K Number
K983399

Validate with FDA (Live)

Date Cleared
1999-01-11

(105 days)

Product Code
Regulation Number
866.3220
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This is an immunochromatographic assay for the simultaneous qualitative detection and distinguishing of Giardia and Cryptosporidium specific antigens in aqueous extracts of fecal specimens.

It is intended for professional laboratory use.

For In Vitro Diagnostic Use.

Device Description

Immunoassay for Giardia and Cryptosporidium Antigens

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Genzyme Contrast® Giardia/Cryptosporidium Combo Rapid Assay:

Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance
SensitivityHigh sensitivity to detect GiardiaGiardia: 100% (Retrospective & Prospective)
High sensitivity to detect CryptosporidiumCryptosporidium: 97.3% (Retrospective), 100% (Prospective)
SpecificityHigh specificity to correctly identify negative cases for GiardiaGiardia: 100% (Retrospective), 99.1% (Prospective)
High specificity to correctly identify negative cases for CryptosporidiumCryptosporidium: 100% (Retrospective), 99.5% (Prospective)
Agreement with Predicate DeviceHigh agreement with a legally marketed predicate device (Alexon ProSpecT)Giardia: 97.2% agreement (Relative Agreement)
Cryptosporidium: 100% agreement (Relative Agreement)
Within-run Precision100% agreement100% agreement
Between-run Precision100% agreement100% agreement

Note: The document does not explicitly state numerical acceptance criteria thresholds like "Sensitivity > 95%". Instead, the performance is presented as the outcome of the study, implying that the observed high sensitivity and specificity values were considered acceptable for demonstrating substantial equivalence. The precision studies explicitly state 100% agreement as the specification.

Study Details

  1. Sample Size used for the test set and the data provenance:

    • Retrospective Analysis vs Microscopic Examination:
      • Giardia: 33 positive, 109 negative (total 142)
      • Cryptosporidium: 37 positive, 105 negative (total 142)
    • Prospective Analysis vs Microscopic Examination:
      • Giardia: 50 positive, 452 negative (total 502)
      • Cryptosporidium: 73 positive, 429 negative (total 502)
    • Retrospective Analysis vs Rapid EIA (Predicate):
      • Giardia: 142 samples (agreement reported for 138/142)
      • Cryptosporidium: 142 samples (agreement reported for 142/142)
    • Provenance: The document does not explicitly state the country of origin. Both "Retrospective Analysis" and "Prospective Analysis" are mentioned, indicating a mix of historical and newly collected samples for the microscopic comparison. The comparison against the predicate EIA is stated as "Retrospective Analysis."
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not specify the number or qualifications of experts. Ground truth was established by "microscopic examination," which is typically performed by trained laboratory personnel (e.g., medical technologists or microbiologists) but no specific expert details are provided.
  3. Adjudication method for the test set:

    • The document does not specify any adjudication method (e.g., 2+1, 3+1). It simply states "microscopic examination" as the reference method, implying a single definitive read or a standard laboratory process.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This study is for an in vitro diagnostic (IVD) assay, not an AI-assisted device for human readers. Therefore, an MRMC comparative effectiveness study comparing human readers with and without AI assistance is not applicable and was not performed.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, effectively. The "Genzyme Giardia/Cryptosporidium Assay" performance (sensitivity and specificity) is reported as a standalone device performance when compared against the reference method (microscopic examination) and the predicate method (Alexon ProSpecT). As an IVD, it is essentially an "algorithm only" in the sense that the assay itself produces the result, without direct human cognitive input to interpret an image or signal beyond reading the immunoassay result.
  6. The type of ground truth used:

    • Expert Consensus / Reference Method: "Microscopic examination" is stated as the "Reference method." This implies that microscopic identification of Giardia and Cryptosporidium organisms in fecal samples serves as the gold standard for establishing the ground truth.
  7. The sample size for the training set:

    • The document does not provide any information regarding a training set sample size. This is typical for traditional IVD assays, where method development and optimization (analogous to training) involve internal studies and reagent formulation, which are distinct from the formal clinical performance studies presented here.
  8. How the ground truth for the training set was established:

    • As no training set is explicitly mentioned or detailed, there is no information provided on how its ground truth was established.

{0}------------------------------------------------

zyme Diagnostics/Medix Biotech subsidiary

1983399 Contrast® GiardialCryptosporidium Combo Rapid Assay Reference No. K983399

December 23, 1998

San Carlos, CA 94070

JAN 1 1 1999

Industrial Road

CONFIDENTIAL

ATTACHMENT 1

510(k) Summary Of Safety and Effectiveness Information

Trade or Proprietary Name:Giardia/Cryptosporidium Combo Rapid Assay
Common or Usual Name:Immunoassay for Giardia and Cryptosporidium Antigens
Manufacturer:Genzyme DiagnosticsMedix Biotech subsidiary1531 Industrial RoadSan Carlos, CA 94070
Contact Person:Barbara Pizza, Manager Regulatory Affairs, (617) 252-7953Genzyme DiagnosticsOne Kendall Square, Building 200Cambridge, MA 02139-1562

The use of the Genzyme Contrast@ Giardia/Cryptosporidium Combo Rapid Assay in the clinical laboratory setting is substantially equivalent to a currently marketed method for Giardia and Cryptosporidium for the detection of Giardia and Cryptosporidium antigen in aqueous fecal specimens.

PERFORMANCE STUDIES:

Comparative Performance Studies

A Comparative performance study was conducted using the Genzyme GiardialCryptosporidium Assay, a Reference method (microscopic examination) and predicate method (Alexon ProSpecT). The summary of the sensitivity and specificity are provided below.

SensitivitySpecificity
Retrospective Analysis Genzyme vs Microscopic examination
Giardia33/33 = 100%109/109 = 100%
Cryptosporidium36/37 = 97.3%105/105 = 100%
Prospective Analysis Genzyme vs Microscopic examination
Giardia50/50 = 100%448/452 = 99.1%
Cryptosporidium73/73 = 100%427/429 = 99.5%
Retrospective Analysis Genzyme vs Rapid EIARelative Agreement
Giardia138/142 = 97.2%
Cryptosporidium142/142 = 100%

The Genzyme method yielded acceptable relative agreement with the predicate method.

{1}------------------------------------------------

Agreement

Both within-run and between-run studies were performed using Giardia and Cryptosporidium positive and neqative stool specimens. Five replicates of each specimen stool pool (Giardia positive, Cryptosporidium positive and GiardialCryptosporidium negative) was tested five times in one batch using the Contrast Giardial Cryptosporidium Combo Rapid Assay over five days. The Agreement was calculated for each stool specifications for within and between-run precision were 100% agreement.

Conclusion

Based on the results of the performance studies described above, the Genzyme Contrast Giardial Cryptosporidium Combo Rapid Assay is substantially equivalent in performance to the reference method (Standard microscopic examination) and to the predicate (Alexon ProSpecT) a legally marketed method for the detection of Giardia and Cryptosporidium antigen in aqueous fecal specimens.

In lieu of a 510(k) statement under 513(i) of the Act, this information is provided as a 510(k) summary for disclosure to any other persons/companies without the specific written authorization from Genzyme Corporation.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple and recognizable, and it is often used to represent the U.S. Department of Health and Human Services.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JAN 1 1 1999

Barbara Pizza Manager, Regulatory Affairs Genzyme Diagnostics One Kendall Square Cambridge, MA 02139-1562

K983399 Re:

Trade Name: Contrast® Giardia/Cryptosporidium Combo Rapid Assay Regulatory Class: II Product Code: MHI Dated: December 23, 1998 Received: December 24, 1998

Dear Ms. Pizza:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours.

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page__________________________________________________________________________________________________________________________________________________________________________ of 1

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name:4 Contrast® Giardia/Cryptosporidium Combo Rapid Assay

Indications For Use:

This is an immunochromatographic assay for the simultaneous qualitative detection and distinguishing of Giardia and Cryptosporidium specific antigens in aqueous extracts of fecal specimens.

It is intended for professional laboratory use.

For In Vitro Diagnostic Use.

Woody Dubois

vision of Clinical Laboratory Devices 510(k) Number.

PRESCRIPTION USE X

§ 866.3220

Entamoeba histolytica serological reagents.(a)
Identification. Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies toEntamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identifyEntamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasiteEntamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.