(84 days)
Not Found
Not Found
No
The summary describes a lyophilized human serum calibrator for clinical chemistry procedures, with no mention of AI or ML technology.
No
Explanation: The device is a calibrator for clinical chemistry procedures, used to establish reference points for measuring substances in human specimens. It does not directly treat or prevent any condition, which is the definition of a therapeutic device.
No
This device is a calibrator used to establish points of reference for clinical chemistry procedures, not to directly diagnose a patient's condition.
No
The device description explicitly states it consists of lyophilized human serum, which is a physical substance, not software.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use explicitly states it's for "quantitative calibration of WIENER LAB's clinical chemistry procedures" and is "used in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens." This directly aligns with the definition of an IVD, which is used to examine specimens derived from the human body to provide information for diagnostic purposes.
- Device Description: The description mentions it's "Ivophilized human serum containing the compounds usually determined in clinical chemistry laboratories." This confirms it's a reagent or material used in a laboratory setting for testing human specimens.
- Predicate Device: The mention of a predicate device, "ROCHE Calibrator for Automated systems (C.f.a.s.)", which is also a calibrator used in clinical chemistry, further supports its classification as an IVD.
Therefore, based on the provided information, Calibrador A plus fits the definition of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
For the quantitative calibration of WIENER LAB's clinical chemistry procedures. Calibrador A plus is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens.
For the quantitative calibration of Wiener lab's clinical chemistry procedures. The "Calibrador A plus" is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens.
Product codes
JIX
Device Description
Calibrador A plus consists of Ivophilized human serum containing the compounds usually determined in clinical chemistry laboratories in the appropriate concentrations to ensure optimum calibration of clinical chemistry procedures. Such concentrations are lot-specific and are provided in product insert.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
ROCHE Calibrator for Automated systems (C.f.a.s.) (Cat. Nº759350)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 862.1150 Calibrator.
(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
0
MAR 1 8 2003
Calibrador A Plus Wiener lab.
Image /page/0/Picture/2 description: The image is a circular seal or logo, likely representing a certification or quality standard. The text "Wiener lab." appears at the top of the circle, and "SISTEMA DE CALIDAD CERTIFICADO" is written along the bottom. In the center of the seal, "ISO 9001" is prominently displayed above a logo that includes the letters "TUV".
Image /page/0/Picture/3 description: The image shows the logo for Wiener lab. The logo consists of a stylized "W" inside of a circle on the left, followed by the text "Wiener lab." in a bold, sans-serif font. Below the company name is the text "Especialidades para Laboratorios Clínicos" in a smaller font.
WIENER LABORATORIOS S.A.I.C. - Riobamba 2944 - 2000 Rosario - Argentina
Phone +54 (341) 432-9191/6 - Fax +54 (341) 432-5454/5555
Internet:
Section 6 - Summary
510(k) Summary "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR 807.92"
"The assigned 510(k) number is: _ K 0 % 4 305 33
According to the requirements of 21 CFR 862.1150, the following Introduction information provides sufficient details to understand the basis of a determination of substantial equivalence.
| 6-1 Submitter
Name, Address,
Contact | Wiener Laboratorios S.A.I.C.
Riobamba 2944
2000 - Rosario - Argentina
Tel: 54 341 4329191
Fax: 54 341 4851986
Contact person: Viviana Cétola
Date Prepared: September 15, 2002 |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 6-2 Device Name | Proprietary name: Wiener lab. Calibrador A Plus.
Common name: Multianalyte Calibrator.
Classification name: Calibrator, Multi-Analyte Mixture.
Device Class II |
1
- 6-3 Predicate We claim substantial equivalence to the currently marketed ROCHE Calibrator for Automated systems (C.f.a.s.) (Cat. Device Nº759350).
Calibrador A plus consists of Ivophilized human serum 6-4 Device Description containing the compounds usually determined in clinical chemistry laboratories in the appropriate concentrations to ensure optimum calibration of clinical chemistry procedures. Such concentrations are lot-specific and are provided in product insert.
For the quantitative calibration of WIENER LAB's clinical 6-5 Intended Use chemistry procedures. Calibrador A plus is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens.
The WIENER LAB. Calibrador A Plus calibrator is substantially 6-6 Equivalencies equivalent to other products in commercial distribution intended and Differences for similar use. Most notably it is substantially equivalent to the currently marketed ROCHE Calibrator for Automated systems (C.f.a.s.)
The following table illustrates the similarities and differences between the WIENER LAB. Calibrador A Plus calibrator and the currently marketed ROCHE Calibrator for Automated systems (C.f.a.s.)
| | ROCHE
C.f.a.s. | WIENER LAB.
Calibrador A Plus |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Intended Use | For use in the calibration
of Roche methods on
clinical chemistry
analyzers. | For the quantitative
calibration of WIENER
LAB's clinical chemistry
procedures. |
| Format | Lyophilized pooled human sera with constituents
added as required to obtain desired components
levels. | |
| Continued on next page | | |
| | ROCHE
C.f.a.s. | WIENER LAB.
Calibrador A Plus |
| Stability | Provided Reagents: stable in refrigerator (2-10°C)
until expiration date printed on label.
Reconstituted calibrator: stable for 8 hours at room
temperature, 2 days refrigerated or 1 month frozen
(-20°C), with exceptions noted in label. | |
| Levels | Single Level. | |
| | Albumin | Albumin |
| | Direct bilirubin | Direct bilirubin |
| | Total bilirubin | Total bilirubin |
| | Calcium | Calcium |
| | Cholesterol | Cholesterol |
| | Creatinine | Creatinine |
| | HDL Cholesterol | HDL Cholesterol |
| Constituent
Analytes | Glucose | Glucose |
| | Iron | Iron |
| | Magnesium | Magnesium |
| | Phosphorus | Phosphorus |
| | Total proteins | Total proteins |
| | Triglycerides | Triglycerides |
| | Uric acid | Uric acid |
| | Urea | Urea |
| | Enzymes | |
| | Electrolytes | |
2
'
3
Above mentioned data show substantial equivalency to the 6-7 Conclusion predicate device.
4
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a bird-like figure, composed of three curved lines that suggest wings and a head.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR 1 8 2003
Dr. Viviana Cetola QC/QA Manager Weiner Laboratorios S.A. I. C. Riobamba 2944 2000 Rosario, Santa Fe Argentina
Re: K024305
Trade/Device Name: Weiner lab. Calibrador A plus Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIX Dated: December 13, 2002 Received: December 24, 2002
Dear Dr. Cetola:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
5
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
6
CDRH ODE
Page | of____________________________________________________________________________________________________________________________________________________________________________ |
---|---|
------ | -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
510(k) Number (if known): | K024305 |
---|---|
Device Name: | Wiener lab. |
Calibrador A plus |
Indications For Use:
- ·
For the quantitative calibration of Wiener lab's clinical chemistry procedures. The "Calibrador A plus" is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) | OR | Over-The-Counter Use (Optional Format 1-2-96) |
---|---|---|
--------------------------------------- | ---- | ----------------------------------------------- |
Iran Cooper
(Division Sign-Off) | |
---|---|
Division of Clinical Laboratory Devices | |
510(k) Number | K024305 |