K Number
K013992

Validate with FDA (Live)

Date Cleared
2001-12-13

(9 days)

Product Code
Regulation Number
862.1545
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nichols Advantage® Bio-Intact PTH (1-84) immunometric assay is intended for use with the Nichols Advantage® Specialty System to measure the levels of parathyroid hormone in serum and EDTA plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.

Device Description

The Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay is a two-site immuno-chemiluminometric IVD device with sufficient reagents for 100 tests, that are performed entirely on the Nichols Advantage® Specialty System (K961142; cleared February 18, 1997). Two goat polyclonal antibodies directed at epitopes on intact human PTH are used. One antibody ("capture reagent") is chemically labeled with biotin, while the second antibody ("detection reagent") is chemically labeled with acridinium ester for subsequent quantitative measurements. A sample of patient serum (preferred) or plasma is added to an assay cuvette, followed by addition of the two goat anti-PTH antibodies and the streptaviding outed magnetic particles. The reaction mixture is allowed to incubate for 30 minutes at 37°C. Because of the high affinity interaction between biotin-labeled antibody and streptavidin, the sandwich complex is captured onto the streptavidin-coated magnetic particles. The captured complex bound to the magnetic particle passes to the Nichols Specialty System for a wash to remove unbound components and/or patient substances. The washed captured complex bound to the magnetic particles within the cuvette wells are analyzed Nichols Specialty System's luminometer via the automatic injection of reagents that initiate the acridinium-based chemiluminescence quantitative reaction. The light is measured by the Nichols Specially System's luminometer and expressed as Relative Light Units (RLU). The amount of PTH bound-labeled antibody is directly proportional to the concentration of intact PTH in the serum or plasma sample.

AI/ML Overview

The Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay is intended for use with the Nichols Advantage® Specialty System to measure parathyroid hormone levels in serum and EDTA plasma. This is used for the differential diagnosis of hypercalcemia and hypocalcemia resulting from disorders of calcium metabolism. The device performance was compared to a previously FDA-cleared predicate device, the Nichols Intact PTH Immunoassay.

Here is a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay)
Precision
Within-Run PrecisionNot greater than that of the predicate device.Not greater than 4% (at doses > 5 pg PTH/mL)
Total PrecisionNot greater than that of the predicate device.Not greater than 8.5% (at doses > 5 pg PTH/mL)
Limit of DetectionComparable to the predicate device.At or below 1.5 pg/mL
RecoveryAcceptable range, comparable to predicate.93% to 103%
ParallelismAcceptable range, comparable to predicate.92% to 111%
High Dose Hook EffectNo significant hook effect.No High Dose Hook Effect observed up to 100,000 pg PTH/mL
Interference (PTH fragments)Minimal interference.PTH fragment 7-84 did not interfere at 3000 pg/mL; PTH fragments 39-68, 53-84, 44-68, 39-84 gave minimal interference.
Correlation with PredicateHigh correlation (e.g., Pearson's r > 0.95).Pearson's correlation coefficient (r) = 0.97 (95% CI: 0.96 to 0.98)
Regression with PredicateSlope and intercept close to 1 and 0, respectively, or within an acceptable range for clinical equivalence.Passing Bablok regression: y = 0.66x - 0.6 (95% CI slope: 0.64-0.68; intercept: -1.1 to -0.2)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 305 patient serum samples.
  • Data Provenance: The document does not specify the country of origin. It can be inferred that the samples were collected retrospectively as they were "patient serum samples" assayed by both methods without modifications, suggesting existing samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided in the document. The "ground truth" for the test set was essentially the results from the predicate device.

4. Adjudication Method for the Test Set

  • Not applicable in this context. The comparison was directly between the values generated by the new device and the predicate device, not against an adjudicated "ground truth" from human experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This submission is for an in vitro diagnostic immunoassay, not an AI-powered image analysis or diagnostic assist device that would typically involve human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, this was a standalone performance study. The Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay operates automatically on the Nichols Advantage® Specialty System, and its performance was assessed directly against the predicate device. There is no "human-in-the-loop" component to the analytical performance of the assay itself, beyond routine laboratory procedures.

7. The Type of Ground Truth Used

  • The "ground truth" for evaluating the new device's performance was the results obtained from an FDA-cleared predicate device, the Nichols Intact PTH Immunoassay. This is a common approach for establishing substantial equivalence for new IVD devices.

8. The Sample Size for the Training Set

  • This information is not provided. The term "training set" is typically associated with machine learning or AI algorithms. For this immunoassay, performance characteristics (precision, LOD, recovery etc.) are established through internal validation studies, not a distinct "training set" in the AI sense.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no mention of a "training set" in the context of an AI/ML algorithm. The calibration of the immunoassay system would be established using calibrators with known PTH concentrations, which are themselves traceable to a primary reference standard (though this specific traceability is not detailed in the provided text).

{0}------------------------------------------------

510(k) Submission for

Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay

Nichols Institute Diagnostics, Inc.

November 5, 2001

Kol3992

SUMMARY STATEMENT OF SAFETY AND EFFECTIVENESS

The sponsor, Nichols Institute Diagnostics, Inc., has developed, manufactured and tested under current Good Manufacturing Practices/Quality System Regulations (cGMP/QSR) an in vitro diagnostic (IVD) device to quantitatively assay for parathyroid hormone (PTH) in serum and EDTA plasma.

The trade name is: Nichols Advantage® Chemiluminescence Bio-Intact Parathyroid Hormone (1-84) Immunoassay, having a common name of: Bio-Intact PTH (1-84) Immunoassay. This is a Class II in vitro diagnostic (IVD) medical device, as per 21 CFR 862.1545 with a Product Code of 75 CEW. This Bio-Intact PTH (1-84) Immunoassay is intended for use with the Nichols Advantage® Specialty System for measurement of parathyroid hormone levels in patient serum (preferred) or plasma, for differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.

The Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay is a two-site immuno-chemiluminometric IVD device with sufficient reagents for 100 tests, that are performed entirely on the Nichols Advantage® Specialty System (K961142; cleared February 18, 1997). Two goat polyclonal antibodies directed at epitopes on intact human PTH are used. One antibody ("capture reagent") is chemically labeled with biotin, while the second antibody ("detection reagent") is chemically labeled with acridinium ester for subsequent quantitative measurements. A sample of patient serum (preferred) or plasma is added to an assay cuvette, followed by addition of the two goat anti-PTH antibodies and the streptaviding outed magnetic particles. The reaction mixture is allowed to incubate for 30 minutes at 37°C. Because of the high affinity interaction between biotin-labeled antibody and streptavidin, the sandwich complex is captured onto the streptavidin-coated magnetic particles. The captured complex bound to the magnetic particle passes to the Nichols Specialty System for a wash to remove unbound components and/or patient substances. The washed captured complex bound to the magnetic particles within the cuvette wells are analyzed Nichols Specialty System's luminometer via the automatic injection of reagents that initiate the acridinium-based chemiluminescence quantitative reaction. The light is measured by the Nichols Specially System's luminometer and expressed as Relative Light Units (RLU). The amount of PTH bound-labeled antibody is directly proportional to the concentration of intact PTH in the serum or plasma sample.

In-house testing of the Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay demonstrated Within-Run precision not greater than 4% at a dose higher than 5 pg PTH/mL, and a Total Precision of not greater than 8.5% at a dose higher than 5 pg PTH/mL, with a throughput of about 180 results per hour. The Limit of Detection for this Bio-Intact (1-84) PTH Immunoassay was estimated to be at or below 1.5 po/mL. with Recovery ranging from 93% to 103%, and Parallelism at 92% to 111%. No High Dose Hook Effect was observed up to 100,000 pg PTH/mL. The PTH fragment 7-84 did not interfere in this assay at a dose up to 3000 pg/mL; PTH fragments 39-68, 53-84, 44-68, 39-84 gave minimal interference. Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay (y) was compared to predicate Nichols Intact PTH Immunoassay (x), previously FDA cleared and currently marketed for Intact PTH assay. Three hundred five (305) patient serum samples were assayed by both methods using insert directions without modifications. The range of values observed with the predicate Intact PTH Immunoassay was 5.0 to 1387 pg/mL, while the range with the Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay was 3.0 to 746 pg/mL. Passing Bablok regression analysis of these data yielded an equation of y = 0.66x - 0.6 (95% confidence intervals of the slope and intercept were 0.64 to 0.68, and -1.1 to -0.2 respectively). Pearson's correlation coefficient (r) of the paired data was 0.97 (95% confidence interval was 0.96 to 0.98), demonstrating that this Bio-Intact PTH (1-84) Immunoassay provided essentially equivalent results to an approved predicate Intact PTH (1-84) Immunoassay.

Additional information on this Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay submission may be obtained by contacting Mr. Robert Schmidt, Vice President of Quality Systems, Nichols Institute Diagnostics, Inc. by telephone at 1-949-240-5417 or by telefax at 1-949-240-5271.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three curved lines representing its wings.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Nichols Institute Diagnostics c/o Raymond Wilson, Pharm.D. California Department of Health Medical Device Safety Section Food and Drug Branch P.O. Box 942732 (MS-357) Sacramento, CA 94234

DEC 1 3 2001

Re: K013992

Trade/Device Name: Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay Regulation Number: 21 CFR 862.1545 Regulation Name: Parathyroid hormone test system Regulatory Class: Class II Product Code: CEW Dated: November 29, 2001 Received: December 4, 2001

Dear Dr. Wilson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket I his letter will and w you do organ substantial equivalence of your device to a legally marketed nonification. THE I Driving of incation for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 If you desire specific active diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (201) 594-1566. Practicely) , 1981) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small information on your respond and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

INDICATIONS FOR USE STATEMENT 4.0

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

Device Name: Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay

Indications For Use:

The Nichols Advantage® Bio-Intact PTH (1-84) immunometric assay is intended for use with the Nichols Advantage® Specialty System to measure the levels of parathyroid hormone in serum and EDTA plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.

Shoren C. Grotte for Jean Cooper


(Division Sign-Off)

ivision of Clinical Laboratory Devices 510(k) Number

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ﺳ Prescription Use

OR

Over-The-Counter Use (Optional Format 1-2-96)

(Per 21 CFR 801.109)

§ 862.1545 Parathyroid hormone test system.

(a)
Identification. A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.(b)
Classification. Class II.