K Number
K982196

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-10-06

(106 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Digital Radiographic Imaging System is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammographic applications.

Device Description

The Digital Radiographic Imaging System is designed to perform radiographic examinations. The detector is comprised of amorphous silicon with a cesium iodide scintillator. The resulting digital image can be sent through a DICOM network for applications such as printing, viewing and storage. The Digital Radiographic Imaging System consists of a wallstand, tubestand, x-ray tube, collimator, system controller, generator and digital detector.

AI/ML Overview

The provided text describes a 510(k) summary for a Digital Radiographic Imaging System by GE Medical Systems. The study aims to demonstrate substantial equivalence to a predicate device, specifically the Advantx Radiographic System combined with an SG-60 vertical wallstand. The primary acceptance criterion revolves around achieving equivalent or better imaging performance compared to film/screen images for diagnostic purposes.

Here is a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Diagnostic equivalence to radiographic film/screen systems.Digital images had equivalent or better image quality than paired film/screen images, as evaluated by radiologists.
No new safety concerns compared to the predicate device.The Digital Radiographic Imaging System presents no new safety concerns.
Compliance with x-ray requirements of 21CFR.Will comply with the x-ray requirements of 21CFR.
Compliance with safety requirements of UL2601-1, IEC601-1, and collateral standards.Will comply with safety requirements of UL2601-1, IEC601-1, and collateral standards.

(Note: The primary focus of the provided text is on image quality for diagnostic equivalence.)

2. Sample size used for the test set and the data provenance:

  • Sample Size: 30 paired digital and film/screen images.
  • Data Provenance: The text does not explicitly state the country of origin. It mentions studies were conducted at "three different research hospitals," suggesting prospective data collection for the purpose of the study. The phrasing "representative anatomical areas" implies a selection of cases.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: Two radiologists at each of three hospitals, totaling six radiologists.
  • Qualifications of Experts: The text identifies them simply as "radiologists." No further details on their experience level (e.g., years of experience, subspecialty) are provided.

4. Adjudication method for the test set:

  • The text states that radiologists "evaluated" the images and "found that the digital images had equivalent or better image quality." It does not specify a formal adjudication method like 2+1 or 3+1. It implies a consensus or individual assessment resulting in the stated conclusion.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • A MRMC comparative effectiveness study was not done as described for AI assistance. The study described is a comparison of digital images to film/screen images by human readers, not a comparison of human readers with vs. without AI assistance. The device is the digital imaging system, not an AI assisting human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • No standalone (algorithm-only) performance study was described. The study explicitly involved human radiologists evaluating the images.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The ground truth was based on expert assessment/consensus of the image quality for diagnostic purposes, comparing the digital images directly to the corresponding film/screen images for the same anatomical areas.

8. The sample size for the training set:

  • The text does not provide information about a specific training set or its sample size. The description focuses on the evaluation data. The device itself is an imaging system, not an AI with a distinct training phase in the context usually meant by "training set" for AI/machine learning.

9. How the ground truth for the training set was established:

  • As no training set is described in the provided text in the context of an AI/machine learning model, this information is not available.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circular shape. The letters and the circle are filled with a dark color, contrasting with the white background. The logo is positioned to the right side of the image.

OCT

e lass

K982196

GE Medical Systems

General Linten Company P ( Box 11: Aldy Juker: 31: 3 . 81-9412

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Pat 807.87(h).

Identification of Submitter:Larry A. Kroger, Ph.D.Senior Regulatory Programs ManagerGE Medical SystemsTel. (414) 544-3894Summary prepared: 19 June, 1998
Identification of Product:Digital Radiographic Imaging System

Stationary X-ray System Classification Name: Manufacturer: GE Medical Systems 3000 N. Grandview Blvd. Waukesha, WI 53188

Marketed Devices: The SG60 vertical wallstand was pre-Amendment and the SCX radiographic system was introduced in 1986 (K862120). Subsequently, these were used with the Ultranet-SA Collimator (K894142), the Maxiray 100 Radiographic Tube (K812915), and the SCPU Generator (K940277).

  • Device Description: The Digital Radiographic Imaging System is designed to perform radiographic examinations. The detector is comprised of amorphous silicon with a cesium iodide scintillator. The resulting digital image can be sent through a DICOM network for applications such as printing, viewing and storage. The Digital Radiographic Imaging System consists of a wallstand, tubestand, x-ray tube, collimator, system controller, generator and digital detector.
  • Indications for Use: The Digital Radiographic Imaging System is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film / screen systems in all general purpose diagnostic procedures. This device is not intended for mammographic applications.

{1}------------------------------------------------

Comparison with It is the opinion of GE Medical Systems that the Digital Predicate: Radiographic Imaging System is of comparable type and substantially equivalent to the Advantx Radiographic System (originally cleared as the SCX K862120) combined with a SG-60 vertical wallstand with respect to intended use, radiation characteristics and image quality. The Digital Radiographic Imaging System presents no new safety concerns. It will comply with the x-ray requirements of 21CFR as well as safety requirements of UL2601-1, IEC601-1 and collateral standards. Summary of Studies: Two radiologists at each of three hospitals evaluated 30 paired digital and film / screen images of representative anatomical

Conclusions: GE considers the Digital Radiographic Imaging System to be equivalent with the predicate device. Digital Radiographic Imaging System provides radiograms that result in equivalent or better imaging performance than film / screen images. The potential hazards, e.g., wrong measurements and misdiagnosis, are controlled by a risk management plan including:

  • · A Hazard Analysis
    image quality.

  • · A Software Development and Validation Process

  • · External validations of paired sets of film / screen image and digital images by three different research hospitals to assess the diagnostic equivalence of the digital images.

areas, and found that the digital images had equivalent or better

:
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health and Human Services. The logo includes the department's seal on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The text is in a sans-serif font and is blue.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager General Electric Company PO Box 414 MILWAUKEE WI 53201

AUG 20 2013

Re: K982196

Trade/Device Name: GE Digital Radiographic Imaging System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR and MQB Dated: September 4. 1998 Received: September 8, 1998

Dear Dr. Kroger:

This letter corrects our substantially equivalent letter of October 6. 1998.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

STATEMENT OF INTENDED USE

510(k) Number (if known):

510(k) Number (if known): K982/960

Device Name: Digital Radiographic Imaging System

Indications for Use

The Digital Radiographic Imaging System is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammographic applications.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801-109)
OR Over-The-Counter Use_

OR Over-The-Counter Use_

Eirid a. Sijmm
(Division Sign-Off)

(Division Sign-Off) Division of Reproductive, Abdominal, El and Radiological Devi 510(k) Number