(19 days)
VITROS Progesterone Reagent Pack
For in vitro diagnostic use only.
The Vitros Progesterone Reagent Pack quantitatively measures progesterone concentration in human serum and plasma.
VITROS Progesterone Calibrators
For in vitro use in the calibration of the VITROS Immunodiagnostic System for the quantitative measurement of progesterone in human serum and plasma (EDTA or heparin).
The VITROS Immunodiagnostic System uses luminescence as the signal in the quantitative and semi-quantitative determination of selected analytes in human body fluids, commonly serum and plasma. Coated microwells are used as the solid phase separation system.
The system is comprised of three main elements:
- The VITROS Immunodiagnostic Products range of immunoassay products (in this case VITROS Immunodiagnostic Products Progesterone Reagent Pack, VITROS Immunodiagnostic Products Progesterone Calibrators, which are combined by the VITROS Immunodiagnostic System to perform the VITROS Progesterone assay).
- The VITROS Immunodiagnostic System instrumentation, which provides automated use of the immunoassay kits. The VITROS Immunodiagnostic System was cleared for market by a separate 510(k) premarket notification (K962919/S1).
- Common reagents used by the VITROS System in each assay. The VITROS Immunodiagnostic Products Signal Reagent and VITROS Immunodiagnostic Products Universal Wash Reagent were cleared as part of the VITROS Immunodiagnostic Products Total T3 Reagent Pack and VITROS Immunodiagnostic Products Total T3 Calibrators 510(k) premarket notification (K964310).
The VITROS System and common reagents are dedicated specifically for use only with the VITROS Immunodiagnostic Products range of immunoassay products.
This submission describes the VITROS Immunodiagnostic Products Progesterone Reagent Pack and VITROS Immunodiagnostic Products Progesterone Calibrators (modified), which is a progesterone assay. The device is intended for in vitro diagnostic use to quantitatively measure progesterone concentration in human serum and plasma. The study presented aims to demonstrate substantial equivalence to a legally marketed predicate device (VITROS Immunodiagnostic Products Progesterone Reagent Pack and VITROS Immunodiagnostic Products Progesterone Calibrators).
1. Table of Acceptance Criteria and Reported Device Performance
The provided document details a comparison of assay characteristics between the predicate and the new device. It does not explicitly state quantitative acceptance criteria or detailed performance metrics. Instead, it focuses on substantial equivalence based on comparable characteristics.
| Device Characteristic | Predicate Device (Current) | New Device (Modified) | Acceptance Criteria/Performance |
|---|---|---|---|
| Number of Calibrators | 3 | 3 | Matches predicate |
| Nominal Calibrator values | 0.0, 4.25, 120 nmol/L | 0.0, 4.25, 120 nmol/L | Matches predicate |
| Physical State of Calibrators | Liquid | Freeze-dried | Change noted; likely evaluated for equivalent performance (details not provided) |
| Reconstitution Volume | Not required | 1 mL | Change noted; likely evaluated for user impact (details not provided) |
| Storage Temperature of Calibrators | 2-8°C | 2-8°C | Matches predicate |
| Calibration range | 0 to 178 nmol/L | 0 to 178 nmol/L | Matches predicate |
| Basic principle | Solid phase immunoassay | Solid phase immunoassay | Matches predicate |
| Tracer | Enzyme labeled | Enzyme labeled | Matches predicate |
| Instrumentation | VITROS Immunodiagnostic System | VITROS Immunodiagnostic System | Matches predicate |
| Sample volume | 25µL | 25µL | Matches predicate |
| Incubation time and temperature | 16 minutes at 37°C with shaking | 16 minutes at 37°C with shaking | Matches predicate |
2. Sample Size Used for the Test Set and the Data Provenance
The provided document does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the study). The submission focuses on comparing the modified device's characteristics to a predicate device, rather than providing detailed clinical study results with a test set.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Not applicable. This is an in vitro diagnostic device for quantitative measurement, and the ground truth would typically be established by established reference methods or primary analytical measurements, not by expert interpretation in the same way imaging or diagnostic algorithms are validated. The document does not mention the use of experts for ground truth establishment.
4. Adjudication Method for the Test Set
Not applicable. As noted above, the assessment is based on analytical performance and comparison to a predicate, not on subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an in vitro diagnostic device, not an AI-assisted diagnostic tool that involves human readers interpreting results.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
The device itself is a standalone immunoassay system. The document implies that the performance of the VITROS Progesterone assay (modified) was evaluated on its own analytical capabilities to demonstrate substantial equivalence to the predicate. There is no mention of human interaction beyond the operation of the instrument.
7. The Type of Ground Truth Used
The ground truth for this type of in vitro diagnostic device would typically be established through highly accurate and precise analytical reference methods, or by comparing agreement with the predicate device's established performance, rather than pathology, outcomes data, or expert consensus in an observational or interpretive sense. The document states that "the information presented in the pre-market notification demonstrates that the performance of the VITROS Progesterone assay (modified) for use with human serum and plasma is substantially equivalent to the cleared predicate device." This implies that the 'ground truth' or benchmark for performance comparison is the predicate device's established performance using recognized analytical methods.
8. The Sample Size for the Training Set
The document does not provide details on a "training set" in the context of machine learning. This device is an immunoassay, not an AI/ML-based algorithm that typically requires a training set. The development of such assays involves analytical validation, linearity studies, precision, accuracy, and interference studies, but these do not fall under the conventional definition of a "training set" for AI.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as this device does not utilize a "training set" in the AI/ML sense. The "ground truth" for developing and validating immunoassay performance would be derived from rigorous analytical testing using known concentrations of analytes, reference materials, and comparison with established methods.
{0}------------------------------------------------
Summary Information
AUG 1 3 2003
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: KO32296.
-
- Submitter Ortho-Clinical Diagnostics, Inc. name, 100 Indigo Creek Drive address, Rochester, New York 14626-5101 contact (585) 453-4469
Contact Person: Susan M. Werner
- Submitter Ortho-Clinical Diagnostics, Inc. name, 100 Indigo Creek Drive address, Rochester, New York 14626-5101 contact (585) 453-4469
-
- Preparation Date Special 510(k) prepared: July 24, 2003 date
-
- Device Trade or Proprietary Name: name VITROS Immunodiagnostic Products Progesterone Reagent Pack VITROS Immunodiagnostic Products Progesterone Calibrators
Common Name: Progesterone assay Classification Name: Progesterone test system (21 CFR 862.1620).
Continued on next page
{1}------------------------------------------------
510(k) Summary, Continued
- The VITROS Immunodiagnostic Products Progesterone Reagent Pack and 4. Predicate device VITROS Immunodiagnostic Products Progesterone Calibrators (modified) are substantially equivalent to the VITROS Immunodiagnostic Products Progesterone Reagent Pack and VITROS Immunodiagnostic Products Progesterone Calibrators (current).
- The VITROS Immunodiagnostic System uses luminescence as the signal in 5. Device description the quantitative and semi-quantitative determination of selected analytes in human body fluids, commonly serum and plasma. Coated microwells are used as the solid phase separation system.
The system is comprised of three main elements:
-
- The VITROS Immunodiagnostic Products range of immunoassay products (in this case VITROS Immunodiagnostic Products Progesterone Reagent Pack, VITROS Immunodiagnostic Products Progesterone Calibrators, which are combined by the VITROS Immunodiagnostic System to perform the VITROS Progesterone assay).
-
- The VITROS Immunodiagnostic System instrumentation, which provides automated use of the immunoassay kits. The VITROS Immunodiagnostic System was cleared for market by a separate 510(k) premarket notification (K962919/S1).
-
- Common reagents used by the VITROS System in each assay. The VITROS Immunodiagnostic Products Signal Reagent and VITROS Immunodiagnostic Products Universal Wash Reagent were cleared as part of the VITROS Immunodiagnostic Products Total T3 Reagent Pack and VITROS Immunodiagnostic Products Total T3 Calibrators 510(k) premarket notification (K964310).
The VITROS System and common reagents are dedicated specifically for use only with the VITROS Immunodiagnostic Products range of immunoassay products.
Continued on next page
{2}------------------------------------------------
510(k) Summary, Continued
-
VITROS Progesterone Reagent Pack 6. Device intended For in vitro diagnostic use only. use The VITROS Progesterone Reagent Pack quantitatively measures Progesterone concentration in human serum and plasma.
VITROS Progesterone Calibrators For in vitro use in the calibration of the VITROS Immunodiagnostic System for the quantitative measurement of progesterone in human serum and plasma (EDTA or heparin). -
The VITROS Immunodiagnostic Products Progesterone Reagent Pack and 7. Comparison to predicate VITROS Immunodiagnostic Products Progesterone Calibrators (modified) are device substantially equivalent to VITROS Immunodiagnostic Products Progesterone Reagent Pack and Calibrators cleared by the FDA (K022901) for in vitro diagnostic use.
Table 1 lists the characteristics of the modified VITROS Progesterone assay and the current VITROS Progesterone assay.
Table 1: List of Assay Characteristics: Comparison to Predicate Device
| DeviceCharacteristic | Predicate DeviceVITROS Progesterone assay(Current) | New DeviceVITROS Progesterone assay(Modified) |
|---|---|---|
| Number of Calibrators | 3 | 3 |
| Nominal Calibrator values | 0.0, 4.25, 120 nmol/L | 0.0, 4.25, 120 nmol/L |
| Physical State of Calibrators | Liquid | Freeze-dried |
| Reconstitution Volume | Not required | 1 mL |
| Storage Temperature ofCalibrators | 2-8°C | 2-8°C |
| Calibration range | 0 to178 nmol/L | 0 to178 nmol/L |
| Basic principle | Solid phase immunoassay | Solid phase immunoassay |
| Tracer | Enzyme labeled | Enzyme labeled |
| Instrumentation | VITROS Immunodiagnostic System | VITROS Immunodiagnostic System |
| Sample volume | 25µL | 25µL |
| Incubation time andtemperature | 16 minutes at 37°C with shaking | 16 minutes at 37°C with shaking |
Continued on next page
{3}------------------------------------------------
510(k) Summary, Continued
The information presented in the pre-market notification demonstrates that the 8. Conclusions performance of the VITROS Progesterone assay (modified) for use with human serum and plasma is substantially equivalent to the cleared predicate device.
The information presented in the premarket notification provides a reasonable assurance that the VITROS Progesterone assay (modified) for use with human serum and plasma is safe and effective for the stated intended use.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a circular logo with the words "U.S. HEALTH & HUMAN SERVICES" written around the top half of the circle. The bottom half of the circle contains the word "DEPARTMENT". Inside the circle is a symbol that looks like three curved lines stacked on top of each other. The lines are thicker at the top and thinner at the bottom, and they appear to be flowing or waving.
Public Health Service
AUG 1 3 2003
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Susan M. Werner Regulatory Affairs Manager Ortho-Clinical Diagnostics 100 Indigo Creek Drive Rochester, NY 14626-5101
Re: K032296
Trade/Device Name: VITROS Immunodiagnostic Progesterone Reagent Pack VITROS Immunodiagnostic Products Progesterone Calibrators Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIS Dated: July 24, 2003
Received: August 4, 2003
Dear Ms. Werner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use Statement
Page 1 of 1
| 510(k) Number (if known): | K032296 |
|---|---|
| Device Name: | VITROS Immunodiagnostic Products Progesterone Reagent PackVITROS Immunodiagnostic Products Progesterone Calibrators |
| Indications for Use: | VITROS Progesterone Reagent PackFor in vitro diagnostic use only.The Vitros Progesterone Reagent Pack quantitatively measuresprogesterone concentration in human serum and plasma.VITROS Progesterone CalibratorsFor in vitro use in the calibration of the VITROSImmunodiagnostic System for the quantitative measurement ofprogesterone in human serum and plasma (EDTA or heparin). |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Over-The-Counter Use __
(Optional Format 1-2-96)
Carol C. Benary/s Jean Cooper, DVM
Division Sign-Off
OR
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K032296
Ortho-Clinical Diagnostics VITROS Immunodiagnostic Products Progesterone Reagent Pack VITROS Immunodiagnostic Products Progesterone Calibrators
Page 7 of 47
§ 862.1150 Calibrator.
(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.