K Number
K021976
Date Cleared
2002-10-25

(130 days)

Product Code
Regulation Number
864.5400
Panel
HE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fibron-1 is a photo-optical instrument used for the performance of in-vitro diagnostic clotting testing of citrated plasma samples in the clinical laboratory. The Fibrin-1, which uses clot formation as an endpoint, may be used for the performance of the Prothrombin Time Test (PT) and the Activated Partial Thromboplastin Test (APTT).

Device Description

The Fibron-1 is a photo-optical instrument used for the performance of in-vitro diagnostic clotting testing of citrated plasma samples in the clinical laboratory. The instrument utilizes photo-optical principles for clot detection. The light source is a high intensity photodiode. The incubator block is temperature regulated to 36.5 - 37.5℃ and contains four measuring positions, 4 reagent and 4 cuvette pre-warming positions.

AI/ML Overview

Here's a summary of the acceptance criteria and study findings for the Fibron-1 Coagulation Analyzer, based on the provided 510(k) summary:

Acceptance Criteria and Reported Device Performance

Acceptance CriterionMetricFibron-1 PerformanceNotes
Instrument Correlation (PT)Correlation (vs ACL 100)0.972Strong positive correlation
Instrument Correlation (PT)Slope (vs ACL 100)1.24
Instrument Correlation (PT)Intercept (vs ACL 100)-2.46
Instrument Correlation (APT)Correlation (vs ACL 100)0.976Strong positive correlation
Instrument Correlation (APT)Intercept (vs ACL 100)6.68
Instrument Correlation (APT)Slope (vs ACL 100)0.763
Instrument Correlation (APT)Correlation (vs Electra 900C)0.981Strong positive correlation
Instrument Correlation (APT)Intercept (vs Electra 900C)2.05
Instrument Correlation (APT)Slope (vs Electra 900C)0.892
Within-run Precision (PT, Control Level 1)Average ± %CV12.0 ± 2.0 % (n=20)Compared favorably to MLA 900C (11.5 ± 2.0%)
Within-run Precision (PT, Control Level 2)Average ± %CV20.0 ± 1.9 % (n=19)Compared favorably to MLA 900C (18.7 ± 1.2%)
Day-to-Day Precision (PT, Normal)Average ± %CV11.6 ± 3.2 %Compared favorably to MLA 900C (11.5 ± 1.2%)
Day-to-Day Precision (PT, Low Abnormal)Average ± %CV19.6 ± 2.1 %Compared favorably to MLA 900C (18.3 ± 2.0%)
Within-run Precision (PT, Normal Control)Average %CV of Duplicates1.28% (n=18 duplicates)
Within-run Precision (PT, Low Abnormal)Average %CV of Duplicates1.12% (n=17 duplicates)
Within-run Precision (PT, High Abnormal)Average %CV of Duplicates1.01% (n=18 duplicates)
Within-run Precision (APTT, Normal Control)Average %CV of Duplicates1.90% (n=18 duplicates)
Within-run Precision (APTT, Low Abnormal)Average %CV of Duplicates0.86% (n=16 duplicates)
Within-run Precision (APTT, High Abnormal)Average %CV of Duplicates2.45% (n=18 duplicates)

Study Details

  1. Sample size used for the test set and the data provenance:

    • Instrument Correlation Studies (PT & APTT):

      • The exact sample size for the correlation studies (Fibron-1 vs. ACL 100 and Fibron-1 vs. Electra 900C) is not explicitly stated as a total number of patients, but rather states "Specimens were evaluated from healthy individuals and from patients with different pathological conditions." The clotting time range observed was 10 to 39 seconds.
      • Data Provenance: "in-house and at a community hospital." This suggests a mix of controlled laboratory environments and a real-world clinical setting. The country of origin is not specified but given the submitter's address (Newton, MA) and the FDA submission, it's likely US-based, at least for the community hospital. The study appears to be prospective, as specimens were evaluated as part of this specific testing.
    • Precision Studies:

      • Within-run Precision:
        • PT Control Level 1: n = 20
        • PT Control Level 2: n = 19
      • Day-to-Day Precision:
        • PT Normal Control: Day 1 (n=20), Day 2 (n=4), Day 3 (n=4), Day 4 (n=4), Day 5 (n=4)
        • PT Low Abnormal: Day 1 (n=19), Day 2 (n=4), Day 3 (n=4), Day 4 (n=4), Day 5 (n=4)
      • Precision (Duplicates):
        • PT Normal Control: 18 duplicates
        • PT Low Abnormal: 17 duplicates
        • PT High Abnormal: 18 duplicates
        • APTT Normal Control: 18 duplicates
        • APTT Low Abnormal: 16 duplicates
        • APTT High Abnormal: 18 duplicates
      • Data Provenance: "in-house" based on the context of the document.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This device is a laboratory instrument that provides quantitative measurements of clotting times. The "ground truth" for the test set is established by the measurements from the predicate devices (ACL 100, MLA-900C, Electra 900C) and established laboratory control materials. There is no mention of human experts establishing ground truth for the test set in the way one would for diagnostic imaging.
  3. Adjudication method for the test set:

    • Not applicable. This is a comparison of instrument performance against predicate devices and known control values, not a subjective interpretation task requiring adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was performed. This device is a standalone laboratory instrument, not an AI-assisted diagnostic tool for human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the entire study focuses on the standalone performance of the Fibron-1 instrument. Its accuracy and precision are evaluated against predicate devices and control materials without any human interpretive component beyond operating the machines.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for evaluating the Fibron-1's performance is derived from:
      • Measurements from legally marketed predicate devices (ACL 100, MLA-900C, Electra 900C).
      • Known values of laboratory control materials (Normal, Low Abnormal, High Abnormal levels).
      • Actual clotting times from patient and healthy individual samples.
  7. The sample size for the training set:

    • This is a traditional medical device (a photo-optical instrument), not an AI/ML-based device that would typically have a separate "training set" for an algorithm. Therefore, no training set is mentioned or applicable in the context of this 510(k) summary.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this type of device.

§ 864.5400 Coagulation instrument.

(a)
Identification. A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.(b)
Classification. Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.