K Number
K990554

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-03-22

(28 days)

Product Code
Regulation Number
862.1150
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dimension® Special Protein Calibrator is intended to be used to calibrate the Immunoglobulin G (IGG), Immunoglobulin A (IGA), and Immunoglobulin M (IGM) methods on the Dimension® clinical chemistry system.
The Special Protein Calibrator for the Dimension® Clinical Chemistry System is an in vitro diagnostic device intended for use in an in vitro test system to establish points of reference that are used in determination of values in the measurement of human substances.

Device Description

The Dimension® Special Protein Calibrator is a liquid human serum-based product. The kit consists of ten vials; two at each of five levels.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (Dimension® Special Protein Calibrator). It focuses on establishing substantial equivalence to a predicate device rather than presenting a performance study with acceptance criteria.

Therefore, the document does not contain the information requested to fill out the table regarding acceptance criteria and a study proving the device meets those criteria.

Specifically:

  • No acceptance criteria are mentioned. The document is focused on comparing the new calibrator to a predicate calibrator.
  • No specific study results (performance, accuracy, precision, etc.) are reported for the Dimension® Special Protein Calibrator itself. The comparison is limited to intended use, analytes, matrix, form, volume, levels, and reference.
  • There is no discussion of sample sizes, data provenance, expert panels, adjudication methods, MRMC studies, standalone performance, or ground truth establishment. These are all elements typically found in a study designed to demonstrate performance against set acceptance criteria, which is not what this 510(k) summary provides.

{0}------------------------------------------------

3/22/99

K990554

Dade Behring

DADE BEHRING INC. P.O. Box 6101 Newark, DE 19714

Summary of Safety and Effectiveness Information

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter's Name:Cathy P. CraftDade Behring Inc.P.O. Box 6101Newark, DE 19714-6101
Date of Preparation:February 3, 1999
Name of Product:Dimension® Special Protein Calibrator
FDA Classification Name:Calibrator
Predicate Device:Beckman CAL I

Device Description: The Dimension® Special Protein Calibrator is a liquid human serum-based product. The kit consists of ten vials; two at each of five levels.

Intended use: The Dimension® Special Protein Calibrator is intended to be used to calibrate the Immunoglobulin G (IGG), Immunoglobulin A (IGA), and Immunoglobulin M (IGM) methods on the Dimension® clinical chemistry system.

Comparison to Predicate Device:

ItemBeckman CAL IDimension® Special Protein Calibrator
Intended UseCalibratorCalibrator
AnalytesImmunoglobulin G (IgG)Immunoglobulin A (IgA)Immunoglobulin M (IgM)Alpha1-acid glycoprotein (AAG)Complement C3 (C3)Complement C4 (C4)Haptoglobin (HPT)Kappa light chain (KAP)Lambda light chain (CAM)Transferrin (TRF)Immunoglobulin G (IgG)Immunoglobulin A (IgA)Immunoglobulin M (IgM)
MatrixHuman serum baseHuman serum base
FormLiquidLiquid
Volume3.0 mL per vial, reconstituted1.5 mL per vial
Levels1 level5 levels
ReferenceIFCC, CRM 470IFCC, CRM 470

{1}------------------------------------------------

Comments on Substantial Equivalence: Both the Beckman CAL I and the Dimension® Special Protein Calibrator are intended to be used as calibrators for IGG, IGA, and IGM methods.

Conclusion: The Dimension® Special Protein Calibrator is substantially equivalent to the Beckman Immunoglobulin Calibrator based on the comparison discussed above.

Cathy P. Craft

Regulatory Affairs and Compliance Manager Date: February 8, 1999

{2}------------------------------------------------

MAR 2 2 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Cathy P. Craft Regulatory Affairs and Compliance Manager Dade Behring Inc. P.O. Box 6101 Newark, DE 19714-6101

Re: K990554

Trade Name: Dimension® Special Protein Calibrator Regulatory Class: II Product Code: JIT Dated: February 8, 1999 Received: February 22, 1999

Dear Ms. Craft:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications For Use Statement

Device Name: Dimension® Special Protein Calibrator

Indications for Use:

The Special Protein Calibrator for the Dimension® Clinical Chemistry System is an in vitro diagnostic device intended for use in an in vitro test system to establish points of reference that are used in determination of values in the measurement of human substances.

Cathy P. Craft Regulatory Affairs and Compliance Manager

February 8, 1999

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Peter E. Makim

ion Sign-Off 510(k) Num

Prescription Use
(Per 21 CFR 801.109)

OR

Over-the-counter Use_

. ﺍﻟﻤﺘﺎﺯ

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.