AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters patients.

Intended use: For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in health care facilities.

EASI 12-lead ECG is only for use on adult and pediatric patients.

ST Segment monitoring is restricted to adult patients only.

The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.

Device Description

The name of this device is the Philips Medical Systems, M1275B Component Compact Monitor. The modification is a new display unit and a modified rack.

AI/ML Overview

The M1275B Component Compact Monitor does not contain AI and is thus not an AI/ML powered device. Hence the information requested will not be available in the provided document.

However, the document does contain information about the device's acceptance criteria and the study conducted to prove its compliance in a general sense.

1. Table of Acceptance Criteria and Reported Device Performance:

The document broadly states:

Acceptance CriteriaReported Device Performance
Pass/Fail criteria based on specifications cleared for the predicate device.Test results showed substantial equivalence. The results demonstrate that the Component Compact Monitor meets all reliability requirements and performance claims.

2. Sample Sizes and Data Provenance:

  • Test Set Sample Size: Not specified. The document states "Testing involved the new device, performance tests, and safety testing from hazard analysis." It does not provide details on the number of devices or scenarios used for these tests.
  • Data Provenance: Not explicitly stated but implied to be from internal testing conducted by Philips Medical Systems in Andover, MA, USA. It is a prospective study as it involved testing a new device.

3. Number of Experts and Qualifications:

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified. The testing would likely have been conducted by engineers and quality assurance personnel at Philips Medical Systems.

4. Adjudication Method:

  • Not applicable as this is not a study involving human interpretation of data for ground truth establishment. The testing would have followed established engineering and quality assurance protocols.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • Not applicable. This is not an AI/ML powered device and therefore no MRMC study would be conducted to assess human reader improvement with AI assistance.

6. Standalone Performance Study:

  • Yes, a standalone performance study was conducted. The document states "Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new device with respect to the predicate. Testing involved the new device, performance tests, and safety testing from hazard analysis." This refers to testing the device itself, without human-in-the-loop performance assessment.

7. Type of Ground Truth Used:

  • The ground truth in this context is the specifications and performance characteristics of the legally marketed predicate device. The new device's performance was measured against these established standards.

8. Sample Size for Training Set:

  • Not applicable. This is not an AI/ML powered device, so there is no "training set."

9. How Ground Truth for Training Set Was Established:

  • Not applicable.

{0}------------------------------------------------

NOV 2 8 2001

Image /page/0/Picture/1 description: The image shows the text "K013199 p.1/a" in black font. The text appears to be handwritten. The text is likely a document identifier or page number.

9.0 510 (k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR §807.92.

  1. The submitter of this premarket notification is: Dave Osborn Quality Program Manager Philips Medical Systems 3000 Minuteman Road Andover, MA 01810-1085

978 659 3178 Tel: 978 685 5624 Fax : Email: dosborn@hsgmed.com

This summary was prepared on 21 September, 2001

    1. The name of this device is the Philips Medical Systems, M1275B Classification names are as follows: Component Compact Monitor.
Device PanelClassificationProCodeDescription
Panel 73Anesthesiology868.1400, IICCKAnalyzer, Gas, Carbon-Dioxide, Gaseous-Phase
868.2375, IIBZQMonitor, Breathing Frequency
868.2480, IILKDMonitor, Carbon-Dioxide, Cutaneous
868.2500, IIKLKMonitor, Oxygen, Cutaneous, for Infant not under Gas Anesthesia
Panel 74Cardiovascular870.1025, IIIDSIDetector and Alarm, Arrhythmia
870.1025, IIIMLDMonitor, ST Segment with Alarm
870.1025, IIIMHXMonitor, Physiological, Patient (with arrhythmia Detection or alarms)
870.1100, IIDSJAlarm, Blood-Pressure
870.1110, IIDSKComputer, Blood-Pressure
870.1130, IIDXNSystem, Measurement, Blood-Pressure, Non-Invasive
870.1435, IIDXGComputer, Diagnostic, Pre-Programmed, Single-Function
870.1915, IIKRBProbe, Thermodilution
870.2060, IIDRQAmplifier and Signal Conditioner, Transducer Signal
870.2300, IIDRTMonitor, Cardiac (incl. Cardiotachometer & Rate Alarm)
870.2340, IIDPSElectrocardiograph
870.2340, IIMLCMonitor, ST Segment
870.2370, IIKRCTester, Electrode, Surface, Electrocardiograph
870.2450, IIDXJDisplay, Cathode-Ray Tube, Medical
870.2600, IDRJSystem, Signal Isolation

{1}------------------------------------------------

Device PanelClassificationProduct CodeDescription
870.2700, IIDQAOximeter
870.2770, IIDSBPlethysmograph, Impedance
870.2800, IIDSHRecorder, Magnetic Tape,Medical
870.2810, IDSFRecorder, Paper Chart
-MSXSystem, Network andCommunication,Physiological Monitors
Panel 80GeneralHospital880.2910, IIFLLThermometer, electronic,clinical
    1. The new device is substantially equivalent to the previously cleared M3/M4 (M3046/3000A) device marketed pursuant to K971910, creared , and K990972, K991773, K992273, K993383, K000822, R981370 , und K903621, as well as M1175/76A pursuant to
      K001057, K001333, and K003621, as well 2011, x000135 R001037, K0015337, Kra K088, K923682, K925910, K941811, K990125, к990476, к992595, к001722.
  • The modification is a new display unit and a modified rack. 4 .
    বা
  • The new device has the same Indications for Use, for use by 5. health care professionals whenever there is a need for monitoring the physiological parameters patients, as the legally marketed predicate device.
    1. The new device has the same technological characteristics as the legally marketed predicate device.
    1. Verification, validation, and testing activities establish the verformance, functionality, and reliability characteristics of the new device with respect to the predicate. Testing involved che new device me, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that Component Compact Monitor meets all reliability requirements and performance claims.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three horizontal lines representing its wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 2 8 2001

Mr. Dave Osborn Phillips Medical Systems, Inc. Cardiac and Monitoring Systems 3000 Minuteman Rd. Andover, MA 01810-1099

K013199 Re:

Trade Name: M1275B Component Compact Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm Regulatory Class: Class III (three) Product Code: MHX Dated: November 12, 2001 Received: November 13, 2001

Dear Mr. Osborn:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Dave Osborn

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Denta Tille

James E. Dilla Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

_ of _________________________________________________________________________________________________________________________________________________________________________ Page _

Over-The-Counter

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Philips Medical Systems, M1275B Component Compact Monitor, Release A.01.

Indicated for use by health care professionals Indications for Use: whenever there is a need for monitoring the physiological parameters patients.

Intended use: For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in health care facilities.

EASI 12-lead ECG is only for use on adult and pediatric patients.

ST Segment monitoring is restricted to adult patients only.

The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

Prescription Use Usc (Per 21 CFR 801.109) (Optional Format 1-2-96)

Telh
Division of Cardiovascular & Respiratory Devices
510(k) Number K013199

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.