ALCON MULTI-PURPOSE DISINFECTING SOLUTION ID 90746
K974624 · Alcon Laboratories · LPN · Mar 4, 1998 · Ophthalmic
Device Facts
| Record ID | K974624 |
| Device Name | ALCON MULTI-PURPOSE DISINFECTING SOLUTION ID 90746 |
| Applicant | Alcon Laboratories |
| Product Code | LPN · Ophthalmic |
| Decision Date | Mar 4, 1998 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 886.5928 |
| Device Class | Class 2 |
Intended Use
Alcon Multi-Purpose Disinfecting Solution ID 90746 is indicated for the daily cleaning, rinsing, disinfecting, and storing of soft (hydrophilic) contact lenses as recommended by your eye care practitioner. For use in chemical (not heat) disinfection. Alcon Multi-Purpose Disinfecting Solution ID 90746 can also be used as a diluent for OPTI-FREE® SUPRACLENS® Daily Protein Remover.
Device Story
Alcon Multi-Purpose Disinfecting Solution ID 90746 is a sterile, buffered, isotonic, aqueous chemical solution for soft contact lens maintenance. It performs daily cleaning, rinsing, disinfection, and storage. It also serves as a diluent for OPTI-FREE SUPRACLENS Daily Protein Remover. Used by patients at home; prescribed or recommended by eye care practitioners. The solution utilizes chemical disinfection (not heat). It benefits patients by maintaining lens hygiene and comfort through simultaneous enzymatic cleaning and disinfection. Clinical evaluation confirmed safety and efficacy for soft contact lens wearers.
Clinical Evidence
Clinical study evaluated the OPTI-FREE SUPRACLENS/Alcon Multi-Purpose Disinfecting Solution ID 90746 regimen. Study duration: 3 months. Sample size: 65 patients (130 eyes). Endpoints: safety and efficacy for daily simultaneous enzymatic cleaning and disinfection of soft (hydrophilic) contact lenses. Results demonstrated the regimen is safe and effective.
Technological Characteristics
Sterile, buffered, isotonic, aqueous solution. Ingredients: sodium citrate, sodium chloride, boric acid, sorbitol, AMP-95, tetronic 1304. Preservatives: edetate disodium 0.05%, Polyquaternium-1 0.001%, AL-6289 0.0005%. Chemical disinfection principle. Compatible with soft (hydrophilic) contact lens Groups I, II, III, and IV.
Indications for Use
Indicated for daily cleaning, rinsing, disinfecting, and storage of soft (hydrophilic) contact lenses (Groups I-IV). For use in chemical disinfection only. Compatible as a diluent for OPTI-FREE SUPRACLENS Daily Protein Remover.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
- OPTI-FREE EXPRESS Multi-Purpose Solution (PMA 830034)
- OPTI-ONE Multi-Purpose Solution (PMA 830034)
- OPTI-FREE Rinsing, Disinfecting and Storage Solution (PMA 830034)
Related Devices
- K973332 — ALCON MULTI-PURPOSE DISINFECTING SOLUTION ID 90746 · Alcon Laboratories · Dec 2, 1997
- K983780 — MULTI-PURPOSE DISINFECTING SOLUTION ID 90746 · Alcon Laboratories · Dec 18, 1998
- K983973 — OPTI-ONE MULTI-PURPOSE SOLUTION · Alcon Laboratories · Dec 18, 1998
- K984575 — MULTI-PURPOSE DISINFECTING SOLUTION ID 90746 · Alcon Laboratories · Jan 14, 1999
- K031193 — SOFT (HYDROPHILIC) CONTACT LENS CARE SOLUTION · Alcon, Inc. · Jul 3, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
December 8, 1997
K974624
# 510(K) SUMMARY
Submitted by:
MAR - 4 1998
Ralph H. Larsen Manager, Regulatory Affairs Alcon Laboratories, Inc. 6201 South Freeway Fort Worth, Texas 76134-2099 (817) 551-4702 (Phone) (817) 551-4630 (Fax)
Device Name:
Common Name: Contact Lens Care Multi-Purpose Solution
Proprietary Name: Alcon Multi-Purpose Disinfecting Solution ID 90746
### Indications for Use:
Alcon Multi-Purpose Disinfecting Solution ID 90746 is indicated for the daily cleaning, rinsing, disinfecting, and storing of soft (hydrophilic) contact lenses as recommended by your eye care practitioner. For use in chemical (not heat) disinfection.
Alcon Multi-Purpose Disinfecting Solution ID 90746 can also be used as a diluent for OPTI-FREE® SUPRACLENS® Daily Protein Remover.
### Description:
Alcon Multi-Purpose Disinfecting Solution ID 90746 is a sterile, buffered, isotonic, aqueous solution containing sodium citrate, sodium chloride, boric acid, sorbitol, AMP-95, tetronic 1304, with edetate disodium 0.05%, Polyquaternium-1) 0.001% and AL-6289 0.0005% as preservatives.
### Substantial Equivalence:
Alcon Multi-Purpose Disinfecting Solution ID 90746 is substantially equivalent, in terms of its actions and indications for use, to OPTI-FREE® EXPRESS Multi-Purpose Solution, OPTI-ONE® Multi-Purpose Solution and OPTI-FREE® Rinsing, Disinfecting and Storage Solution, all approved under PMA 830034. Alcon Multi-Purpose Disinfecting Solution ID 90746 meets the
{1}------------------------------------------------
guidelines set forth in FDA's May 1, 1997 Guidance for Industry; Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.
### Safety and Effectiveness:
#### Non-Clinical Data A.
## Microbiological Studies
During simultaneous enzymatic cleaning and disinfection, the product was evaluated for microbiological safety and effectiveness using the FDA Guidelines for contact lens solutions. The formulation meets the Stand-Alone criteria for disinfection of contact lens against bacteria, yeast and mold.
## Preclinical
Preclinical toxicology tests have been conducted to substantiate the safety of the product combination for use with all soft (hydrophilic) contact lens (Group I, II, III, and IV). The studies include: (1) cytoxicity (agar overlay); (2) ocular safety (irritation) evaluations.
## Compatibility/Cleaning Efficacy
Studies were conducted to determine product compatibility with soft contact lenses and its ability to clean laboratory deposited lenses. The studies demonstrated the compatibility and cleaning efficacy of the OFTI-FREE® SUPRACLENS® Daily Protein Remover/Alcon Multi-Purpose Disinfecting Solution ID 90746 regimen.
#### Clinical B.
A study was conducted to clinically evaluate the safety and efficacy of OPTI-FREE® SUPRACLENS® Daily Protein Remover/Alcon Multi-Purpose Disinfecting Solution ID 90746 regimen for cleaning, and disinfection of all soft (hydrophilic) contact lenses (65 patients/130 eyes - 3 months). This clinical study demonstrated the OPTI-FREE® SUPRACLENS® Daily Protein Remover/Alcon Multi-Purpose Disinfecting Solution ID 90746 regimen is safe and effective for the daily simultaneous enzymatic cleaning and disinfection of soft (hydrophilic) contact lenses.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a bird. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 4 1998
Mr. Ralph H. Larsen Manager, Regulatory Affairs Alcon Laboratories, Inc. 6201 South Freeway Fort Worth, Texas 76134-2099
Re: K974624
> Trade Name: Alcon Multi-Purpose Disinfecting Solution ID 90746 Regulatory Class: II Product Code: 86 LPN Dated: December 8, 1997 Received: December 11, 1997
Dear Mr. Larsen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Ralph H. Larsen
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Roerl Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):_K974 62Y
Device Name: Alcon Multi-Purpose Disinfecting Solution ID 90746
Indications for Use:
For the daily cleaning, rinsing, disinfecting, and storing of soft (hydrophilic) contact lenses as recommended by your eye care practitioner. For use in chemical (not heat) disinfection.
Alcon Multi-Purpose Disinfecting Solution ID 90746 can also be used as a diluent for OPTI-FREE® SUPRACLENS® Daily Protein Remover.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Myra Smith
(Division Sign-Off)
Division of Ophthalmic Devices
10(k) Number K974624
Prescription Use_ (Per 21 CFR 801.109)
Over-The-Counter Use X