Intended Use

The catheter is intended to be used for the controlled selective infusion of physician specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuro anatomy.

Device Description

The Flow Rider Flow Directed Micro Catheter is a single lumen infusion catheter designed to be flow directed into the distal vasculature. The Flow Directed Micro Catheter has a hydrophilic coating to provide lubricity for navigation of vessels. A standard luer lock adapter on the proximal hub is used for the attachment of accessories. The catheter tip is indicated by a radiopaque marker to facilitate fluoroscopic visualization.

AI/ML Overview

The provided text describes the 510(k) Summary of Safety and Effectiveness for the Flow Rider™ Flow Directed Micro Catheter. This is a medical device submission, and the "acceptance criteria" and "study" are interpreted in the context of demonstrating substantial equivalence to predicate devices, rather than a clinical trial proving specific performance metrics against a predefined threshold. Therefore, the information requested will be framed within the context of regulatory approval for a medical device.

Here's the breakdown of the information requested:

1. A table of acceptance criteria and the reported device performance

For a 510(k) submission, "acceptance criteria" are not typically presented as quantitative thresholds for clinical performance but rather as successful completion of various tests demonstrating the device's safety, effectiveness, and substantial equivalence to predicate devices. The "reported device performance" refers to the successful outcomes of these tests.

Acceptance Criteria CategoryReported Device Performance (Summary of Study Results)
BiocompatibilityPassed Biocompatibility testing in accordance with ISO 10993-1:1992(E). Suitable for its application.
Physical TestingAll testing yielded acceptable results for:
- Dimensional inspection
- Tensile strength tests
- Burst pressure tests
- Flow rate tests
- Torque tests
- Performance under simulated conditions
Animal StudiesDemonstrated that the Flow Rider Catheter is substantially equivalent to the predicate devices (Target Spinnaker™ and Balt Magic™).
Substantial EquivalenceFound to be substantially equivalent to predicate devices (Target Spinnaker™ Flow Directed Infusion Catheter, K965189; Balt Magic™ Infusion Catheter, K923368) in intended use and principles of operation.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: The document does not specify exact sample sizes for each physical test (e.g., how many catheters were burst tested or torqued). For the animal studies, the number of animals used or the number of procedures performed is not explicitly stated.
  • Data Provenance: The document does not specify the country of origin for the testing data. Given the submitting company (Micro Therapeutics, Inc.) is based in San Clemente, CA, USA, it's highly probable the testing was conducted in the USA. The studies mentioned (biocompatibility, physical, animal) are prospective in nature, as they are performed specifically to evaluate the new device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This type of information is not applicable to a 510(k) submission for a catheter based on the provided text. The "ground truth" for this device is established through objective engineering tests (physical properties), standardized biocompatibility assays, and in vivo animal studies demonstrating functional equivalence to existing devices. There is no mention of human expert evaluation or consensus for establishing "ground truth" for the performance characteristics reported.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable. This device's evaluation relies on objective measurements and observations in laboratory and animal settings, rather than human review with an adjudication process.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The Flow Rider™ Flow Directed Micro Catheter is a physical medical device, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study or evaluation of AI assistance is irrelevant to this submission.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" in this context is established through:

  • Engineering specifications and standards: For dimensional inspection, tensile strength, burst pressure, flow rate, and torque.
  • Standardized biological assays: For biocompatibility testing (ISO 10993-1:1992(E)).
  • Physiological and functional observations during animal studies: To demonstrate the catheter's ability to be flow-directed, infuse agents, and navigate vasculature, in comparison to predicate devices.

8. The sample size for the training set

Not applicable. This device does not involve machine learning or AI, so there is no concept of a "training set."

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K980104

PR 10 1998

Image /page/0/Picture/2 description: The image shows the logo for Micro Therapeutics, Inc. The logo consists of the letters "MTI" in a bold, sans-serif font. To the left of the letters, there is a series of small squares that appear to be arranged in a grid-like pattern. Below the letters, the words "MICRO THERAPEUTICS, INC." are written in a smaller, serif font.

510(k) Summary of Safety and Effectiveness Flow Rider™ Flow Directed Micro Catheter

Prepared January 8, 1998

Trade Name:Flow RiderTM Flow Directed Micro Catheter
Generic Name:Percutaneous Intravascular CatheterClassification:Class II
Submitted By:Micro Therapeutics, Inc.1062-F Calle NegocioSan Clemente, CA 92673Contact:Tom DaughtersRegulatory Affairs(714) 361-0610

Predicate Devices

Target Spinnaker™ Flow Directed Infusion Catheter, K965189 Balt Magic™ Infusion Catheter, K923368

Device Description

The Flow Rider Flow Directed Micro Catheter is a single lumen infusion catheter designed to be flow directed into the distal vasculature. The Flow Directed Micro Catheter has a hydrophilic coating to provide lubricity for navigation of vessels. A standard luer lock adapter on the proximal hub is used for the attachment of accessories. The catheter tip is indicated by a radiopaque marker to facilitate fluoroscopic visualization.

Intended Use

The catheter is intended to be used for the controlled selective infusion of physician specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuro anatomy.

Testing

Biocompatibility testing was performed on the Flow Directed Micro Catheter in accordance with the International Standard for the Biological Evaluation of Medical Devices. Part 1: Guidance on Selection of Tests (ISO 10993-1:1992(E)). Results of the tests showed that the device passed Biocompatibility testing and is suitable for its application.

Physical testing of the product included dimensional inspection, tensile strength tests, burst pressure tests, flow rate tests, torque tests and performance under simulated conditions. All testing of the product yielded acceptable results.

Animal studies were performed to assess the performance of the Flow Rider Catheter. The studies demonstrated that the Flow Rider Catheter is substantially equivalent to the predicate devices.

Summary of Substantial Equivalence

The Flow Rider Flow Directed Micro Catheter is substantially equivalent to the predicate devices in intended use and principles of operation.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 0 1998

Mr. Tom Daughters Requlatory Affairs Micro Therapeutics, Inc. 1062-F Calle Negocio San Clemente, CA 92673

Re: K980104 Flow Rider™ Flow Directed Micro Catheter Trade Name: Requlatory Class: II Product Code: KRA Dated: January 9, 1998 January 13, 1998 Received:

Dear Mr. Daughters:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device _ can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any

{2}------------------------------------------------

Page 2 - Mr. Tom Daughters

obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):

Device Name:

Flow Rider™ Flow Directed Micro Catheter

Indications for Use:

The catheter is intended to be used for the controlled selective infusion of physician specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuro anatomy.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over the Counter Use _________________________________________________________________________________________________________________________________________________________ OR

(Per 21 CFR 801.109)

(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices

510(k) NumberK980604
------------------------

Exhibit 6, Page 4

§ 870.1210 Continuous flush catheter.

(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).