K Number
K983168

Validate with FDA (Live)

Date Cleared
1998-11-18

(69 days)

Product Code
Regulation Number
872.4200
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Avant handpiece will be used by dentists to remove decayed tooth structure, and to cut or alter the shape of teeth so they may be restored to form and function.

Device Description

The Avant is a high speed dental handpiece, 132 millimeters long, with a lightly textured finish for enhanced grip. It is a hand-held, channeled instrument that is powered by compressed air that is delivered through a hose to an air channel in the handpiece at 32-40 psi. This impels the turbine in the head of the handpiece to revolve at approximately 375.000 rpm. Other internal channels deliver air and water to the head for cleaning and cooling. The handpiece may contain, as an accessory, an optics device to light the area of operation. Any common dental bur for cutting (not supplied by Ora Innovations) is held in place in the handpiece head by a push-button chuck.

The Avant handpiece housing is constructed of a high temperature, medical grade liquid crystal polymer (Vectra A530 by Ticona, Summit, NJ). There are three major parts which are made from metal: a standard turbine (supplied by a manufacturer of dental handpiece turbines), the threaded connector at the back end which attaches to the standard air/water hose in the dental unit, and the internal tubes that carry the chip air and water. These three parts are substantially equivalent to those in the predicate device.

Similar to the predicate device, the Avant handpiece must be cleaned and lubricated with a quality commercial dental handpiece cleaner/lubricant. It is supplied non-sterile, but must be sterilized before use, as with the predicate device. It will be packaged as a single unit or in multiples.

AI/ML Overview

The provided document is a 510(k) summary for a dental handpiece, focusing on establishing substantial equivalence to a predicate device. It is a regulatory submission, not a study report designed to present detailed acceptance criteria and performance data for a device.

Therefore, the document does not contain the detailed information required to fill out the requested table and answer many of the questions. Specifically, it lacks:

  • A table of explicit acceptance criteria with numerical targets.
  • Reported device performance against specific metrics.
  • Details on sample sizes for test sets, data provenance, number/qualifications of experts, adjudication methods, MRMC study results, standalone performance, ground truth types for test/training sets, and training set sample size or ground truth establishment.

Based on the information available:

Acceptance Criteria and Reported Device Performance

The document states that "The following tests were derived from the American Dental Association guideline, ISO 7785-1 and G95-1/ISO-10993." and lists the tests:

  • Air pressure
  • Biocompatibility
  • Polymer material
  • Color
  • Extraction force
  • Rotational speed
  • Sterility validation

It then broadly claims, "The test reports, located in the Appendix of the 510(k) application, show that the Avant handpiece is safe and effective." However, it does not provide a specific table of acceptance criteria or the reported device performance values against these criteria.

Study Details

  1. A table of acceptance criteria and the reported device performance:

    • Not provided in the document. The document lists types of tests performed but does not specify the quantitative acceptance criteria or the numerical results of these tests.
  2. Sample size used for the test set and the data provenance:

    • Not provided. The document states "Performance testing: Comparative performance testing and clinical evaluations are not submitted as part of this 510(k)." This indicates that no separate clinical or comparative performance test set, as might be understood for AI/diagnostic devices, was used or submitted. The "tests" mentioned (air pressure, biocompatibility, etc.) would likely refer to engineering or material verification tests.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not provided. This type of information is typically relevant for diagnostic devices where expert review establishes ground truth. For this mechanical dental handpiece, the "ground truth" would be objective measurements against technical specifications.
  4. Adjudication method for the test set:

    • Not applicable/Not provided.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, not done. The device (a dental handpiece) is a mechanical tool, not an AI or diagnostic tool that would involve human readers or AI assistance in interpretation. The document explicitly states "Comparative performance testing and clinical evaluations are not submitted as part of this 510(k)."
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The device is a mechanical handpiece; there is no "algorithm" in the context of diagnostic or AI performance.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not explicitly stated in detail. For mechanical devices like this, the "ground truth" for the listed tests (air pressure, rotational speed, etc.) would be established by objective engineering measurements and adherence to specified test standards (e.g., ISO 7785-1, G95-1/ISO-10993). Biocompatibility would be assessed against specific material standards and biological reactivity tests.
  8. The sample size for the training set:

    • Not applicable/Not provided. This concept is relevant for AI models, not for a mechanical dental handpiece.
  9. How the ground truth for the training set was established:

    • Not applicable/Not provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date NOV 1 8 1998 at the top left. Below the date is the logo for Iora Innovations. The logo is a stylized letter "i" with a curved line above it, followed by the word "ORA" in bold letters. Below "ORA" is the word "INNOVATIONS" in smaller letters.

K983168

510(k) SUMMARY

Page 1 of 2

Gerald L. Cogan Contact person:

September 8, 1998 Date summary prepared:

Device name:

Proprietary name: Avant™ High Speed Handpiece Common or usual name: High speed dental handpiece Classification name: Dental handpiece and accessories, Class 1, 21 CFR 872.4200

Legally marketed device for substantial equivalence comparison: Midwest Tradition P High Speed Handpiece (K963050) Manufactured by Dentsply Midwest, Des Plaines, IL

Description of the device:

The Avant is a high speed dental handpiece, 132 millimeters long, with a lightly textured finish for enhanced grip. It is a hand-held, channeled instrument that is powered by compressed air that is delivered through a hose to an air channel in the handpiece at 32-40 psi. This impels the turbine in the head of the handpiece to revolve at approximately 375.000 rpm. Other internal channels deliver air and water to the head for cleaning and cooling. The handpiece may contain, as an accessory, an optics device to light the area of operation. Any common dental bur for cutting (not supplied by Ora Innovations) is held in place in the handpiece head by a push-button chuck.

The Avant handpiece housing is constructed of a high temperature, medical grade liquid crystal polymer (Vectra® A530 by Ticona, Summit, NJ). There are three major parts which are made from metal: a standard turbine (supplied by a manufacturer of dental handpiece turbines), the threaded connector at the back end which attaches to the standard air/water hose in the dental unit, and the internal tubes that carry the chip air and water. These three parts are substantially equivalent to those in the predicate device.

Similar to the predicate device, the Avant handpiece must be cleaned and lubricated with a quality commercial dental handpiece cleaner/lubricant. It is supplied non-sterile, but must be sterilized before use, as with the predicate device. It will be packaged as a single unit or in multiples.

Ora Innovations, Inc.

322 NW 5th Avenue, Suite 207

Portland, Oregon USA 97209

503.228.1123 Fax 503.228.1223

{1}------------------------------------------------

CONFIDENTIAL

Intended use of device:

The Avant handpiece will be used by dentists to remove decayed tooth structure, and to cut or alter the shape of teeth so they may be restored to form and function.

The following tests were derived from the American Dental Association guideline, ISO 7785-1 and G95-1/ISO-10993:

Air pressure Biocompatibility Polymer material Color Extraction force Rotational speed Sterility validation

The test reports, located in the Appendix of the 510(k) application, show that the Avant handpiece is safe and effective.

The Avant handpiece is similar to the predicate device in that they both perform the same functions in the same manner, directing air at high pressure to metal turbines which rotate dental cutting burs. Both devices have tubes which direct air and water to the head to clean and cool the area, and they both can be steam sterilized. In addition, they have the option of including an optical light feature.

The principal difference lies in the construction of the housing of the handpieces. The Avant head and handle are made from a high temperature, medical grade polymer that will retain dimensional stability and strength through multiple sterilization cycles. The head and handle of the predicate device are made of stainless steel.

Performance testing: Comparative performance testing and clinical evaluations are not submitted as part of this 510(k).

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized symbol resembling a person with outstretched arms, with three curved lines forming the body and arms. The symbol is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICE, USA" is arranged around the upper portion of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 8 1998

Gerald L. Cogan, DMD Director of Clinical Affairs Ora Innovations, Incorporated 319 S.W. Washington Street, Suite 620 Portland, Oregon 97204

K983168 Re : Trade Name: Avant High Speed Handpiece, Avant Fiber Optic High Speed Handpiece Requlatory Class: I Product Code: EFB September 8, 1998 Dated: September 10, 1998 Received:

Dear Dr. Cogan:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਮੈ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Dr. Cogan

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours

Timothy A. Ulatowski
Director

Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

{4}------------------------------------------------

Avant" High Speed Handpiece 510(k) Notification Page 3

Indications for Use Statement

510(k) Number: _______________________________________________________________________________________________________________________________________________________________

Device name: Avant High Speed Handpiece

Indications for Use:

The Avant handpiece will be used by dentists to remove decayed tooth structure, and to cut or alter the shape of teeth so they may be restored to form and function.

(Please do not write below this line)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription UseOR(Division Sign-Off)Division of Dental, Infection Control,and General Hospital Devices510(k) NumberOver-The-Counter Use

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.