K Number
K011863
Date Cleared
2002-02-15

(246 days)

Product Code
Regulation Number
870.4450
Panel
RA
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VCS-A Series Clamp is indicated for use in the temporary occlusion of blood vessels during vascular surgical procedures.

Device Description

The VCS-A Series Clamp with Ultrasound Doppler is a ring handled mechanical surgical instrument with a ratchet closure to adjust the tension required to occlude the target vessel. The Clamp has an integrated Doppler probe, which allows for blood flow monitoring by using a conventional portable Doppler Ultrasound unit. The VCS-A Series Clamp is made of surgical grade stainless steel and is available in various sizes. The clamp is a sterile, single use device. The method of sterilization will be Gamma Radiation with a SAL of 106.

AI/ML Overview
  1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document states that "The performance and functional testing demonstrated that the vascular clamp is substantially equivalent to the predicate devices. The VCS-A series Clamp with Ultrasound Doppler does not raise any new safety, effectiveness, or performance issues. Data are on file at Vascular Control Systems, Inc."

    Since the document doesn't explicitly list numerical acceptance criteria or specific performance metrics, the table below reflects this general statement of substantial equivalence to predicate devices.

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Overall PerformanceSubstantial equivalence to predicate devices regarding safety, effectiveness, and performance.The VCS-A series Clamp with Ultrasound Doppler demonstrated substantial equivalence to predicate devices, raising no new safety, effectiveness, or performance issues.
    Functionality(Implicit: Temporary occlusion of blood vessels and integrated Doppler functionality)Demonstrated functionality consistent with intended use and predicate devices.
    Safety(Implicit: No new safety concerns compared to predicate devices)No new safety issues were identified.
    Effectiveness(Implicit: Achieves intended purpose effectively)No new effectiveness issues were identified.
    Biocompatibility(Implicit: Suitable for contact with blood vessels)(Not explicitly stated, but implied by regulatory clearance and lack of new safety issues)
    SterilizationGamma Radiation with SAL of 10-6Sterilization method (Gamma Radiation with SAL of 10-6) is specified. Performance of this method implies it met acceptance criteria for sterility.
    Material Properties(Implicit: Suitable for surgical use)(Not explicitly stated, but implied by regulatory clearance and lack of new safety issues)
  2. Sample size used for the test set and the data provenance:

    The document states that "Data are on file at Vascular Control Systems, Inc." without providing specific details regarding the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature).

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The study focuses on "performance and functional testing" rather than expert-driven ground truth assessment for image interpretation or diagnosis.

  4. Adjudication method for the test set:

    This information is not provided in the document.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    An MRMC comparative effectiveness study was not done. The device is a mechanical surgical clamp with an integrated Doppler, not an AI-assisted diagnostic tool for human readers.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable as the device is a physical surgical instrument, not an algorithm. Therefore, a standalone algorithm-only performance study would not be relevant.

  7. The type of ground truth used:

    The "ground truth" for this device would be its ability to physically perform its intended function (temporary occlusion of blood vessels) and the accurate function of the integrated Doppler. This would be assessed through engineering and functional testing rather than expert consensus, pathology, or outcomes data in the traditional sense of a diagnostic device. The performance and functional testing implicitly served as the "ground truth" verification against engineering specifications and predicate device performance.

  8. The sample size for the training set:

    This information is not applicable. The VCS-A Series Clamp with Ultrasound Doppler is a mechanical device, not a machine learning or AI algorithm that requires a training set.

  9. How the ground truth for the training set was established:

    This is not applicable, as there is no training set for this type of device.

§ 870.4450 Vascular clamp.

(a)
Identification. A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.(b)
Classification. Class II (performance standards).