(58 days)
The DePuy Duraloc Acetabular Cup Liners are to be used with the DePuy Duraloc Acetabular Cup System's metal acetabular shells to resurface the acetabular socket in cemented or cementless total hip arthroplasty.
The DePuy Duraloc Acetabular Liners are manufactured from ultra high molecular weight polyethylene (UHMWPE) and are available in two designs. standard and Charnley Bore (CB), Both designs are available in two styles, a neutral and 10° lip, with an inner diameter (I.D.) of 22.225mm and are available in sizes to fit Duraloc metal acetabular shells with outer diameters of 48mm-74mm. The Duraloc liners are locked into the Duraloc metal acetabular shells with a metal wire locking ring supplied with the shells.
Here's an analysis of the provided text regarding acceptance criteria and the supporting study, structured as requested.
It's important to note that the provided text is a 510(k) summary for a medical device (DePuy Duraloc® Acetabular Cup System - Enduron Liner). 510(k) submissions typically focus on demonstrating substantial equivalence to a predicate device rather than presenting a detailed new performance study against specific acceptance criteria for a novel device. Therefore, many of the requested fields related to a performance study will be "Not Applicable" or "Not Provided" in this context.
Acceptance Criteria and Device Performance Study Analysis
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Substantial Equivalence | Design: Similar to predicate devices (K900832, K900891, K961186) with minor changes (smaller I.D., larger O.D. range, additional 2mm PE for CB liners). | The standard liners are "identical in design" to previously cleared devices except for smaller inner diameter. CB liners are "identical in design" except for smaller inner diameter and additional 2mm PE. |
| Material: Identical material (UHMWPE, ASTM F-648) to predicate devices. | UHMWPE, ASTM F-648. | |
| Intended Use: Identical intended use to predicate devices. | "Intended use... has not changed from the cleared... UHMWPE liners of the DePuy Duraloc Acetabular Cup System." | |
| Technological Characteristics: Fundamental scientific technology has not changed. | "Fundamental scientific technology... has not changed." | |
| Key Specifications Comparison | Inner Diameters: Proposed 22.225mm. | Matches current submission, comparable but smaller than some predicates (26, 28, 32mm) and within range of others (22.225, 26, 28mm). |
| Outer Diameters: Proposed 48-74mm. | Matches current submission, larger range than some predicates (38-46mm) and comparable to others (46-74mm). | |
| Design Type: Proposed Standard and Charnley Bore (CB). | Matches current submission, predicate devices were primarily "Standard." | |
| Lip: Proposed Neutral and 10°. | Matches current submission, predicate devices had Neutral, 10°, and 20°. | |
| Locking Mechanism: Proposed Metal Wire Ring. | Matches current submission and all predicate devices. |
2. Sample size used for the test set and the data provenance
- Sample Size: Not applicable. This submission is based on demonstrating substantial equivalence through design and material comparison, not a performance study involving a test set of data.
- Data Provenance: Not applicable.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. Ground truth for a performance study is not relevant to a substantial equivalence demonstration based on design, material, and intended use comparison.
4. Adjudication method for the test set
- Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is a component of a hip prosthesis, not an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable. This device is a physical medical implant, not an algorithm.
7. The type of ground truth used
- Not applicable. For a substantial equivalence claim, the "ground truth" is largely the established safety and effectiveness of the predicate devices based on their prior clearances and clinical history.
8. The sample size for the training set
- Not applicable. No training set is relevant for this type of device and submission.
9. How the ground truth for the training set was established
- Not applicable.
{0}------------------------------------------------
2 1998 DEC
SUMMARY OF SAFETY AND EFFECTIVENESS
| NAME OF SPONSOR: | DePuy®, Inc.P.O. Box 988Warsaw, Indiana 46581-0988 |
|---|---|
| 510(k) CONTACT: | Sally FoustSenior Regulatory Submissions AssociateDePuy Orthopaedics, Inc.1 (219) 372-7455; FAX: 1 (219) 267-7098E-Mail: Sally_Foust@ccgate.depuy.com |
| TRADE NAME: | DePuy Duraloc® Acetabular Cup System - EnduronLiner |
| COMMON NAME: | Hip Cup Liner |
| CLASSIFICATION: | 888..3358 - Hip joint metal/polymer semi-constrainedcementless prosthesis |
DEVICE PRODUCT CODE: 87 LPH and JDM
SUBSTANTIALLY EQUIVALENT DEVICE:
Modified DePuy Acetabular Cup System (K900832) Modified AML® + Acetabular Cup Prosthesis (K900891) DePuy Duraloc® Cementless Acetabular Cup System (K961186)
DEVICE DESCRIPTION AND INTENDED USE:
The DePuy Duraloc Acetabular Liners are manufactured from ultra high molecular weight polyethylene (UHMWPE) and are available in two designs. standard and Charnley Bore (CB), Both designs are available in two styles, a neutral and 10° lip, with an inner diameter (I.D.) of 22.225mm and are available in sizes to fit Duraloc metal acetabular shells with outer diameters of 48mm-74mm. The Duraloc liners are locked into the Duraloc metal acetabular shells with a metal wire locking ring supplied with the shells.
The DePuy Duraloc Acetabular Cup Liners are to be used with the DePuy Duraloc Acetabular Cup System's metal acetabular shells to resurface the acetabular socket in cemented or cementless total hip arthroplasty.
BASIS OF SUBSTANTIAL EQUIVALENCE:
The intended use as described in its labeling, fundamental scientific technology, and material of the 22.225mm I.D. UHMWPE liners of the DePuy Duraloc Acetabular Cup System has not changed from the cleared (K900832, K900891, and K961186) 22.225mm, 26mm, 28mm and 32mm I.D. UHMWPE liners of the DePuy Duraloc Acetabular Cup System.
Image /page/0/Picture/12 description: The image shows the number 000005. The numbers are in a simple, sans-serif font. The numbers are all black and are on a white background. The number 5 is slightly tilted to the right.
{1}------------------------------------------------
The standard liners are identical in design to the previously cleared (K900832. K900891) DePuv Duraloc acetabular cup liners except for the smaller (22.225mm) inner diameter. The standard liners are also identical in design to the previously cleared (K961186) DePuy Duraloc Bantam acetabular cup liners except the subject liners are available in larger sizes to fit Duraloc metal acetabular shells with outer diameters of 48mm-74mm.
The Charnley Bore (CB) liners are identical in design to the previously cleared (K900832, K900891) DePuy Duraloc acetabular cup liners except for the smaller (22,225mm) inner diameter and the additional 2mm of polyethylene added on the face of the liner wall. The Charnley Bore (CB) liners are also identical in design to the previously cleared (K961186) DePuy Duraloc Bantam acetabular cup liners except that the subject liners are available in larger sizes to fit Duraloc metal acetabular shells with outer diameters of 48-74mm and have an additional 2mm of polyethylene added on the face of the liner wall.
Based on similarities of design, material and intended use, DePuy believes that the subject DePuy Duraloc Acetabular Liners are substantially equivalent to the cleared liners of the DePuy Duraloc Acetabular Cup System.
| CurrentSubmission | ModifiedDePuyAcetabular CupSystem | Modified AMLAcetabular CupProsthesis | DePuy DuralocCementlessAcetabular CupSystem | |
|---|---|---|---|---|
| K900832 | K900891 | K961186 | ||
| Materials | UHMWPEASTM F-648 | UHMWPEASTM F-648 | UHMWPEASTM F-648 | UHMWPEASTM F-648 |
| InnerDiameters | 22.225mm | 26, 28and 32mm | 26, 28and 32mm | 22.225, 26and 28mm |
| OuterDiameters | 48-74mm | 46-74mm | 46-74mm | 38-46mm |
| Design | Standard andCharnley Bore(CB) | Standard | Standard | Standard |
| Lip | Neutraland 10° | Neutral,10° and 20° | Neutral,10° and 20° | 10° |
| LockingMechanism | Metal Wire Ring | Metal Wire Ring | Metal Wire Ring | Metal Wire Ring |
Refer to the following table for a summary of the similarities:
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
2 1998 DEC
Ms. Sally Foust Regulatory Submissions Associate DePuy, Inc. P.O. Box 988 700 Orthopedic Drive Warsaw, Indiana 46581-0988
Re: K983491 DePuy Duraloc® Acetabular Trade Name: Cup System Euduron Liner Regulatory Class: II Product Codes: LPH and JDM Dated: October 2, 1998 October 5, 1998 Received:
Dear Ms. Foust:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Ms. Sally Foust
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
ia M. Witten, Ph.D., M.D. Cel Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known)_K983491
Device Name: DePuy Duraloc Acetabular Cup Liners: Additional Liners
Indications for Use:
The DePuy Duraloc Acetabular Cup Liners are to be used with the DePuy Duraloc Acetabular Cup System's metal acetabular shells to resurface the acetabular socket in cemented or cementless total hip arthroplasty.
Concurrence of CDRH, Office of Device Evaluation
OR
Prescription Use_ X (Per 21 CFR 801.109)
Over-The-Counter Use_
Division Sign Off
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number. K983491
..
N/A