K Number
K992320
Date Cleared
2000-05-18

(311 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Megabello Balloon Dilatation Catheters are recommended for percutaneous transluminal angioplasty (PTA) of the iliac and femoral vessels whose lumens are obstructed by atherosclerotic plaque. Megabello catheters are not indicated for use in coronary arteries nor in the neurovasculature. Any other use than those indicated is not recommended.

Device Description

The Megabello Percutaneous Transluminal Angioplasty (PTA) Catheter consists of a multilumen catheter with a controlled-compliance balloon mounted at the distal tip. The balloon is designed to inflate to a known diameter and length at a specific pressure and is inflated and deflated via the side port. A second lumen allows access to the distal tip of the catheter for guidewire insertion. Two marker bands are positioned under the balloon at the proximal and distal tapers to aid in accurate placement.

AI/ML Overview

The provided document is a 510(k) summary for the Megabello Percutaneous Transluminal Angioplasty (PTA) Catheter. It focuses on demonstrating substantial equivalence to a predicate device rather than providing detailed acceptance criteria and a study report for an AI/ML powered device.

Therefore, I cannot provide the requested information as it pertains to AI/ML powered medical devices and is not present in the provided text. The document describes a traditional medical device (balloon catheter) and its non-clinical performance testing.

Here's why the specific points in your request cannot be answered from the provided text:

  • Acceptance criteria and reported device performance (table): The document lists technical characteristics of the device and the predicate but does not present acceptance criteria in a quantitative, pass/fail manner for performance metrics. It generally states that testing was performed "similar to those in the guidance" and found acceptable.
  • Sample size and data provenance: Not applicable in the context of AI/ML evaluation. The document mentions non-clinical (i.e., bench) testing, not a clinical study with a test set of data.
  • Number of experts and qualifications: Not applicable. There's no mention of expert review for establishing ground truth as there's no diagnostic output being generated by an AI.
  • Adjudication method: Not applicable for the same reason as above.
  • MRMC comparative effectiveness study: Not applicable, as this is not an AI/ML device being evaluated for its impact on human reader performance.
  • Standalone performance: Not applicable, as it's a physical catheter, not a standalone algorithm.
  • Type of ground truth used: Not applicable, as there's no diagnostic task or ground truth to establish. Testing focused on physical properties (e.g., burst strength, compliance).
  • Sample size for training set: Not applicable, as there's no AI/ML model to train.
  • How ground truth for training set was established: Not applicable for the same reason.

The document primarily demonstrates substantial equivalence through:

  • Comparison of intended use.
  • Comparison of basic construction and technical characteristics.
  • Non-clinical performance testing according to recognized guidance documents (e.g., "Guidance for the Submission of Research and Marketing Applications for Interventional Cardiology Devices"). These tests, such as Balloon Minimum Burst Strength, Balloon Compliance, and Bond Strength, are standard for evaluating the physical and mechanical properties of catheters. The document states these tests were "performed similar to those in the guidance" and the catheter was "found acceptable," implying it met internal specifications or industry standards, but specific numerical acceptance criteria are not detailed.

{0}------------------------------------------------

MAY 1 8 2000

K992320

510(k) Summary
----------------

Date Prepared

February 18, 2000

Submitter

SubmitterAddress:Boston Scientific CorporationSchneider (USA) Inc5905 Nathan LaneMinneapolis, MN 55442
Phone :Fax :(612) 694-5500(612) 694-5858
Contact PersonRonald W. BennettRegulatory Affairs Project Manager
Device Name and Classification
Trade NameMegabello Percutaneous Transluminal Angioplasty(PTA) Catheter

Catheter

Class II

Device Description

Predicate Device

Common Name

Classification

The Megabello Percutaneous Transluminal Angioplasty (PTA) Catheter consists of a multilumen catheter with a controlled-compliance balloon mounted at the distal tip. The balloon is designed to inflate to a known diameter and length at a specific pressure and is inflated and deflated via the side port. A second lumen allows access to the distal tip of the catheter for guidewire insertion. Two marker bands are positioned under the balloon at the proximal and distal tapers to aid in accurate placement.

Percutaneous Transluminal Angioplasty (PTA)

SMASH TM Percutaneous Transluminal

Angioplasty (PTA) Catheter - K972512

Intended Use

Megabello Balloon Dilatation Catheters are recommended for percutaneous transluminal angioplasty (PTA) of the iliac and femoral vessels whose lumens are obstructed by atherosclerotic plaque. Megabello catheters are not indicated for use in coronary arteries nor in the neurovasculature. Any other use than those indicated is not recommended.

{1}------------------------------------------------

Technical Characteristic Comparison to Predicate

The Megabello Percutaneous Transluminal Angioplasty (PTA) Catheter has the same intended use and basic construction as the predicate device, the SMASH™ Percutaneous Transluminal Angioplasty (PTA) Catheter. The following table compares the technical characteristics of the two devices:

FeatureMEGABELLOPercutaneous TransluminalAngioplasty (PTA) CatheterSmash™ PercutaneousTransluminal Angioplasty(PTA) Catheter
ConfigurationMultilumenMultilumen
Shaft Length (cm)75-12060-120
Shaft French Size75
Balloon Length (mm)20-4020-80
Balloon Diameter (mm)12, 143-12
Rated Burst Pressure (atm)77-15
Guidewire Size (inches)0.0350.035
Balloon Markers2 (90/10 Platinum/ Iridium)2 (Gold)
Balloon MaterialNylonNylon
Manifold MaterialPebaxABS
Shaft MaterialPebax 6333Nylon

Performance Data

The Megabello Percutaneous Transluminal Angioplasty (PTA) Catheter had non-clinical performance testing, as did the predicate device, the SMASH™M Percutaneous Transluminal Angioplasty (PTA) Catheter.

Testing of the Megabello Percutaneous Transluminal Angioplasty (PTA) Catheter generally followed the PTCA Balloon Catheters section in Part II of the "Guidance for the Submission of Research and Marketing Applications for Interventional Cardiology Devices", May 1994. The following types of tests were performed similar to those in the guidance:

Balloon Minimum Burst Strength Balloon Compliance (Distensibility) Balloon Inflation/Deflation Performance Balloon Fatigue (Repeated Balloon Inflation) Bond Strength Catheter Diameter and Balloon Profile Catheter Body Burst Pressure

{2}------------------------------------------------

The following additional evaluations were also performed:

Sheath Removal Profile

The biocompatibility of the catheter was also tested and found acceptable.

Summary

In summary, Boston Scientific Corporation has demonstrated the Megabello Percutaneous Transluminal Angioplasty (PTA) Catheter is substantially equivalent to the SMASHIM Percutaneous Transluminal Angioplasty (PTA) Catheter based on design and test results for the indications for use given, and is therefore acceptable for commercialization.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white circular logo. The logo contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle. Inside the circle is a stylized emblem consisting of three curved lines that resemble a person with outstretched arms.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 8 2000

Mr. Ronald W. Bennett Regulatory Affairs Project Manager Boston Scientific Corporation Scneider (USA) Inc. 5905 Nathan Lane Minneapolis, MN 55442

K992320 Re: Megabello Balloon Dilatation Catheter Regulatory Class: II (two) Product Code: LIT February 18, 2000 Dated: February 22, 2000 Received:

Dear Mr. Bennett:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{4}------------------------------------------------

Page 2 - Mr. Ronald W. Bennett

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Bin E. Hawr
James E. Dillard III

Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):

Device Name:

Megabello Percutaneous Transluminal Angioplasty (PTA) Catheter

Indications for Use:

Megabello Balloon Dilatation Catheters are recommended for percutaneous transluminal angioplasty (PTA) of the iliac and femoral vessels whose lumens are obstructed by atherosclerotic plaque. Megabello catheters are not indicated for use in coronary arteries nor in the neurovasculature. Any other use than those indicated is not recommended.

Tom E. Harvey


(Division Sign-Off)
Division of Cardiovascular, Respiratory,
5/11/10

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

and Neurological Devi

510(k) Number

OR

Over-The-Counter Use

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).