K Number
K020355

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-05-14

(99 days)

Product Code
Regulation Number
864.7750
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The i-STAT PT is a prothrombin time test cartridge and is an in vitro diagnostic test intended for quantitative prothrombin time testing for the monitoring of oral anticoagulation therapy using fresh venous or capillary whole blood samples. The i-STAT PT test is not intended for evaluating individual factor deficiencies. The cartridge is to be used with the i-STAT Portable Clinical Analyzer with thermal control (Models 200 and 300), but will not run on the Philips Medical Systems (formerly Agilent Technologies) Blood Analysis Module (BAM). As part of the i-STAT System, the PT test is to be used by trained and certified health care professionals in accordance with a facility's policies and procedures.

The i-STAT PT, a prothrombin time test, is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or warfarin.

Device Description

The i-STAT PT test is contained in a single test cartridge. In use, approximately 40 microliters of fresh whole blood are placed in the cartridge as described below. The cartridge is inserted into the thermally controlled i-STAT Portable Clinical Analytical steps are performed automatically. Patient and user information may be entered into the analyzer via a keypad during the automated analysis cycle.

In the i-STAT PT test the endpoint is indicated by the appearance of an electroactive marker generated by the thrombin-mediated conversion of a synthetic substrate included in the reagent. Detection of the marker indicates generation of thrombin and therefore complete activation of the coagulation cascade. The result is reported as an International Normalized Ratio (INR) and, optionally, in seconds. The optionally displayed seconds is intended to reflect a typical plasma prothrombin time.

The PT test cartridge is assembled from plastic components that provide the conduits for fluid handling r no r r toot our negor chips. The coagulation test is identified to the user through the name and color code on the cartridge label and by the analyzer through features integral to the cartridge.

During the test the blood sample is mixed with reagents which are coated on the cartridge cover in a segment of the sensor channel. The reagent layer includes tissue thromboplastin as an activating agent, the thrombin substrate, a heparin-neutralizing enzyme, and inert matrix components. These reagents allow activation of the coagulation cascade and detection of clot formation.

Whole blood is introduced into the sample well of the cartridge at the sample port and the cartridge is closed and inserted into the analyzer. Insertion of the cartridge initiates a controlled and monitored sequence of steps in the instrument. These are:

  • Electrical contact is made between the analyzer electronic input circuits and the cartridge. . The analyzer identifies the type of cartridge being used and the tests contained in the cartridge.
  • The dry chips and sensor channel are heated to 37°C. .
  • The blood is then moved forward. Feedback from the fluid position sensor is used to allow . controlled oscillation of the blood segment resulting in dissolution of the reagent layer.
  • During the course of testing, the position of the blood segment is actively controlled to . maintain the length of the blood containing the reagent coincident with the endpoint detector.
  • Calculation of the sample clot time is performed and displayed. .
AI/ML Overview

The provided text describes the i-STAT PT Test, a prothrombin time test cartridge for monitoring oral anticoagulation therapy. The acceptance criteria and supporting studies are detailed in the "Summary of Non-Clinical Performance in Support of Substantial Equivalence" and "Summary of Clinical Test Performance is Support of Substantial Equivalence Claims" sections.

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a quantitative, pass/fail manner. Instead, it demonstrates performance in various conditions and compares it to a predicate device, the Coaguchek™ S System. The implicit acceptance criteria appear to be the demonstrated insensitivity to certain factors and comparable performance to the predicate device within reasonable clinical limits.

Implicit Acceptance Criteria (Demonstrated Performance)Reported Device Performance (i-STAT PT Test)Predicate Device Performance (Coaguchek™ S System) (where available)
Insensitivity to heparinInsensitive up to 1.0 U/mLSensitive to levels over 0.15 U/mL
Insensitivity to fibrinogen (manipulated samples)Insensitive to levels as low as 70 mg/dLNot explicitly stated for specific levels, but endpoint detection relies on fibrinogen conversion.
Insensitivity to fibrinogen (clinical samples)Insensitive between 142 and 528 mg/dLNot explicitly stated.
Insensitivity to hematocritHematocrits in the range of 24-52% do not significantly affect resultsNot explicitly stated.
Sensitivity to warfarin-impacted factorsSensitive to all factors impacted by warfarin therapyImplied by its intended use for anticoagulation monitoring.
Within-sample reproducibility5.4% (from duplicate sample testing in method comparison)5.9-9.8% in whole blood samples in therapeutic range
Imprecision (Plasma Controls, Level 1)%CV 4.5 at INR 1.1Not explicitly stated.
Imprecision (Plasma Controls, Level 2)%CV 6.9% at INR 2.5Not explicitly stated.
Total imprecision (whole blood controls)Not explicitly stated, but plasma control data is given.10.2% at INR 1.5, 15.1% at INR 3.7
Correlation with laboratory plasma instruments (Dade Innovin reagent)Correlation coefficients of 0.898, 0.943, 0.948 across three sitesComparison is against a lab method, not the predicate directly for this statistic.
Correlation between i-STAT capillary and venous samplesCorrelation coefficient of 0.962Not applicable, as this is an internal comparison for the i-STAT device.

2. Sample size used for the test set and the data provenance:

  • Method Comparison Study (i-STAT PT vs. Dade Innovin lab instrument):
    • Sample Sizes: N = 183 (Site 1), N = 180 (Site 2), N = 177 (Site 3). Total = 540 patient samples.
    • Data Provenance: The samples were from "patients undergoing routine monitoring of oral anticoagulation therapy." This implies prospective clinical data collected at "three external sites." The country of origin is not explicitly stated but is implicitly the USA given the FDA submission.
  • Capillary vs. Venous Sample Comparison Study:
    • Sample Size: N = 59.
    • Data Provenance: From "patients undergoing routine monitoring of oral anticoagulation therapy." This implies prospective clinical data from one site. The country of origin is not explicitly stated.
  • Other non-clinical studies: The sample sizes for heparin-spiked, manipulated fibrinogen, and manipulated hematocrit samples are not specified, only that "studies established" these characteristics. These are likely laboratory-controlled, prospective studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The ground truth for the clinical method comparison study was established by "laboratory plasma instruments using Dade Innovin reagent (x)". The document does not specify the number of experts or their qualifications for performing these reference laboratory tests. It's assumed to be standard clinical laboratory practice.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

No adjudication method is described. The comparison is directly between the i-STAT device and a reference laboratory method (Dade Innovin reagent) for patient samples, and between i-STAT capillary and venous samples.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not an AI-assisted diagnostic device for human readers. It's an in vitro diagnostic (IVD) device. Therefore, an MRMC comparative effectiveness study involving human readers with/without AI assistance is not applicable and was not performed or described.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The i-STAT PT test is a standalone device that automatically performs the test and reports results (INR and seconds). The analysis steps are "performed automatically." While "trained and certified health care professionals" use the device, their role is operational (sample collection, cartridge insertion, data entry), not interpretive in the sense of a diagnostic image or signal. The clinical studies described represent standalone performance of the device compared to a reference method.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The primary ground truth for the device's performance claims against routine laboratory testing is comparison to a recognized laboratory reference method (Dade Innovin reagent on plasma instruments). For other claims (heparin, fibrinogen, hematocrit sensitivity), the ground truth was established through manipulated samples and clinical patient studies where the actual levels of these components could be measured and correlated with i-STAT results.

8. The sample size for the training set:

The document describes studies for validation and performance comparison for regulatory submission. It does not explicitly mention a separate "training set" for an algorithm. This suggests that if any internal algorithm development or calibration occurred, the data used for that is not disclosed here as a distinct training set. The clinical and non-clinical data presented are for demonstrating performance and substantial equivalence.

9. How the ground truth for the training set was established:

As no explicit "training set" is described for an algorithm, the method for establishing its ground truth is not applicable based on the provided text.

{0}------------------------------------------------

VAY 1 4. 2002

Image /page/0/Picture/1 description: The image shows a handwritten string of characters, "K020355". The characters are written in a dark ink, and the handwriting appears somewhat cursive. The numbers are easily readable, and the letter K is capitalized.

Image /page/0/Picture/2 description: The image shows the logo for i-STAT Corporation. The logo is in black and white and features a stylized "i" followed by "-STAT" in a larger font. Below the main text is the word "CORPORATION" in a smaller, sans-serif font.

510(k) Summary

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1900 and CFR 807.92.

The assigned 510(k) number is:

Summary prepared on: January 30, 2002

Submitted by:

i-STAT Corporation 104 Windsor Center Dr. E. Windsor. NJ 08520 Phone: 609-443-9300 FAX: 609-443-9310

Contact:

Paul VanDerWerf, Ph.D. Vice-President Regulatory Affairs and Quality Assurance i-STAT Corporation 104 Windsor Center Drive East Windsor, NJ 08520 Phone: 609-443-9300 Fax: 609-443-9310

Establishment Registration Number: 2245578

ldentification of Device:

Device Name: Prothrombin Time (PT) Test Proprietary/Trade Name: i-STAT® PT Test Common Name: PT Device Classification: II Requlation Number: CFR § 864.7750 Panel: Hematology Kits and Packages Product Code: GJS

Identification of the Predicate Device:

Device Name: Coaguchek™ S System

Intended Use of the Device:

The i-STAT PT is a prothrombin time test cartridge and is an in vitro diagnostic test intended for quantitative prothrombin time testing for the monitoring of oral anticoagulation therapy using fresh venous or capillary whole blood samples. The i-STAT PT test is not intended for evaluating individual factor deficiencies. The cartridge is to be used with the i-STAT Portable Clinical Analyzer with thermal control (Models 200 and 300), but will not run on the Philips Medical Systems (formerly Agilent Technologies) Blood Analysis Module (BAM). As part of the i-STAT System, the PT test is to be used by trained and certified health care professionals in accordance with a facility's policies and procedures.

Description of the Device:

The i-STAT PT test is contained in a single test cartridge. In use, approximately 40 microliters of fresh whole blood are placed in the cartridge as described below. The cartridge is inserted into the thermally controlled i-STAT Portable Clinical Analytical steps are performed automatically. Patient and user information may be entered into the analyzer via a keypad during the automated analysis cycle.

{1}------------------------------------------------

In the i-STAT PT test the endpoint is indicated by the appearance of an electroactive marker generated by the thrombin-mediated conversion of a synthetic substrate included in the reagent. Detection of the marker indicates generation of thrombin and therefore complete activation of the coagulation cascade. The result is reported as an International Normalized Ratio (INR) and, optionally, in seconds. The optionally displayed seconds is intended to reflect a typical plasma prothrombin time.

The PT test cartridge is assembled from plastic components that provide the conduits for fluid handling r no r r toot our negor chips. The coagulation test is identified to the user through the name and color code on the cartridge label and by the analyzer through features integral to the cartridge.

During the test the blood sample is mixed with reagents which are coated on the cartridge cover in a segment of the sensor channel. The reagent layer includes tissue thromboplastin as an activating agent, the thrombin substrate, a heparin-neutralizing enzyme, and inert matrix components. These reagents allow activation of the coagulation cascade and detection of clot formation.

Whole blood is introduced into the sample well of the cartridge at the sample port and the cartridge is closed and inserted into the analyzer. Insertion of the cartridge initiates a controlled and monitored sequence of steps in the instrument. These are:

  • Electrical contact is made between the analyzer electronic input circuits and the cartridge. . The analyzer identifies the type of cartridge being used and the tests contained in the cartridge.
  • The dry chips and sensor channel are heated to 37°C. .
  • The blood is then moved forward. Feedback from the fluid position sensor is used to allow . controlled oscillation of the blood segment resulting in dissolution of the reagent layer.
  • During the course of testing, the position of the blood segment is actively controlled to . maintain the length of the blood containing the reagent coincident with the endpoint detector.
  • Calculation of the sample clot time is performed and displayed. .

Comparison to Technological Features of the Predicate Device:

The following technological difference between the i-STAT and CoaguChek S Systems is noteworthy:

Endpoint detection in both the CoaguChek S and i-STAT Systems relies upon detecting the action of thrombin, the final enzyme activated in the coagulation cascade, on a substrate within the sample. In the case of the CoaguChek S test, thrombin converts its natural substrate, fibrinogen, to fibrin which then crosslinks and causes localized or extended clotting throughout the sample. The instrument detects clot formation as the resulting restriction of the iron particles motion through the sample. In the i-STAT test, the generated thrombin converts an added substrate to a species that can be detected electrochemically. The signal for the appearance of this marker is used to assign the endpoint time. This is described in detail in section 5.3.2. Since fibrinogen is not a factor affected by warfarin, sensitivity to depleted levels is not a practical limitation for the intended use of monitoring of oral anticoagulant therapy. See section 10.4 for data characterizing the effect of fibrinogen concentration on the i-STAT PT test.

Summary of Non-Clinical Performance in Support of Substantial Equivalence:

  • Studies using whole blood heparin-spiked samples established that the i-STAT PT test is insensitive . to heparin up to 1.0 U/mL. The package insert for the Coaguchek S System states that it is sensitive to levels of heparin over 0.15 U/mL.
  • Studies using manipulated samples of various fibrinogen levels established that the i-STAT PT test is . insensitive to fibrinogen levels as low as 70 mg/dL.
  • Clinical patient studies demonstrated that the i-STAT PT test is insensitive to fibrinogen levels ● between 142 and 528 mg/dL.
  • Clinical studies established that hematocrits in the range of 24 52% do not significantly affect the ● results of the i-STAT PT test.
  • Studies using manipulated samples established that the i-STAT PT test is sensitive to all factors ● impacted by warfarin therapy.

{2}------------------------------------------------

  • From duplicate sample testing in the method comparison study, it can be seen that the within-sample . reproducibility of the i-STAT PT test is 5.4%. The Coaguchek S System package insert indicates imprecision of 5.9.8% in whole blood samples in the therapeutic range.
  • The imprecision of the i-STAT PT test in plasma controls was established using in-house and user . studies. The Level 1 Control %CV was 4.5 at INR of 1.1, and the Level 2 Control %CV was 6.9% at INR of 2.5. This includes within-lot, lot, vial-to-vial, analyzer-to-analyzer, and operator-tooperator components of the imprecision. The total imprecision in whole blood controls reported for the Coaguchek S System 10.2% at INR of 1.5 and 15.1% at INR of 3.7.

Summary of Clinical Test Performance is Support of Substantial Equivalence Claims:

Studies conducted at three external sites compared the results of the i-STAT PT test (y) to those of laboratory plasma instruments using Dade Innovin reagent (x) for samples from patients undergoing routine monitoring of oral anticoagulation therapy. Samples collected in plastic tubes were analyzed routhe monite hightem while plasma from citrated samples collected during the same venipuncture were analyzed on the laboratory methods were compared using Deming regression analysis. The results are summarized in the table below:

StatisticDefinitionSite 1Site 2Site 3
NThe number of patient samples used inthe comparison183180177
Mean, INRThe average of the comparative methodresult over the sample population2.32.32.5
Range, INRThe range of comparative method resultsobtained over the sample population1.0 - 3.91.0 - 4.31.0 - 4.6
Sx, INRThe standard deviation of the comparativemethod results across the samplepopulation0.7290.7770.779
SlopeThe Deming regression estimate of theslope0.9221.0130.914
Intercept,INRThe least squares linear regressionestimate of the intercept0.4020.0120.054
CorrelationThe correlation coefficient calculated fromlinear regression0.8980.9430.948
Sy.x, INRThe standard error of the estimate of theregression of y on x0.3220.2720.191

A comparison of the i-STAT capillary samples and i-STAT venous samples was performed on patients undergoing routine monitoring of oral anticoagulation therapy. The results are presented in the table below:

StatisticCapillarystudy site
N samples59
Mean, INR2.5
Range, INR1.3 - 6.4
Sx, INR0.855
Slope1.076
Intercept, INR-0.131
Correlation0.962
Sy,x, INR0.195

{3}------------------------------------------------

Conclusions:

Based on clinical and non-clinical data the i-STAT PT test is insensitive to heparin up to 1.0 U/mL, hematocrit level from 24 – 52%, and fibrinogen concentrations from 70 – 528 mg/dL; and sensitive to all factors impacted by warfarin therapy. Studies using plasma controls and whole blood indicate adequate precision for normal and therapeutic results. Clinical data indicates acceptable correlation to the predicate device.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, overlaid on three human profiles. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAY 1 4 2002

Paul VanDerWerf, Ph.D. Vice President Regulatory Affairs and Quality Assurance i-STAT Corporation 104 Windsor Center Drive East Windsor, New Jersey 08520

Re: K020355

Trade/Device Name: i-STAT® Prothrombin Time Test Regulation Number: 21 CFR § 864.7750 Regulation Name: Test, Time, Prothrombin Regulatory Class: II Product Code: GJS Dated: May 3, 2002 Received: May 6, 2002

Dear Dr. VanDerWerf:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

3 Indications for use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Prothrombin Time test.

The i-STAT PT, a prothrombin time test, is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or warfarin.

(Please do not write below this line—continue on another page if needed.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseOROver-The-Counter-Use
-------------------------------------------------------------------------------------------------------------------------------------------------
(Per 21 CFR 801.109)(Optional Format 1-2-96)
------------------------------------------------

Josephine Rantida.

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K020355

§ 864.7750 Prothrombin time test.

(a)
Identification. A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).(b)
Classification. Class II (performance standards).