K Number
K012128
Device Name
PROTECTIV ACUVANCE IV CATHETER SYSTEM
Date Cleared
2001-08-01

(23 days)

Product Code
Regulation Number
880.5200
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
A properly placed I.V. catheter provides access to a vein or artery. The I.V. Catheter System is designed for single use and has a needlestick protection feature. The risk of accidental needlesticks is reduced by a self-blunting needle system activated automatically as the catheter is threaded into the vessel. These catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. 14 to 22 gauge catheters may be used with power injectors up to 300 psi.
Device Description
Intravascular catheters are single use devices which provide access to veins or arteries.
More Information

PROTECTIV* ACUVANCE IV Catheter System

Not Found

No
The summary describes a mechanical device for IV access with a safety feature, and there is no mention of AI or ML in the intended use, device description, or performance studies.

No.
The device is used to provide access to a vein or artery for infusion, which is a diagnostic or interventional procedure, not a therapeutic procedure performed by the device itself.

No
The description states the device "provides access to a vein or artery" and "may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy." This indicates it's used for therapeutic purposes (infusion) rather than diagnosis. It's an I.V. catheter system.

No

The device description and intended use clearly describe a physical I.V. catheter system with a needlestick protection feature, which is a hardware device. There is no mention of software as the primary or sole component.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use/Indications for Use: The description clearly states the device's purpose is to provide access to a vein or artery for infusion. This is a direct medical intervention on the patient's body.
  • Device Description: It describes the device as providing access to veins or arteries, again indicating a direct interaction with the patient's circulatory system.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens (like blood, urine, or tissue) in vitro (outside the body) to provide information about a physiological state, health, or disease.

IVDs are used to perform tests on samples taken from the body, not to directly access or treat the body itself. This device is a medical device used for direct patient care.

N/A

Intended Use / Indications for Use

A properly placed I.V. catheter provides access to a vein or artery. The I.V. Catheter System is designed for single use and has a needlestick protection feature. The risk of accidental needlesticks is reduced by a self-blunting needle system activated automatically as the catheter is threaded into the vessel. These catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. 14 to 22 gauge catheters may be used with power injectors up to 300 psi.

Product codes

FOZ

Device Description

Intravascular catheters are single use devices which provide access to veins or arteries.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

vein or artery

Indicated Patient Age Range

any patient population with consideration given to patient size

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Bench testing demonstrates that the 14 to 22 gauge ACUVANCE catheters are appropriate for use with power injectors up to 300 psi.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

PROTECTIV* ACUVANCE IV Catheter System

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).

0

Ethicon Endo-Surgery, Inc. Ethicon Blude Barger J, inci

K012/28

PROTECTIV* ACUVANCE IV Catheter System 510(k) Summary of Safety and Effectiveness

Company

Ethicon Endo-Surgery, Inc. 4545 Creek Rd. Cincinnati, OH 45242

Contact

Katie Fordyce Regulatory Affairs Associate

Date Prepared:

July 6, 2001

Name of Device

Trade Name: PROTECTIV* ACUVANCE IV Catheter System Classification Name: Catheter, Intravascular (short-term)

Predicate Devices: PROTECTIV* ACUVANCE IV Catheter System

Device Description: Intravascular catheters are single use devices which provide access to veins or arteries.

Intended Use: A properly placed I.V. catheter provides access to a vein or artery. The I.V. Catheter System is designed for single use and has a needlestick protection feature. The risk of accidental needlesticks is reduced by a self-blunting needle system activated automatically as the catheter is threaded into the vessel. These catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. 14 to 22 gauge catheters may be used with power injectors up to 300 psi.

Technological Characteristics: The technological characteristics of the new device are the same as those of the predicate device. No changes (materials, specifications, manufacturing or sterilization processes) have been made to the currently marketed catheters. The catheters differ from the currently marketed catheters only in the instructions for use which now include use of the 14 to 22 gauge catheters with power injectors up to 300 psi. The use with power injectors claim is the same as the claim on other Ethicon Endo-Surgery (previously Johnson & Johnson Medical) catheters with regard to use with power injectors.

Performance Data: Bench testing demonstrates that the 14 to 22 gauge ACUVANCE catheters are appropriate for use with power injectors up to 300 psi.

1

Image /page/1/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's logo in the center. The logo is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.

AUG 1 2001

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Katie Fordyce Regulatory Affairs Associate Ethicon Endo-Surgery, Incorporated 4545 Creek Road Cincinnati, Ohio 45242-2839

K012128 Re : PROTECTIV ACUVANCE IV Catheter Trade/Device Name: System 880.5200 Regulation Number: Regulatory Class: II Product Code: FOZ Dated: July 6, 2001 Received: July 9, 2001

Dear Ms. Fordyce:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the marke is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate abe beaud in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action.

2

Page 2 - Ms. Fordyce

concerning your device in the Federal Register. Please note: this response to your premarket notification submission does curs response co figation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on compreated as (advertising of your device, please contact the office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.gov/cdrh/dsma/dsmamain.html".

sincerely yours,

hy A. Ulatowski Time Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

3

510(k) Number (if known): K 012128

Device Name: PROTECTIV* ACUVANCE IV Catheter System_

Indications for Use:

A properly placed I.V. catheter provides access to a vein or artery. The I.V. Catheter System is designed for single use and has a needlestick protection feature. The risk of accidental needlesticks is reduced by a self-blunting needle system activated automatically as the catheter is threaded into the vessel. These catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. 14 to 22 gauge catheters may be used with power injectors up to 300 psi.

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurence of CDRH, Office of Device Evaulation (ODE)

(Optional Format 3-10-98)

Pattria Cucenote

Division of Dental. and General Hospital 510(k) Number .