K Number
K021428

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-07-05

(63 days)

Regulation Number
862.1225
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CLINITEK ATLAS PRO 12 Reagent Pak system is intended for use in at-risk patient groups to assist diagnosis in the following areas: kidney function, carbohydrate metabolism, urinary tract infections and liver function. The system also measures physical characteristics, including acid-base balance and urine concentration. Test results can be used along with other diagnostic information to rule out certain disease states. The CLINITEK ATLAS PRO Reagent Pak system is for professional use in centralized laboratory locations such as in hospitals and reference laboratories.

Device Description

The CLINITEK ATLAS PRO 12 Reagent Pak is a roll of firm plastic to which are affixed 490 reagent strips. It is for use with the CLINITEK ATLAS® Automated Urine Chemistry Analyzer. Each reagent strip contains 11 separate reagent areas to test for low level protein (15 mg/dL albumin), high level protein (>30 mg/dL protein), creatinine, occult blood, glucose, ketone (acetoacetic acid), leukocytes, nitrite, pH, bilirubin and urobilinogen. A protein-to-creatinine ratio is also determined. In addition, each strip contains a nonreactive 12th pad that is used for the determination of the specimen color.

AI/ML Overview

The provided text describes a 510(k) submission for the CLINITEK ATLAS PRO 12 Reagent Pak System. However, it does not include specific quantitative acceptance criteria or detailed results of a study designed to prove the device meets such criteria.

The document generally states: "The performance of the CLINITEK ATLAS PRO 12 Reagent Pak System was studied internally and in clinical settings. The studies demonstrated that typical users in centralized laboratories could obtain clinical test results that are comparable to commonly used laboratory tests." And, "Studies show that the product provides clinical results comparable to other test methods in current clinical practice."

Without specific acceptance criteria and an actual study report with raw data or summary statistics, it's impossible to completely fill out the requested table and answer many of the study-specific questions.

Here's an attempt to answer based only on the provided text, highlighting what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

Note: The provided text does not specify quantitative acceptance criteria for the CLINITEK ATLAS PRO 12 Reagent Pak System. It only makes a general statement about performance being "comparable to commonly used laboratory tests." Therefore, the "Acceptance Criteria" column cannot be filled with specific numbers or thresholds. The "Reported Device Performance" is also broad.

Performance Metric (for Creatinine and Low Level Protein/Albumin Detection, and Protein-to-Creatinine Ratio)Acceptance CriteriaReported Device Performance
Creatinine Test Accuracy/ComparabilityNot specifiedComparable to commonly used laboratory tests
Low Level Protein (Albumin at 15 mg/dL) Test Accuracy/ComparabilityNot specifiedComparable to commonly used laboratory tests
Protein-to-Creatinine Ratio Accuracy/ComparabilityNot specifiedComparable to commonly used laboratory tests
Other analytes (Occult blood, glucose, ketone, leukocytes, nitrite, pH, bilirubin, urobilinogen, high level protein)Not specifiedComparable to commonly used laboratory tests

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified.
  • Data Provenance: "studied internally and in clinical settings." This implies a mix, but specific countries or retrospective/prospective nature are not described.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not specified. The document refers to "commonly used laboratory tests" as the comparison, implying these tests served as a reference standard, but the process of establishing ground truth for individual samples using experts is not detailed.

4. Adjudication method for the test set

  • Not specified.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This device is an automated urine chemistry analyzer system, not an AI-assisted diagnostic imaging or interpretation tool for human readers. Therefore, an MRMC study related to human reader improvement with AI assistance is not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, effectively. The "device" in this context (CLINITEK ATLAS PRO 12 Reagent Pak System used with the CLINITEK ATLAS® Automated Urine Chemistry Analyzer) operates as an automated system. The performance assessment describes its results as comparable to other laboratory methods. This fits the description of a "standalone" or "algorithm only" performance, as the system itself produces the results without direct human interpretive input being evaluated. The document states "typical users in centralized laboratories could obtain clinical test results," suggesting the system's output is directly used.

7. The type of ground truth used

  • The ground truth was established by "commonly used laboratory tests" or "other test methods in current clinical practice." This implies comparison to existing, validated diagnostic methods. Specifics of these methods (e.g., gold standard chemical assays for creatinine, micro-albuminuria tests for low-level protein) are not detailed.

8. The sample size for the training set

  • Not applicable as this is not a machine learning/AI model described to be "trained." This is a diagnostic reagent system.

9. How the ground truth for the training set was established

  • Not applicable (see point 8).

{0}------------------------------------------------

'JUL

510(k) SAFETY AND EFFECTIVENESS SUMMARY

Prepared:April 25, 2002
Submitter:Bayer Diagnostics
Address:Bayer Diagnostics511 Benedict Ave.Tarrytown, NY 10591phone: (914)-524-2446
Contact:Kenneth T. Edds, Ph.D.Manager, Regulatory Affairs
Device:
Trade/Proprietary Name:CLINITEK ATLAS® PRO™ 12 ReagentPak System
Common/Usual Name:Reagent Strips for UrinalysisTest for Creatinine in Urine
Document Control Number:K02/428
Classification:
Name:Creatinine test system
Class:II
CFR:21 CFR §862. 1225
Code:75 JFY

Predicate Devices:

Multistix PRO™ 11 Reagent Strips

Multistix PRO™ 10LS Reagent Strips

Multistix PRO™ 7G Reagent Strips

CLINITEK® 50, CLINITEK® 100, CLINITEK® 200+,

CLINITEK® 500 Urine Chemistry Analyzers

Manufacturer: Bayer Diagnostics

Page 1 of 3

{1}------------------------------------------------

Device Description:

The CLINITEK ATLAS PRO 12 Reagent Pak is a roll of firm plastic to which are affixed 490 reagent strips. It is for use with the CLINITEK ATLAS® Automated Urine Chemistry Analyzer. Each reagent strip contains 11 separate reagent areas to test for low level protein (15 mg/dL albumin), high level protein (>30 mg/dL protein), creatinine, occult blood, glucose, ketone (acetoacetic acid), leukocytes, nitrite, pH, bilirubin and urobilinogen. A protein-to-creatinine ratio is also determined. In addition, each strip contains a nonreactive 12th pad that is used for the determination of the specimen color.

Intended Use:

The CLINITEK ATLAS PRO 12 Reagent Pak system is intended for use in at-risk patient groups to assist diagnosis in the following areas: kidney function, carbohydrate metabolism, urinary tract infections and liver function. The system also measures physical characteristics, including acid-base balance and urine concentration. Test results can be used along with other diagnostic information to rule out certain disease states. The CLINITEK ATLAS PRO 12 Reagent Pak system is for professional use in centralized laboratory locations such as in hospitals and reference laboratories.

Technological Characteristics:

The CLINITEK ATLAS PRO 12 Reagent Pak system includes a new "Protein Low" reagent that allows "low level" detection of urine albumin at 15 mg/dL in addition to the current protein reagent used for detection of "high level" protein at 30 mg/dL. A new creatinine reagent is used for the estimation of the protein-to-creatinine ratio of the urine specimen based on all three tests. Use of the protein-to-creatinine ratio can assist in the diagnosis of kidney function by minimizing the impact of changes in the protein result due to exercisc, orthostatic proteinuria, diuresis and urine concentration. The ratio allows for the use of single-void specimens in the discrimination of normal and abnormal levels of protein. Ratio results of >150 mg/g are considered to be abnormal.

The assay for low level protein (albumin) is based on dye binding using the high affinity dye, Bis(3',3"-diiodo-4',4"-dihydroxy-5',5"-dinitrophenyl)-3,4,5,6tetrabromosulfonephthalein). At a constant pH, the development of a blue color is owing to the presence of albumin; the intensity of the color is proportional to the albumin concentration. The assay for creatinine is based on the peroxidase-like activity of copper creatinine complexes that catalyze the reaction of diisopropylbenzene dihydroperoxide and 3,3',5,5'-tetramethyl-benzidine.

{2}------------------------------------------------

Assessment of Performance:

The performance of the CLINITEK ATLAS PRO 12 Reagent Pak System was studied internally and in clinical settings. The studies demonstrated that typical users in centralized laboratories could obtain clinical test results that are comparable to commonly used laboratory tests.

Conclusion:

The CLINITEK ATLAS PRO 12 Reagent Pak System has been developed for use with the CLINITEK ATLAS Automated Urine Chemistry Analyzer. New tests for the determination of low levels of albumin and urine creatinine have been added to the CLINITEK ATLAS Automated Urine Chemistry Analyzer menu to allow for a determination of the protein-to-creatinine ratio of the urine specimen. Studies show that the product provides clinical results comparable to other test methods in current clinical practice.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

5 2002 JUL

Kenneth T. Edds, Ph.D. Manager, Regulatory Affairs Bayer Diagnostics 511 Benedict Avenue Tarrytown, NY 10591

K021428 Re:

Trade/Device Name: Clinitek Atlas® Pro™ 12 Reagent Pak System Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: Class II Product Code: JFY; JIP; JIP; JIL; LJX Dated: April 25, 2002 Received: May 3, 2002

Dear Dr. Edds:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): K091428

Device Name:

CLINITEK ATLAS® PRO™ 12 Reagent Pak System

Indications for Use:

The CLINITEK ATLAS PRO 12 Reagent Pak is a roll of firm plastic to which are affixed 490 reagent strips. It is for use with the CLINITEK ATLAS® Automated Urine Chemistry Analyzer. Each reagent strip contains 11 separate reagent areas to test for low level protein (15 mg/dL albumin), high level protein (>30 mg/dL protein), creatinine, occult blood, glucose, ketone (acetoacetic acid), leukocytes, nitrite, pH, bilirubin and urobilinogen. A protein-tocreatinine ratio is also determined. In addition, each strip contains a nonreactive 12" pad that is used for the determination of the specimen color.

The CLINITEK ATLAS PRO 12 Reagent Pak system is intended for use in at-risk patient groups to assist diagnosis in the following areas: kidney function, carbohydrate metabolism, urinary tract infections and liver function. The system also measures physical characteristics, including acid-base balance and urine concentration. Test results can be used along with other diagnostic information to rule out certain disease states. The CLINITEK ATLAS PRO Reagent Pak system is for professional use in centralized laboratory locations such as in hospitals and reference laboratories.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) NumberK021428
Prescription Use(Per 21 CFR 801.109)OROver-The-Counter Use(Optional Format 1-2-96)
------------------------------------------------------------------------------------------------

§ 862.1225 Creatinine test system.

(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.