K Number
K034057
Device Name
LACRYTEST
Manufacturer
Date Cleared
2004-03-09

(70 days)

Product Code
Regulation Number
866.5510
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Lacrytest is a rapid immunoassay for the detection of total IgE in tears, Semi-quantitative detection of total IgE In tears (< 2.5 KIU/liter, 2.5 - 10 kIU/liter, 10 - 40 klU/liter, and > 40 klU/liter) indicates local IgE production associated with allergic conjunctivitis. This test is used in the physician office and professional clinical laboratories.

Device Description

Lacrytest (marked CE) allows the rapid and specific determination of total IgE in tears (in vivo/in vitro). This test is a simple and rapid immuno-chromatographic test, which enables the identification of allergic conjunctivitis in one single step. Its main advantage is that it can be done in vivo without being invasive. The speed and simplicity of this unique test enables a better and faster treatment of the patient. Lacrytest is a ready-to-use strip, with all the necessary reagents already incorporated into it. No instrumentation or prior provision of tear samples is required: Lacrytest uses monoclonal and polyclonal antibodies to detect total IgE in tears. Lacrytest gives results from just one teardrop, and is placed in contact with the patient's eye for less than three minutes, causing minimum discomfort when compared to the Schirmer assay. Doctors do not have to take tear sample, and an indicator appears on the strip when enough tear fluid has been absorbed. The test strip is then soaked with sterile water and the evaluation of amount of total IgE detected. The results appear in the form of easily visible red-violet lines in the presence of total IgE levels greater than 2,5 KUI/L (the detection limit). These results are stable and can be archived. The test can be used in the laboratory, in the doctor's surgery or on a home visit.

AI/ML Overview

Here's an analysis of the provided text to extract the requested information about the Lacrytest device:

Acceptance Criteria and Device Performance

The acceptance criteria for the Lacrytest device are implicitly established by comparison to the predicate device, UniCAP total IgE FEIA Assay. The document primarily focuses on demonstrating comparable sensitivity and specificity.

Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Implied by Predicate UniCAP)Reported Device Performance (Lacrytest)
SensitivityNot explicitly stated but UniCAP is 91.7%91.75% (comparable to UniCAP)
SpecificityNot explicitly stated but UniCAP is 98.5%98.53% (equal to UniCAP)
Detection Limit< 2 kIU/L2.5 kIU/L
Precision (Within-assay CV)Low levels: 2.8% (at 13 kIU/L) Mid/High levels: 2.0% (at 75 kIU/L), 2.8% (at 640 kIU/L)Low levels: 10% (at 5 kIU/L) Mid/High levels: 4% (at 20 kIU/L), 0% (at 50 kIU/L)
Precision (Between-assay CV)Low levels: 8.9% (at 13 kIU/L)Not explicitly provided in the same format for Lacrytest
Agreement of Negative ResultsNot explicitly stated for UniCAP87.79%
Agreement of Positive ResultsNot explicitly stated for UniCAP84.21%
Accuracy of Semi-quantitative ResultsNot explicitly stated for UniCAP100% (except near Intensity 2 upper limit)
Normal Healthy Subjects < 2.5 kIU/LNot explicitly stated for UniCAP87%

Note on Acceptance Criteria: The document describes the Lacrytest as having "slightly lower sensitivity than UniCAP and an equal specificity." This implies that the acceptance criteria for sensitivity were flexible enough to accommodate a minor difference. The precision data comparison also highlights areas where Lacrytest's performance differs (e.g., higher CV at low levels).

Study Details

2. Sample size used for the test set and the data provenance

  • Sample Size: 165 human serum samples were used for comparative studies against UniCAP.
  • Data Provenance: The document does not explicitly state the country of origin for these 165 human serum samples, nor does it specify if the data was retrospective or prospective. It only mentions "human serums."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • The document does not mention the use of experts to establish ground truth for the test set in the context of the comparative study. The ground truth appears to be based on the results obtained from the predicate device, UniCAP.

4. Adjudication method for the test set

  • No adjudication method is described for the test set. The comparison is directly made against the UniCAP results. The "human reading of the device" is cited as a factor for some agreement differences, but no specific adjudication process for discrepancies is detailed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study was done. This device is a rapid immunoassay test, not an AI-assisted diagnostic tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • The Lacrytest is a standalone immuno-chromatographic test. Its performance metrics (sensitivity, specificity, precision) are measured for the device itself.
  • However, the interpretation of the results involves "human reading," as noted for the agreement of results ("mainly explained by the human reading of the device"). This suggests that while the test is standalone, its ultimate performance in practice can be influenced by human interpretation of the visual lines.

7. The type of ground truth used

  • The primary ground truth for the performance study (sensitivity and specificity) was established by comparison to the UniCAP total IgE FEIA Assay, which served as the reference method. This is a form of reference standard comparison.
  • For the "normal healthy subjects" section, the ground truth was presumably based on clinical diagnosis or established healthy status.

8. The sample size for the training set

  • The document does not mention a "training set" as would be relevant for machine learning algorithms. The performance data provided is for the device as evaluated, implying a validation or verification set against a reference. This is a traditional immunoassay, not an AI-based system that undergoes a separate training phase.

9. How the ground truth for the training set was established

  • As there is no mention of a training set, the establishment of ground truth for such a set is not applicable to the information provided.

{0}------------------------------------------------

K034057

MAR - 9 2004

SECTION 7

Premarket notification summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of SMDA 1990 21 CFR 807.92

The assigned 510(k) number is:

Date of summary preparation: Submitter's name: Address:

Phone: Fax : Contact : Device name: Trade name: Classification name:

July 15 2002 Mr. Joël Lesser 1 Place Alexis Ricordeau 44 093 NANTES Cedex 1 FRANCE 33 (0)2 51 72 11 40 33 (0)2 51 72 37 73 Mr. Joël Lesser Lacrytest

Lacrytest Immunological test

7- PREMARKET NOTIFICATION 510(K) SUMMARY. .

7-1 Predicate device.

Device Classification NameIGE, ANTIGEN, ANTISERUM,CONTROL
Regulation Number866.5510
510(K) NumberK964152
Device NameUNICAP TOTAL IGEASSAY/PHARMACIA TOTAL IGECONTROL

{1}------------------------------------------------

ADIATEC SA Company asserts that the UniCAP total IgE FEIA Assay is substantially equivalent to the Lacrytest. Each product is designed to measure the level of total IgE. Each product is an Immuno Assay.

7-2 Intended use.

Lacrytest is a rapid immunoassay for the total IgE detection in tears. Normal values are below 2,5 kUI/l, (3 ng/ml). This assay indicates in a qualitative way the presence of total class E immunoglobulins in tears, with level above normal value.

Lacrytest offers help for ocular allergy diagnosis in a simple and fast way.

Lacrytest is to be used with IgE mediated allergic ocular disorders.

It allows knowing if an ocular inflammation, conjunctivitis, or dry eyes symptom could be linked to allergic ocular disorder.

This test is dedicated to the physicians and clinical Laboratory.

7-3 General description

LACRYTEST (marked CE) allows the rapid and specific determination of total IgE in tears (in vivo/in vitro).

This test is a simple and rapid immuno-chromatographic test, which enables the identification of allergic conjunctivitis in one single step. Its main advantage is that it can be done in vivo without being invasive. The speed and simplicity of this unique test enables a better and faster treatment of the patient.

7-4 Description of the device.

Lacrytest, a simple and rapid immuno-chromatographic test, enables the identification of allergic conjunctivitis in one single step. The speed and simplicity of this unique test enables better and faster care and treatment of the patient.

Lacrytest is a ready-to-use strip, with all the necessary reagents already incorporated into it. No instrumentation or prior provision of tear samples is required: Lacrytest uses monoclonal and polyclonal antibodies to detect total IgE in tears. Lacrytest gives results from just one teardrop, and is placed in contact with the patient's eye for less than three minutes, causing minimum discomfort when compared to the Schirmer assay. Doctors do not have to take tear sample, and an indicator appears on the strip when enough tear fluid has been absorbed. The test strip is then soaked with sterile water and the evaluation of amount of total IgE detected.

The results appear in the form of easily visible red-violet lines in the presence of total IgE levels greater than 2,5 KUI/L (the detection limit). These results are stable and can be archived. The test can be used in the laboratory, in the doctor's surgery or on a home visit.

{2}------------------------------------------------

7-5 Qualitative constitution of the reagent:

Chemical products:

  • · trihydrated tetrachloroauric acid (III): HAu Cl4 3H20
  • · Natrium Azide: NaN3
  • · Polyethylen glycol 20000: HO (C2H4O)n H
  • · Hydrolyzed Polyvinyl Alcohol, 80%
  • · Sucrose 98%: D(+) saccharose
  • · Thimérosal to 0,001%

Materials:

  • · Nitrocellulose
  • · Polyester resin fibers,
  • · Cellulose,
  • · Adhesive band
  • · The band in contact with the eye is a membrane constituted of polyester fibers and a surfactant. Those two components, "Pall Profile Filter" and "Bioinert Filter", are unprovided with toxicity as specified in the toxicity tests certificates enclosed in the annex.

Biological products:

  • · Monoclonal Antibodies and mouse and goat polyclonal antibodies, purified with affinity chromatography, specific of type E human immunoglobulins (hIgE) and type G.mouse immunoglobulins
  • · Cream milk Gloria Nestlé
  • · Purified Bovine Serum Albumin: animal origin

7- 6 Performance equivalence: Technology comparison.

uses the ELISA method whereas ADIATEC develops immuno-UniCAP chromatographic assay. UniCAP can give total IgE levels accurate measurement and uses 48 different parameters.

Lacrytest is a ready-to-use strip, with all the necessary reagents already incorporated into it. No instrumentation or prior provision of tear is necessary. The total IgE level is given on the strip.

Contrary to that, UniCAP needs blood sampling and so human serum, or a high volume of tears. The results are given with UniCAP 100 Instrument which is programmed to automatically calculate data from UniCAP total IgE assay. The total IgE values for controls and patient samples are automatically calculated and printed out

  • Like the LACRYTEST, UniCAP total IgE Assay is an in vitro Assay for the measurement of total IgE. An other common point is the fact that the both devices are to be used in clinical laboratories.

{3}------------------------------------------------

But contrary to UniCAP, Lacrytest is furthermore to be used in medical offices.

7-7 Performance Equivalence-method comparison.

  • Comparative studies, carried out on 165 human serums with reference to the UniCAP, have shown that it has a sensitivity of 91.7% and a specificity of 98.5%.

  • We can make a parallel between the precision of the Lacrytest and the precision of the UniCAP:

Precision. Lacrylest
Intra-runLO Ll12ﺩ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍ
Number of determinations30303030
Total IgE levels KUI/L02050
Ranges (intensities)00 - 1 2 - 34
Mean value (intensity)0
Variance %O
Inter-runL0L1L2L3
Number of determinations20202020
Total IgE levels KUI/L052050
Ranges (intensities)00 - 12 - 33 - 3
Mean value (intensity)0123
Variance %01040

Procision: I acruast

Precision: UniCAP

Serum levelCoefficients of variation (%)
KIU/IWithin assayBetween
132.88.9
752.05.3
6402.89.1

The coefficient of variation of UniCAP is similar at low levels (13 kIU/l) than high levels. Lacrytest has a higher coefficient of variation for low levels (5 kIU/l), but its coefficient of variation is similar for level of 20 kIU/1.

  • We can make a parallel between the sensitivity and specificity of the Lacrytest and those of the UniCAP:

Lacrytest has a detection limit of 2,5 kFU/1 and the UniCAP detection limit is < 2 kIU/L.

UniCAP mean recovery is 98+/-5%

The sensitivity of Lacrytest is 91.75% and its specificity 98.53% Then, Lacrytest has a slightly lower sensitivity than UniCAP and an equal specificity.

{4}------------------------------------------------

7-8 Quality of semi-quantitative Lacrytest results:

Agreement of negative results is 87,79% and agreement of positive results is 84,21%, mainly explained by the human reading of the device.

Accuracy determination of semi-quantitative results is of 100% except near the upper limit of the Intensity 2 (30 kIU/l) where upper estimation can occur.

7-9 Comparing results obtained by professional Lab technicians with persons at the point of care facilities:

No significant differences occurred between the different sites regarding the negative samples.

Some differences have been observed concerned the agreement of IgE concentration categories, mainly concerning an upper estimation of the IgE levels

7-10 Percentage of normal healthy subjects with a Lacrytest response less than 2.5kIU/liter

87% of normal healthy subjects give a Lacrytest response negative, i.e. less than 2,5 kIU/l. 13% of normal healthy subjects give an Intensity 1 Lacrytest results (low positive).

Further studies are in course to determine the prognostic value of this Lacrytest range intensity regarding the presence of an allergic field.

7-11 6-9-13 Manufacturing information :

Manufacturing quality control are performed on final packaging strips using the NIBSC Standard 75/502.

Samples of each batch are kept in room temperature storage for a follow up of the batches.

CONCLUSION

The data presented in the summary of safety and effectiveness is the data that the food and drug administration used in granting DPC substantial equivalence for the LACRYTEST.

Mr Joël LESSER

President of DIATEC SA

DECEMBER 19/2003
Dated

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling three stylized stripes or lines, possibly representing a stylized caduceus or a similar emblem associated with health and medicine.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Joel Lesser President Adiatec SA 5 Passage Douard 4Rue Des Olivettes Nantes France 44000

MAR - 9 2004

K034057 Re: Trade/Device Name: Lacrytest Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A, G, M, D, and E immunological test system Regulatory Class: Class II Product Code: DGC Dated: December 20, 2003 Received: December 30, 2003

Dear Mr. Lesser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Joseph L. Hackett

Joseph L. Hackett, Ph.D. Acting Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: LACRYTEST

Indications For Use:

Lacrytest is a rapid immunoassay for the detection র্তা total IgE in tears, Semi-quantitative detection of total IgE In tears (< 2.5 KIU/liter, 2.5 - 10 kIU/liter, 10 - 40 kl(J/liter, and > 40 klU/liter) indicates local IgE production associated with allergic conjunctivitis. This test is used in the physician office and professional clinical laboratories.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

ia chan

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K034057

Page 1 of

§ 866.5510 Immunoglobulins A, G, M, D, and E immunological test system.

(a)
Identification. An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.(b)
Classification. Class II (performance standards).