K Number
K024032
Manufacturer
Date Cleared
2002-12-30

(24 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery is intended for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.

Device Description

The Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery features one (1) Muli-purpose shaped guide catheter for passing balloon catheters or appropriate leads and one (1) guide catheter dilator to facilitate guide catheter passage.

The Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery contains a dilator that is cleared for market distribution via 510(k).

The key change being made to the catheter is the change in the shape of the distal section in order to respond to the needs of physician preferences and to address subtle anatomical differences between patients. Other changes include the elimination of the existing system components (i.e. adjustable hemostasis valves, guidewires, guidewire clip, guide catheter slitters) which are commercialised with the Attain™ Access 6218A Left-heart delivery system, with the exception of the guide catheter dilator. The revised configuration of one (1) guide catheter and one (1) guide catheter dilator per package has necessitated changes to the associated packaging and labelling. In addition, the modified device will be given a new model number, Attain™ 6216A-MP.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device (Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery). It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through clinical studies or standalone algorithm performance.

Therefore, many of the requested categories are not applicable as this type of submission generally relies on bench testing, material comparisons, and a comparison of intended use to predicate devices, rather than extensive clinical performance studies with specific statistical endpoints.

Here's a breakdown based on the provided text:


  1. Table of acceptance criteria and the reported device performance

    This information is not explicitly provided in the document. The submission states: "In vitro testing, packaging and sterilization validation testing was performed to support substantial equivalence to the predicate devices. The results of this testing indicate that the Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery meets all of its design and performance requirements." However, the specific acceptance criteria (e.g., tensile strength, flexibility, lubricity) and their corresponding performance values are not detailed. The performance is summarized as "meets all of its design and performance requirements."

  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified. The document mentions "In vitro testing" but does not provide details on the number of units tested.
    • Data Provenance: Not specified. "In vitro testing" typically implies laboratory testing, not human subject data.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a catheter, and its performance evaluation for a 510(k) submission generally involves physical and mechanical testing, not interpretation by medical experts to establish "ground truth."

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this type of testing does not involve adjudication of expert opinions.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a physical catheter, not an AI or diagnostic imaging device that would involve human readers or AI assistance.

  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The device is a physical catheter, not an algorithm.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for this device's evaluation would be engineering specifications and standards for physical and mechanical properties, not expert consensus, pathology, or outcomes data, which are relevant for diagnostic or treatment efficacy studies.

  8. The sample size for the training set

    Not applicable. This device is not an AI algorithm that requires a training set.

  9. How the ground truth for the training set was established

    Not applicable. This device is not an AI algorithm, so there is no training set or associated ground truth establishment process.

{0}------------------------------------------------

DEC 3 0 2002

510(K) Summary of Substantial Equivalence

Date preparedDecember 3, 2002
Submitter:Medtronic IrelandParkmore Business Park WestGalway, Ireland.
Submission Correspondent:Lynn JensenSr. Regulatory Affairs Specialist7000 Central Avenue N.E.Minneapolis, MN 55432
Telephone:(763) 514-4459
Fax:(763) 514-6424
E-Mail:lynn.a.jensen@medtronic.com
Proprietary Name:Attain ™ 6216A-MP Multi-purpose GuideCatheter for Left-heart delivery
Common Name:Catheter, Percutaneous
Device Classification:Class II, 21 CFR § 870.1250
Product Code:74 DQY

Performance Standard

Performance standards do not currently exist for these devices. None established under Section 514.

Device Description

The Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery features one (1) Muli-purpose shaped guide catheter for passing balloon catheters or appropriate leads and one (1) guide catheter dilator to facilitate guide catheter passage.

The Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery contains a dilator that is cleared for market distribution via 510(k).

The key change being made to the catheter is the change in the shape of the distal section in order to respond to the needs of physician preferences and to address subtle anatomical differences between patients. Other changes include the elimination of the existing system components (i.e. adjustable hemostasis valves, guidewires, guidewire clip, guide catheter slitters) which are commercialised with

{1}------------------------------------------------

CONFIDENTIAL

May not be reproduced without written permission from Medtronic, Inc.

the Attain™ Access 6218A Left-heart delivery system, with the exception of the guide catheter dilator. The revised configuration of one (1) guide catheter and one (1) guide catheter dilator per package has necessitated changes to the associated packaging and labelling. In addition, the modified device will be given a new model number, Attain™ 6216A-MP.

Indications for Use

The intended use of the Medtronic Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery is for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus

Substantially Equivalent Device

Attain LDS 6216A and Attain Access 6218A Left-heart Delivery System Predicate Devices

Attain™ 6216A-MP Multi-purposeGuide Catheterfor Left-heartdeliveryPredicate DevicePredicate Device ManufacturerPredicate 510(k)
6216A-MPMulti-purposeGuide CatheterMedtronic Attain LDS6216A Left-heart DeliverySystemMedtronic, Inc.Minneapolis, MN 55432K021587
Medtronic Attain Access6218A Left-heart DeliverySystemMedtronic, Inc.Minneapolis, MN 55432K021589
Guide CatheterDilatorMedtronic Attain Access6218A Left-heart DeliverySystemMedtronic, Inc.Minneapolis, MN 55432K021589

Summary of Studies

In vitro testing, packaging and sterilization validation testing was performed to support substantial equivalence to the predicate devices. The results of this testing indicate that the Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery meets all of its design and performance requirements.

Biocompatibility Information

The Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery is biocompatible and meets the requirements of ISO 10993-1. The materials used to fabricate the Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery are identical to those used to fabricate the guide catheters of the predicate devices.

{2}------------------------------------------------

Sterilization Validation

The Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery is sterilized using a 100% Ethylene Oxide (ETO) sterilization process. Appropriate processes for sterilizing the devices were validated.

Conclusion (Statement of Equivalence)

Through the data and information presented, as well as similarities to a legally marketed device, Medtronic, Inc. considers the Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery to be substantially equivalent to the previously discussed legally marketed predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Medtronic, Inc. c/o Ms. Lynn Jensen Senior Regulatory Affairs Specialist Cardiac Rhythm Management 7000 Central Avenue, N.E. Minneapolis, MN 55432

Re: K024032

Trade Name: Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-Heart Delivery Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: DOY Dated: December 3, 2002 Received: December 6, 2002

Dear Ms. Jensen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special ('ontrols) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You musi comply with all the Act's requirements, including, but not limited to: registration and lisung (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in vour Section 510(k) premarket notification. The FDA finding of substantial equivalence of vour device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

J. QUTTA
Bram D. Zuckerman, M.D.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

. CONFIDENTIAL May not be reproduced without written permission from Medtronic, Inc. .

.

:

. :

:.

ing the .

. . . . .

Indications for Use Statement

510(k) Number (if known):N/A
Device Name:Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery
Indications For Use:The Attain™ 6216A-MP Multi-purpose Guide Catheter for Left-heart delivery is intended for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseX
(Per 21 CFR 801.109)

OR

Over-The-Counter Use_________________
-----------------------------------------

Signature

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Numbe

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).