AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The JoWire Neo's Guide Wires are not to be used in the cerebral blood vessel.

Device Description

The JoWire Asahi PTCA Guide Wires are steerable guide wires with a maximum diameter of 0.014" and available in a 300 cm length. The extension wire is connected to the end of the guide wire outside the body. The wire is constructed from stainless steel core wire with varying core lengths and diameters for each design. The core wire and core are soldered. The distal end of the guide wire has a radiopaque tip that is available straight and is made soft to easily bend with the vessel curve. The coating (hydrophilic or silicone) is applied to the distal portion of the wire guide wire. The proximal section of the guide wire is coated with PTFE.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the JoWire Asahi PTCA Guide Wires, which aims to demonstrate substantial equivalence to a predicate device. The document states that performance data was collected through bench and biocompatibility testing.

Here's an analysis of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Not explicitly stated in the document"Met the acceptance criteria and performed similarly to the predicate devices."

2. Sample size used for the test set and the data provenance:

  • Sample size: Not specified.
  • Data provenance: Not specified, but based on the nature of bench and biocompatibility testing, it would be in vitro (bench testing) and potentially in vivo (biocompatibility, if animal studies were performed, though not explicitly stated for this 510(k)). There is no indication of human clinical data or geographical origin.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable as the testing described is bench and biocompatibility, not expert-adjudicated clinical data.

4. Adjudication method for the test set:

  • Not applicable for bench and biocompatibility testing. Performance is assessed against pre-defined engineering and biological criteria.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a medical guidewire, not an AI-assisted diagnostic tool.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used:

  • For bench testing: Engineering specifications, performance standards, and comparisons to predicate device characteristics.
  • For biocompatibility testing: Established biological safety standards and guidelines.

8. The sample size for the training set:

  • Not applicable. This is a physical medical device, not a machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable.

Summary from the provided text:

The submission highlights that bench and biocompatibility testing were conducted. The crucial statement is: "Bench and biocompatibility testing were conducted according to the recommendations from relevant guidances to demonstrate that the JoWire Asahi PTCA Guide Wires met the acceptance criteria and performed similarly to the predicate devices." This indicates that the study performed was the bench and biocompatibility testing, and its outcome was that the device successfully met the predefined acceptance criteria and showed performance comparable to the predicate device. However, the specific details of these criteria (e.g., tensile strength, lubricity targets, specific biological response limits) and the quantitative results are not provided in the summary.

{0}------------------------------------------------

April 2002MAY 14 2003Special 510(k)K031277
510K) Summary of Safety and Effectiveness
Date Prepared:April 21, 2003
Submitted by:Jomed Inc.2870 Kilgore Rd.Rancho Cordova, CA 95670
Contact Person:Terry SchultzRegulatory Affairs Manager
Phone Number:(858) 673-0189 Ext. 286
Fax Number:(858) 673-3837
Device Trade Name:JoWire Asahi PTCA Guide Wire
Device ClassificationName and Class:Catheter Guide Wire, Class II (21 CFR 870.1330)
Predicate Devices:JoWire Asahi PTCA Guide Wire (K022762)

MAY 1 4 2003

JoWire Asahi PTCA Guide Wires

Special 510(k)

Device Description:

Jomed Inc.

April 2002

The JoWire Asahi PTCA Guide Wires are steerable guide wires with a maximum diameter of 0.014" and available in a 300 cm length. The extension wire is connected to the end of the guide wire outside the body. The wire is constructed from stainless steel core wire with varying core lengths and diameters for each design. The core wire and core are soldered. The distal end of the guide wire has a radiopaque tip that is available straight and is made soft to easily bend with the vessel curve. The coating (hydrophilic or silicone) is applied to the distal portion of the wire guide wire. The proximal section of the guide wire is coated with PTFE.

Intended Use:

The JoWire Asahi PTCA Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The JoWire Asahi PTCA Guide Wires are not to be in the cerebral blood vessel.

Device Technological Characteristics and Comparison to Predicate Device:

The JoWire Asahi PTCA Guide Wires are made of the same materials, available in the same diameters, have a similar design, and have the same indications for use as the predicate devices and other currently marketed PTCA Guide Wires.

{1}------------------------------------------------

510(k) Summary (cont'd)

Performance Data:

Bench and biocompatibility testing were conducted according to the recommendations from relevant guidances to demonstrate that the JoWire Asahi PTCA Guide Wires met the acceptance criteria and performed similarly to the predicate devices. No new safety or effectiveness issues were raised during the testing.

Conclusion:

The JoWire Asahi PTCA Guide Wires are substantially equivalent to the claimed predicate devices and other currently marketed PTCA Guide Wires.

14031277 Premarket Notification [510(k)] Number

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle head facing left, with three parallel lines extending from the back of the head, representing the agency's mission to promote health and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 4 2003

Ms. Terry Schultz Regulatory Affairs Manager JOMED. Inc. 15330 Avenue of Science, Suite 200 San Diego CA 92128

Re: K031277

Trade Name: JoWire Asahi PTCA Guide Wires Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: April 21, 2003 Received: April 22, 2003

Dear Ms. Schultz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Terry Schultz

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

.
Bram D. Zuckerman, M.D.

ram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Jomed Inc. April 2003 JoWire Asahi PTCA Guide Wires Special 510(k)

510(k) Number (if known): ___ K 031277

Device Name: JoWire Asahi PTCA Guide Wires

Indications for Use:

To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The JoWire Neo's Guide Wires are not to be used in the cerebral blood vessel.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.19)

Over-the Counter Use __________________________________________________________________________________________________________________________________________________________________________

Division Sign-Off) 510(k) N

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.