(26 days)
No
The description focuses on standard digital image processing tools and volumetric analysis, with no mention of AI or ML techniques.
No.
The device is an image analysis software package designed to support physicians in evaluating, documenting, and following up on lung lesions. It provides tools for volumetric analysis and classification, which are diagnostic and monitoring functions, not therapeutic.
Yes
The device is designed to support physicians in confirming the presence or absence of lung lesions and evaluating, documenting, and following up on such lesions using volumetric analysis, which directly aids in diagnosis and disease management.
Yes
The device is explicitly described as a "self-contained image analysis software package" and its description focuses solely on software functionalities and algorithms for processing CT data. There is no mention of accompanying hardware components or hardware-specific validation.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The description clearly states that Lung CARE CT is an image analysis software package that evaluates CT volume data sets. This data comes from medical imaging, not from biological samples.
- Intended Use: The intended use is to support physicians in evaluating and following up on lung lesions identified in CT scans. This is a diagnostic aid based on imaging, not a test performed on a biological sample.
The device processes medical images to provide information to a physician, which falls under the category of medical imaging software or a diagnostic imaging device accessory, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
Lung CARE CT is a self-contained image analysis software package for evaluating CT volume data sets. Combining enhanced commercially available digital image processing tools with optimized workflow and reporting tools, the software is designed to support the physician in confirming the presence of absence of physician-identified lung lesions (eg. nodules) in addition to evaluation, documentation and follow-up of any such lesions using standard or low-dose spiral CT scanning. This evaluation tool allows for volumetric analysis of pulmonary nodule or lesion size over time, helping the Physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue unambiguously, with respect to their size, dimensions, shape and position.
Product codes
90 JAK
Device Description
This premarket notification covers Siemens LungCARE CT software package. It is based on Siemens syngo software platform.
Lung CARE CT is a self-contained image analysis software package for cvaluating CT volume data sets. Combining enhanced commercially available digital image processing tools (MIP, MPR, SSD, VRT), evaluation tools (volumetric estimation using consistent standardized measurement protocol, comparator tool for nodule matching by synchronization of two datasets, classification of nodules using configurable descriptors) and reporting tools (targeted presets, saved lesion) with optimized workflow palette, the software package is designed to support the physician in confirming the presence of physician identified lung lesions (eg. nodulcs) in addition to evaluation, documentation and follow-up of any such lesions using standard or low-dose spiral CT scanning. This visualization tool allows for volumetric analysis of pulmonary nodule or lesion size over time, helping the Physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue unambiguously, with respect to their size, dimensions, shape and position.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
CT
Anatomical Site
lung
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Physician
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Kohl et al., performed a Lung Phantom bench testing study to determine the influence of the scanning and reconstruction parameters on reproducibility of volume measurements using LungCARE CT. The results indicate that the reproducibility of volumetric mcasurements is independent of reconstructed field of view but is slightly dependent on the reconstruction kernel used to perform the evaluation. The results also clearly demonstrate the advantage of using thin slice collimations for better reproducibility of volume measurements. Comparison of volument measurements performed at various dose levels clearly demonstrates that the usage of normal dose versus low dose brings no additional benefit to the volume estimation. In conclusion, the authors propose that the end-user should use thin slice collimations, medium kernel, low radiation dose with a full field of view reconstruction and last but not the least consistently use the same protocol for the best reproducibility of volumetric estimation.
Wormanns et al., performed a clinical evaluation of the reproducibility of volumetric measurements using LungCARE CT in 10 patients with pulmonary metastatic disease. In total 150 pulmonary nodules were manually marked and then evaluated. Volume and diameter of the nodules was calculated with the LungCARE CT software package. The proposed segmentation results, provided by the software package, were not modified by the user. The results indicate that the clearly defined compact type of pulmonary nodules have a better volume reproducibility, than the ill-defined types of nodules with multiple connections to pleura and/or vessels without user intervention by modifying the segmentation results. The authors concluded that they had evaluated a method that allows for the reliable estimation of volume growth but the user is cautioned to carry out the evaluation carefully and always critically assess the visual representation of the software proposed segmentation results especially for ill-defined nodules.
Key Metrics
Not Found
Predicate Device(s)
K013381, K941546, K923524/S2, K990426, K973010, K010938
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.
0
510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
I. GENERAL INFORMATION
Device Name and Classification
Product Name: | LungCARE CT software package |
---|---|
Common Name | 3D CT Reconstruction Software |
Classification Name: | Accessory to Computed Tomography System |
Classification Panel: | Radiology |
CFR Section: | 21 CFR §892.1750 |
Device Class: | Class II |
Product Code: | 90 JAK |
Establishment:
Importer/Distributor:
Siemens Medical Solutions, Inc. 186 Wood Avenue South Iselin, NJ 08830
Registration Number: 2240869
Manufacturing Facility:
Siemens AG Medical Solutions Henkestrasse 127 D-91052 Erlangen, Germany syngo is a registered trademark of Siemens AG
Contact Person: Praveen P Nadkami Technical Specialist Telephone: (732) 321-4950 Fax: (732) 321-4841
Date of Preparation of Summary: May 10th 2002
II. SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING THE SUBSTANTIAL EQUIVALENCE DETERMINATION
Device Description and Intended Use:
This premarket notification covers Siemens LungCARE CT software package. It is based on Siemens syngo software platform.
1
Lung CARE CT is a self-contained image analysis software package for cvaluating CT volume data sets. Combining enhanced commercially available digital image processing tools (MIP, MPR, SSD, VRT), evaluation tools (volumetric estimation using consistent standardized measurement protocol, comparator tool for nodule matching by synchronization of two datasets, classification of nodules using configurable descriptors) and reporting tools (targeted presets, saved lesion) with optimized workflow palette, the software package is designed to support the physician in confirming the presence of physician identified lung lesions (eg. nodulcs) in addition to evaluation, documentation and follow-up of any such lesions using standard or low-dose spiral CT scanning. This visualization tool allows for volumetric analysis of pulmonary nodule or lesion size over time, helping the Physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue unambiguously, with respect to their size, dimensions, shape and position.
General Safety and Effectiveness Concerns:
The device labeling contains instructions for use and any necessary cautions and warning, to provide for safe and effective use of the device.
Risk management is cosured via a hazard analysis, which is used to identify potential hazards. These potential hazards arc controlled via software development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
Summary of Pre-clinical and Clinical Information:
Kohl et al., performed a Lung Phantom bench testing study to determine the influence of the scanning and reconstruction parameters on reproducibility of volume measurements using LungCARE CT.
The results indicate that the reproducibility of volumetric mcasurements is independent of reconstructed field of view but is slightly dependent on the reconstruction kernel used to perform the evaluation. The results also clearly demonstrate the advantage of using thin slice collimations for better reproducibility of volume measurements. Comparison of volument measurements performed at various dose levels clearly demonstrates that the usage of normal dose versus low dose brings no additional benefit to the volume estimation.
In conclusion, the authors propose that the end-user should use thin slice collimations, medium kernel, low radiation dose with a full field of view reconstruction and last but not the least consistently use the same protocol for the best reproducibility of volumetric estimation.
Wormanns et al., performed a clinical evaluation of the reproducibility of volumetric measurements using LungCARE CT in 10 patients with pulmonary metastatic disease. In total 150 pulmonary nodules were manually marked and then evaluated. Volume and diameter of the nodules was calculated with the LungCARE CT software package. The proposed segmentation results, provided by the software package, were not modified by the user.
2
The results indicate that the clearly defined compact type of pulmonary nodules have a better volume reproducibility, than the ill-defined types of nodules with multiple connections to pleura and/or vessels without user intervention by modifying the segmentation results. The authors concluded that they had evaluated a method that allows for the reliable estimation of volume growth but the user is cautioned to carry out the evaluation carefully and always critically assess the visual representation of the software proposed segmentation results especially for ill-defined nodules.
Substantial Equivalence:
The LungCARE CT software package, addressed in this premarket notification, is substantially equivalent to the following commercially available software package:
Manufacturer | Product | 510(k) | Clearance date |
---|---|---|---|
GE Medical | Advanced Lung Analysis-I | K013381 | 10/26/01 |
The LungCARE CT software package described in this 510(k) has the same intended use and similar technical characteristics as the commercially available software listed above.
In addition, the image processing, display and evaluation components of LungCARE are currently available software components on CT systems like Siemens Somatom Plus 4 with Volume Zoom option, K941546, cleared on September 20th, 1994, software options like the Volume Rendering Technique option, K923524/S2, cleared on May 17th 1994 and the Calcium Scoring Software package, K990426 cleared on 04/30/99, and Workstations like the Siemens RealTime 3D Diagnostic Workstation (3D Virtuoso), K973010, which received clearance on Nov 10th 1997 and the syngo Multimodality Workstation. K010938, cleared on 26th June 2001. LungCARE packages these image processing and image display components in an optimized workflow palette.
In summary, Siemens is of the opinion that LungCARE CT software package does not introduce any new potential safety risks and is substantially equivalent to and performs as well as the predicate software components and the predicate device.
3
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three heads, representing the department's focus on health, human services, and the well-being of the population. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Public Health Service
JUL 16 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Siemens Medical Solutions, Inc. % Mr. Wolfram Gmelin Technical Manager TÜV Rheinland of North America 12 Commerce Road NEWTON CT 06470
Re: K022013
Trade/Device Name: LungCARE CT Software Package Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II
Product Code: 90 JAK Dated: June 27, 2002 Received: July 2, 2002
Dear Mr. Gmelin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
4
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
8xx.1xxx | (301) 594-4591 |
---|---|
876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
Indication for use
510(k) Number (if known):
Device Name:
Lung CARE CT Software Package
Lung CARE CT is a self-contained image analysis software package for evaluating CT volume data sets. Combining enhanced commercially available digital image processing tools with optimized workflow and reporting tools, the software is designed to support the physician in confirming the presence of absence of physician-identified lung lesions (eg. nodules) in addition to evaluation, documentation and follow-up of any such lesions using standard or low-dose spiral CT scanning. This evaluation tool allows for volumetric analysis of pulmonary nodule or lesion size over time, helping the Physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue unambiguously, with respect to their size, dimensions, shape and position.
(Please do not write below this line - continue on another page if needed)
Concurrence of the CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) | |
---|---|
Division of Reproductive, Abdominal, and Radiological Devices | |
510(k) Number | K020013 |
Prescription Use (Per 21 CFR 801.109) | OR Over-The-Counter Use |
---|---|
--------------------------------------- | ------------------------- |