K Number
K994291

Validate with FDA (Live)

Manufacturer
Date Cleared
2000-03-01

(71 days)

Product Code
Regulation Number
862.1150
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Special Protein Calibrator for the Dimension® clinical chemistry system is an in vitro diagnostic device intended to establish points of reference that are used in determination of human IgA, IgG, IgM, complement C3, complement C4, and Transferrin values.

Device Description

The Dimension® Special Protein Calibrator is a liquid human serum-based product supplied as a kit of ten vials; two at each of five levels.

AI/ML Overview

The provided text is a 510(k) summary for the Dade Behring Dimension® Special Protein Calibrator. This document describes a calibrator device and its comparison to a predicate device, rather than a study on a diagnostic device's performance. Therefore, most of the requested information regarding acceptance criteria, study design, expert involvement, and ground truth is not applicable.

Here's an breakdown of the relevant and non-relevant sections based on your request:

1. A table of acceptance criteria and the reported device performance

  • Not applicable as traditionally defined for a diagnostic device. This submission focuses on establishing substantial equivalence for a calibrator. The "performance" being demonstrated is that the calibrator functions comparably to a legally marketed predicate calibrator for its intended use, which is to calibrate specific assays. The document doesn't present specific performance metrics (like sensitivity, specificity, accuracy) for a diagnostic outcome.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable. There is no "test set" in the sense of patient samples being evaluated by an algorithm or device to generate a diagnostic output. This is a calibrator, and its equivalence is based on its intended use, analytes, matrix, form, volume, levels, and reference to a standard (IFCC, CRM 470).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. No ground truth for a test set is established by experts for this type of submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. No test set or adjudication method is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not a study involving human readers or AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a calibrator, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Not applicable in the typical diagnostic sense. The "ground truth" for calibrators relates to their traceable reference standards. The document explicitly states the calibrator uses "IFCC, CRM 470" as a reference, which are international standards for protein measurements.

8. The sample size for the training set

  • Not applicable. There is no training set mentioned for this product.

9. How the ground truth for the training set was established

  • Not applicable. There is no training set mentioned.

Summary of the Acceptance Criteria and Study (as presented in the 510(k) context):

The "acceptance criteria" in this 510(k) submission are implicitly tied to demonstrating substantial equivalence to a predicate device. For a calibrator, this means showing that the new device is as safe and effective as the predicate and does not raise new questions of safety or effectiveness.

Table of Acceptance Criteria (based on substantial equivalence comparison):

Feature/AspectAcceptance Criteria (based on comparison to Predicate Device: Beckman CAL 1)Reported Device Performance (Dimension® Special Protein Calibrator)
Intended UseMust be a calibrator for the same or a subset of clinically relevant analytes in a comparable context.Intended to be used to calibrate C3, C4, Transferrin (TRNF), Immunoglobulin G (IGG), Immunoglobulin A (IGA), and Immunoglobulin M (IGM) methods on the Dimension® clinical chemistry system. (Matches predicate for listed analytes).
AnalytesMust calibrate the same analytes or a subset of the predicate's analytes.Calibrates Complement C3 (C3), Complement C4 (C4), Transferrin (TRNF), Immunoglobulin G (IGG), Immunoglobulin A (IGA), and Immunoglobulin M (IGM). (Matches predicate for these six analytes; predicate has four additional).
MatrixHuman serum base.Human serum base. (Matches predicate).
FormLiquid.Liquid. (Matches predicate).
Volume (per vial)Comparable or acceptable for intended use. (Predicate: 3.0 mL reconstituted).1.5 mL per vial. (Different volume, but implied to be acceptable for intended use).
LevelsSuitable number of levels for calibration. (Predicate: 1 level).5 levels. (Different number of levels, but potentially an improvement or different approach to calibration, implied to be acceptable).
ReferenceTraceable to recognized international standards.IFCC, CRM 470. (Matches predicate).

Study Proving Device Meets Acceptance Criteria:

The "study" in this context is the comparison to a predicate device as presented in the "Summary of Safety and Effectiveness Information" section. The document explicitly states:

  • "Both the Beckman CAL 1 and the Dimension® Special Protein Calibrator are intended as calibrators for C3, C4, Transferrin, IgG, IgA, and IgM methods." This addresses the primary functional sameness.
  • The table systematically compares the two devices across key attributes (Intended Use, Analytes, Matrix, Form, Volume, Levels, Reference).
  • Conclusion: "The Dimension® Special Protein Calibrator is substantially equivalent to the Beckman CAL 1 calibrator based on the comparison discussed above."

This form of demonstration, by detailed direct comparison to a legally marketed predicate device, is the standard method for establishing substantial equivalence for many Class I and Class II devices in 510(k) submissions where performance data from clinical studies are not typically required for calibrators. No specific clinical performance study measuring sensitivity, specificity, etc., was conducted or described in this submission, as the device's role is to ensure accurate calibration of other assays.

{0}------------------------------------------------

K994291

MAR - 1 2000

DADE BEHRING

DADE BEHRING INC. P.O. Box 6101 Newark, DE 19714

Summary of Safety and Effectiveness Information

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR$807.92.

Submitter's Name:Richard M. VaughtDade Behring Inc.P.O. Box 6101Newark, DE 19714-6101
Date of Preparation:December 20, 1999
Name of Product:Dimension® Special Protein Calibrator
FDA Classification Name:Calibrator
Predicate Device:Beckman CAL 1
Device Description:The Dimension® Special Protein Calibrator is a liquid human serum-based product supplied as a kit of ten vials; two at each of five levels.
Intended Use:The Special Protein Calibrator is intended to be used to calibrate thecomplement component, C3 (C3), complement component,C4 (C4),Transferrin (TRNF), Immunoglobulin G (IGG), Immunoglobulin A(IGA), and Immunoglobulin M (IGM) methods on the Dimension®clinical chemistry system.

Comparison to Predicate Device:

Dimension®
DeviceBeckman CAL 1Special Protein Calibrator
Intended Use:CalibratorCalibrator
Analytes:Complement C3 (C3)Complement C4 (C4)Transferrin (TRF)Immunoglobulin G (IgG)Immunoglobulin A (IgA)Immunoglobulin M (IgM)Alpha₁ -acid glycoprotein (AAG)Haptoglobin (HPT)Kappa light chain (KAP)Lambda light chain (CAM)Complement C3 (C3)Complement C4 (C4)Transferrin (TRNF)Immunoglobulin G (IGG)Immunoglobulin A (IGA)Immunoglobulin M (IGM)
Matrix:Form:Volume:Levels:Reference:Human serum baseLiquid3.0 mL per vial, reconstituted1 levelIFCC, CRM 470Human serum baseLiquid1.5 mL per vial5 levelsIFCC, CRM 470

{1}------------------------------------------------

Comments on Substantial Equivalence:

Both the Beckman CAL 1 and the Dimension® Special Protein Calibrator are intended as calibrators for C3, C4, Transferrin, IgG, IgA, and IgM methods.

Conclusion:

The Dimension® Special Protein Calibrator is substantially equivalent to the Beckman CAL 1 calibrator based on the comparison discussed above.

K.M. Knight

Richard M. Vaught Regulatory Affairs and Compliance Manager Date: December 20, 1999

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and head. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

MAR = 1 2000

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Richard M. Vaught Regulatory Affairs and Compliance Manager Dade Behring, Inc. P.O. Box 6101 Newark, Delaware 19714-6101

Re: K994291 Trade Name: Dimension® Special Protein Calibrator Regulatory Class: II Product Code: JIT Dated: December 20, 1999 Received: December 21, 1999

Dear Mr. Vaught:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE STATEMENT

Device Name:

Dimension® Special Protein Calibrator

Indications for Use:

The Special Protein Calibrator for the Dimension® clinical chemistry system is an in vitro diagnostic device intended to establish points of reference that are used in determination of human IgA, IgG, IgM, complement C3, complement C4, and Transferrin values.

Richard M. Vaught Regulatory Affairs and Compliance Manager

在新闻网易是民营商品名誉的地址:老宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝宝

February 17, 2000

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Peter E. Mahison

(Division Sign-Division of Clinical Laboratory Dev 510(k) Numbe

Prescription Use
(Per 21 CFR 801.109)


OR

Over-the-counter Use_

(Optional format 1-2-96)

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.