization canisters with pre and post filtration to purify water for hemodialysis.
Device Description
The PEDI systems are primary or temporary devices to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62:2001 .
The systems consist of a booster pump / pressure tank assembly, sediment filter, carbon filtration, mixed bed deionizers, resin trap filter, endotoxin filter and system monitoring with flow isolation/divert and remote alarm functions. Refer to Operating Manuals for flow diagrams.
The booster pump / pressure tank assembly (central system only) consists of a multistage pump, The 000ster punp / pressure with. The multistage pump construction has minimum 304 SST wetted
{1}------------------------------------------------
KCA-2C:8
PAGE 2 of 2
000010
parts with no copper or iron wetted parts. The pressure (bladder) tank construction is an airtight shell patis with no copper of non world parks. Subtragm to separate water from tark inside of coul folled steel with a 1 27 upploved battle tax proved starsing 115V/1P/60Hz power to operate a multistage pump to raise pressure up to 80 psig.
The sediment filter consists of a polypropylene housing with pressure relief valve and a replaceable I he scumical the filter cartridge consists of spun polypropylene fiber wound on a polypropylene core and with a filtration rating of 10 microns. (Back-up system uses existing sediment filtration).
The carbon filtration consists of ABS lined fiberglass pressure tank containing granular activated carbon and PVC distributors. The size and number of tanks is dependent on flowrate required with carbon and I VC dictricated in Appendix C. Back-up system uses facilities existing carbon filters if available, Acute system single tank.
The mixed bed deionizers consist of ABS lined fiberglass pressure tank with PVC distributors containing mixed bed ion exchange resin. The size and number of tanks is dependent on flowrate required and inlet water conductivity maximum values listed in Operating Manual.
The resin trap filter consists of a polypropylene housing with pressure relief valve and a replaceable filter cartridge. The filter cartridge consists of spun polypropylene fiber wound on a polypropylene core and has a filtration rating of 5 microns. (Not used on acute systems).
The endotoxin filter consists of a polypropylene housing with a replaceable filter cartridge. The filter cartridge consists of an element with 222 o-ring/closed type seals and has a reduction factor of 10' for bacteria and 10 for endotoxins.
System monitoring consists of pressure gauges, quality control indicator and resistivity monitor. Pressure gauges prior to deionization minimum brass body with phosphorous bellows, gauge after deionization 304 SST stem with SST bellows. The resistivity monitor has a digital readout, indicators for above and below water quality, push buttons for alarm setpoint indication and alarm function test. Connections to the monitor provide under low water quality: remote alarm (horn), isolation of flow to use (valve) and diversion of flow to drain (valve). Inline sensor construction stainless element with polypropylene body. Divert to drain not supplied for acute systems located in patient room. Monitor and divert to drain not supplied for back-up systems with existing monitor/alarm/divert. Refer to Operating Manuals for Control Diagram.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows a logo with a stylized bird in flight. The bird is composed of thick, curved lines, giving it a modern and abstract appearance. The logo is encircled by text, which is partially obscured but appears to follow the curve of the circle. The overall design is simple yet recognizable, likely representing a company or organization associated with flight, freedom, or a similar concept.
APR 1 5 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Raymond Kowalec Simply Clean Air & Water 28 Shepard Drive P.O. Box 310962 NEWINGTON CT 06131-0962
Re: K042018
Trade/Device Name: Acute PEDI System, Central PEDI System, and Back-up PEDI System Regulation Number: 21 CFR §876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: FIP Dated: March 4, 2005 Received: March 22, 2005
Dear Mr. Kowalec:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter will and in yourse FDA finding of substantial equivalence of your device to a legally prematication: " The PDF missification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrree for your and of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the 001:37). " Ou may overnment and Consumer Assistance at its toll-free number (800) DV liston or other Manazactares ins Internet address http://www.fda.gov/cdrh/dsmamain.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
K042018 510(k) Number (if known):
Device Name:
Acute PEDI System Central PEDI System Back-up PEDI System
Indications For Use:
Systems are intended for use with a hemodialysis system to remove organic and inorganic substances Dystelling in contaminants from water used to dilute dialysis concentrate to form dialysate, reprocessing of hemodialyzers, equipment rinse and disinfection.
The Acute PEDI System consists of a sediment filter, carbon filtration, mixed bed deionizers, The route I LL roystem monitoring with remote alarm functions. It is intended for use in a location with a portable dialysis machine for the dialysis of a single patient.
The Central PEDI System consists of a booster pump / pressure tank assembly, sediment filter, I he Contain I Do reystem conson or a resin trap filter, endotoxin filter and system monitoring with Caroon million/divert and remote alarm functions. It is intended for use in a central location for the dialysis of multiple patients.
The Back-up PEDI System consists of mixed bed deionizers, intermediate water quality (QC) indicator (light) and connection hoses. It is intended for use to provide a "back-up" for reverse marcator (tight) and connish" the product outlet of a RO to meet AAMI RD62:2001 requirements. osmobile filtration and system monitoring with flow isolation / divert and remote alarm functions are provided when required to meet ANSI/AAMI RD62:2001 requirements.
Over-The-Counter Use AND/OR X Prescription Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ylance Bradin
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K042018
§ 876.5665 Water purification system for hemodialysis.
(a)
Identification. A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.(b)
Classification. Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.