(146 days)
No
The summary describes a standard immunoassay for measuring parathyroid hormone levels and does not mention any AI or ML components. The performance studies focus on method comparison and analytical metrics, not AI/ML model performance.
No
This device is an immunoassay used to measure parathyroid hormone levels for diagnostic purposes and to aid in monitoring therapeutic intervention but does not directly provide therapy.
Yes
The "Intended Use / Indications for Use" section explicitly states that "Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia...and hypocalcemia...resulting from disorders of calcium metabolism." It also mentions its use "as an aid in monitoring therapeutic intervention of secondary hyperparathyroidism." These directly relate to diagnosing and monitoring medical conditions.
No
The device description explicitly states it is a "two-site chemiluminescence assay for use with the Nichols Advantage® Specialty System," indicating it is a physical assay kit used with a hardware system, not a standalone software device.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use explicitly states that the device is used to "measure the levels of parathyroid hormone in serum, EDTA plasma and heparinized plasma." It also describes how these measurements are used in the "differential diagnosis of hypercalcemia and hypocalcemia" and as an "aid in monitoring therapeutic intervention." These are all diagnostic purposes performed on biological samples in vitro (outside the body).
- Device Description: The device is described as a "two-site chemiluminescence assay," which is a common type of laboratory test performed on biological samples.
- Performance Studies: The document includes performance studies comparing the device to another assay using biological samples (serum and renal dialysis samples).
- Predicate Device: The predicate device listed is also an "Immunoassay," further indicating its use in laboratory testing of biological samples.
All of these points align with the definition and characteristics of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The Nichols Advantage® Bio-Intact PTH (1-84) immunoassay is intended for use with the Nichols Advantage Specialty System to measure the levels of parathyroid hormone in serum, EDTA plasma and heparinized plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Measurements of parathyroid hormone levels with the Bio-Intact PTH (1-84) Immunoassay are also used as an aid in monitoring therapeutic intervention of secondary hyperparathyroidism that frequently occurs in chronic kidney disease. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
Product codes (comma separated list FDA assigned to the subject device)
CEW
Device Description
The Bio-Intact PTH (1-84) Assay is a two-site chemiluminescence assay for use with the Nichols Advantage® Specialty System.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Precision within Run:
- Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay (K013992): Not greater than 4% at dose greater than 5 pg/mL
- Bio-Intact PTH (1-84) Assay (Modified): Not greater than 6% at a dose greater than 5 pg/mL
Precision Total:
- Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay (K013992): Not greater than 9.5% at dose greater than 34 pg/mL
- Bio-Intact PTH (1-84) Assay (Modified): Not greater than 11% at a dose greater than 5 pg/mL
Method Comparison 1 - Serum Analysis Nichols Advantage Intact PTH Assay Vs. Bio-Intact PTH (1-84) Assay:
- Sample Size: 305
- Range of Results:
- Nichols Advantage Intact PTH Assay: 5.0 to 1387 pg/mL
- Nichols Advantage Bio-Intact PTH (1-84) Assay: 3.0 to 746 pg/mL
- Passing Bablok Regression Equation: y = 0.66x - 0.6
- Least Squares Linear Regression Equation: y = 0.60x + 4.2
- Pearson's Correlation Coefficient (r): 0.97
Method Comparison 2 - Renal Dialysis Samples Nichols Advantage Intact PTH Assay Vs. Bio-Intact PTH (1-84) Assay:
- Sample Size: 3187
- Range of results:
- Nichols Advantage Intact PTH Assay: 7 - 1797pg/mL
- Nichols Advantage Bio-Intact PTH (1-84) Assay 4-998
- Least Squares Linear Regression Equation: Y = 0.52 x + 1.3
- Pearson's Correlation Coefficient (r): 0.97
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1545 Parathyroid hormone test system.
(a)
Identification. A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.(b)
Classification. Class II.
0
12.0 510(k) SUMMARY
K033727
This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of CFR 807.92.
510(k) Number: Not Known
1. Name of Submitter, Contact Person and Date Summary Prepared:
Nichols Institute Diagnostics 1311 Calle Batido San Clemente, CA 92673 Phone: 949-940-7417 Fax: 949-940-7440
Contact Person: Robert L. Schmidt Date Prepared: November 25, 2003
2. Device Name
Trade/Proprietary Name: | Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay |
---|---|
Common/Usual Name: | Bio-Intact PTH (1-84) Assay |
Classification Name: | Radioimmunoassay, Parathyroid Hormone |
3. Predicate Device:
The device that is the subject of this submission is substantially equivalent to the Nichols Advantage® Bio-Intact (1-84) PTH Immunoassay (K013992; Cleared 12/31/01). This 510(k) is for labeling changes to the predicate device.
4. Device Description:
The Bio-Intact PTH (1-84) Assay is a two-site chemiluminescence assay for use with the Nichols Advantage® Specialty System.
5. Intended Use
The Nichols Advantage® Bio-Intact PTH (1-84) immunoassay is intended for use with the Nichols Advantage Specialty System to measure the levels of parathyroid hormone in serum, EDTA plasma and heparinized plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Measurements of parathyroid hormone levels with the Bio-Intact PTH (1-84) Immunoassay are also used as an aid in monitoring therapeutic intervention of secondary hyperparathyroidism that frequently occurs in chronic kidney disease. Assay results should be
1
used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
6. Comparison to predicate device:
The Bio-Intact PTH (1-84) Assay is substantially equivalent to the Nichols Advantage Bio-Intact PTH(1-84) Immunoassay previously submitted as K013992. The following tables compare the Bio-Intact PTH (1-84) Assay (modified) with the predicate device, the Nichols Advantage Bio-Intact PTH (1-84) Immunoassay (K013992).
Similarities:
- Intended Use: For the quantitative determination of intact PTH in human serum or EDTA . plasma.
- Both assays use specific antibodies to bind intact PTH. .
- Both assays can use human serum or EDTA plasma for the test sample. .
- Both assays use chemiluminometric technology based on acridinium esters. .
- Both assays use 150 microliters of sample. .
- Both assays use two point calibration every two weeks (maximum) of stored working . calibration curve; or when controls out of range.
- Both assays use Streptavidin-coated magnetic particles and streptavidin-biotin separation . technology.
- Both assays' total incubation period is 30 minutes at 37℃. .
- The sensitivity of both assays is 1.5 pg/mL. (analytical) and 4.0 pg/mL (functional). ♥
- . The throughput is the same for both assays at 180 test/hour.
- . The cross reactives are the same between assays.
- . The recovery is the same between assays.
- The parallelism is the same between assays. .
- The high dose hook effect is the same between assays. .
Differences:
Intended Use: The Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay is intended for use with the Nichols Advantage® Specialty System to measure the levels of parathyroid hormone in serum, EDTA plasma and heparinized plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Measurements of parathyroid hormone levels with the Bio-Intact PTH (1-84) Immunoassay are also used as an aid in monitoring therapeutic intervention of secondary hyperparathyroidism that frequently occurs in chronic kidney disease. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
- The modified device includes data on preanalytical variables for serum, EDTA plasma, ● heparinized plasma and EDTA whole blood.
2
Performance Characteristics: | |||
---|---|---|---|
FEATURE | Nichols Advantage® | ||
Bio-Intact PTH (1-84) | |||
Immunoassay | |||
(K013992) | Bio-Intact PTH (1-84) Assay | ||
(K033302)* | Bio-Intact PTH (1-84) Assay | ||
(Modified) | |||
Precision | |||
within | |||
Run | Not greater than 4% at | ||
dose greater than 5 | |||
pg/mL | Not greater | ||
than 6% at a | |||
dose greater | |||
than 5 pg/mL | Not greater | ||
than 6% at a | |||
dose greater | |||
than 5 pg/mL | |||
Precision | |||
Total | Not greater than 9.5% | ||
at dose greater than 34 | |||
pg/mL | Not greater | ||
than 11% at a | |||
dose greater | |||
than 5 pg/mL | Not greater | ||
than 11% at a | |||
dose greater | |||
than 5 pg/mL | |||
*K033302 was submitted October 14, 2003 and is under review. | |||
Method Comparison 1 - Serum Analysis Nichols Advantage Intact PTH Assay Vs. Bio-Intact PTH (1- |
- Assay | | | |
| FEATURE | Nichols Advantage® Bio-Intact
PTH (1-84) Immunoassay
(K013992) | Bio-Intact PTH (1-84) Assay
(Modified) | |
| Sample Size: | 305 | 305 | |
| Range of
Results: | Nichols Advantage Intact PTH
Assay: 5.0 to 1387 pg/mL
Nichols Advantage Bio-Intact
PTH (1-84) Assay: 3.0 to 746
pg/mL | Nichols Advantage Intact PTH Assay: 5.0 to
1387 pg/mL
Nichols Advantage Bio-Intact PTH (1-84) Assay: 3.0
to 746 pg/mL | |
| Passing Bablok
Regression
Equation: | y = 0.66x - 0.6 | y = 0.66x - 0.6 | |
| Least Squares
Linear
Regression
Equation: | y = 0.60x + 4.2 | y = 0.60x + 4.2 | |
| Pearson's
Correlation
Coefficient (r): | 0.97 | 0.97 | |
| Method Comparison 2 - Renal Dialysis Samples Nichols Advantage Intact PTH Assay Vs. Bio-Intact
PTH (1-84) Assay | | | |
| FEATURE | Nichols Advantage® Bio-Intact
PTH (1-84) Immunoassay
(K013992) | Bio-Intact PTH (1-84) Assay
(Modified) | |
| Sample Size: | N/A | 3187 | |
| Range of results: | N/A | Nichols Advantage Intact PTH Assay:
7 - 1797pg/mL
Nichols Advantage Bio-Intact PTH (1-84) Assay 4-998 | |
| Least Squares Linear Regression Equation: | N/A | $Y = 0.52 x + 1.3$ | |
| Pearson's Correlation Coefficient (r): | N/A | 0.97 | |
3
4
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Public Health Service
APR 2 2 2004
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Robert L. Schmidt Director of Quality Nichols Institute Diagnostics 1311 Calle Batido San Clemente, CA 92673
K033727 Re:
Trade/Device Name: Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay Regulation Number: 21 CFR 862.1545 Regulation Name: Parathyroid hormone test system Regulatory Class: Class II Product Code: CEW Dated: March 18, 2004 Received: March 19, 2004
Dear Mr. Schmidt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, ITDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a dctermination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practicc requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
5
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Jean M. Cooper, MS, DVM.
Yean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
6
INDICATIONS FOR USE STATEMENT )
INDICATIONS FOR USE STATEMENT
K033727 510(k) Number (if known):
Device Name: Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay
Indications For Use:
The Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay is intended for use with the Nichols Advantage® Specialty System to measure the levels of parathyroid hormone in serum, EDTA plasma and heparinized plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Measurements of parathyroid hormone levels with the Bio-Intact PTH (1-84) Immunoassay are also used as an aid in monitoring therapeutic intervention of secondary hyperparathyroidism that frequently occurs in chronic kidney disease. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
Prescription Use
(Per 21 CFR 801.109) ✓
OR
Over-The-Counter Use (Optional Format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Albert C
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K033721
Page 6 of 119