K Number
K012449
Manufacturer
Date Cleared
2001-10-26

(86 days)

Product Code
Regulation Number
866.5660
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diamedix Is anti-Cardiolipin IgG/IgM Test Kit is an indirect enzyme immunoassay (EIA) for the semi-quantitative measurement of IgG or IgM antibodies to cardiolipin in human serum as an aid in the assessment of the risk of thrombosis in patient with SLE or SLE-like disorders. The assay can be used either manually or in conjunction with the MAGO® Plus Automated EIA Processor.

Device Description

The Is anti-Cardiolipin IgG/IgM Test System is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative measurement of IgG or IgM antibodies to cardiolipin in human serum

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Diamedix Is-anti-Cardiolipin IgG/IgM Test System, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined quantitative acceptance criteria with specific numerical targets. Instead, it describes performance characteristics and then presents the results. The implicit acceptance criteria are that the device demonstrates comparable performance to a commercially available predicate device and satisfactory analytical and clinical performance.

Performance MetricIndication of Acceptance Criteria (Implicit)Reported Device Performance
Calibration Equivalence3-point vs. 6-point calibration should show strong correlation.IgG: Y = 2.4036 + 0.9996X (r = 0.9861, n=172). IgM: Y = 0.5333 + 0.9411X (r = 0.9925, n=162). (Strong linear correlation demonstrated for both.)
Relative Sensitivity (IgG)Should be comparable to commercially available ELISA kit.91.5% (95% CI: 79.6-97.6%)
Relative Specificity (IgG)Should be comparable to commercially available ELISA kit.100.0% (95% CI: 96.6-100.0%)
Overall Agreement (IgG)Should be comparable to commercially available ELISA kit.97.4% (95% CI: 93.5-99.3%)
Relative Sensitivity (IgM)Should be comparable to commercially available ELISA kit.78.4% (95% CI: 67.3-87.1%)
Relative Specificity (IgM)Should be comparable to commercially available ELISA kit.100.0% (95% CI: 95.8-100.0%)
Overall Agreement (IgM)Should be comparable to commercially available ELISA kit.90.1% (95% CI: 84.4-94.2%)
Clinical Specificity (IgG)High specificity in normal population, acceptable specificity in other autoimmune diseases.Normals: 99.5% (213/214), RPR Positive: 73.3% (11/15)
Clinical Specificity (IgM)High specificity in normal population, acceptable specificity in other autoimmune diseases.Normals: 98.5% (211/214), RPR Positive: 73.3% (11/15)
Clinical Sensitivity (IgG)Detectable in patients with APS and SLE.APS: 82.5% (47/57), SLE: 21.2% (7/33), Other Autoimmune Diseases: 8.6% (3/35)
Clinical Sensitivity (IgM)Detectable in patients with APS and SLE.APS: 47.4% (27/57), SLE: 21.2% (7/33), Other Autoimmune Diseases: 14.2% (5/35)
Cross ReactivityShould show minimal cross-reactivity with other common autoantibodies.Out of 36 samples tested (SSA/SSB, Scl-70, Jo-1, dsDNA, RF), one sample positive for dsDNA showed a positive result. The remaining 34 samples were negative. (Suggests minimal cross-reactivity with most, but potential with dsDNA.)
Linearity (IgG)Demonstrates a linear response over a range of concentrations.Y = -4.9960 + 147.5101 X (r = 0.9957, R² = 0.9915).
Linearity (IgM)Demonstrates a linear response over a range of concentrations.Y = 4.0837 + 87.9931 X (r = 0.9849, R² = 0.9701).
Manual vs. Automated (IgG)Manual and MAGO Plus methods should show strong correlation.Y = 1.1058 + 1.1976 X (r = 0.9883, n=172). 6-point (automated vs. manual): (automated) = 1.0696 (manual) + 4.0821; 95% CI slope [1.0174 to 1.1218], intercept [1.3042 to 6.8600]. (Strong linear correlation demonstrated.)
Manual vs. Automated (IgM)Manual and MAGO Plus methods should show strong correlation.Y = 1.3732 + 1.1510 X (r = 0.9917, n=162). 6-point (automated vs. manual): (automated) = 1.0169 (manual) + 2.2121; r=0.9772; 95% CI slope [0.9824 to 1.0514], intercept [1.1343 to 3.2899]. (Strong linear correlation demonstrated.)
Precision (CV%)Coefficients of Variation (CV%) for intra-assay and interassay should be within acceptable limits (typically <15-20% for immunoassays).Manual IgG: Interassay CV% range: 4.42-13.93%. MAGO Plus IgG: Interassay CV% range: 10.17-29.69%. Manual IgM: Interassay CV% range: 4.84-20.03%. MAGO Plus IgM: Interassay CV% range: 6.79-17.62%. (Generally acceptable, with some higher CVs for very low and high range samples, particularly for MAGO Plus IgG. Smallest values often have higher CVs inherently.)
Limit of Detection (Analytical Sensitivity)Low detection limit for clinical relevance.Determined as 0.4 GPL or MPL U/ml.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Calibration Equivalence:
    • IgG: 172 samples
    • IgM: 193 samples
    • Provenance: Retrospective sera (not specified by country).
  • Relative Sensitivity and Specificity:
    • IgG: 172 frozen retrospective sera
    • IgM: 194 frozen retrospective sera
    • Provenance: Retrospective sera (not specified by country).
  • Clinical Sensitivity and Specificity:
    • Total: 354 frozen retrospective, clinically characterized sera
      • 214 normal sera
      • 57 sera from patients with Antiphospholipid Syndrome (APS)
      • 33 sera from patients with Systemic Lupus Erythematosus (SLE)
      • 35 sera from patients with other autoimmune diseases (Sjögren's, scleroderma, polymyositis/dermatomyositis, rheumatoid arthritis)
      • 15 sera from RPR-positive patients
    • Provenance: Retrospective sera (not specified by country).
  • Cross Reactivity: 36 samples (positive for SSA/SSB, Scl-70, Jo-1, dsDNA, and RF antibodies). Provenance: Not specified.
  • Linearity: "Several highly positive samples" were serially diluted. (Number not specified). Provenance: Not specified.
  • Correlation of Manual and MAGO Plus Results:
    • IgG: 172 serum samples
    • IgM: 162 serum samples
    • Provenance: Not specified.
  • Precision: 6 serum samples of varying reactivity tested. Provenance: Not specified.
  • Expected Values (Normal Population): 148 healthy blood donors from South Florida. Provenance: Prospective, South Florida, USA.
  • Expected Values (Clinical Population): 57 sera from patients with lupus anticoagulant antibodies. Provenance: Not specified.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The studies described are in vitro diagnostic (IVD) performance studies, not typically requiring human readers or expert radiologists to establish ground truth.

  • For Relative Sensitivity and Specificity, a "commercially available ELISA kit for detecting IgG and/or IgM cardiolipin antibodies" and a "referee EIA method" were used as the comparator/ground truth. No human expert adjudication is mentioned as part of this.
  • For Clinical Sensitivity and Specificity, samples were "clinically characterized sera" from specific patient groups (Normals, APS, SLE, Other Autoimmune Diseases, RPR Positive). The characterization of these patient groups (e.g., diagnosis of APS or SLE) would have been established by clinical experts (e.g., rheumatologists) based on diagnostic criteria, but the document does not specify how many or the qualifications of these clinicians, as it's assumed to be part of the standard diagnostic process for collecting such samples. The ground truth here is the patient's clinical diagnosis.

4. Adjudication Method for the Test Set

No explicit adjudication method (like 2+1 or 3+1) is mentioned or appears applicable, as the ground truth is established either by:

  1. Comparison to another legally marketed (predicate/referee) assay.
  2. Pre-existing clinical diagnosis of patient samples.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

No. This device is an in-vitro diagnostic (ELISA) kit. It is designed to directly measure antibody levels in human serum, not to assist human readers in interpreting medical images or other complex data. Therefore, an MRMC study and the concept of "human readers improving with AI assistance" are not applicable to this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, in essence. The primary performance evaluations (Sensitivity, Specificity, Linearity, Precision, Cross-reactivity) were performed on the device itself, either manually or using the automated MAGO® Plus system. The results are quantitative measurements from the assay, not interpretations by an algorithm that would then be compared to human interpretation. The "algorithm" in this context is the assay's biochemical reaction and detection system.

7. The Type of Ground Truth Used

  • Relative Sensitivity and Specificity: Comparison to a "commercially available ELISA kit" and a "referee EIA method."
  • Clinical Sensitivity and Specificity: "Clinically characterized sera" (i.e., outcomes/clinical diagnosis data for patient groups like APS, SLE, Normals).
  • Calibration Equivalence, Linearity, Precision, Manual vs. Automated Correlation: The ground truth is the expected output based on known standards or a well-established reference measurement method (e.g., higher concentration samples, manual method as a reference for automated).
  • Cross Reactivity: Known positive samples for other autoantibodies.
  • Expected Values: Clinically healthy or diagnosed patient populations.

8. The Sample Size for the Training Set

The document does not explicitly describe a "training set" in the context of machine learning or AI models. This is a traditional IVD device, not an AI/ML product. The samples used for validation were used to characterize the device's performance directly.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no explicitly defined "training set" for an AI/ML model. The ground truth for the various performance characteristic studies was established as described in point 7.

{0}------------------------------------------------

OCT 2 6 2001

K0124449

510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K012449

Applicant Information:

Date Prepared:October 17, 2001
Name:Diamedix Corporation
Address:2140 N. Miami AvenueMiami, FL 33127
Contact Person:Dr. Lynne Stirling
Phone Number:305-324-2354
Fax Number:305-324-2388

Device Information:

Trade Name:Is anti-Cardiolipin IgG/IgM Test System
Common Name:Anti-Cardiolipin ELISA test
Classification Name:Anticardiolipin immunological test system

Equivalent Device:

Orgentec Anti-cardiolipin ELISA Assay

Device Description: The Is anti-Cardiolipin IgG/IgM Test System is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative measurement of IgG or IgM antibodies to cardiolipin in human serum

Intended Use: The assay is intended for the semi-quantitative measurement of IgG or IgM antibodies to cardiolipin in human serum. The results of the assay can be used as an aid in the assessment of the risk of thrombosis in patients with SLE or SLE-like dosorders.

Principle of the Procedure:

The Is-anti-Cardiolipin IgG/IgM Test System is an indirect solid-phase enzyme immunoassay. Highly purified cardiolipin is coated onto plastic microwells and saturated with highly purified human (32-Glycoprotein I. Standards, controls and diluted patient samples are added to the wells. Any patient IgG or IgM antibodies in the sample bind to the well. Anti-human IgG or IgM horseradish peroxidase conjugate is then added After incubation and washing, a substrate solution is then added to each well. In the presence of bound enzyme, the substrate is converted to a blue colored product. After acid additon to stop the reaction, a yellow end product is formed that is read spectrophotometrically at 450 nm (reference 600-630 nm) and is directly proportional to the concentration of cardiolipin IgG or IgM antibodies in the sample.

{1}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS

Performance Characteristics

Non-clinical studies were performed using the manual method and 6-point calibration unless otherwise indicated.

A. 3-point vs 6-point calibration

To demonstrate the equivalence of both calibration methods the results of 172 samples tested using the Is-anti-Cardiolipin IgG and 193 samples tested using the Is-anti-cardiolipin IgM calculated using either the 3-point or 6-point calibration systems were subjected to linear regression analysis. Scattergrams and regression lines of the results obtained with 95% confidence intervals are shown in FIGURES 1 and 2. Also included are the regression statistics.

Image /page/1/Figure/5 description: The image contains two scatter plots comparing 3-point and 6-point result correlations for anti-Cardiolipin IgG and IgM. Figure 1 shows the correlation for IgG, with the equation Y = 2.4036 + 0.9996X, a sample size of 172, and a correlation coefficient of 0.9861. Figure 2 displays the correlation for IgM, with the equation Y = 0.5333 + 0.9411X, a sample size of 162, and a correlation coefficient of 0.9925.

B. Relative Sensitivity and Specificity

One hundred and seventy-two frozen retrospective sera were tested for IgG antibodies and one hundred and ninety--four frozen retrospective sera were tested for IgM antibodies using the Is-anti-Cardiolipin IgG/IgM Test Kit and a commercially available ELISA kit for detecting IgC and/or IgM cardiolipin antibodies. Based on the results of this testing the relative sensitivity, specificity and overall agreement were calculated. The resultsobtained are shown in TABLES 1 and 2. For anti-cardiolipin IgG, further resolution of the discordant samples showed that the four samples that were negative in the Is-anti-Cardiolipin IgG and positive by the other EIA were also negative by a referee EIA method. For anti-cardiolipin IgM, further resolution of the discordant samples showed that of the 16 samples negative in the Is-anti-Cardiolipin IgM and positive in the other EIA, thirteen were negative and three were positive by a referee method.

TABLE 1

TABLE 2

Is-anti-Cardiolipin IgG
PositiveNegativeEquivocal
OtherEIAPositive4340
Negative01070
*Equivocal3150
**95% CI
Relative Sensitivity$43/47 = 91.5 %$79.6-97.6%
Relative Specificity$107/107 = 100.0 %$96.6-100.0%
Overall Agreement$150/154 = 97.4 %$93.5-99.3%
PositiveNegativeEquivocal
Positive58166
Negative0870
*Equivocal0270
Relative Sensitivity58/74= 78.4 %**95% CI767.3-87.1%
Relative Specificity87/87= 100.0%95.8-100.0%
Overall Agreement145/161= 90.1%84.4-94.2%

Is-anti-Cardiolipin IgM

{2}------------------------------------------------

    • Equivocal results were excluded from calculations. ** 95% Confidence Intervals (CI) calculated by the Exact Method .
      NOTE : Please be advised that relative' refers to the comparison of the assay's on the of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgement can be made on the comparison's accuracy to predict disease.

Linear regression analyses and scattergrams for the correlation studies with the comparative method are shown in FIGURES 3 and 4.

Image /page/2/Figure/3 description: The image contains two scatter plots comparing two different methods. The first scatter plot, titled "FIGURE 3: Is anti-Cardiolipin IgG Correlation with Comparative Method", plots the comparative method GPL U/ml against Is-anti-Cardiolipin IgG GPL U/ml. The equation of the regression line is Y = 1.0977 + 0.8910 X, with a correlation coefficient of 0.9600 and a coefficient of determination of 0.9216. The second scatter plot, titled "FIGURE 4: Is anti-Cardiolipin IgM Correlation with Comparative Method", plots the comparative method MPL U/ml against Is-anti-cardiolipin IgM MPL U/ml, with a regression equation of Y = 1.6035 + 0.8449 X, a correlation coefficient of 0.9687, and a coefficient of determination of 0.9383.

C. Clinical Sensitivity and Specificity

A total of three hundred and fifty-four frozen retrospective, clinically characterized sera were assayed wing the A total of mree hundred and inty-four reassess both the clinical sensitivity and clines and estincity of the Is anti-Cardiding in These samples consisted of 214 normal sera, 57 sera from patients with disamotions of assay System. These samples Consisted of 214 notinal organ of themators (SE), 35 sera from patients with lipid syndrome (APS), 35 Sell from patchs with systems, polymyositisdematomyositis and theumaother autominate uiseasss suer as SJOgiors bynershop or success. Results are summarized in TABLE 3.

Note that the analytical sensitivity, or limit of detection, calculated by assaying Standard A. 20 times and taking Note that the allarytical Schsirvity, or mint of economined as being 0.4 GPL or MPL U/ml.

{3}------------------------------------------------

TABLE 3
IgGIgM
Patient GroupTotalPositiveNegative / Equiv.PositiveNegative / Equiv.
Normals21412133211
APS5747102730
SLE33726726
Other AutoimmuneDiseases35332530
RPR Positive15411411
Clinical Specificity:IgGIgM
# Neg or Equiv./Total ## Neg or Equiv./Total #
Normals213/214 = 99.5%211/214 = 98.5%
RPR Positive11/15 = 73.3%11/15 = 73.3%
Clinical Sensitivity :IgGIgM
# Pos/Total ## Pos/Total #
APS47/57 = 82.5%27/57 = 47.4%
SLE7/33 = 21.2%7/33 = 21.2%
Other AutoimmuneDiseases3/35 = 8.6%5/35 = 14.2%

D. Cross Reactivity

To assess the potential for positive results due to cross reactive antibition the Is antil Cardicalism Incl various autoantibodies (SSA/SSB, Scl-70, J-1, dsDNA and RF) were tested using the Is-anti-Cardiolini 2gG/Lyl various autoanthodies (SSASSB, SCF-70, Jo-1, used and one sample positive for dobby for doDNA Test Kit. In all, 36 sample were tested. One sample positive for veraining 34 samples were negative.

E. Linearity

To assess the linearity of the Is-anti-Cardiolipin IgG/IgM Test Kir several highty positive samples were serially To assess the linearity of the Is-ant-Calition was the tested in the respectively of or IDM assay systems. Repress sentative linear regression graphs and scategrams with 95% confidence intervals are presented in FIGURES 5 and 6.

Image /page/3/Figure/5 description: The image contains two scatter plots, labeled as Figure 5 and Figure 6. Figure 5 shows the linearity of Is anti-Cardiolipin IgG, with the equation Y = -4.9960 + 147.5101 X. The intercept is -4.99597, the slope is 147.51010, the coefficient of determination is 0.9915, the correlation coefficient r is 0.9957, and the 95% CI for r is 0.9756 to 0.9993. Figure 6 shows the linearity of Is anti-Cardiolipin IgM, with the equation Y = 4.0837 + 87.9931 X, and the intercept is 4.08374, the slope is 87.99309, the coefficient of determination is 0.9701, the correlation coefficient r is 0.9849, and the 95% CI for r is 0.9275 to 0.9969.

{4}------------------------------------------------

F. Correlation of Manual and MAGO Plus results

The Is-anti-Cardiolipin IgG/IgM Test Kit has been developed for automated as well as manual use. To demonstrate the equivalence of the manual and MAGO Plus procedures, the results of 172 serum samples tested for sative and of antibodies and 162 sera tested for anti-cardiolipin IgM by both the manual and automated methods were plotted. Scatter rams and regression lines of the results obtained with 95% confidence intervals mentown in FIGURES 7 and 8. The data indicate good correlation with a Correlation Coefficients (r) of 0.983 for anti-cardiolipin IgG and 0.9917 for anti-cardiolipin IgM.

Image /page/4/Figure/2 description: The image contains two scatter plots comparing manual and MAGO Plus correlation for anti-Cardiolipin IgG and IgM. The left plot, labeled "FIGURE 7: Is anti-Cardiolipin IgG," shows a linear relationship with the equation Y = 1.1058 + 1.1976 X, a sample size of 172, a coefficient of determination of 0.9768, and a correlation coefficient of 0.9883. The right plot, labeled "FIGURE 8: Is anti-Cardiolipin IgM," also shows a linear relationship with the equation Y = 1.3732 + 1.1510 X, a sample size of 162, a coefficient of determination of 0.9835, and a correlation coefficient of 0.9917.

With the 6-point calibration, linear regression of the IgG results showed (automated) = 1.0696 (manual) + 4.0821; with the o point date most, are 1.0174 to 1.1218 and 1.3042 to 6.8600 respectively. For gM results (automated) = 1.0169 (manual) + 2.2121; r = 0.9772. 95% CI for the slope and the intercept are 0.9824 to 1.0514 and 1.1343 to 3.2899 respectively.

G. Precision

To assess the precision of the Is anti-Cardiolipin IgG/IgM Test Kit six serum samples of varying reactivity were tested in triplicate in three separate runs. Precision was assessed both manually and using the MAGO Plus Automated EIA Processor. Precision was assessed for both IgG and IgM antibody types. The results obtained using 6-point Calibration are shown in TABLES 4-7.

TABLE 4 : Manual Intra-Assay and Interassay Precision for Is-anti-Cardiolipin IgG

SERUMINTRA-ASSAY DAY 1INTRA-ASSAY DAY 2INTRA-ASSAY DAY 3INTERASSAY (n=9)
MEANGPLSDCV%MEANGPLSDCV%MEANGPLSDCV%MEANGPLSDCV%
A1.60.000.001.60.063.691.60.127.071.60.074.42
B1.10.065.411.10.065.411.10.065.091.10.065.52
C17.40.452.5917.61.699.6318.61.538.2317.91.287.17
D25.42.118.3024.81.395.6027.21.214.4625.81.786.89
E35.30.722.0529.00.581.9931.21.595.1031.82.929.17
F58.12.193.7774.62.002.6873.510.1513.8268.79.5713.93

TABLE 5 : MAGO Plus - Intra-Assay and Interassay Precision for Is-anti-Cardiolipin IgG

SERUMINTRA-ASSAY DAY 1INTRA-ASSAY DAY 2INTRA-ASSAY DAY 3INTERASSAY (n=9)
MEAN GPLSDCV%MEAN GPLSDCV%MEAN GPLSDCV%MEAN GPLSDCV%
A2.50.4718.903.20.5817.863.91.2331.513.20.9529.69
B1.30.1511.752.00.125.871.90.105.261.70.3520.59
C29.11.153.9631.81.063.3340.06.3215.7933.65.9017.56
D39.77.4518.7642.91.182.7649.24.519.1644.06.0713.80
E41.50.902.1645.22.685.9251.51.763.4146.14.6910.17
F95.32.302.4281.15.006.1671.37.049.8782.611.3413.73

{5}------------------------------------------------

TABLE 6 : Manual Intra-Assay and Interassay Precision for Is-anti-Cardiolipin IgM

SERUMINTRA-ASSAY RUN 1INTRA-ASSAY RUN 2INTRA-ASSAY RUN 3INTERASSAY (n=9)
MEANMPLSDCV%MEANMPLSDCV%MEANMPLSDCV%MEANMPLSDCV%
A0.80.1518.331.10.000.000.80.1215.060.90.1820.03
B1.10.1210.191.40.000.000.90.066.191.20.2118.41
C26.00.351.3526.20.612.3221.22.1810.2924.52.7311.15
D35.30.531.5036.11.504.1731.50.852.7034.32.296.67
E61.42.604.2365.61.862.8355.50.350.6260.84.677.67
F63.33.876.1165.33.234.9662.62.604.1663.73.084.84

TABLE 7 : MAGO Plus - Intra-Assay and Interassay Precision for Is-anti-Cardiolipin IgM

SERUMINTRA-ASSAY RUN 1INTRA-ASSAY RUN 2INTRA-ASSAY RUN 3INTERASSAY (n=9)
MEANMPLSDCV%MEANMPLSDCV%MEANMPLSDCV%MEANMPLSDCV%
A2.60.259.681.90.062.991.80.063.152.10.3717.62
B2.80.217.432.20.156.842.00.125.872.30.3816.52
C29.42.056.9830.81.304.2225.00.953.7928.42.9510.39
D37.71.604.2440.62.315.6936.72.536.9138.32.606.79
E65.83.254.9470.40.650.9259.52.814.7365.25.228.01
F85.99.3010.8380.22.002.4966.72.874.3077.69.8812.73

{6}------------------------------------------------

Expected Values

The prevalence of anti-cardiolipin IgG and/or IgM antibodies may vary depending on anumber of factors such I he prevaler, geographical location, race, type of test used and clinical history of individual patients. as age, generalizar to and in absent, tace, type of two a very low incident in the normal healthy population. Increased incidence can occur in the elderly population. A published study has shown a prevacy population population. Increased includence can occur in the encorry popular on 2% for a younger population. In the encrying for addition, anti-cardiolipin antibodies were detected in 23% of elderly individuals who were also positive for anti-nuclear antibodies (13).

In the present study, the expected values for a normal, healthy population were assessed by testing sera from In In the present study, the expected values to a normal, notal.of per-Cardiolipin IgC/I/DM Test Sitting both Igor one hundred and lorty-eight 3. Fronta blood dones in and forty-seven seral (99.3%) were negative for antibod ies, one serum (0.7%) was positive and none were equivocal. For IgM antibodies, and forty-six ies, one serum (0.7%) was positive and none were equivocal. The age distribution and (90.0%) weevalences for this population are shown in TABLE 8.

The expected values for a clinical population were assessed by testing fifty-seven sera from patients with antibody The expected values for a Cilincal population were assessed of artify of the Arman Matthody types. Forty-seven (82.5%) were positive, nine (15.8%) were regative and one (1.7%) was equivocal for IgCly was types. Forty-seven (47.3%) were positive, twenty-nine (50.9%) were negative, and one (1.7%) was equivocal for IgM antibodies.

equired to oget and antibodies are shown in FIGURES 9-12.

Total NumberNumber of DonorsPrevalence
148IgGIgM
GeographicLocation:South Florida : 1480.7%1.4%
Age
10-1970.0%0.0%
20-29360.0%0.0%
30-39730.0%2.7%
40-49224.5%0.0%
50-5980.0%0.0%
60-6920.0%0.0%

TABLE 8: Age Distribution and Prevalence of anti-Cardiolipin IgG and IgM Antibodies in a Normal S. Florida Population

{7}------------------------------------------------

FIGURE 9 Distribution of anti-Cardiolipin IgG in a Normal Population

FIGURE 10 Distribution of anti-Cardiolipin IgM in a Normal Population

Image /page/7/Figure/2 description: This image is a histogram showing the frequency of GPL U/ml. The x-axis represents GPL U/ml, ranging from 0 to 55, while the y-axis represents frequency, ranging from 0 to 160. The histogram shows a high frequency around 0-5 GPL U/ml, with the frequency being around 140. There are also some frequencies around 45-55 GPL U/ml, but they are much lower.

Image /page/7/Figure/3 description: This image is a histogram showing the frequency of MPL U/ml. The x-axis represents MPL U/ml, ranging from 0 to 18, while the y-axis represents the frequency, ranging from 0 to 100. The histogram shows that the highest frequency occurs at the lowest MPL U/ml values, with a tall bar at the beginning, and the frequency decreases as the MPL U/ml values increase.

FIGURE 11 Distribution of anti-Cardiolipin IgG in a Clinical Population

Image /page/7/Figure/5 description: The image shows the title of a figure. The figure is labeled as "FIGURE 12". The title of the figure is "Distribution of anti-Cardiollpin IgM in a Clinical Population".

Image /page/7/Figure/6 description: This image is a histogram showing the frequency of GPL U/ml. The x-axis represents GPL U/ml, ranging from 0 to 200, while the y-axis represents frequency, ranging from 0 to 16. The histogram shows the distribution of GPL U/ml values, with the highest frequency occurring between 0 and 20, with a frequency of approximately 15.

Image /page/7/Figure/7 description: The image is a histogram showing the frequency of MPL U/ml. The x-axis represents MPL U/ml, ranging from 0 to 120, while the y-axis represents frequency, ranging from 0 to 30. The histogram shows a high frequency at the lower end of MPL U/ml, with the highest frequency around 0, and a smaller peak around 100 MPL U/ml.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a stylized caduceus, a symbol often associated with healthcare. The caduceus is depicted with a staff entwined by a serpent, representing healing and medicine.

OCT 2 6 2001

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Lynne Stirling, Ph.D. Vice President, Regulatory Affairs Diamedix Corporation 2140 North Miami Avenue Miami. Florida 33127

Re: K012449

Trade/Device Name: Diamedix Is-anti-Cardiolipin IgG/IgM Test System Regulation Number: 21 CFR § 866.5660 Regulation Name: Multiple Autoantibodies Immunological Test System Regulatory Class: Class II Product Code: MID Dated: September 28, 2001 Received: October 1, 2001

Dear Dr. Stirling:

We have reviewed your Section 510(k) premarket notification of intent to market the device wo nave ro rowed your we determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 as stated in the encreations of the enactment date of the Medical Device Amendments, or to conincered prices that have been reclassified in accordance with the provisions of the Federal Food, Drug, de necs that have been resuire approval of a premarket approval application (PMA). and Cosmetic Frev (110) inst the device, subject to the general controls provisions of the Act. The r ou may, diereleve, mains of the Act include requirements for annual registration, listing of general voltaron provisions practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that I Dri has intact a and regulations administered by other Federal agencies. You must or any I oderal statutes and securements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OF R Fat 6077, accems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality by events (Sections 531-542 of the Act); 21 CFR 1000-1050.

{9}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nouncation. The I Dr Imanig of cassification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFF Part 801 and II you desire specific acin vitto diagnostic devices), please contact the Office of Compliance at additionally 607.10 for in This diagliestions on the promotion and advertising of your device, (201) 594-1566. Traditional (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small monifacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Appendix G. Indications for Use Statement

INDICATIONS FOR USE STATEMENT

510(K) NUMBER : _ K O/2449

DEVICE NAME : Is anti-Cardiolipin IgG/IgMTest System

Indications for Use : The Diamedix Is anti-Cardiolipin IgG/lgM Test Kit is multations for USC : The Blains (EIA) for the semi-quantitative meaan indirect enzyme immanoaocay (2.1%) in human serum as an
surement of IgG or IgM antibodies to cardiolipin in human situs of Engl surement of ige of ight antiboution to the risk of thrombosis in patient with SLE or ald in the assessment of the not of the of all be used either manually or in conjunction with the MAGO® Plus Automated EIA Processor

Szusan Altaie

(Division Sign-Off) Division of Clinical Laboratory Devices

510(k) Number K012449

For prescription Use X

§ 866.5660 Multiple autoantibodies immunological test system.

(a)
Identification. A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).(b)
Classification. Class II (performance standards).