(30 days)
For use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature.
Utilizing a proprietary process, LRM produces a PTFE coated stainless steel steerable core. The proximal portion of the core wire is coated with PTFE to provide lubricity and improve wire handling. A platinum alloy coil, placed within the confines of the outer coil and over-wrapping the distal core section, provides radiopacity in the distal tip. The coils are secured in their location by solder, which is attached to the core. The proximal end of the outer coil is attached to the core with braze, and the distal end has a mid-point of the outer coil which is attached to the core with solder (applicable designs). The product is offered with a shapeable straight tip or in a preshaped configuration. The guidewires are coated with MDX (silicone) or with hydrophilic coating. The guidewires are bound by the following parameters: Outside Diameter: .014", Lengths: 175cm - 300cm, Tips: Straight or shaped with various tip flexibilities, Flexibility: Floppy to Extra Support.
Here's a breakdown of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Feature/Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Visual/Tactile Attributes | Met established specific inspection criteria requirements. | The results of these tests demonstrated the functionality and performance characteristics of these guidewires are comparable to the currently marketed devices. |
| Dimensional Attributes | Met established specific inspection criteria requirements. | The results of these tests demonstrated the functionality and performance characteristics of these guidewires are comparable to the currently marketed devices. |
| Mechanical Attributes | Met established specific inspection criteria requirements. | The results of these tests demonstrated the functionality and performance characteristics of these guidewires are comparable to the currently marketed devices. |
| Hydrophilic Properties | Met criteria established for the hydrophilic properties of the guidewire (in addition to basic design criteria). | The results of these tests demonstrated the functionality and performance characteristics of these guidewires are comparable to the currently marketed devices. (Implied that hydrophilic properties were also found comparable given the overall statement). |
| Functional Equivalence | Device functionality and performance characteristics must be comparable to currently marketed devices. | The results of these tests demonstrated the functionality and performance characteristics of these guidewires are comparable to the currently marketed devices. |
| Biocompatibility | Acceptable biocompatibility for the device. | Biocompatibility testing has been performed on the material components of this device. This testing, along with a market history of proven biocompatibility, establishes acceptable biocompatibility for this device. |
| Design Control (21 CFR 820.30) | Conformance with design control procedure requirements. | LRM is in conformance with the design control procedure requirements as specified in 21 CFR 820.30. |
| Risk Analysis (FMEA) | All verification and validation activities result in the ability to demonstrate that the predetermined acceptance criteria were met. | Risk analysis was completed by means of a Failure Mode and Effect Analysis (FMEA) and all verification and validation activities resulted in the ability to demonstrate that the predetermined acceptance criteria were met. |
| Material/Product/Process Qualification | Assure that each product manufactured with the hydrophilic coating remains equivalent to the predicate products, and that the changes will not have an adverse affect on the safe and effective use of the product. | LRM has formal quality systems in place to assure that each product manufactured with the hydrophilic coating remains equivalent to the predicate products, and that the changes will not have an adverse affect on the safe and effective use of the product. The quality systems include Engineering Change Order Review, Material Qualification, Product Qualification, and Process Qualification. These controls are applied to each product size/group. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document does not specify an exact numerical sample size for the test set used in non-clinical tests. It refers to "Test pieces" being tested.
- Data Provenance: The data provenance is not explicitly stated. The testing was conducted by Lake Region Manufacturing, Inc. (LRM), based in Chaska, MN, USA. It is implied to be internal company testing and is retrospective to the submission date.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- This information is not applicable to this submission. The "acceptance criteria" and "ground truth" discussed here refer to objective engineering and material performance tests for a medical device (guidewire), not a diagnostic or AI-driven interpretative device that would require expert consensus for ground truth.
4. Adjudication Method for the Test Set
- This information is not applicable. Adjudication methods are relevant for subjective assessments, typically found in clinical trials or diagnostic studies where multiple experts interpret data. The tests described are objective non-clinical performance and material tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- No, a MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of an AI system on human reader performance, which is not the scope of this guidewire submission.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- No, a standalone algorithm performance study was not done. This submission is for a physical medical device (guidewire), not an algorithm or AI system.
7. The Type of Ground Truth Used
- The "ground truth" for the test set was based on established specific inspection criteria requirements for visual/tactile, dimensional, and mechanical attributes, as well as criteria established for the hydrophilic properties of the guidewire. For biocompatibility, it was based on biocompatibility testing performed on material components and a market history of proven biocompatibility. For design control and risk analysis, it was based on conformance with 21 CFR 820.30 and results of FMEA verification and validation activities. For material/product/process qualification, it was based on achieving equivalence to predicate products and ensuring no adverse effect on safe and effective use.
8. The Sample Size for the Training Set
- This information is not applicable. This submission describes the testing of a physical medical device (guidewire) to demonstrate substantial equivalence, not the development or training of an AI algorithm.
9. How the Ground Truth for the Training Set Was Established
- This information is not applicable. As stated above, this is not an AI/algorithm submission.
{0}------------------------------------------------
JUL 1 6 2004
SECTION 2.0 - SUMMARY OF SAFETY AND EFFECTIVENESS
June 15, 2004
2.1 General Information
- 2.1.1 Company Name, Address, and Telephone Number
Lake Region Manufacturing, Inc. (LRM) 340 Lake Hazeltine Drive Chaska, MN 55318 Telephone: (952) 448-5111 Fax: (952) 448-3441
Contact Name: Karen Mortensen Senior Regulatory Compliance Specialist
2.1.2 Device Trade Name/Proprietary Name
LRM produces guidewires on an OEM basis for other manufacturers, kit assemblers, and distributors. Consequently there are a large number of trade and proprietary names not including or associated with LRM. LRM has no proprietary names of its own to be included with this submission.
-
2.1.3 Device Common Names/Usual Names and Classification Names
These devices are commonly known as guides, guidewires, or spring guidewires. The current classification names, and product codes are Angiographic Guidewire (74HAP), Catheter Guidewire (74DQX), and Radiological Catheter Guidewire (74JAJ). -
2.1.4 Establishment Registration Number: 2126666
-
2.1.5 Classification of Devices
The classification names listed above were originally classified as Class II devices by the Neurology (84HAD), Cardiovascular (74DOX) and Radiology (90JAJ) Review Panels, respectively.
- 2.1.6 Applicability of Performance Standards
LRM has determined that no mandatory performance standards have been established for these devices under Section 514 of the Medical Amendments to Federal Food, Drug, and Cosmetic Act or by any subsequent regulatory action. LRM has also determined that there are no applicable voluntary standards.
{1}------------------------------------------------
-
2.2 Labels, Labeling, and Advertising
LRM produces cardiovascular and vascular guidewires on an OEM basis for other manufacturers, kit assemblers, and distributors. There is no direct distribution by Changes to the customer controlled labels, labeling, or promotional LRM. material are at their discretion, including the resolution of any resulting regulatory obligations. A fraction of the total production bears LRM controlled labels and labeling. -
2.3 Statement of Availability
This summary is being included in the Premarket Notification submission in lieu of a statement of availability. -
2.4 Device Description
- Utilizing a proprietary process, LRM produces a PTFE coated stainless 2.4.1 steel steerable core. The proximal portion of the core wire is coated with PTFE to provide lubricity and improve wire handling. A platinum alloy coil, placed within the confines of the outer coil and over-wrapping the distal core section, provides radiopacity in the distal tip. The coils are secured in their location by solder, which is attached to the core. The proximal end of the outer coil is attached to the core with braze, and the distal end has a mid-point of the outer coil which is attached to the core with solder (applicable designs). The product is offered with a shapeable straight tip or in a preshaped configuration. The guidewires are coated with MDX (silicone) or with hydrophilic coating. The guidewires are bound by the following parameters:
| Outside Diameter: | .014" |
|---|---|
| Lengths: | 175cm - 300cm |
| Tips: | Straight or shaped with various tip flexibilities |
| Flexibility: | Floppy to Extra Support |
- 2.4.2 Engineering Specifications
The design specifications are the same for the proposed device as they are for the LRM predicate devices [reference 510(k)s K970376 and K011968]. The finished device with hydrophilic coating must meet the same basic design criteria and also meet the criteria established for the hydrophilic properties of the guidewire.
{2}------------------------------------------------
2.5 Substantial Equivalence Data
Background Information 2.5.1
The table below lists the differences between the predicate device and the proposed device. Testing was done to ensure the changes to the device met the predetermined acceptance criteria.
| Item | Proposed Device Differences from LRM Predicate cleared under510(k) K970376 |
|---|---|
| Raw Materials | Core:No change to raw materialCoil:No change to raw materialCoating:Addition of hydrophilic coating as alternate coating(reference K991898 Microvena Corporation HyTekTMGuidewire for predicate of a legally marketed device withinthe same classification for the same intended use thatemploys the hydrophilic coating in its design.) |
| Assembly Process | No change to assembly processes |
| Physical Characteristics | No change |
| Labeling/IFU | The only change to the label or IFU will be wording specific to thehydrophilic properties of the guidewire. |
| Intended Use | No change to intended use |
| Anatomical Sites | No change |
| Target Population | No change |
| Performance Testing | No change |
| Safety Characteristics | No change |
| Biocompatibility | No change |
| Risk Analysis | No change |
2.5.2 Non-Clinical Tests
In order to demonstrate equivalence of the proposed device, LRM performed testing to established requirements. Test pieces were tested and inspected according to established specific inspection criteria requirements for visual/tactile, dimensional and mechanical attributes. The results of these tests demonstrated the functionality and performance characteristics of these guidewires are comparable to the currently marketed devices.
2.6 Qualification and Biocompatibility Test Data
- 2.6.1 Design Control
LRM is in conformance with the design control procedure requirements as specified in 21 CFR 820.30. Risk analysis was completed by means of a Failure Mode and Effect Analysis (FMEA) and all verification and validation activities resulted in the ability to demonstrate that the predetermined acceptance criteria were met.
{3}------------------------------------------------
2.6.2 Material/Product/Process Qualification
LRM has formal quality systems in place to assure that each product manufactured with the hydrophilic coating remains equivalent to the predicate products, and that the changes will not have an adverse affect on the safe and effective use of the product. The quality systems include Engineering Change Order Review, Material Qualification, Product Qualification, and Process Qualification. These controls are applied to each product size/group.
2.6.3 Biocompatibility Testing
Biocompatibility testing has been performed on the material components of this device. This testing, along with a market history of proven biocompatibility, establishes acceptable biocompatibility for this device.
- 2.7 Packaging and Sterilization Information
LRM produces guidewires on an OEM basis for other manufacturers, kit assemblers, and distributors. There is no direct distribution by LRM. A portion of the production is private label, sterile packaged to customer specifications, a fraction of that product is provided sterile to the customer.
The single packaged hydrophilic coated guidewire is placed in a dispenser and then into a Tyvek®/poly pouch. The packaged product may be packaged as five or ten pouchs in a shelf carton, which are typical packaging configurations.
There will be no changes to the sterilization process for the portion of packaged product shipped sterile to the customer. For the product that is shipped bulk, the packaging design and sterilization process parameters are the customer's responsibility. LRM will not recommend that its customers modify their packaging or sterilization procedures as a result of this submission.
- 2.8 Intended Use Statement
For use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature.
NOTE: The modification of this device does not alter its intended use.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with outstretched wings, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is simple, using black ink on a white background.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 6 2004
Lake Region Manufacturing, Inc. c/o Ms. Karen Mortensen Senior Regulatory Compliance Specialist 340 Lake Hazeltine Drive Chaska, MN 55318-1029
Re: K041624 Trade Name: PTCA Steerable Hydrophilic Guide Regulation Number: 21 CFR 870.1330 Regulation Name: Guide, Catheter Regulatory Class: Class II (two) Product Code: DQX Dated: June 15, 2004 Received: June 16, 2004
Dear Ms. Mortensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Ms. Karen Mortensen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Duna R. Kirchner
(J) Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name:_ PTCA Steerable Hydrophilic Guidewire
Indications For Use:
For use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dmna R. Vúchues
sion Sign-Off) sion of Cardiovascular Devices
(k) Number K041624
§ 870.1330 Catheter guide wire.
(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.