K Number
K981029

Validate with FDA (Live)

Date Cleared
1998-07-20

(123 days)

Product Code
Regulation Number
866.5660
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This assay is designed for the in-vitro masourement of specific IgG autoantibodies against proteinase 3 (PR3) present in human serum, as an aid in the diagnosis of Wegener's granulomatosis and certain forms of autoimmune vasculitis in conjunction with other clinical findings.

Device Description

Not Found

AI/ML Overview

I am sorry, but the provided text does not contain the information requested to describe the acceptance criteria and the study that proves the device meets those criteria. The document is a 510(k) clearance letter from the FDA for a device called "Bindazyme Anti-PR3 Enzyme Immunoassay Kit," which confirms that the device is substantially equivalent to legally marketed predicate devices.

The letter mentions the "Indications for Use Statement" where the device's purpose is described, but it does not detail specific acceptance criteria (like sensitivity, specificity, accuracy thresholds), nor does it provide information about a study proving these criteria were met. It primarily focuses on regulatory approval and compliance.

Therefore, I cannot provide:

  1. A table of acceptance criteria and reported device performance.
  2. Sample size used for the test set or data provenance.
  3. Number of experts or their qualifications for ground truth.
  4. Adjudication method.
  5. Information on a multi-reader multi-case (MRMC) comparative effectiveness study.
  6. Information on a standalone algorithm performance study.
  7. The type of ground truth used.
  8. Sample size for the training set.
  9. How ground truth for the training set was established.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized emblem. The emblem depicts a stylized eagle with three human profiles incorporated into its wing design.

JUL 201998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

The Binding Site, LTD. c/o Jay H. Geller East Tower, Suite 600 2425 West Olymic Boulevard Santa Monica, CA 90404

K981029 Re :

Trade Name: Bindazyme Anti-PR3 Enzyme Immunoassay Kit Requlatory Class: II Product Code: MOB Dated: June 11, 1998 Received: June 15, 1998

Dear Mr. Geller:

We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labelinq, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) , it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{1}------------------------------------------------

Paqe 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and * advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html"

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

INDICATIONS FOR USE STATEMENT

Anti-PR3 Enzyme Immunoassay Kit Device Name:

This assay is designed for the in-vitro Indications for Use: masourement of specific IgG autoantibodies against proteinase 3
(PR3) present in human serum, as an aid in the diagnosis of Wegener's granulomatosis and certain forms of autoimmune vasculitis in conjunction with other clinical findings.

Ceta E. Madim

(Division Si 510

Image /page/2/Picture/6 description: The image contains a signature in black ink on a white background. The signature appears to be cursive and somewhat stylized. The signature is oriented diagonally across the image.

§ 866.5660 Multiple autoantibodies immunological test system.

(a)
Identification. A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).(b)
Classification. Class II (performance standards).