K Number
K091428

Validate with FDA (Live)

Date Cleared
2010-01-27

(258 days)

Product Code
Regulation Number
868.1890
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ndd EasyOne Pro Respiratory Analysis System is designed for conducting lung function measurements in general or specialist practices or in hospitals. The EasyOne Pro Respiratory Analysis System can also be used outside of the laboratory when performing lung function screenings or measurements in occupational medicine. The EasyOne Pro Respiratory Analysis System is used to conduct slow and forced spirometry on adults and children starting at age 4. Measurement of Diffusing Capacity of the lung based on CO (DLCO) can be performed on adults and children starting at age 6.

Device Description

The ndd EasyOne Pro Respiratory Analysis System is a portable device for performing lung function measurements such as spirometry and diffusion capacity (DLCO) tests. Measured lung function parameters can be compared to predicted normal values that are computed based on gender, age, and body height of the patient; the applied equations are available in literature and/or published standards. The device consists of a compact main unit and a hand-held sensor. Spirometry tests can be performed by connecting the hand-held sensor to the main unit. In order to perform DLCO tests a valve unit must additionally be connected to the hand-held sensor. The valve unit is connected with the main unit with a gas supply tube. The spirette respiratory tube is a mouthpiece for single patient use, which is inserted in the sensor. In the valve unit there is also an accessory for single patient use inserted, which is called barriette. The barriette prevents the passage of microorganisms into EasyOne Pro Respiratory Analysis System. A touch screen is integrated in the main unit that provides the user interface. The main power switch, different connections as well as different ports are located on the rear panel of the main unit. The EasyOne Pro can be used as a stand-alone system and can be connected to a network. The device has built-in quality control to assure correct test performance and equipment function.

AI/ML Overview

The provided text describes the ndd EasyOne Pro Respiratory Analysis System, a device for lung function measurements, and details its substantial equivalence to predicate devices. It mentions testing performed but lacks specific quantifiable acceptance criteria and detailed study outcomes.

Here's an analysis based on the provided text sections {0-4}:

1. Table of Acceptance Criteria and Reported Device Performance:

The document broadly states conformance to standards and recommendations rather than specific numerical acceptance criteria.

Acceptance CriteriaReported Device Performance
SpirometryMeets ATS recommendations for accuracy and precision during dynamic wave form testing.
DLCO (Diffusion Capacity)Meets ATS recommendations for accuracy and precision, demonstrated by comparative measurements with another DLCO device and DLCO simulator testing.
Electrical SafetyConforms with IEC 60601-1 and IEC 60601-1-2 requirements.
BiocompatibilityMaterials used meet requirements in accordance with ISO 10993.
SoftwareSoftware verification and validation revealed that the EasyOne Pro Respiratory Analysis System meets the specified criteria.

2. Sample Size for the Test Set and Data Provenance:

The document states "Dynamic wave form testing was performed regarding the diagnostic spirometry tests" and "The DLCO performance was tested by using a DLCO simulator. In addition comparative measurements with another DLCO testing device". However, the specific sample sizes for these tests (e.g., number of waveforms or DLCO measurements) are not provided. The data provenance is not explicitly mentioned, but the nature of the tests (simulator and comparative measurements) suggests a controlled laboratory or clinical setting rather than solely retrospective data from human patients. The company is based in Zurich, Switzerland, so the testing might have occurred there or at a testing facility.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

The document does not specify the number of experts used for ground truth or their qualifications. The ground truth for spirometry and DLCO is based on simulator outputs and comparison with a predicate device, as well as adherence to ATS recommendations, which are established by expert consensus in respiratory medicine.

4. Adjudication Method for the Test Set:

The document does not describe an adjudication method involving multiple human readers for the test set. The validation appears to be based on direct measurement comparisons against simulators and a predicate device as per established standards (ATS recommendations).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or performed according to this document. The device is for lung function measurement, not for interpreting medical images or data that typically require human readers with and without AI assistance in an MRMC study. The comparison is between the device's measurements and established standards/predicate devices, not between human readers.

6. If a Standalone Performance Study Was Done:

Yes, a standalone performance study was done. The document states:

  • "Dynamic wave form testing was performed regarding the diagnostic spirometry tests."
  • "The DLCO performance was tested by using a DLCO simulator."
  • "comparative measurements with another DLCO testing device demonstrate that the EasyOne Pro Respiratory Analysis System meets the ATS recommendations for accuracy and precision for DLCO testing and the intended diagnostic spirometry tests."

These describe the testing of the EasyOne Pro's performance on its own, relative to reference methods (simulators, predicate devices) and standards (ATS recommendations), without direct human-in-the-loop interaction to evaluate its output.

7. The Type of Ground Truth Used:

The ground truth used for testing appears to be:

  • Established standards/guidelines: Specifically, ATS (American Thoracic Society) recommendations for lung function measurements.
  • Simulator outputs: For DLCO, "a DLCO simulator" was used.
  • Predicate device measurements: For DLCO, "comparative measurements with another DLCO testing device" were performed, implying the predicate device served as a reference standard.

8. The Sample Size for the Training Set:

The document does not provide any information regarding a "training set" or its sample size. This type of device (a measurement instrument rather than an AI/machine learning diagnostic tool) would typically be validated through engineering and performance testing against physical standards and known reference devices, rather than through a machine learning training/test set paradigm.

9. How the Ground Truth for the Training Set Was Established:

As no training set is mentioned (see point 8), this information is not applicable and not provided in the document.

{0}------------------------------------------------

K091425

JAN 2 7 2010

510(k) Summary 5

In accordance with 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitter Information:

Submitter name:ndd Medizintechnik AGTechnoparkstrasse 18005 Zurich, Switzerland
United statescontact person:Mr. Jerry MasielloOperations Manager
ndd Medical Technologies, Inc
Phone:(877) 904.0090 x11
Fax:(978) 824.3853
Mailto:jmasiello@nddmed.com
Date prepared:May 11, 2009

Device Name:

Proprietary name:EasyOne Pro Respiratory Analysis System
Common name:Pulmonary function testing device
Class:Class II according to 868.1890
Classification name:Calculator, Predicted Values, Pulmonary Function
Product code:BTY

Predicate Device:

Substantial Equivalence is claimed with the SensorMedics Vmax 229 Series Pulmonary Function Analysis Instrument, K942211. The ndd Medical Technologies EasyOne Pro Respiratory Analysis System is a successor of the EasyOne Spirometer, K993921. Compared to the EasyOne Spirometer the EasyOne Pro Respiratiory Analysis System provides DLCO (Diffusion Capacity) options in addition to spirometry tests.

Device Description:

The ndd EasyOne Pro Respiratory Analysis System is a portable device for performing lung function measurements such as spirometry and diffusion capacity (DLCO) tests. Measured lung function parameters can be compared to predicted normal values that are computed based on gender, age, and body height of the patient; the applied equations are available in literature and/or published standards. The device consists of a compact main unit and a hand-held sensor. Spirometry tests can be performed by connecting the hand-held sensor to the main unit. In order to perform DLCO tests a valve unit must additionally be connected to the hand-held sensor. The valve unit is connected with the main unit with a gas supply tube.

{1}------------------------------------------------

The spirette respiratory tube is a mouthpiece for single patient use, which is inserted in the sensor. In the valve unit there is also an accessory for single patient use inserted, which is called barriette. The barriette prevents the passage of microorganisms into EasyOne Pro Respiratory Analysis System. A touch screen is integrated in the main unit that provides the user interface. The main power switch, different connections as well as different ports are located on the rear panel of the main unit. The EasyOne Pro can be used as a stand-alone system and can be connected to a network. The device has built-in quality control to assure correct test performance and equipment function.

Intended Use:

The ndd EasyOne Pro Respiratory Analysis System is designed for conducting lung function measurements in general or specialist practices or in hospitals. The EasyOne Pro Respiratory Analysis System can also be used outside of the laboratory when performing lung function screenings or measurements in occupational medicine. The EasyOne Pro Respiratory Analysis System is used to conduct slow and forced spirometry on adults and children starting at age 4. Measurement of Diffusing Capacity of the lung based on CO (DLCO) can be performed on adults and children starting at age 6.

Comparison of Technological Characteristics:

The device has the same technological characteristics as the predicate devices. Testing was conducted to demonstrate that lung function measurements are as accurate and precise as the methods used in predicate devices. The new as well as the predicate devices meet the ATS recommendations for lung function measurements. The EasyOne Pro Respiratory Analysis System is therefore substantially equivalent to the predicate devices.

Summary of Testing:

Dynamic wave form testing was performed regarding the diagnostic spirometry tests. The DLCO performance was tested by using a DLCO simulator. In addition comparative measurements with another DLCO testing device demonstrate that the EasyOne Pro Respiratory Analysis System meets the ATS recommendations for accuracy and precision for DLCO testing and the intended diagnostic spirometry tests.

The device was tested to demonstrate conformance with IEC 60601-1 and IEC 60601-1-2 requirements for electrical safety.

The materials used meet the requirements for biocompatibility in accordance with ISO 10993.

Software verification and validation revealed that the EasyOne Pro Respiratory Analysis System meets the specified criteria.

Conclusion:

Based on the above, ndd Medical Technologies concluded that the EasyOne Pro Respiratory Analysis System is substantially equivalent to the legally marketed predicate devices and is safe and effective for its intended use.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font and is horizontally aligned. The text is likely part of a document or sign.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JAN 2 7 2010

ndd Medizintechnik AG C/O Mr. Jerry Masiello Operations Manager ndd Medical Technologies, Incorporated Two Dundee Park Andover, Massachusetts 01810

Re: K091428

Trade/Device Name: EasyOne Pro Respiratory Analysis System Regulation Number: 21 CFR 868.1890 · Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: II Product Code: BTY Dated: January 21, 2010 Received: January 27, 2010

Dear Mr. Masiello:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings or feathers. The eagle faces to the left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Page 2 - Mr. Masiello

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/cdrh/comp/ for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

h for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K091428

EasyOne Pro Respiratory Analysis System Device Name:

Indications for Use:

The ndd EasyOne Pro Respiratory Analysis System is designed for conducting lung function measurements in general or specialist practices or in hospitals. The EasyOne Pro Respiratory Analysis System can also be used outside of the laboratory when performing lung function screenings or measurements in occupational medicine. The EasyOne Pro Respiratory Analysis System is used to conduct slow and forced spirometry on adults and children starting at age 4. Measurement of Diffusing Capacity of the lung based on CO (DLCO) can be performed on adults and children starting at age 6.

Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Schultheis

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K091428

§ 868.1890 Predictive pulmonary-function value calculator.

(a)
Identification. A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.(b)
Classification. Class II (performance standards).