K Number
K981270

Validate with FDA (Live)

Date Cleared
1998-05-01

(24 days)

Product Code
Regulation Number
862.1170
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IRMA Blood Analysis System is intended for professional use in those settings where direct measurement of blood such as blood gases (pCO2 and pO2), pH, Na , K , iCa BUN. CI'. and Hct. in whole blood are performed such as the clinical laboratory or the patient bedside.

The pH, pCO2, pO2 measurements, and their associated calculated values are used to assess acid-base status and state of oxygenation. Common causes of acid-base disturbances include: cardiopulmonary disease, metabolic abnormalities, drugs and poisons, and fluid imbalance.

The electrolyte measurements (Na . K . Cl) are used to assess hydrational status, aid in the diagnosis of respiratory and metabolic acid-balance, and prevention of cardiac arrhythmia. Common disease states which utilize these measurements for diagnosis are acid-base disturbances, dehydration, diarrhea, ketoacidosis, alcoholism and other toxicities.

The measurement of ionized calcium is used to assess disease states such as thyroid abnormalities, renal failure or transplant, and to monitor dialysis patients.

The measurement of blood urea nitrogen is used to monitor renal disease, dialysis --patients, and hyperalimentation.

The measurement of hematocrit is used to assess anemia, blood loss such as in an accident or during surgical procedures, and polycythemia.

With the addition of BUN and Cr, the IRMA Blood Analysis System will measure: blood gases (pCO2 and pO2), pH, sodium, potassium, ionized calcium, blood urea nitrogen, chloride, and hematocrit. The sensor arravs of the cartridge will also be packaged in various combinations.

Device Description

The current IRMA Blood Analysis System comprises a system of an electronic instrument and disposable cartridges (single-use or multi-use) intended for the measurement of blood gases (pCO2 and pO2), pH, potassium, sodium, ionized calcium, and hematocrit in blood.

The new system will measure the above mentioned analytes plus blood urea nitrogen (BUN) and chloride (Cl) in blood on the single-use cartridge. Except for the addition of these two analytes and their associated changes, the system will remain the same as the current system.

The IRMA analyzer can use either battery or AC power. The system's operation utilizes a microprocessor which is controlled by internal electronics and diagnostics. The microprocessor controls the touch screen, analog electronics which collect the digital signals from the sensors and the controls the printer. The printer provides a hard copy of the measured and calculated values.

Samples are introduced via syringe or capillary injections with the IRMA Capillary Collection Device. The minimum sample volumes are 200uL from a syringe injection and 125uL from a Capillary Collection Device injection. Other capillary collection devices which require aspiration are not compatible with the system.

The cartridges utilize microelectrode technology for the measurement of the following blood analytes: pH, pCO2, pO2, sodium, potassium, ionized calcium, blood urea nitrogen, chloride, and hematocrit.

The principles of measurement are similar to traditional electrode methodologies for blood gas and electrolyte measurements. The pH, pCO2, Na*, K*, iCa**, BUN, and Cli utilize ion-selective potentiometric electrodes including a reference electrode. The pO2 electrode is an amperometric Clark electrode. The hematocrit sensor utilizes a conductivity electrode.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the IRMA Blood Analysis System's BUN and Chloride sensors, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

AnalyteAcceptance Criteria (Implicit)Reported Device Performance (Correlation to Reference Method)
Chloride (Cl-)Not explicitly stated as a numerical threshold, but implied to be "substantially equivalent" to predicate device.IRMA Cl = 0.9262 * Cl- conc (mM) - 22.43 Correlation Coefficient (r) = 0.9976 Sy.x = 4.0
Blood Urea Nitrogen (BUN)Not explicitly stated as a numerical threshold, but implied to be "substantially equivalent" to predicate device.IRMA BUN = 1.0831 * BUN conc (mg/dL) + 3.7837 Correlation Coefficient (r) = 0.9996 Sy.x = 1.5

Note: The acceptance criteria are "not explicitly stated" in numerical form within the provided text. They are implied by the statement "indicate substantial equivalence between Cl- and BUN sensors on the IRMA Blood Analysis System and the predicate devices" and the presentation of strong correlation coefficients.

2. Sample Size Used for the Test Set and Data Provenance

  • Chloride (Cl-) Test Set Sample Size: n = 99
  • BUN Test Set Sample Size: n = 97
  • Data Provenance:
    • Country of Origin: Not explicitly stated, but the submission is to the US FDA and the company is based in Roseville, MN, USA.
    • Retrospective or Prospective: Not explicitly stated, but the description "split sample studies were conducted using whole blood samples measured by both the IRMA system and a reference method" and "heparinized whole blood samples were prepared by spiking with varying concentrations of electrolytes to allow testing throughout the reportable range" suggests a prospective study where samples were prepared and then tested on both systems.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: The text states, "Testing was conducted by three laboratory personnel."
  • Qualifications of Experts: Not explicitly stated beyond "laboratory personnel." No specific professional certifications (e.g., medical technologist) or years of experience are mentioned. The document notes that the IRMA system was previously cleared for use by non-laboratory personnel, and this submission's data does not include such personnel.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not explicitly described. The study appears to rely on direct comparison of the IRMA system's results against a reference method, rather than an adjudication process between multiple readers/interpretations. "The method of least squares was used to determine the best fit line."

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

  • MRMC Study Done?: No, an MRMC comparative effectiveness study was not done. This study focuses on the analytical accuracy and precision of a diagnostic device (blood analyzer) compared to reference methods, not on human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Standalone Performance Done?: Yes, the accuracy and precision data presented for Chloride and BUN represent the standalone performance of the IRMA Blood Analysis System's sensors, measuring the analytes in blood and aqueous samples. It's an instrument-based measurement, not an AI or human interpretation task.

7. The Type of Ground Truth Used

  • Type of Ground Truth: Reference methods, specifically a Vitros Chemistry System DT6011 (for BUN) and a Buchler Digital Chloridometer (for Chloride) for accuracy studies. For precision, aqueous standards and control materials with known concentrations were used.

8. The Sample Size for the Training Set

  • Training Set Sample Size: Not applicable. The IRMA Blood Analysis System is a medical device based on established electrochemical and enzymatic methodologies, not a machine learning or AI model that requires a "training set" in the traditional sense. Its "training" is in its engineering design, calibration, and manufacturing process.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth for Training Set Was Established: Not applicable, as this is not a machine learning device. The "ground truth" for its operation is inherent in the design and calibration using NIST traceable gases and salt standards, as described in the "Calibration" section.

{0}------------------------------------------------

K981270

MAY 1 1998

510(k) Summary

1. SubmitterDiametrics Medical, Inc.
2658 Patton Road
Roseville, MN 55113
2. Establishment Registration Number:2183953
3. Device Names
Proprietary Name:IRMA Blood Analysis System

Common Name:

Blood Gas/Electrolyte/Hematocrit Analyzer

Classification Name:

Blood gases (pCO2, pO2), pH, sodium, potassium, ionized calcium, blood urea nitrogen, chloride, and hematocrit test systems

    1. Device Classification
      The blood gases (pCO2 and pO2), pH, sodium, potassium, ionized calcium, chloride, blood urea nitrogen and hematocrit systems have been classified under Clinical Chemistry Test Systems 21 CFR 862.1120, 862.1665, 862.1600, 862.1145, 862.1170, 862.1770 and under Hematology and Pathology Devices, 864.6400. All are Class II devices.
    1. Compliance to Classification Requirements No performance standards have been established for the above referenced systems. Diametrics Medical intends to comply with any standards applicable to this product developed in the future.
    1. Proposed labels, labeling, and advertisements The proposed labeling appears in Appendix B (Pouch and cartridge label), Appendix C (Package insert) and Appendix D (User manual). There are no required changes to the current device label. Diametrics Medical, Inc. does not intend to advertise the product at this time.

{1}------------------------------------------------

7 . System Description and Statement of Substantial Equivalence

System Description

The current IRMA Blood Analysis System comprises a system of an electronic instrument and disposable cartridges (single-use or multi-use) intended for the measurement of blood gases (pCO2 and pO2), pH, potassium, sodium, ionized calcium, and hematocrit in blood.

The new system will measure the above mentioned analytes plus blood urea nitrogen (BUN) and chloride (Cl) in blood on the single-use cartridge. Except for the addition of these two analytes and their associated changes, the system will remain the same as the current system. A description of the new system follows.

The IRMA analyzer can use either battery or AC power. The system's operation utilizes a microprocessor which is controlled by internal electronics and diagnostics. The microprocessor controls the touch screen, analog electronics which collect the digital signals from the sensors and the controls the printer. The printer provides a hard copy of the measured and calculated values.

Samples are introduced via syringe or capillary injections with the IRMA Capillary Collection Device. The minimum sample volumes are 200uL from a syringe injection and 125uL from a Capillary Collection Device injection. Other capillary collection devices which require aspiration are not compatible with the system.

The cartridges utilize microelectrode technology for the measurement of the following blood analytes: pH, pCO2, pO2, sodium, potassium, ionized calcium, blood urea nitrogen, chloride, and hematocrit.

The principles of measurement are similar to traditional electrode methodologies for blood gas and electrolyte measurements. The pH, pCO2, Na*, K*, iCa**, BUN, and Cli utilize ion-selective potentiometric electrodes including a reference electrode. The pO2 electrode is an amperometric Clark electrode. The hematocrit sensor utilizes a conductivity electrode. Further description of the principles of operation for the electrodes can be found in the draft labeling of the user manual.

Calibration

The IRMA sensors are calibrated prior to each test using a calibrant prepackaged over the sensors. The calibrant is manufactured and tested with NIST traceable gases and salt standards. Calibration of the cartridge is completed when information determined at the factory for each lot of cartridges is combined with measurements taken during the calibration process. Factory derived calibration parameters are input into the analyzer by calibration code entry.

Throughout the calibration and analysis process, signals from the sensors are analyzed. If any abnormal conditions are detected, an error message is generated and the test will be terminated. If there are no abnormal conditions, then the sample results (measured and calculated) are displayed after successful calibration and analysis. In addition, the user has the option to print a hard copy of the results.

The user may also output the stored data via a RS232 serial port to IDMS ( IRMA Data Management System) or to a LIS.

{2}------------------------------------------------

Cartridge operation

Upon insertion of the cartridge, the user will enter a calibration code and the software's checksums will determine whether the code is appropriate for that cartridge type. Screen prompts will guide the user through the calibration process and sample analysis

For a more detailed description of the sample sequence refer to Section 2 of the User Manual Draft Labeling which is located in Appendix D.

Intended Use Comparison

The IRMA Blood Analysis System is intended for professional use in those settings where direct measurement of blood such as blood gases (pCO2 and pO2), pH, Na* , K* , iCa* , Cl; BUN, and Hct in whole blood are performed such as the clinical laboratory or the patient bedside.

Substantial Equivalence

This notification demonstrates that the IRMA Blood Analysis System's BUN and Cl sensors are substantially equivalent to several other commercially cleared products: Johnson and Johnson Vitros Chemistry System DT6011 (BUN) and the Buchler Digital Chloridometer (CI).

The IRMA and Vitros analysis systems are similar in that both systems are intended for the measurement of blood urea nitrogen (BUN) by enzymatic methods which utilize an urease reaction. The IRMA and Buchler Digital Chloridometer systems are similar in that both systems are intended for the measurement of chloride by electrochemical methods.

The predicate systems and proposed IRMA Blood Analysis System are compared in Tables #1-2.

{3}------------------------------------------------

IRMAVitros
Detection Methodenzymatic electrochemicalurease reactionenzymatic colorimetricurease reaction
Analytes measuredpH, pCO2, pO2, Na+, K+, iCa++,BUN, CI-, Hct*BUN and Cl- not available onmulti-useBUN, ammonia, amylase,cholesterol, creatinine,glucose, HDL cholesterol,hemoglobin, lactate,magnesium, neonatal bilirubin,phosphorus, total bilirubin,total protein, triglycerides,uric acid
Measuring RangeBUN 0 - 150 mg/dLBUN 1 - 100 mg/dL
Operating Temp.15-30°C (59-86°F)15.5-29.4℃ (60-85°F)
Operating Humidity0-80%15-75%**Upper RH value dependent oftemperature.
SampleWhole blood0.2 - 3.0 mL, syringe0.125 mL from capillarycollection deviceSerum/plasma10 µL per test
Power7.2 V NiCAD rechargeablebatteryorAC adapter120 VAC 1 amp240 VAC 0.5 amp
ReagentsSupplied in self-containeddisposable cartridgeSupplied in self-containeddisposable slide
Weight5 lbs.19 lbs.
ResultsDisplay and printer on boardDisplay and printer on board
CalibrationAutomatic with each sampleUser initiated process uponinstallation, changing of slidelots, when QC is out, or asneeded for system verification.
Sensorsdisposable single-useormulti-usedisposable single-use

Table #1 Comparison of IRMA to Vitros System Features

ﺰ ﺳﻴﺒﻴﺎ

:

:

{4}------------------------------------------------

IRMABuchler Chloridometer
Detection Methodelectrochemicalelectrochemical
Analytes measuredpH, pCO2, pO2, Na+, K+, iCa++,BUN, Cl-, Hct*BUN and Cl- not available onmulti-useCl-
Measuring RangeCl- 30-150 mM or mEq/LCl- 1.0-999.9 mEq/L
Operating Temp.15-30°C (59-86°F)Room temperature (range notspecifically defined)
Operating Humidity0-80%Not defined
Blood SampleWhole blood0.2 - 3.0 mL, syringe0.125 mL from capillarycollection deviceserum/plasma10 μL on low range100 μL on high range
Power7.2 V NICAD rechargeablebatteryorAC adapter115 VAC, 60 Hz230 VAC, 50 Hz
ReagentsSupplied in self-containeddisposable cartridgeChloridometer acid reagentwhich is manually measured
Weight5 lbs.10 lbs.
ResultsDisplay and printer on boardDisplay on board
CalibrationAutomatic with each sampleManual blank adjustment ofreagents and check withchloride standard
Sensorsdisposable single-useormulti-usereusable silver wire

1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1

Table, #2, Comparison of IRMA to Buchler Chloridometer System Features

:

{5}------------------------------------------------

8 . Performance Characteristics: Supporting data for Substantial Equivalence

This section provides blood and aqueous data generated in-house using the IRMA Blood Analysis System's chloride and BUN sensors. The samples used in the study were spiked heparinized blood and prepared aqueous standards.

  • A. Accuracy Data
    To determine chloride and BUN accuracy of the IRMA Blood Analysis, split sample studies were conducted using whole blood samples measured by both the IRMA system and a reference method (refer to Table #3). The method of least squares was used to determine the best fit line. Testing was conducted by three laboratory personnel. The IRMA system was cleared in a previous 510k submission for use by non-laboratory personnel therefore this submission does not include data by non-laboratory personnel such as nurses. The heparinized whole blood samples were prepared by spiking with varying concentrations of electrolytes to allow testing throughout the reportable range.

The data shown below in Table #4 and the x-y plots located in Appendix A indicate substantial equivalence between CI- and BUN sensors on the IRMA Blood Analysis System and the predicate devices.

Table #3 Reference methods

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.Company of the company of the clife of the closed to the closed to the closes of the children the collection of the children the collection of the children the collection of------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Comments of the program and the program and the many of the many of the many of the many of the.-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Comments of the Research and Concession Comes of the Comments of Concession of Concession Come of Concession Come of Concession Come of Concession Come of Concession Come ofA case a manage controllation and manufacturer and consisted--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Company.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A LEW SHOW HOME AND CONSULT A LE REALSHIP SHEATERS CHARACT CHARACT LAND CONSULTER COLLECTION COLLECTION COLLECTION COLLECTION COLLEGION COLLECTION COLLEGION COLLEGION COLLECTION COLLECTION CONNOMER COLLECTION CO
Complete of Children Company of the commend and any and well and with a program and any and any and the comments of the comments of the comments of the comments of the commen------------------.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------10.00 000 000 000------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------LE R. R. BE LEW LEASE .------------------------------------------------------------------------------------------------------------------------------------------------------------------------------1

Table #4 Blood Accuracy: The method of least squares was used to determine the best fit line.

.of the control of the closed of the closed on the consideration of the consistedсоданного всего в составления с в ставетского с своемника с седание с продобращие содательного в собставляется в собставляется в собство на свъем в об объем в об объем в об о-------------------------------------------------------------------------------100 CONSULT LINERAL LINE SECTION CONSULTION ALL CONTRACT ALL CONTRACT ALL CONTRACT ALL CONTRACT ALL CONTRACT AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND A.Comprehensive and Children and Children and Children Children Children Children Children Children Children Children Children Children Children Children Children Children ChilNo. of Recommens and Commens and Canada and------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------AN AND A CR IN LE AND LEVEL FOR CONSULTION FOR LEARNER LE CONSULE-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------I a proposant I a picture and any and the managerIN LEARNELIES MITCompany of Children Comments of ChildrenStandard States andCommend of a c the Children assessment consisted as any and the different of the commend and the different of the commend and the different of the commend and the different o
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A LEA BE COLORIESWe was and the commend of the commend of the commend of the comments of the contribution of the contribution of the contribution of the contribution of the contribution of th------------------------------------------------------------------------------------------------------------------------------------------------------------------------------LELE FREE LE SELECTRANCIAL LULL FRIEN (SELECTRICAL LULL LIGHTS COLLECTION CONSULTER-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------1on the most of the state of the first and the first and the contributed to the country ofCompany of Concession in a consideration of the contribution of the comments of the comments of the comments of the comments of the comments of the comments of the comments o------------------------------------------------------------------------------------------------------------------------------------------------------------------------------CONSULT CONSULT CONSULTER CONSULTERS OF THE LES CLOUDS OF------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Annual described for a former of the commend to the commend to the commend to the commend to the commend to the commend to the commend to the commend to the commend to the coMarket on the cases of the called to the called to the children the---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------I I I I l t t l t t t t a see a separate and second and second and the second of the second of the second of the second of the second of the second the second the second the------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Comments of Concession of Concession of Concession of Children of Children of Children of Children of Children of Children of Children of Children of Children of Children ofChildren annual considered the re a more considered as a manufacture as a manufacturer and considered as a commendI SURNER IN ------------------------------------------------------------------------------------------------------------------------------------------------------------------Canada Call Annual Property ofCommens of the consisted in the consideration of the first---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Commenss of the Resident of the Children of Children of Children of Children of Children of Children of Children of Children of Children of Children of Children of Children oStatus of Canadian Company of1Complete of the first and consistence of consumers of the first the programments of the first- House of the program and the management on and the comments of the first for the first and the first and the first and the first andCommend on the country of can and control of cannot of can----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------BBC C C EXE EXPLOSE------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

r = correlation coefficient

{6}------------------------------------------------

B. Precision Data

r recision Data
To determine precision of the Cl' and BUN sensors of the IRMA system, samples at a given level ro determine problem of the samples used in the study consisted of three levels of aqueous were testion reportionly. The our problement of the control materials. Each day of testing, each solutions similar to commercially a familians from each of the three levels. Data from this study are shown in Tables #5-6.

DayOperatornLevelMeanSDCV
Day 115130.60.762.5
25130.20.270.9
35130.80.943.0
Combined15130.50.722.4
Day 215130.50.943.1
25130.20.632.1
34130.50.170.6
Combined14130.40.652.1
Day 115256.60.220.4
25257.10.280.5
35257.00.430.8
Combined15256.90.360.6
Day 215257.00.300.5
25257.10.150.3
35257.10.300.5
Combined15257.10.240.4
Day 115385.91.782.1
25387.10.610.7
34387.00.420.5
Combined14386.61.201.4
Day 214386.01.591.9
25386.21.451.7
35385.21.902.2
Combined14385.81.601.9

Table #5 Cl Aqueous Precision

Combined = Results from all users combined for a day of testing.

{7}------------------------------------------------

Table #6 BUN Aqueous Precision
--------------------------------------
DayOperatornLevelMeanSDCV
Day 115117.52.4013.7
25118.01.729.6
35117.41.367.8
Combined15117.61.7610.0
Day 215117.70.965.4
25116.31.438.8
35117.31.066.1
Combined15117.11.257.3
Day 115233.81.404.1
25234.82.988.6
35234.23.9311.5
Combined15234.32.778.1
Day 215233.32.196.6
25233.71.293.8
35234.52.848.2
Combined15233.82.096.2
Day 115364.14.567.1
24364.81.812.8
35365.44.817.4
Combined14364.73.825.9
Day 215368.71.462.1
25367.43.795.6
35367.62.553.8
Combined15367.92.633.9

Combined = Results from all users combined for a day of testing.

{8}------------------------------------------------

  • D. Display range
    Graphs #1-2 show the tested linear range of the product. This data along with the blood accuracy data which were shown earlier justify the display ranges shown in Table #8.
Table #8 IRMA Blood Analysis System Display Rangeand and the commend of the commend of the many and any and any and the many of the may
---------------------------------------------------------------------------------------------------------------------------------------------
THE RESEARCE A RESEARCE LA CONSULT LA CONSULTION COLLECTION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLECTION COL------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------The program and a superior of the program and any and the provinced and the commend of the comments of the contribution of the contribution of the contribution of the contrib------------------------------------------------------------------------------------------------A CONTRACTOR COLLECTION COLLEGION OF CHILIP OF CLEASE OF CHEARTH OF CHEARTH OF CHEARTH AND THE CHARGE OF CHEARTH AND THE CHARGE OF CHEARTH AND THE CHARGE OF CHEARTH AND THE CN.J. BREAK REVENUES & A Company of the Children Comments------------------------------------------------------------------------------------------------------------------------------------------------------------------------------I--------------------The first and the first of the control of the control of the control of the commend of the commend of the many of the many of the may be any and the many of the many of the m----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------and a first program court procession in of experience in an any andCollection of the later the station of the control of the promote of the program and interesting of any a----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------. I . I . LE LEASINGLE AREA A MILIE . IN MINISTER LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAS-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Comment of Children and Children and Children and Children and Children------------------------------------------------------------------------------------------------------------------------------------------------------------------------------of the may be and the world of the times and in the bigger will be the first and-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------been and and the complex would with a provinced and a film the film theLA & LEALERS COLLEGE COLLECTION COLLECTION COLLECTION COLLECTION COLLEGIONA------------------------------------------------------------------------------------------------------------------------------------------------------------------------------CONNECT & CAR COLLECTION CONSULTION COLLECTION COLLECTION CONTRACT COLLEGION CONTRACT COLLEGION CONTRACT COLLEGION CONTRACT COLLEGION CONTRACT COLLEGION CONTRACT COLLEGION CO.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------.------------------------------------------------------------------------------------------------------------------------------------------------------------------------------1---------------------------------------------------Acres of the are a comments100 100
---------1Annual Property of11 4 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------LENGINEER I BARRENER A LA BERTAIN AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Company of the contract and the many of promote of procession for program and the management of theCompany of the program and consideration andI I a loss possibles i concession i consistent consistent in consistent in controlled in the contribution in the contribution in the contribution in------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------1------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------CA READING . IL BREAK ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Graph #1

Image /page/8/Figure/5 description: The image is a graph titled "Mean(IRMA Cl- mM) By Mean(Cl- conc (mM))". The graph shows a linear relationship between the mean concentration of chloride (Cl-) and the mean IRMA Cl- mM. The x-axis represents the Mean(Cl- conc (mM)) and ranges from 0 to 200, while the y-axis represents the Mean(IRMA Cl- mM) and ranges from 0 to 200. A linear fit line is plotted on the graph, showing a positive correlation between the two variables.

IRMA CI- Linear Range

IRMA Cl = 0.9262 Cl- conc (mM) -22.43 n = 99 samples - with mean of each level shown r = 0.9976 Sy.x =4.0 Range evaluated = 10.6 - 146.8 mM

{9}------------------------------------------------

Graph #2

Image /page/9/Figure/1 description: The image is a graph titled "Mean IRMA BUN (mg/dL) By Mean BUN conc (mg/dL)". The graph shows a linear relationship between Mean IRMA BUN and Mean BUN concentration. The x-axis represents Mean BUN concentration in mg/dL, ranging from 0 to 150, while the y-axis represents Mean IRMA BUN in mg/dL, also ranging from 0 to 150. A line of best fit, labeled as "Linear Fit", is plotted on the graph.

I RMA BUN Linear Range

IRMA BUN= 1.0831 BUN conc (mg/ dL) + 3.7837 N = 97 samples - with mean of each level shown r = 0.9996 Sv.x = 1.5 Range evalutated = 7.3 - 124.5 mg/ dL

11. Safety and Effectiveness

The addition of the CI- and BUN sensors to the IRMA Blood Analysis System has not been found to interfere or cause any adverse effects with the operation of the IRMA instrument or other nearby equipment. There are no additional risks to the patient or user when cartridges containing these sensors are utilized with the IRMA system.

The IRMA Blood Analysis System has already been certified to the required electrical and EMI safety specifications of medical equipment such as UL 2601-1 and UL 2601-2 (or equivalent).

{10}------------------------------------------------

11. Software Information

The IRMA Blood Analysis System's software has been developed, tested and will be validated in accordance with Diametrics' software development procedures and where applicable, the FDA's Reviewer's Guide to Computer Controlled Medical Devices (August, 1991). Based on the criteria of this document, our product is perceived to be of a low level of concern. Below is a description of the software development process.

Software requirements are developed based on approved product design requirements. All design specifications are reviewed for consistency and accuracy with respect to one another and a software hazard analysis is performed. The software hazard analysis for this product is located in Table #9. The software code is developed from the approved software requirements. Software revisions are maintained under a software version control system.

Software test plans and procedures are developed based on the software requirements specifications. Related software requirements are grouped to form the basis of test procedures. Each functional requirement (or group of requirements) is translated into a functional test and is included in a test procedure. These procedures are designed to ensure that functional requirements are met, software related system specifications fulfilled and software safety features adequately tested.

Code inspections and walk-throughs are performed by software engineers not involved in the code development in order to detect design flaws and validate software logic where functional testing is insufficient.

Written test procedures are executed and the results are reviewed to ensure that acceptance criteria have been met.

Software testing is also conducted in a less structured environment (free-form testing) by technicians and users under a variety of conditions in an attempt to "break" the software,

Considerable attention is applied to system safety throughout the verification. validation and testing of the software. A system hazard analysis which is done in the early development phase of the product is reviewed and updated as required. Safety requirements are identified in the hazard analysis and incorporated into the design specifications and tests via functional procedures. Potential error and exception conditions are also identified and tested. Stress conditions are also tested during the verification and validation phase.

All events identified as potential hazards for the IRMA Blood Analysis System have been addressed through hardware, software, or product labeling.

Changes to software requirements are documented, reviewed and approved. Each revision of the software is reviewed and evaluated for complete functionality and impact to the hazard analysis. New revisions are subjected to the same tests as the previous version. Revisions to areas that do not affect calculations may be regression tested. Each version of the software released to manufacturing will be archived with a list of the changes and validation data.

{11}------------------------------------------------

Table #9 Software Hazard Analysis

ﺍ ﺗﻮﺻﻌﻬﺎ ﻭﺍﻟﺘﻲ ﺗﻪ ﺍﻟ

Potential Hazard:Inaccurate AlgorithmsAction to MinimizeHazardAdd'IInformationLikelihoodSeverity
CalibrationTest & ValidationplansDevelopment PhaseQC checkImprobableMarginal
Sample CalculationsTest & ValidationplansDevelopment PhaseQC checkImprobableMarginal
System operationTest & ValidationplansDevelopment PhaseQC checkImprobableMarginal
Incorrect IR readingTest & ValidationplansDevelopment PhaseQC checkImprobableMarginal
Temperature &Barometric checksTest & ValidationplansDevelopment PhaseQC checkImprobableMarginal

Table #10 Categories

Table #10 Categories
LikelihoodSeverity
FrequentNegligible
ProbableMarginal
OccasionalCritical
RemoteCatastrophic
Improbable
Incredible

{12}------------------------------------------------

Software Certification

I certify that, in my capacity as Director of Quality Assurance and Regulatory Affairs of Diametrics Medical, Inc., prior to market release of the product, the software development process will be completed. The final documentation will include official signoff of the software requirements specifications, test plans and test completion criteria, and their results which demonstrate the system specifications and functional requirements were met along with release of the software revision.

We have shown substantial equivalence between the IRMA Blood Analysis System utilizing the multi-use cartridge and the predicate devices even though the software development process has not yet been completed.

Name of Company Official Correspondent: Steve Boeh

Signature:

e: Steve Burch

Title: Director of QA/RA

Company Name: Diametrics Medical Inc.

Date of Signature: Month/Day/Year

{13}------------------------------------------------

13. Attachments

The proposed labeling appears in Appendix B (Pouch and Cartridge Labels), Appendix C (Cartridge Package Insert) and Appendix D (User Manual). There are no required changes to the IRMA Blood Analysis System's device label. The draft user manual which is shown contains the information which is pertinent to the addition of the new CI and BUN sensors. Any screen changes associated with these new sensors will also be made to the pictures in the user manual. This information will either be incorporated into the current IRMA SL Series 2000 User Manual or it will become an addendum to it. Diametrics Medical, Inc. does not intend to advertise the product at this time.

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is an abstract image of an eagle with three stylized human profiles incorporated into its design.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

1 1998 MAY

Steve Boeh . Director, Quality Assurance and Regulatory Affairs Diametrics Medical, Inc. 2658 Patton Road Saint Paul, Minneapolis 55113 Minnesota K981270 Re :

IRMA® Blood Analysis System Regulatory Class: II Product Code: CGZ, CDS, GKG, JFP, JGS, CHL, CEM Dated: April 2, 1998 Received: April 7, 1998

Dear Mr. Boeh:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਕੇ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Register. this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{15}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{16}------------------------------------------------

Indications for Use and Intended Use Statement

tr 981271 510(k) Number: IRMA Blood Analysis System with additional menu: blood urea nitrogen and Device Name: chloride

Indications for Use

Statement of Intended Use

The IRMA Blood Analysis System is intended for professional use in those settings where direct measurement of blood such as blood gases (pCO2 and pO2), pH, Na , K , iCa BUN. CI'. and Hct. in whole blood are performed such as the clinical laboratory or the patient bedside.

The pH, pCO2, pO2 measurements, and their associated calculated values are used to assess acid-base status and state of oxygenation. Common causes of acid-base disturbances include: cardiopulmonary disease, metabolic abnormalities, drugs and poisons, and fluid imbalance.

The electrolyte measurements (Na . K . Cl) are used to assess hydrational status, aid in the diagnosis of respiratory and metabolic acid-balance, and prevention of cardiac arrhythmia. Common disease states which utilize these measurements for diagnosis are acid-base disturbances, dehydration, diarrhea, ketoacidosis, alcoholism and other toxicities.

The measurement of ionized calcium is used to assess disease states such as thyroid abnormalities, renal failure or transplant, and to monitor dialysis patients.

The measurement of blood urea nitrogen is used to monitor renal disease, dialysis --patients, and hyperalimentation.

The measurement of hematocrit is used to assess anemia, blood loss such as in an accident or during surgical procedures, and polycythemia.

With the addition of BUN and Cr, the IRMA Blood Analysis System will measure: blood gases (pCO2 and pO2), pH, sodium, potassium, ionized calcium, blood urea nitrogen, chloride, and hematocrit. The sensor arravs of the cartridge will also be packaged in various combinations.

(Please DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use(Per 21 CFR 801.109) ✓OR Over-The-Counter Use(Optional Format 1-2-96)

Division Sign-Off)

Division of Clinical Laboratory Devices

510(k) Number 9812

§ 862.1170 Chloride test system.

(a)
Identification. A chloride test system is a device intended to measure the level of chloride in plasma, serum, sweat, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.(b)
Classification. Class II.