K Number
K021589
Manufacturer
Date Cleared
2002-05-30

(15 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Attain™ Access 6218A Left-heart delivery system is intended for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.

Device Description

The left-heart delivery system features two guide wires to facilitate venous access, adjustable hemostasis valves to reduce blood loss during the implant procedure, two guide catheters for passing venogram balloon catheters or appropriate leads, a guide catheter dilator to facilitate guide catheter passage, guide catheter slitters for removing guide catheters, and a guide wire clip to help contain the guide wire in the sterile field.

The Attain Access 6218A Left-heart delivery system combines devices that are either cleared for market distribution via 510(k) or are exempt from premarket notification because of Class I designation.

The key changes being made to the system are the addition of Tungsten Carbidefilled polyether block amide material to the guide catheter distal tip segment and the addition of extra slitters and adjustable hemostasis valves in a revised system configuration.

AI/ML Overview

The provided text describes a 510(k) submission for the Medtronic Attain™ Access 6218A Left-heart delivery system. This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets specific performance acceptance criteria through a standalone study with defined endpoints.

Here's a breakdown based on the request, highlighting what is and isn't available in the provided text:

1. A table of acceptance criteria and the reported device performance:

Acceptance Criteria (Implied)Reported Device Performance
BiocompatibilityFound to be biocompatible (according to ISO 10993-1)
Sterilization EfficacyAppropriately validated for 100% Ethylene Oxide (ETO) process
Design and Performance Requirements (General)Meets all design and performance requirements
Functional Equivalence to Predicate Device (e.g., Guide Catheters, Dilator, Hemostasis Valve, Guide Wires, Slitters, Guide Wire Clip)Substantially equivalent to Medtronic Attain Access 6218 Left-heart Delivery System (K012083) and Class I exempt device (for Guide Wire Clip)
Mechanical/Physical Properties (e.g., material integrity, functionality of components)In vitro testing performed to support substantial equivalence.

Explanation: The document does not explicitly state numerical acceptance criteria or specific performance metrics as would be found in a clinical trial or a detailed engineering specification report. Instead, it relies on demonstrating that the device performs as well as or similarly to a legally marketed predicate device, and that it meets general design and performance requirements supported by in vitro and biocompatibility testing.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified for any particular test. The document mentions "representative samples" for biocompatibility testing.
  • Data Provenance: Not explicitly stated. The testing was performed internally by Medtronic.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This submission does not involve clinical data requiring expert review for ground truth establishment. The evaluation is based on device characteristics, in vitro testing, and comparison to a predicate device.

4. Adjudication method for the test set:

  • Not applicable. No adjudication method is mentioned as there's no expert review of cases or outcomes to adjudicate.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is not an AI/software device and no MRMC study was performed or mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a physical medical device (delivery system), not an algorithm or AI.

7. The type of ground truth used:

  • Not explicitly stated as "ground truth" in the context of clinical outcomes or pathology. For biocompatibility, the "ground truth" is adherence to ISO 10993-1 standards. For sterilization, it's successful validation of the ETO process. For substantial equivalence, the "ground truth" is the established characteristics and performance of the predicate device.

8. The sample size for the training set:

  • Not applicable. There is no "training set" as this is not a machine learning or AI device.

9. How the ground truth for the training set was established:

  • Not applicable.

Summary of the Study that Proves the Device Meets Acceptance Criteria:

The "study" described is a set of in vitro testing, biocompatibility testing, and sterilization validation.

  • Biocompatibility Testing: Performed on "representative samples" of the Attain product. The criteria for acceptance were based on adherence to ISO 10993-1, and the results indicated that "all materials were found to be biocompatible."
  • Sterilization Validation: The device is sterilized using a 100% Ethylene Oxide (ETO) process. "Appropriate processes for sterilizing the devices were validated," implying that the validation study confirmed the sterilization method effectively renders the device sterile.
  • In vitro Testing: Mentioned as being performed "to support substantial equivalence to the predicate device." The results indicated that the "Attain Access 6218A meets all of its design and performance requirements." Specific parameters or acceptance criteria for this in vitro testing are not detailed in the provided text, but it would likely include mechanical strength, flow characteristics, and component inter-compatibility, all compared against the predicate device or internal specifications.

The overall conclusion is based on demonstrating substantial equivalence to a previously cleared predicate device (Medtronic Attain Access 6218 Left-heart Delivery System, K012083) and adherence to recognized standards (ISO 10993-1 for biocompatibility). The 510(k) process accepts this approach for devices that are similar to existing ones.

{0}------------------------------------------------

MAY 3 0 2002

K021589

CONFIDENTIAL

May not be reproduced without written permission from Medtronic, Inc.

510(K) Summary of Substantial Equivalence

Date prepared Submitter:

Submission Correspondent:

Telephone: Fax:

E-Mail:

Proprietary Name:

Common Name: Device Classification: Product Code:

May 13, 2002

Medtronic, Inc. Parkmore Business Park West Galway, Ireland.

Karen Reidt, RAC Sr. Product Regulation Manager 7000 Central Avenue N.E. Minneapolis. MN 55432

(763) 514-3914

(763) 514-6424

karen.reidt@medtronic.com

Attain™ Access 6218A Left-heart delivery system

Catheter, Percutaneous

Class II, 21 CFR § 870.1250

74 DQY

Performance Standard

Performance standards do not currently exist for these devices. None established under Section 514.

Device Description

The left-heart delivery system features two guide wires to facilitate venous access, adjustable hemostasis valves to reduce blood loss during the implant procedure, two guide catheters for passing venogram balloon catheters or appropriate leads, a guide catheter dilator to facilitate guide catheter passage, guide catheter slitters for removing guide catheters, and a guide wire clip to help contain the guide wire in the sterile field.

The Attain Access 6218A Left-heart delivery system combines devices that are either cleared for market distribution via 510(k) or are exempt from premarket notification because of Class I designation.

The key changes being made to the system are the addition of Tungsten Carbidefilled polyether block amide material to the guide catheter distal tip segment and

{1}------------------------------------------------

CONFIDENTIAL

May not be reproduced without written permission from Medtronic, Inc.

the addition of extra slitters and adjustable hemostasis valves in a revised system configuration.

Indications for Use

The intended use of the Medtronic Attain Access 6218A Left-heart delivery system is for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus

Substantially Equivalent Device

Attain Access6218A Left-heartDelivery SystemDevicePredicate DevicePredicate Device ManufacturerPredicate 510(k)
45 cm and 50 cmGuide CathetersMedtronic Attain Access6218 Left-heart DeliverySystemMedtronic, Inc.Minneapolis, MN 55432K012083
Guide CatheterDilatorMedtronic Attain Access6218 Left-heart DeliverySystemMedtronic, Inc.Minneapolis, MN 55432K012083
AdjustableHemostasis ValveMedtronic Attain Access6218 Left-heart DeliverySystemMedtronic, Inc.Minneapolis, MN 55432K012083
120 cm GuideWiresMedtronic Attain Access6218 Left-heart DeliverySystemMedtronic, Inc.Minneapolis, MN 55432K012083
4 Fr and 6 FrGuide CatheterSlittersMedtronic Attain Access6218 Left-heart DeliverySystemMedtronic, Inc.Minneapolis, MN 55432K012083
Guide Wire ClipClass I device, exempt from premarket notification

Attain Access 6218A Left-heart Delivery System Predicate Device

Summary of Studies

In vitro testing and biocompatibility/sterilization testing was performed to support substantial equivalence to the predicate device. The results of this testing indicates that the Attain Access 6218A meets all of its design and performance requirements.

Biocompatibility Information

Biocompatibility testing was performed on representative samples of Attain product. Complete testing according to ISO 10993-1 was conducted and all materials were found to be biocompatible.

{2}------------------------------------------------

Sterilization Validation

The Attain Access 6218A Left-heart delivery system is sterilized using a 100% Ethylene Oxide (ETO) sterilization process. Appropriate processes for sterilizing the devices were validated.

Conclusion (Statement of Equivalence)

Through the data and information presented, as well as similarities to a legally Through the data and microniation the Attain Access 6218A Left-heart marketed device, medically equivalent to the previously discussed legally marketed predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, representing the department's mission to protect the health of all Americans and provide essential human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 3 0 2002

Ms. Karen Reidt Principal Regulatory Affairs Specialist Cardiac Rhythm Management Medtronic, Inc. 7000 Central Avenue N.E. Minneapolis, MN 55432

Re: K021589

Trade Name: Attain™ Access 6218A Left-Heart Delivery System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: DQY Dated: May 13, 2002 Received: May 15, 2002

Dear Ms. Reidt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Karen Reidt

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Reala Well

Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

CONFIDENTIAL May not be reproduced without written permission from Medtronic, Inc.

Indications for Use Statement

NA KO21589 510(k) Number (if known): Attain™ Access 6218A Left-heart delivery system Device Name:

Indications For Use:

The Attain™ Access 6218A Left-heart delivery system is intended for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-The-Counter Use_

Division of Cardiovascular & Respiratory Devices
510(k) Number K021589

OR

Special 510(k): Device Modification Premarket Notification Attain™ Access 6218A Left-heart delivery system

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).