K Number
K011958
Manufacturer
Date Cleared
2001-12-04

(165 days)

Product Code
Regulation Number
866.5270
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CRP-Latex (II) SEIKEN High Sensitivity Assay kit is a quantitative assay for use with an Automated Chemical Analyzer for the determination of C-reactive protein in human serum and in heparinized and EDTA-plasma. Measurement of C-reactive protein is useful in the detection and evaluation of infection, tissue injury, and inflammatory disorders.

Device Description

The CRP-Latex (II) SEIKEN High Sensitivity Assay Kit is a latex in vitro diagnostic immunoassay for the quantitative determination of C-reactive protein in human serum. Antigen in the sample bonds to the specific anti-CRP antibody, which has been adsorbed to latex particles, and agglutinates. The agglutination is detected as an absorbance change when read on an automated analyzer (the Hitachi 717 was used for these studies), with the magnitude of the change being proportional to the quantity of CRP in the sample. The actual concentration is then determined by interpolation from a calibration curve prepared from calibrators of known concentration.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the DENKA SEIKEN CO.,LTD. CRP-Latex (II) SEIKEN High Sensitivity Assay Kit, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Correlation with Predicate Device (Donor Samples)r = 0.999 (Correlation Coefficient)
Slope (Least Squares)1.034
Intercept (Least Squares)-0.173
Slope (Passing/Bablock)1.014
Intercept (Passing/Bablock)-0.094
Correlation with Predicate Device (Normal Adult Donor Samples)r = 0.996 (Correlation Coefficient)
Precision (Within Run & Between Day)
Level 1 (% CV)Did not exceed 1.81%
Level 2 (% CV)Did not exceed 1.65%
Level 3 (% CV)Did not exceed 1.3%
Lower Level of Detection (Sensitivity)0.05 mg/L
Assay RangeUp to 10.0 mg/L

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • 70 donor samples for comparative performance studies against the predicate device.
    • 388 normal adult donor samples for supportive clinical data.
  • Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. It refers to "donor samples" and "normal adult donor samples," suggesting a clinical setting for sample collection.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not mention the use of experts to establish ground truth for the test set. The study compares the new device against a legally marketed predicate device, using the predicate device's results as the reference for comparison.

4. Adjudication Method for the Test Set

Not applicable. The study involved a direct comparison of a new device's readings with a predicate device's readings, rather than an expert review process requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This study is focused on the analytical performance of an in vitro diagnostic assay, comparing its results to a predicate device, not on the impact of human readers with or without AI assistance.

6. Standalone Performance Study

Yes, a standalone performance study was done in the sense that the device's analytical performance (precision, sensitivity, assay range) was evaluated independently. However, the primary "standalone" output as commonly understood for AI/CAD devices (algorithm-only performance) is not directly applicable here. The device itself is an assay, and its performance is measured by its output values. The comparison with the predicate device also represents a form of standalone performance evaluation in relation to an established method.

7. Type of Ground Truth Used

The "ground truth" for the comparative performance study was established by the results obtained from the legally marketed predicate device (N High Sensitivity CRP Assay [Dade Behring Inc.]). For analytical performance metrics like precision, the "ground truth" would be derived from the inherent analytical properties of the assay and control materials.

8. Sample Size for the Training Set

Not applicable. This is an in-vitro diagnostic assay based on immunological principles, not an AI/machine learning device that requires a training set in the conventional sense. The device's mechanism is based on antigen-antibody reactions, not on learning from a dataset.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set in the context of an AI/machine learning algorithm for this device.

{0}------------------------------------------------

DENKA SEIKEN CO.,LTD

3-4-2, Nihonbashi kayabacho, Chuo-ku, Tokyo, Japan 103-0025

I. 510(k) Summary

DEC 0 4 2001

This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510 (k) number is:

(A)(1) Submitter's name: Denka Seiken Co., Ltd.

Submitter's address: 3-4-2, Nihonbashi kayabacho, Chuo-ku Tokyo, Japan 103-0025

Submitter's telephone number: (03) 3669-9421

Contact Person: Mr. Yousuke Meguro Assistant Manager International Sales and Business Development Dept.

Date Summary Prepared: June 20, 2001

(2) Trade or proprietary device name: CRP-Latex (II) SEIKEN High Sensitivity Assay Common or usual name: C-reactive protein (II) Assay Classification Name: C-reactive protein immunological test system Panel: Immunology Class: II

(3) Legally marketed predicate device: N High Sensitivity CRP Assay Dade Behring Inc. .

(4) Subject device description:

The CRP-Latex (II) SEIKEN High Sensitivity Assay Kit is a latex in vitro diagnostic immunoassay for the quantitative determination of C-reactive protein in human serum. Antigen in the sample bonds to the specific anti-CRP antibody, which has been adsorbed to latex particles, and agglutinates. The agglutination is detected as an absorbance change when read on an automated analyzer (the Hitachi 717 was used for these studies), with the magnitude of the change being proportional to the quantity of CRP in the sample. The actual concentration is then determined by interpolation from a calibration curve prepared from calibrators of known concentration.

{1}------------------------------------------------

DENKA SEIKEN CO.,LTD.

3-4-2, Nihonbashi kayabacho, Chuo-ku, Tokyo, Japan 103-0025

The CRP-Latex (II) SEIKEN High Sensitivity Assay Kit is a quantitative assay for use with an Automated Analyzer for the determination of C-reactive protein in human serum and plasma. Measurement of C-reactive protein is useful in the detection and evaluation of infection, tissue injury, and inflammatory disorders.

(6) Performance data:

The CRP-Latex (II) Seiken High Sensitivity Assay and the predicate device, N High Sensitivity CRP Assay have only minor difference that do not affect the performance, safety or effectiveness of the measurement.

Comparative performance studies conducted on 70 donor samples yielded a high correlation coefficient upon comparison of the CRP -Latex (II) Seiken High Sensitivity and the predicate device. N High Sensitivity CRP. The correlation coefficient r = 0.999; slope = 1.034. v intercept = - 0.173 (Least squares): slope = 1.014. v intercept = - 0.094 (Passing/Bablock). Supportive clinical data from 388 normal adult donor samples confirm this finding, with correlation coefficient r = 0.996.

Precision studies, both within run and between day studies, were performed using three levels of control material. % CV for Level 1 did not exceed 1.81%; for Level 2, % CV did not exceed 1.65%; and for Level 3, % CV did not exceed 1.3%.

The lower level of detection (sensitivity of the assay) is at 0.05mg/L, with the assay range up to 10.0 mg/L.

These findings serve to demonstrate that the performance of the CRP-Latex (II) Seiken High Sensitivity Assay kit is substantially equivalent to the predicate device, N High Sensitivity CRP (Dade Behring).

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of an eagle or other bird with outstretched wings. The image is in black and white.

De.Seiya Sato, Ph.D. Plant Director Denka Seiken Co., Ltd. 3-4-2, Nihonbashi-Kayabacho, Chuo-Ku, Tokyo, Japan 103-0025

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 0 4 2001

Re: K011958

Trade/Device Name: CRP-Latex (II) SEIKEN High Sensitivity Assay Kit Regulation Number: 21 CFR 866.5270 Regulation Name: C-reactive protein immunological test system Regulatory Class: Class II Product Code: DCN Dated: September 28, 2001 Received: October 1, 2001

Dear Dr. De. Seiya Sato:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Denka Seiken Co., Ltd. Pre-market Notification CRP-Latex (II) SEIKEN High Sensitivity Assay Kit

C. Indications for use of the Device

Page 1 of 1

510(k) Number): ______________________________________________________________________________________________________________________________________________________________

CRP-Latex (II) SEIKEN High Sensitivity Assay Kit High Sensitivity Assay Device Name: Kit

Indications for Use:

The CRP-Latex (II) SEIKEN High Sensitivity Assay kit is a quantitative assay for use with an Automated Chemical Analyzer for the determination of C-reactive protein in human serum and in heparinized and EDTA-plasma. Measurement of C-reactive protein is useful in the detection and evaluation of infection, tissue injury, and inflammatory disorders.

(Please do not write below this line-continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices

510(k) Number

Prescription Use_X or (Per 21 CFR 801.109) (Optional Format 1-2-96)

Over-the-Counter Use

§ 866.5270 C-reactive protein immunological test system.

(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).