K Number
K992601

Validate with FDA (Live)

Date Cleared
1999-10-27

(85 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is an electronic clinical thermometer using a thermistor to detect body temperature from the oral, ampit, and rectal in the neonatal, pediatric and adult population used in the clinical and home testing.

Device Description

The OSTI Temp - Teller Digital Thermometer, (Models DT-101, DT-102, DT-103, DT-201, DT-202, and DT-203). is an electronic thermometer using a thermistor as the temperature sensor. The sensor's electric signal is then calculated and displayed by an ASIC (Application Specific IC) Models DT-101 and DT-201 display the temperature within a 100" of a degree. Models DT-101, DT-102 and DT-103 employ a 1.5 times larger LCD display than competitive products for easier reading. Models DT-101, DT-102, DT-201 and DT-202 have a water-resistant function. The digital thermometer comprises: a thermistor for temperature sensing, a reference resistor for comparing the resistance of the thermistor, a buzzer for sounding effect, an ASIC and a LCD display for calculating and displaying the target temperature digitally which the thermistor is immersed. The system uses a 1.5V DC battery for the power supply and the battery power is displayed in LCD if the battery is exhausted.

AI/ML Overview

This document describes a 510(k) submission for the Temp Teller-Digital Thermometer. The information provided heavily emphasizes comparison to predicate devices and adherence to standards rather than a detailed study for performance. Therefore, many of the requested fields cannot be extracted or are explicitly stated as not applicable.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly define "acceptance criteria" in terms of specific performance thresholds for a study. Instead, it refers to compliance with voluntary standards. The "device performance" is primarily stated as being "similar" to predicate devices.

Acceptance Criteria (Implied)Reported Device Performance
Compliance with ASTM E1112Stated compliance with ASTM E1112
Compliance with IEC 60601-1Stated compliance with IEC 60601-1
Compliance with IEC 60601-1-2Stated compliance with IEC 60601-1-2
Similar design and intended use to predicate devices"Similar in design and intended use to the predicates"
No new questions of safety or effectiveness compared to predicates"differences in their technological characteristics do not raise any new questions of safety or effectiveness."
Functional operation (e.g., power supply, display)Described features like 1.5V DC battery, LCD display, buzzer, water-resistant function (for some models).

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not applicable. The document explicitly states: "Controlled human clinical studies were not conducted using the OSTI Temp Teller Digital Thermometer and predicate devices." The substantiation relies on "bench testing" and "similarity to predicate devices."
  • Data Provenance: Not applicable, as detailed clinical testing was not performed.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

Not applicable, as controlled human clinical studies were not conducted.

4. Adjudication Method for the Test Set:

Not applicable, as controlled human clinical studies were not conducted.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Controlled human clinical studies were not conducted."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

Based on the context of a digital thermometer, which is a physical device, the concept of an "algorithm only" performance study is not directly applicable in the same way it would be for AI software. The performance testing described is likely intrinsic to the device's hardware and embedded logic. The document indicates "bench testing" was performed, which would be a standalone evaluation of the device's physical and electronic properties.

7. The Type of Ground Truth Used:

For the "bench testing," the ground truth would have been established by calibrated reference thermometers and controlled environmental conditions, according to the standards (e.g., ASTM E1112 for electronic thermometers). The document does not specify the exact methods or reference standards used for setting this ground truth beyond referencing the ASTM standard itself.

8. The Sample Size for the Training Set:

Not applicable. This device is a physical electronic thermometer, not a machine learning algorithm that requires a "training set" in the conventional sense. Its design and calibration are based on established engineering principles and compliance with standards.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no "training set" for this type of device.

{0}------------------------------------------------

OCT 27 1999

K 992601

EXHIBIT # 1

510(K) SUMMARY

This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 5l0(k) number is: ______________

Submitter's Identification: 1.

Oriental System Technology Inc. 2F No. 23, Industry E. Road 9" Science Based Industrial Park Hsinehu, Taiwan, R.O.C.

Contact:

Mr. Herman Lee General Manager

Date Summary Prepared: July 1999

2. Name of the Device:

Temp Teller-Digital Thermometer, Models DT-101, DT-102, DT-103, DT-201, DT-202, and DT-203,

3. Predicate Device Information:

    1. Q-Tips®, Faichney, Private Label Clinical Electronic Thermometers K# 962497
    1. Wallgreens PAP-WD

4. Device Description:

The OSTI Temp - Teller Digital Thermometer, (Models DT-101, DT-102, DT-103, DT-201, DT-202, and DT-203). is an electronic thermometer using a thermistor as the temperature sensor. The sensor's electric signal is then calculated and displayed by an ASIC (Application Specific IC) Models DT-101 and DT-201 display the temperature within a 100" of a degree. Models DT-101, DT-102 and DT-103 employ a 1.5 times larger LCD display than competitive products for easier reading. Models DT-101, DT-102, DT-201 and DT-202 have a water-resistant function.

The digital thermometer comprises: a thermistor for temperature sensing, a reference resistor for comparing the resistance of the thermistor, a buzzer for sounding effect, an ASIC and a LCD display for calculating and displaying the target temperature digitally which the thermistor is immersed.

{1}------------------------------------------------

The system uses a 1.5V DC battery for the power supply and the battery power is The system uses a 1.3V DC battery for the power Eappy and displayed in LCD if the battery is exhausted.

Intended Use: ട്.

The device is an electronic clinical thermometer using a thermistor to detect body The device is an electionile official the neonatal, pediatric and adult population used in the clinical and home testing.

Comparison to Predicate Devices: 6.

The OSTI Temp Teller – Digital Thermometer, Models DT-101, DT-102, DT-103, DT-The OOT Temp Tollor - Bigkar bother to the following digital thermometers.

  • a. Q-I ips®, Faichney, Private Label Clinical Electronic Thermometers K# 962497
  • b. Wallgreens PAP-WD

The OSTI Temp Teller Digital Thermometer is similar in design and intended use to the rrne OOTT Temp Teller Digiter Front range and resolution. All products use the same predicates differing only in ambient - thermistor as well as the use of a metal cap to protect temperature sensing chomblay, ASIC, a buzzer and a 1.55V battery.

Discussion of Non-Clinical Tests Performed for Determination of Substantial 7. Equivalence are as follows:

Compliance to applicable voluntary standards includes ASTM E1112, and Compliance to uppliedblo velancery -1 and IEC 60601-1-2 requirements.

Guidance Documents included the FDA "Guidance On The Content of Premarket Notification (510(k)) Submissions for Clinical Electronic Thermometers".

Discussion of Clinical Tests Performed: 8.

Controlled human clinical studies were not conducted using the OSTI Temp Teller Digital Thermometer and predicate devices.

ರು Conclusions:

The OSTI Temp Teller -- Digital Thermometer, Models DT-101, DT-102, DT-103, DT-201, DT-202, and DT-203, have the same intended use and similar technological characteristics as predicate devices. Moreover, bench testing toonhologiour onaracterission and clinical testing supplied demonstrate that any differences in their technological characteristics do not raise any new questions affierorioso iffectiveness. Thus, the OSTI Temp Teller Digital Thermometer, Models DT-101, DT-102, DT-103, DT-201, DT-201, and DT-203 is substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is written in a simple, sans-serif font. The words are arranged horizontally, with "Public" and "Health" on the first line and "Service" on the second line. The text appears to be part of a document or sign.

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is in a sans-serif font. The logo is black and white.

OCT 2 7 1999

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Oriental System Technology, Incorporated C/O Ms. Susan D. Goldstein-Falk Official correspondent for Oriental System Technology, Incorporated MDI Consultants, Incorporated 55 Northern Boulevard, Suite 410 Great Neck, New York 11021

Re : K992601 Trade Name: Temp Teller - Digital Thermometer, Models DT-101, DT-102, DT-103, DT-201, DT-202, and DT-203 Regulatory Class : II Product Code: FLL Dated: July 30, 1999 Received: August 3, 1999

Dear Ms. Susan D. Goldstein-Falk:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any

{3}------------------------------------------------

Page 2 - Ms. Susan D. Goldstein-Falk

obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sinderely yours,

Alilut

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Exhibit B

Page __ 1 __ of __ 1 __

510(k) Number (if known):

Device Name: Oriental System Technology Inc. Temp Teller Digital Thermometer, Models DT-101, DT-102, DT-103, DT-201, DT-202, DT-203

Indications For Use:

The device is an electronic clinical thermometer using a thermistor to detect body temperature from the oral, ampit, and rectal in the neonatal, pediatric and adult population used in the clinical and home testing.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-The-Counter Use
(Optional Format 1-2-96)

Patricia Cuemke
(Division sign-off)

OR

(Division sign-off) Division of Dental, Infection control, And General Hospital Devices 510(k) number _ A 992601

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.