K Number
K984099

Validate with FDA (Live)

Date Cleared
1998-12-10

(23 days)

Product Code
Regulation Number
864.8625
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

QBC Centrifugal Hematology Control is intended as a quality control for QBC+, QBC II, QBC II Plus, QBC Reference and QBC AUTOREAD centrifugal hematology systems, to monitor hematocrit, hemoglobin, white blood cell count, granulocyte count, lymphocyte/monocyte count, and platelet count.

Device Description

The product is an in vitro diagnostic reagent composed of mammalian erythrocytes, mammalian leukocytes and simulated platelets in a plasma-like fluid with preservatives. OBC Control is composed of stable materials that provide a means of monitoring the performance of QBC centrifugal hematology systems. Overall system performance of disposable OBC blood tubes, including tube coating and float, is also monitored. QBC control is available in two levels that have varying concentrations of the parameters measured. QBC Control is used and tested in the same manner as patient samples.

AI/ML Overview

The provided text is a 510(k) Summary for the QBC® Centrifugal Hematology Control device. It describes the device, its intended use, and a nonclinical study conducted to demonstrate substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Stability: Remain within the assay range over the life of the product (expiration dating).QBC® Centrifugal Hematology Control "passed the acceptance criteria of remaining within the assay range over the life of the product." Expiration dating established at 21 days (closed vial) and 8 days (open vial) when stored at 2-8°C.
Precision: Demonstrate low standard deviations and %CVs.QBC® Centrifugal Hematology Control "demonstrated precision as indicated by the small standard deviations and %CVs obtained during testing."
Improved platelet performance."Improved platelet performance was also noted." (No specific quantitative metric provided in the summary, but it's stated as an observed improvement).

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The text states "Nonclinical testing of 3 validation lots centered on the performance attributes of stability and precision." While the number of "lots" is provided, the specific number of individual samples or measurements within those lots is not specified.
  • Data Provenance: The text does not explicitly state the country of origin of the data. It is a nonclinical study conducted by R&D Systems, Inc. in Minneapolis, MN, USA. The study design is implied to be prospective as it's a validation study for a new product, though not explicitly stated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

  • Not Applicable. This was a nonclinical performance study of a quality control material, not a study involving human interpretation of clinical data in comparison to expert ground truth. The "ground truth" for a quality control material would be its expected range of values for each parameter, which is typically established by the manufacturer through rigorous testing and characterization.

4. Adjudication Method for the Test Set:

  • Not Applicable. As noted above, this was a nonclinical performance study. No human adjudication of results was required as the evaluation was against pre-defined performance specifications (stability, precision).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No. An MRMC study is not relevant for this type of device (a hematology control). This device is an in vitro diagnostic reagent used to monitor the performance of hematology analyzers, not to be interpreted by human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, in essence. The entire nonclinical study described is effectively a "standalone" performance evaluation of the quality control material itself when used with the intended hematology systems. There is no "human-in-the-loop" component to the performance of the control material, only human operation of the analyzer and interpretation of the analyzer's output in relation to the control's established ranges.

7. The Type of Ground Truth Used:

  • For stability and precision, the "ground truth" is typically the manufacturer-established assay range and statistical performance targets for the control material. These targets are derived from extensive characterization of the product's components and expected performance over time and under various conditions. The text mentions "remaining within the assay range" and "small standard deviations and %CVs," indicating these pre-defined targets as the ground truth.

8. The Sample Size for the Training Set:

  • Not applicable/Not specified. This device is a quality control material, not an AI/ML algorithm that requires a "training set" in the conventional sense. The "training" for such a product involves its initial formulation, characterization, and the establishment of its expected performance characteristics. The 3 validation lots are for testing, not training.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. As above, there's no "training set" in the AI/ML context. The expected performance characteristics (the "ground truth" for a QC material) are established by the manufacturer through their R&D processes, including formulation, optimization, and extensive testing during product development to define its intended assay ranges, stability, and precision.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a logo for R&D Systems. The logo features the letters "R" and "D" in a bold, sans-serif font, with an ampersand (&) symbol placed between them. Below the letters, the word "SYSTEMS" is written in a smaller, sans-serif font. The logo appears to be in black and white.

DEC 1 0 1998

Special 510(k) Summary OBC® Centrifugal Hematology Control

Date of Summary: Company Name:

Contact name:

Classification name: Classification code:

Product name: CFR section: Device Class:

November 16, 1998 R&D Systems, Inc. 614 McKinley Place N.E. Minneapolis, MN 55413 Kenneth T. Edds. Ph.D. 612-379-2956, FAX 612-379-6580

multiparameter hematology control 81JPK Hematology Control mixtures for Quality Control QBC® Centrifugal Hematology Control 864.8625 Class II

Device to which substantial equivalence is claimed:

OBC® Centrifugal Hematology Control, manufactured by R&D Systems, Inc. and currently being sold by Becton Dickinson Primary Care Diagnostics, Sparks, MD. 510(k) number: K954137

The product is an in vitro diagnostic reagent composed of mammalian erythrocytes, mammalian leukocytes and simulated platelets in a plasma-like fluid with preservatives. OBC Control is composed of stable materials that provide a means of monitoring the performance of QBC centrifugal hematology systems. Overall system performance of disposable OBC blood tubes, including tube coating and float, is also monitored. QBC control is available in two levels that have varying concentrations of the parameters measured. QBC Control is used and tested in the same manner as patient samples.

Intended use: OBC Centrifugal Hematology Control is intended as a quality control for OBC+. OBC II. OBC II Plus, QBC Reference and QBC AUTOREAD centrifugal hematology systems, to monitor hematocrit, hemoglobin, white blood cell count, granulocyte count, lymphocyte/monocyte count, and platelet count.

QBC® Centrifugal Hematology Control has an intended use that is similar to the predicate device. The technologies of the two devices are similar.

Nonclinical testing of 3 validation lots centered on the performance attributes of stability and precision. QBC® Centrifugal Hematology Control passed the acceptance criteria of remaining within the assay range over the life of the product. QBC® Centrifugal Hematology Control also demonstrated precision as indicated by the small standard deviations and %CVs obtained during testing. Improved platelet performance was also noted. Expiration dating has been established at 21 days in the customers hands (closed vial) and 8 days open vial when stored at 2-8°C and handled according to instructions for use.

R & D Systems, Inc. 614 McKinley Place N.E. Minneapolis, MN 55413

PHONE: (612) 379-2956 FAX: (612) 379-6580

WATS: (800) 343-7475 E-MAIL: info@rndsystems.com

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread.

Public Health Service

DEC 1 0 1998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Kenneth T. Edds, Ph.D. Requlatory Affairs R&D Systems, Inc. 614 McKinley Place N.E. Minneapolis, Minnesota 55413

Re : K984099 Trade Name: QBC® Centrifuqal Hematology Control Regulatory Class: II Product Code: JPK November 16, 1998 Dated: Received: November 17, 1998

Dear Dr. Edds:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obliqation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Paqe 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbrandinq by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number: K984099

Device Name:QBC Centrifugal Hematology Control

Indications for Use:

QBC Centrifugal Hematology Control is intended as a quality control for QBC+, QBC II, QBC II Plus, QBC Reference and QBC AUTOREAD centrifugal hematology systems, to monitor hematocrit, hemoglobin, white blood cell count, granulocyte count, lymphocyte/monocyte count, and platelet count.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) NumberK984099
Prescription UseOROver-The-CounterUse
---------------------------------------------------------

(Optional Format 1-2-96)

§ 864.8625 Hematology quality control mixture.

(a)
Identification. A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).(b)
Classification. Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.