(56 days)
The "Wiener lab. Bilirrubina Total AA" test system is a quantitative in vitro diagnostic device intended to be used in the quantitative determination of total bilirubin in human sera and heparinized plasmas on both manual and automated systems. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block.
Total bilirubin (both conjugated and free) is measured using a stable dichlorophenyl diazonium salt (DPD) to form an azobilirubin compound with maximal O.D. at 546 nm. Surfactants are used as reaction accelerators. The amount of bilirubin is determined by measuring the absorbance of this pigment.
The provided document describes the predicate device and the new device being submitted for 510(k) clearance, the Wiener lab. BILIRRUBINA TOTAL AA test system. It focuses on establishing substantial equivalence to the predicate device through a comparison of their features and performance characteristics, particularly related to analytical performance (precision, linearity, expected values).
However, the document does not contain information about:
- Specific acceptance criteria used in a study to prove the device meets these criteria. Instead, it presents performance characteristics and compares them to the predicate.
- A dedicated study design with a test set, training set, ground truth establishment, or expert involvement as would be typical for AI/ML-based device evaluation.
- Multi-reader multi-case (MRMC) comparative effectiveness studies.
- Standalone algorithm performance.
The information provided is typical for an IVD (In Vitro Diagnostic) device submission where the focus is on analytical performance characteristics rather than clinical diagnostic accuracy based on image analysis or human interpretation.
Therefore, many of the requested fields cannot be filled from the given text.
Here's a breakdown of what can be extracted and what cannot:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" but presents performance data for the new device and compares it to the predicate. The "acceptance" is implied by the concluding statement of substantial equivalence.
| Performance Characteristic | Predicate Device (DMA Test System) | Wiener LAB. Test System (Reported Performance) |
|---|---|---|
| Intended Use | Quantitative determination of total bilirubin in human serum and EDTA or heparinized plasma. | Quantitative determination of total bilirubin in human serum and heparinized plasma. |
| Test Principle | Total bilirubin measured using a stabilized diazonium salt of 3,5-dichloroaniline. | Total bilirubin measured using a stable dichlorophenyl diazonium salt (DPD). |
| Reagents | 3,5-dichlorophenyldiazonium tetrafluoroborate and surfactants. | R1: hydrochloric acid and surfactant; R2: dichlorophenyl diazonium salt. Sample blank: hydrochloric acid and surfactant. |
| Preparation of Working Reagent | Ready to use. | Reconstitute each Reagent 2 vial with stated volume of Reagent 1. |
| Stability of Final Color | 60 minutes at room temperature (15-30ºC). | 30 minutes at room temperature. |
| Wavelength of Reading | 540 nm. | 546 nm. |
| Calibration | Single point. | (Not specified for Wiener Lab, implied similar if not stated otherwise in comparison table). |
| Linearity | 20 mg/dl. | (Not specified for Wiener Lab, implied similar if not stated otherwise in comparison table). |
| Expected values | 0.0 - 1.5 mg/dl. | Adults: up to 1.0 mg/dl; Newborns: ranging from 2.0 to 16 mg/dl depending on birth condition and days of life. |
| Within-run precision | Low Serum Control: CV = 5.00%; Moderate Serum Control: CV = 2.20%; High Serum Control: CV= 1.56%. | Normal Level Serum: CV = 1.50%; High Level Serum: CV = 0.98%. |
| Run-to-run precision | Low Serum Control: CV = 6.40%; Moderate Serum Control: CV = 0.83%; High Serum Control: CV= 2.10%. | Normal Level Serum: CV = 2.68%; High Level Serum: CV = 1.08%. |
2. Sample size used for the test set and the data provenance:
This information is not provided in the document. The precision data (within-run and run-to-run) would have been generated using a laboratory test, but the specific sample sizes for these tests are not mentioned. There is no information about the country of origin or if the data was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This is an IVD device, and the "ground truth" for analytical performance (precision, linearity) would be based on reference methods or spiked samples, not expert interpretation.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable. As above, this concept applies to human interpretation of diagnostic results, not the analytical performance of an IVD reagent.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No. This is not an AI-assisted diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is an IVD reagent kit; its performance is standalone in the sense that it produces a quantitative value, but there's no "algorithm" in the context of AI/ML.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
For precision, the ground truth would be the known concentration of bilirubin in control samples as measured by a reference method or validated assays. For expected values, it would be based on large-scale population studies.
8. The sample size for the training set:
Not applicable. There is no "training set" in the context of AI/ML for this type of IVD device. The methods are chemical reactions, not learned algorithms.
9. How the ground truth for the training set was established:
Not applicable. See above.
{0}------------------------------------------------
NOV 1 2002
BILIRRUBINA TOTAL AA Wiener lab.
Image /page/0/Picture/3 description: The image shows the logo for Wiener lab. The logo consists of a stylized "W" symbol on the left, followed by the text "Wiener lab." in a bold, sans-serif font. Below the company name, there is the text "Especialidades para Laboratorios Clinicos" in a smaller font, which translates to "Specialties for Clinical Laboratories".
WIENER LABORATORIOS S.A.I.C. - Riobamba 2944 - 2000 Rosario - Argentina
Phone +54 (341) 432-9191/6 - Fax +54 (341) 432-5454/5555
Internet: http://www.wiener-lab.com.ar
Section 6 - Summary
510(k) Summary "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR 807.92"
יי "The assigned 510(k) number is: _
According to the requirements of 21 CFR 862.1110, the following Introduction information provides sufficient details to understand the basis of a determination of substantial equivalence.
| 6-1 Submitter Name, Address, Contact | Wiener Laboratorios S.A.I.C.Riobamba 29442000 - Rosario - ArgentinaTel: 54 341 4329191Fax: 54 341 4851986Contact person: Viviana CétolaDate Prepared: July 07, 2002 |
|---|---|
| 6-2 Device Name | Proprietary name: Wiener lab. BILIRRUBINA TOTAL AACommon name: Bilirubin (total) test system |
| Classification name: Diazo colorimetry, Bilirubin Device Class II,CIG, as per 21 CFR section 862.1110.Device Class II |
{1}------------------------------------------------
We claim substantial equivalence to the currently marketed 6-3 Predicate DMA Total Bilirubin Plus test system (Cat. Nº 1245). Device
Total bilirubin (both conjugated and free) is measured using a 6-4 Device stable dichlorophenyl diazonium salt (DPD) to form an Description azobilirubin compound with maximal O.D. at 546 nm. Surfactants are used as reaction accelerators.
The amount of bilirubin is determined by measuring the absorbance of this pigment.
- The WIENER LAB. BILIRRUBINA TOTAL AA test system is a 6-5 Intended Use quantitative in vitro diagnostic device intended to be used in the quantitative determination of total bilirubin in human sera and heparinized plasmas on both manual and automated systems. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block.
6-6 Equivalencies The WIENER BILIRUBIN TOTAL AA test system is and Differences substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed DMA Total Bilirubin Plus test system.
The following table illustrates the similarities and differences between the WIENER BILIRUBIN test system and the currently marketed DMA Total Bilirubin Plus test system.
| DMATest System | WIENER LAB.Test System | |
|---|---|---|
| Intended Use | Quantitativedetermination of totalbilirubin in human serumand EDTA or heparinizedplasma. | Quantitativedetermination of totalbilirubin in human serumand heparinized plasma. |
| Continued on next page | ||
| DMATest System | WIENER LAB.Test System | |
| Test Principle | Total bilirubin (bothconjugated and free) ismeasured using astabilized diazonium saltof 3,5-dichloroanilinewhich reacts withbilirubin to formazobilirubin withmaximum absorbance at540 nm. Surfactants areused as reactionaccelerators. | Total bilirubin (bothconjugated and free) ismeasured using a stabledichlorophenyl diazoniumsalt (DPD) to form anazobilirubin compoundwith maximal O.D. at 546nm. Surfactants are usedas reaction accelerators. |
| Reagents | Reagent: 3,5-dichlorophenyldiazoniumtetrafluoroborate andsurfactants. | R1: hydrochloric acidand surfactant.R2: dichlorophenyldiazonium salt.Reagent for sampleblank: hydrochloric acidand surfactant. |
| Preparation ofWorking Reagent | Ready to use | Reconstitute eachReagent 2 vial withstated volume ofReagent 1. |
| Stability of FinalColor | 60 minutes at roomtemperature (15-30ºC) | 30 minutes at roomtemperature |
| Wavelength ofReading | 540 nm | 546 nm |
| Calibration | Single point | |
| Linearity | 20 mg/dl | |
| Expected values | 0.0 - 1.5 mg/dl | - Adults: up to 1.0 mg/dl- Newborns: ranging from2.0 to 16 mg/dldepending on birthcondition (full term orpremature) and days oflife. |
| Continued on next page | ||
| DMATest System | WIENER LAB.Test System | |
| Within-runprecision | Low Serum Control:CV = 5.00%Moderate Serum Control:CV = 2.20%High Serum Control:CV= 1.56% | Normal Level Serum:CV = 1.50%High Level Serum:CV = 0.98% |
| Run-to-runprecision | Low Serum Control:CV = 6.40%Moderate Serum Control:CV = 0.83%High Serum Control:CV= 2.10% | Normal Level Serum:CV = 2.68%High Level Serum:CV = 1.08% |
{2}------------------------------------------------
{3}------------------------------------------------
6-7 Conclusion Above mentioned data show substantial equivalency to the predicate device.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 098 Gaither Road Rockville MD 20850
1 2002
Dr. Viviana Cetola OC/OA Manager Wiener Laboratorios S.A.I.C. Riobamba 2944 2000 Rosario, Santa Fe Argentina
K022962 Re:
Trade/Device Name: Wiener Lab Bilirrubina Total AA Regulation Number: 21 CFR 862.1110 Regulation Name: Bilirubin (total or direct) test system Regulatory Class: Class II Product Code: CIG Dated: August 30, 2002 Received: September 6, 2002
Dear Dr. Cetola:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
| 510(k) Number (if known): | K032962 |
|---|---|
| Device Name: | Wiener lab. |
| Bilirrubina Total AA |
Page ofIndications For Use:
The "Wiener lab. Bilirrubina Total AA" test system is a quantitative in vitro diagnostic device intended to be used in the quantitative determination of total bilirubin in human sera and heparinized plasmas on both manual and automated systems. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|---|---|
| Division of Clinical Laboratory Devices | |
| 510(K) Number: K022962 |
| Prescription Use | OR | Over-The-Counter Use |
|---|---|---|
| (Per 21 CFR 801.109) | (Optional Format 1-2-96) |
§ 862.1110 Bilirubin (total or direct) test system.
(a)
Identification. A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.(b)
Classification. Class II.