K Number
K032494
Device Name
PVS 1700 STEERABLE DELIVERY SYSTEM MICROCATHETER
Date Cleared
2003-09-05

(23 days)

Product Code
Regulation Number
870.1210
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The PVS 1700 Steerable Delivery System®-10 Microcatheter is intended to be used to access tortuous vasculature for sub-selective controlled infusion or delivery of diagnostic, embolic, and therapeutic agents into the distal, peripheral, coronary, and neurovasculature, and for guide wire exchange/support during diagnostic or interventional procedures.
Device Description
The PVS 1700 Steerable Delivery System®-10 is a 1.9F/2.4F (nominal distal/proximal) tubular device, 155 cm in length, with a lumen to be used for delivery of contrast, drugs, or embolics. The lumen is constructed from a polymeric material and has an inside diameter of 0.0165". The device is coated on the outer diameter with a lubricious coating over the distal segment of the device. Two radiopaque markers are positioned at the distal tip of the device to aid visualization under fluoroscopy. The proximal end of the device has a standard luer adapter for attachment of accessories and can be used to flush the lumen. The subject device has the ability to access distal, tortuous vasculature over a guide wire, deliver embolics and agents, and has the ability to be steered like a guide wire as needed.
More Information

Not Found

No
The device description and performance studies focus on the mechanical and material properties of a steerable microcatheter, with no mention of AI or ML capabilities.

No.
The device is a delivery system for therapeutic agents, but it does not inherently provide therapy itself.

No

The device is a microcatheter intended for the delivery of agents (diagnostic, embolic, and therapeutic) and for guidewire support, not for diagnosing conditions itself.

No

The device description clearly outlines a physical, tubular device made of polymeric material with a lumen, radiopaque markers, and a lubricious coating. It is a hardware device intended for physical insertion into vasculature.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly describes a device used for accessing vasculature and delivering agents within the body. This is an interventional device, not a device used to examine specimens in vitro (outside the body).
  • Device Description: The description details a catheter designed for insertion into blood vessels, with features like a lumen for delivery, radiopaque markers for visualization under fluoroscopy, and steerability. These are characteristics of an in-vivo device.
  • Lack of IVD Characteristics: There is no mention of analyzing biological specimens (blood, urine, tissue, etc.) or performing tests on samples outside the body.

Therefore, the PVS 1700 Steerable Delivery System®-10 Microcatheter is an interventional medical device, not an IVD.

N/A

Intended Use / Indications for Use

The PVS 1700 Steerable Delivery System®-10 Microcatheter is intended to be used to access tortuous vasculature for sub-selective controlled infusion or delivery of dlagnostic, embolic, and therapeutic agents into the distal, peripheral, coronary, and neurovasculature, and for guldewire exchangelsupport during diagnostic or interventional procedures.

Product codes

KRA

Device Description

The PVS 1700 Steerable Delivery System®-10 is a 1.9F/2.4F (nominal distal/proximal) tubular device, 155 cm in length, with a lumen to be used for delivery of contrast, drugs, or embolics. The lumen is constructed from a polymeric material and has an inside diameter of 0.0165". The device is coated on the outer diameter with a lubricious coating over the distal segment of the device. Two radiopaque markers are positioned at the distal tip of the device to aid visualization under fluoroscopy. The proximal end of the device has a standard luer adapter for attachment of accessories and can be used to flush the lumen. The subject device has the ability to access distal, tortuous vasculature over a guide wire, deliver embolics and agents, and has the ability to be steered like a guide wire as needed.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

fluoroscopy

Anatomical Site

distal, peripheral, coronary, and neurovasculature

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Biocompatibility of the PVS 1700 Steerable Delivery System®-10 has been verified in accordance with ISO 10993-1, Biocompatibility of Medical Devices - Part 1. Test results confirmed biocompatibility of the subject device when tested as an external communicating, blood contact, short duration (

§ 870.1210 Continuous flush catheter.

(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).

0

K032494/

Summary of Safety and Effectiveness

Prepared 11 August 2003
General
ProvisionsSubmitter of 510(k)
Premarket Notification:Precision Vascular
2405 West Orton Circle
West Valley City, UT 84119
Phone: 801.974.1700
Fax: 801.974.1740
Contact Person:Rick Gaykowski
Vice President, Regulatory/Clinical
Affairs & Quality Systems
Device Trade Name:
Device Generic Name:PVS 1700 Steerable Delivery System®-10
Infusion Catheter
The predicate devices are listed in the table below.
DeviceManufacturer510(k) Number,
Concurrence DateProduct
Code
Predicate
DevicesSDS 0.018"PVS, Inc.K020733, 21 March 2002KRA
Excelsior SL-10BSC/TargetK013789, 06 Dec 2001DQY
Rebar-14Micro TherapeuticsK993672, 04 Jan 2000KRA
ClassificationClass II, 21 CFR 870.1210, Continuous Flush Catheter, KRA
Performance
StandardsPerformance standards have not been established by FDA under section 514 of the
Federal Food, Drug and Cosmetic Act.
Intended UseThe PVS 1700 Steerable Delivery System®-10 Microcatheter is intended to be used
to access tortuous vasculature for sub-selective controlled infusion or delivery of
diagnostic, embolic, and therapeutic agents into the distal, peripheral, coronary, and
neurovasculature, and for guide wire exchange/support during diagnostic or
interventional procedures.
Device
DescriptionThe PVS 1700 Steerable Delivery System®-10 is a 1.9F/2.4F (nominal
distal/proximal) tubular device, 155 cm in length, with a lumen to be used for delivery of
contrast, drugs, or embolics. The lumen is constructed from a polymeric material and
has an inside diameter of 0.0165". The device is coated on the outer diameter with a
lubricious coating over the distal segment of the device. Two radiopaque markers are
positioned at the distal tip of the device to aid visualization under fluoroscopy. The
proximal end of the device has a standard luer adapter for attachment of accessories
and can be used to flush the lumen. The subject device has the ability to access distal,
tortuous vasculature over a guide wire, deliver embolics and agents, and has the ability
to be steered like a guide wire as needed.

.

.. . . .

1

| Technological
Characteristics | Technological similarities between the PVS 1700 Steerable Delivery System®-10
and predicate devices include the basal tubular design and dimensions, polymeric
materials and construction, and hydrophilic coating. In instances where the
technological characteristics may differ slightly, it has been demonstrated that there
are no new questions raised regarding safety and efficacy of the PVS 1700 Steerable
Delivery System®-10. |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety and
Performance
Tests | Biocompatibility of the PVS 1700 Steerable Delivery System®-10 has been verified in
accordance with ISO 10993-1, Biocompatibility of Medical Devices - Part 1. Test
results confirmed biocompatibility of the subject device when tested as an external
communicating, blood contact, short duration (