K Number
K033689
Device Name
OSSTELL MENTOR RFA
Date Cleared
2004-02-26

(94 days)

Product Code
Regulation Number
872.4200
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Osstell Mentor is indicated for use in measuring the stability of implants in the oral cavity and craniofacial region. The Osstell Mentor can add important information to the evaluation of implant stability and can be used as part of an overall treatment evaluation program. The final implant treatment decisions are the responsibility of the surgeon.
Device Description
The Osstell Mentor Resonance Frequency Analyzer (RFA) is an updated version of the Osstell™ (K003714), the system designed to measure dental implant stability in the oral cavity and craniofacial region. Similar to K003714, the Osstell Mentor is a portable, handheld instrument that involves the use of the noninvasive technique, Resonance Frequency Analysis. The updated system involves the use of a Smartpeg (aluminum rod) attached to the dental implant by means of a screw. The Smartpeg is excited by a magnetic pulse from the measurement probe on the handheld instrument. The resonance frequency, which is the measure of implant stability, is calculated from the response signal. Results are displayed as the Implant Stability Quotient (ISQ). The ISQ is a measurement of the stability of the implant and is derived from the resonance frequency value obtained from the Smartpeg.
More Information

No
The description focuses on Resonance Frequency Analysis (RFA) and calculating an Implant Stability Quotient (ISQ) based on physical principles, with no mention of AI or ML algorithms for data analysis or interpretation.

No
The device is used for measuring implant stability and providing information for evaluation, but it does not directly treat or alleviate a disease or condition. The final treatment decisions are the responsibility of the surgeon.

Yes
The device measures the stability of implants (which can indicate if there's a problem or to monitor healing) and provides information for an overall treatment evaluation program. While final treatment decisions are with the surgeon, the device provides data (ISQ) to inform those decisions, which aligns with the definition of a diagnostic device.

No

The device description explicitly states it is a "portable, handheld instrument" and involves a "measurement probe on the handheld instrument," indicating it is a hardware device, not software-only.

Based on the provided information, the Osstell Mentor is not an In Vitro Diagnostic (IVD).

Here's why:

  • IVDs are used to examine specimens derived from the human body. The Osstell Mentor directly measures the stability of an implant in situ within the oral cavity and craniofacial region. It does not analyze a biological sample taken from the patient.
  • The device description clearly outlines a physical measurement technique (Resonance Frequency Analysis) applied directly to the implant. This is a mechanical measurement, not a biochemical or biological analysis of a specimen.

Therefore, the Osstell Mentor falls outside the definition of an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The Osstell Mentor RFA is indicated for use to measure the stability of dental implants in the oral cavity and craniofacial region.

The Osstell Mentor is indicated for use in measuring the stability of implants in the oral cavity and craniofacial region. The Osstell Mentor can add important information to the evaluation of implant stability and can be used as part of an overall treatment evaluation program. The final implant treatment decisions are the responsibility of the surgeon.

Product codes

EKX

Device Description

The Osstell Mentor Resonance Frequency Analyzer (RFA) is an updated version of the Osstell™ (K003714), the system designed to measure dental implant stability in the oral cavity and craniofacial region. Similar to K003714, the Osstell Mentor is a portable, handheld instrument that involves the use of the noninvasive technique, Resonance Frequency Analysis. The updated system involves the use of a Smartpeg (aluminum rod) attached to the dental implant by means of a screw. The Smartpeg is excited by a magnetic pulse from the measurement probe on the handheld instrument. The resonance frequency, which is the measure of implant stability, is calculated from the response signal. Results are displayed as the Implant Stability Quotient (ISQ). The ISQ is a measurement of the stability of the implant and is derived from the resonance frequency value obtained from the Smartpeg.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

oral cavity and craniofacial region

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K003714

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.

0

K033689

PREMARKET NOTIFICATION 510(K) SUMMARY

Sponsor:
------------

Integration Diagnostics Ltd. Gamlestadsvagen 3 B 415 02 Goteborg SWEDEN Telephone: +46-31-340 82 51 Fax +46-31-41 31 15

Contact: Anders Petersson

Manufacturer:

Same as above

3004070020 Registration Number:

Contact Person:

Trish Landry or Russ Pagano M Squared Associates, Inc. 719 A Street, NE Washington DC 20002 Telephone: 202-546-1262 Fax: 202-546-3848 E-mail: tlandry@msquaredassociates.com

Trade Name of Device:Osstell Mentor
Common Name:Dental Implant Stabilizer
Classification name:Handpiece, Direct Drive, AC-Powerede
Product Code:EKX
Regulation Class:I
Regulation Number:$870.4200

1

The Osstell Mentor Resonance Frequency Analyzer (RFA) is an Device Description: updated version of the Osstell™ (K003714), the system designed to measure dental implant stability in the oral cavity and craniofacial region. Similar to K003714, the Osstell Mentor is a portable, handheld instrument that involves the use of the noninvasive technique, Resonance Frequency Analysis. The updated system involves the use of a Smartpeg (aluminum rod) attached to the dental implant by means of a screw. The Smartpeg is excited by a magnetic pulse from the measurement probe on the handheld instrument. The resonance frequency, which is the measure of implant stability, is calculated from the response signal. Results are displayed as the Implant Stability Quotient (ISQ). The ISQ is a measurement of the stability of the implant and is derived from the resonance frequency value obtained from the Smartpeg.

The Osstell Mentor RFA is indicated for use to measure the stability of Indications for Use: dental implants in the oral cavity and craniofacial region.

Basis for Substantial Equivalence

Osstell™M Predicate Device:

The Osstell Mentor is substantially equivalent to the previous version of the device, the Osstell (K003714), for the following reasons:

  • The Osstell Mentor has the same indication for use as the Osstell cleared by FDA under 510(k), K003714.
  • The functioning of the Osstell Mentor and the Osstell is similar in that both devices use a form of vibration to assess implant stability.
  • The Osstell Mentor has no unique or new applications, indications or functions.
  • The Osstell Mentor raises no new issues with respect to safety or effectiveness

2

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 6 2004

Integration Diagnostics Limited C/O Ms. Tish Landry Senior Consultant M Squared Associates, Incorporated 719 A Street NE Washington, DC 20002

Re: K033689

Trade/Device Name: Osstell Mentor RFA Regulation Number: 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EKX Dated: February 18, 2004 Received: February 19, 2004

Dear Ms. Landry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

3

Page 2 - Ms. Landry

co.nply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of R Fat 807); adoling (2 (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product lastillion you to begin marketing your device as described in your Section 510(k) I mis letter with and R your e FDA finding of substantial equivalence of your device to a legally premaince noticated in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you attire office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general micronational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Chris

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

Indications for Use

510(k) Number (if known): K033689

Device Name:_Osstell™ Mentor

Indications For Use: The Osstell Mentor is indicated for use in measuring the stability of implants in the oral cavity and craniofacial region. The Osstell Mentor can add important information to the evaluation of implant stability and can be used as part of an overall treatment evaluation program. The final implant treatment decisions are the responsibility of the surgeon.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kevir Mully for MSR

ion Sian Division of Anesthesiology, General Hospital, Infection Control, Dental De

510(k) Number: K033689