K Number
K982711

Validate with FDA (Live)

Device Name
PREMIER GIARDIA
Date Cleared
1998-11-25

(113 days)

Product Code
Regulation Number
866.3220
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Premier Giardia enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Giardia antigens in human stool. Test results are intended to aid in the diagnosis of Giardia infection (giardiasis).

Device Description

The Premier Giardia test utilizes polyclonal anti-Giardia capture antibody adsorbed to microwells.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Premier Giardia device, based on the provided text:

Acceptance Criteria and Device Performance

The core acceptance criteria for the Premier Giardia device, as presented in comparison to its predicate device, are focused on its diagnostic performance metrics:

1. Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Implied by Predicate)Reported Device Performance (Premier Giardia)
Sensitivity98%98%
Specificity98%98%

Note: The document explicitly states that the Premier Giardia performed at 98% sensitivity and 98% specificity, directly matching the performance of the predicate device, the Alexon ProspecT Giardia Microplate Assay. This direct match implies that these performance metrics of the predicate device serve as the acceptance criteria for the new device.


2. Sample Size and Data Provenance for the Test Set

The provided 510(k) summary does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the data).


3. Number and Qualifications of Experts for Ground Truth

The document does not specify the number of experts used to establish the ground truth for the test set or their qualifications. For an in vitro diagnostic device like this, ground truth would typically be established by culture, microscopy, or other established reference methods interpreted by qualified laboratory personnel.


4. Adjudication Method for the Test Set

The document does not describe an adjudication method for the test set.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed or reported in this 510(k) summary. This type of study is more common for imaging devices where human interpretation is a critical component influencing diagnostic accuracy. The Premier Giardia is an in vitro diagnostic assay with a qualitative (positive/negative) output.


6. Standalone (Algorithm Only) Performance

The information provided describes the performance of the Premier Giardia as a standalone device. It's an enzyme immunoassay (EIA) that directly detects Giardia antigens. There is no indication of a human-in-the-loop component or an "algorithm only" performance separate from the assay itself. The "Level of Skill Required" is stated as a "Laboratory Technician," meaning technicians perform the assay according to the specified steps and interpret the results.


7. Type of Ground Truth Used

The document does not explicitly state the type of ground truth used for the performance evaluation (e.g., expert consensus, pathology, other reference methods, or outcomes data). However, for an in vitro diagnostic for infectious disease, the ground truth is typically established by:

  • Microscopy: Direct observation of Giardia cysts or trophozoites in stool samples by a trained microbiologist.
  • Culture: Growing the Giardia parasite in a laboratory setting (though this can be challenging for Giardia).
  • Reference molecular tests: PCR or other nucleic acid-based tests could also be used as a gold standard.

Given the context of a 510(k) for an EIA, it is highly probable that the reference "Performance vs. Reference Methods" implies comparison against microscopy or another established laboratory method for Giardia detection.


8. Sample Size for the Training Set

The provided 510(k) summary does not mention a training set sample size. For an immunoassay like the Premier Giardia, there isn't typically a "training set" in the machine learning sense. The assay itself is developed and optimized, not "trained" on data. The figures for sensitivity and specificity reflect the performance after development and optimization, assessed against a "test set" (though its size is not specified).


9. How the Ground Truth for the Training Set Was Established

Since there is no "training set" in the context of this immunoassay, the question of how its ground truth was established is not applicable. The development of the assay likely involved internal validation and optimization against known positive and negative samples, but these are part of the assay's development, not a separate "training set" with ground truth in the way a machine learning algorithm would have.

{0}------------------------------------------------

NOV 2.5 1998

K982711

Meridian Diagnostics, Inc. Cincinnati, OH 45244

510(k) Notification Premier Giardia

A. 510(k) Summary Identification Information

Submitter's Information:

Submitter's Name and Address:

Allen D. Nickol, PhD Meridian Diagnostics, Inc. 3471 River Hills Drive Cincinnati, OH 45244

Phone Number: 1-800-543-1980

Contact Person: Allen D. Nickol, PhD Director of Clinical and Regulatory Affairs

Date Summary Prepared: August 3, 1998

Name of Device: Premier Giardia

Classification Name: Giardia spp., 83 MHI

Predicate Equivalent Device: Alexon ProspecT Giardia Microplate Assay

Description of Device: The Premier Giardia test utilizes polyclonal anti-Giardia capture antibody adsorbed to microwells.

Intended Use:

The Premier Giardia enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Giardia antigens in human stool. Test results are intended to aid in the diagnosis of Giardia infection (giardiasis).

Comparison with Predicate Device: The following comparison of the use, technology, function and performance supports the Statement of Equivalence between the Premier Giardia test and Alexon Prospect-T Giardia EIA. The differences in technology do not raise additional concerns regarding safety and Safety and effectiveness are demonstrated to be substantially effectiveness. equivalent.

{1}------------------------------------------------

Meridian Diagnostics, Inc. Cincinnati, OH 45244

510(k) Notification Premier Giardia

MethodPremier GiardiaAlexon ProspecT Giardia
Intended UseDetection of Giardia antigens in patient stoolDetection of Giardia Specific Antigen (GSA 65) in aqueous extracts of fecal specimens
ResultsQualitativeQualitative
Specimen RequiredPreserved (Formalin, SAF) and Unpreserved StoolsPreserved (Formalin, SAF, MF), Unpreserved Stool, and stools in C&S (or equivalent) transport media
TechnologySandwich Enzyme ImmunoassayPolyclonal capture, monoclonal detect, polyclonal conjugate, TMB substrateSandwich Enzyme ImmunoassayPolyclonal capture, monoclonal conjugate, TMB substrate
Level of Skill RequiredLaboratory TechnicianLaboratory Technician
Function1. Specimen diluted 1/4 and 200µl added to well containing rabbit anti-Giardia capture Ab.2. Incubate 1 hr at room temperature.3. Wash 5 times.4. Add 2 drops detection Ab and two drops enzyme conjugate per well.5. Incubate 30 minutes at room temperature.6. Wash 5 times.7. Add 4 drops substrate.8. Incubate 10 minutes at room temperature.9. Add two drops stop solution and read visually or spectrophotometrically1. Sample preparation varies with specimen type. Some are diluted 1/4, others are not diluted. Add 0.2ml to wells.2. Incubate 1 hr at room temperature.3. Wash 3 times.4. Add 4 drops Enzyme Conjugate.5. Incubate 30 minutes at room temperature6. Wash 5 times.7. Add 4 drops Substrate8. Incubate 10 minutes at room temperature.9. Add 1 drop Stop Solution and read visually or spectrophotometrically
InterpretationPos/Neg read visually or spectrophotometrically. Fixed cutoff 0.140 single wavelength (450nm) or 0.100 dual wavelength (450-630nm)Pos / Neg read visually or spectrophotometrically. Color chart for visual; single wavelength read pos if ≥0.05 absorbance units above negative control
Performance vs.Reference MethodsSensitivitySpecificity98%98%98%98%

..

Interfering Substances: Blood and Barium Sulfate do not interfere with results.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

NOV 2 5 1998

Allen D. Nickol. Ph.D. Director of Clinical and ·Regulatory Affairs Meridian Diagnostics, Inc. 3471 River Hills Drive Cincinnati, Ohio 45244

Re: K982711 Trade Name: Premier Giardia Regulatory Class: II Product Code: MHI Dated: October 21, 1998 Received: October 22, 1998

Dear Dr. Nickol:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Meridian Diagnostics, Inc. Cincinnati, OH 45244

E. Indications for Use Statement

510(k) Number (if known): K982711

Device Name: Premier Giardia

Indications For Use:

The Premier Giardia enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Giardia antigens in human stool. Test results are intended to aid in the diagnosis of Giardia infection (giardiasis).

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Dubois

Division of Clinical Laboratory Devices 510(k) Number

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

§ 866.3220

Entamoeba histolytica serological reagents.(a)
Identification. Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies toEntamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identifyEntamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasiteEntamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.