(73 days)
The Revolution TX/ Digital Radiographic Table System is indicated for use in generating radiographic images of human anatomy. It is Intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammographic applications.
The Revolution TX/i Digital Radiographic Table System is designed to perform radiographic x-ray examinations. The detector is comprised of amorphous silicon with a cesium iodide scintillator. The resulting digital image can be sent through a DICOM network for applications such as printing, viewing and storage. The Revolution TX/i Digital Radiographic Table System consists of a an elevating radiographic table with integrated digital detector, x-ray tube, x-ray tube hanger, collimator, system controller, and generator. A separate, conventional SG100 vertical bucky stand is provided for chest or other general purpose radiographic procedures.
The provided text is a 510(k) summary for the GE Medical Systems Revolution TX/i Digital Radiographic Table System. It states that the device is substantially equivalent to a predicate device and does not contain detailed acceptance criteria or a comprehensive study report with the requested specifics.
Here's a breakdown of the information that can be extracted versus what is not available from the provided text, based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
- Not available: The document does not specify acceptance criteria in terms of quantitative performance metrics (e.g., sensitivity, specificity, accuracy, image quality scores, etc.) nor does it report specific device performance against such criteria.
2. Sample size used for the test set and the data provenance
- Not available: No information is provided about a specific test set, its sample size, or data provenance (country of origin, retrospective/prospective).
3. Number of experts used to establish the ground truth for the test set and qualifications of those experts
- Not available: Since no specific test set or performance evaluation is described, there's no mention of experts or their qualifications for establishing ground truth.
4. Adjudication method for the test set
- Not available: No information on an adjudication method for a test set is provided.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and its effect size.
- Not available: The document does not mention an MRMC study or any effect size related to human reader improvement with or without AI assistance. The device predates widespread AI integration in such systems.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not available: This is a hardware system, not an AI algorithm, so a standalone performance study in that context is not applicable and not mentioned.
7. The type of ground truth used
- Not available: No ground truth methodology is described, as a formal performance study against a specific clinical condition is not presented.
8. The sample size for the training set
- Not available: Since this is a hardware device submission focusing on substantial equivalence, there is no mention of a "training set" in the context of an algorithm.
9. How the ground truth for the training set was established
- Not available: As above, no training set is discussed.
Summary of the "Study" Mentioned in the Document:
The document states:
- "Summary of Studies: The device has the same detector and acquisition system as the predicate device, and the same intended uses. It will be evaluated for conformance with UL and IEC safety standards. We consider the device to be substantially equivalent to the predicate device."
This indicates that the "study" conducted for this 510(k) submission was primarily a comparison to a predicate device to establish substantial equivalence, focusing on:
- Identical detector and acquisition system.
- Identical intended uses.
- Conformance with UL and IEC safety standards.
It is not a clinical performance study in the sense of demonstrating diagnostic accuracy against a ground truth for specific clinical conditions, but rather a demonstration that the new device functions similarly and is as safe as a previously cleared device. The conclusion states that the TX/i system provides "radiographic images that result in equivalent or better diagnostic capabilities than film / screen images," but this is a general statement of equivalence rather than a result from a specific comparative clinical trial detailed in this document.
{0}------------------------------------------------
GE Medical Systems
AUG 30 1999
f
General Electric Company PO Box 414, Milwaukee, WI 53201
K99 2066
SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Pat 807.87(h).
| Identification of Submitter: | Larry A. Kroger, Ph.D.Senior Regulatory Programs ManagerGE Medical SystemsTel. (414) 544-3894Summary prepared: June 15, 1999 |
|---|---|
| Identification of Product: | Revolution TX/i Digital Radiographic Table System |
| Classification Name: | Stationary X-ray System |
| Manufacturer: | GE Medical Systems3000 N. Grandview Blvd.Waukesha, WI 53118 |
| Marketed Devices: | Digital detector (part of Digital Radiographic ImagingSystem, renamed Revolution XQ/i system) (K982196);SCX (Advantx) radiographic system (K862120);Ultranet-SA Collimator (K894142); Maxiray 100Radiographic Tube (Pre-amendment); SCPU Generator(K940277) and SG100 vertical bucky stand (Pre-amendment). |
| Device Description: | The Revolution TX/i Digital Radiographic Table Systemis designed to perform radiographic x-ray examinations.The detector is comprised of amorphous silicon with acesium iodide scintillator. The resulting digital imagecan be sent through a DICOM network for applicationssuch as printing, viewing and storage. The RevolutionTX/i Digital Radiographic Table System consists of a anelevating radiographic table with integrated digitaldetector, x-ray tube, x-ray tube hanger, collimator,system controller, and generator. A separate,conventional SG100 vertical bucky stand is provided forchest or other general purpose radiographicprocedures. |
| Indications for Use: | The Revolution TX/i Digital Radiographic Table Systemis indicated for use in generating radiographic imagesof human anatomy. It is intended to replace |
{1}------------------------------------------------
radiographic film / screen systems in all general purpose diagnostic procedures. This device is not intended for mammographic applications.
- Comparison with Predicate: The Revolution TX/i Digital Radiographic Table System is substantially equivalent to the Revolution XQ/i system, originally cleared as Digital Radiographic Imaging System in 510(k) K982196. The table used in the Digital Radiographic Table System is substantially equivalent to the Compax 40E table cleared in 510(k) K884930.
Summary of Studies: The device has the same detector and acquisition system as the predicate device, and the same intended uses. It will be evaluated for conformance with UL and IEC safety standards. We consider the device to be substantially equivalent to the predicate device.
GE considers the Revolution TX/i Digital Radiographic Conclusions: Table System to be equivalent with the predicate device. The TX/i Digital Radiographic Table System provides radiographic images that result in equivalent or better diagnostic capabilities than film / screen images. The potential hazards, e.g., wrong measurements and misdiagnosis, are controlled by a risk management plan including:
- · A Hazard Analysis
- · A Software Development and Validation Process
- · Certification to applicable UL and IEC safety standards
- · External evaluations
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
AUG 20 2013
Larry A. Kroger. Ph.D. Senior Regulatory Programs Manager General Electric Company PO Box 414 MILWAUKEE WI 53201
Re: K992066
Trade/Device Name: Revolution TX/I Digital Radiographic Table System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR and MOB Dated: June 17, 1999 Received: June 18, 1999
Dear Dr. Kroger:
This letter corrects our substantially equivalent letter of August 30, 1999.
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
STATEMENT OF INTENDED USE
510(k) Number (if known): K992066
Device Name: Revolution TX/i Digital Radiographic Table System
Indications for Use
The Revolution TX/ Digital Radiographic Table System is indicated for use in generating radiographic images of human anatomy. It is Intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammographic applications.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801-109)
ﮯ۔
OR Over-The-Counter Use_
David G. Ingram
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devices 510(k) Number
N/A